K232831 · Anoxia Medical, Inc. · BTA · Feb 22, 2024 · General, Plastic Surgery
Device Facts
Record ID
K232831
Device Name
Quiver Aspiration Pump
Applicant
Anoxia Medical, Inc.
Product Code
BTA · General, Plastic Surgery
Decision Date
Feb 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Indications for Use
The Quiver Aspiration Pump is intended for general suction use in hospitals or clinics.
Device Story
The Quiver Aspiration Pump is a portable, battery-operated (18VDC, non-rechargeable) device providing vacuum suction for fluid and debris removal. It consists of a diaphragm pump driven by a DC motor, an integrated 500mL waste collection bag, and sterile disposable accessories (aspiration tubing and inline flow control switch). Used in hospitals or clinics, the device is operated by healthcare personnel to provide suction via a standard medical luer connection. The user initiates and terminates suction using the inline flow control switch at the distal end of the tubing. The pump is non-patient contacting; the entire assembly is single-use and disposable. By providing a vacuum source, the device facilitates the aspiration of materials, aiding in general suction applications.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including visual inspection, leak and vacuum tests (air), max vacuum tests (water), simulated use, tensile strength, system comparative testing, packaging/shelf-life testing, and electrical safety/EMC testing.
Technological Characteristics
Portable, battery-operated (18VDC) diaphragm pump. Provides up to -29inHg vacuum pressure. Includes 500mL waste collection bag. Accessories include 72-inch aspiration tubing and inline flow control switch with standard medical luer fittings. Single-use, non-sterile pump; accessories provided sterile. Non-patient contacting. No software or complex algorithms.
Indications for Use
Indicated for general suction use in hospitals or clinics for the removal or aspiration of fluids and debris.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 22, 2024
Anoxia Medical Inc. % Bosmat Friedman-Cox Regulatory Consultant ProMedoss, Inc. 3521 Hatwynn Rd. Charlotte, North Carolina 28269
Re: K232831
Trade/Device Name: Quiver Aspiration Pump Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA, JCX Dated: September 13, 2023 Received: September 13, 2023
Dear Bosmat Friedman-Cox:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling; medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.02.22 08:17:47 -05'00'
Mark Trumbore, Ph.D., GWCPM Assistant Director, THT4A1: Robotically-Assisted Surgical Devices Team
{2}------------------------------------------------
DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K232831
Device Name
Quiver Aspiration Pump
Indications for Use (Describe)
The Quiver Aspiration Pump is intended for general suction use in hospitals or clinics.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Quiver Aspiration Pump Traditional 510(k)
Image /page/4/Picture/1 description: The image shows the logo for Anoxia Medical, Inc. The logo features a stylized sun graphic in red and white, followed by the company name in bold, black letters. The text is all capitalized and clearly legible.
510(k) Summary
# 510(k) Summary [Traditional 510(k)] Quiver Aspiration Pump 510(k) Number K
#### SUBMITTER 1.
### Applicant's Name:
Anoxia Medical, Inc. 3475 Investment Blvd, #9 Hayward, CA 94545
### Primary Contact:
Bosmat Friedman Regulatory Affairs Consultant 3521 Hatwynn Rd. Charlotte, NC 28269 Phone: 980-308-1636 bosmat.f(@promedoss.com
Date Prepared: September 5, 2023
2. DEVICE Trade Name: Quiver Aspiration Pump Device: Pump, Portable, Aspiration (Manual Or Powered) Classification Code: Product Code: BTA Regulation No: 878.4780 Class: 2 Review Panel: General & Plastic Surgery
#### PREDICATE DEVICE 3.
Penumbra Pump MAX, by Penumbra Inc., Product code JCX, cleared Under: K122756.
#### DEVICE DESCRIPTION 4.
The Quiver Aspiration Pump is designed to provide general suction for use in hospitals or clinics. The Quiver Aspiration Pump provides battery powered suction for the removal or aspiration of fluids and debris. The Pump is a portable, single-use, non-sterile and provides up to -29inHg vacuum pressure. For sterile field use, the pump is provided with sterile accessory tubing and a sterile flow control switch. The accessories are also single-use and disposable. The Quiver Aspiration Pump system is comprised of the following:
- Quiver Aspiration Pump: The Quiver Aspiration Pump is battery-operated (18VDC, nonrechargeable), has an ON/OFF button, and serves as a vacuum source for aspirating and suctioning for general use applications. Within the pump assembly, the DC motor drives a
{5}------------------------------------------------
| | Quiver Aspiration Pump |
|--|-----------------------------------|
| | Traditional 510(k) 510(k) Summary |
diaphragm pump that provides high vacuum suction and flows in one direction for material removal only. Affixed to the pump is a waste collection bag (500mL) that collects and contains all fluids and materials that were aspirated. The aspiration pump is non-patient contacting. Sterile aspiration tubing accessories are provided with the pump to facilitate connection to other suction devices with a standard medical luer connection. The inline flow control switch allows the user to initiate and terminate the suction from the distal end of the tubing. The entire pump assembly is disposable following standard biohazard handling procedures.
- Accessories: Included in the pump package are the aspiration tubing and the inline control ● flow switch. Both accessories are individually packaged in sterile pouches. These components are finished sterile devices manufactured and sterilized by Merit Medical (South Jordan, UT) and sold to Anoxia as finished medical devices. The aspiration tubing has a standard female luer fitting. The inline flow control switch is connected to the distal end of the tubing. The tubing is 72 inches long to allow for use within a sterile field if necessary for general suction applications. The flow switch has a standard male luer that can be connected to other devices used for suction that have the same standard medical luer. The two accessories and the pump are packaged in a carton box.
#### 5. INDICATIONS FOR USE
The Quiver Aspiration Pump is intended for general suction use in hospitals or clinics.
#### SUBSTANTIAL EQUIVALENCE 6.
The Quiver Aspiration Pump is substantially equivalent to the predicate device based on the following:
# Intended Use
The intended use of the proposed device is identical to that of the cleared device.
# Technology
Both the Quiver Pump and Penumbra Pump perform aspiration through a single lumen aspiration tube attached to a continuous aspiration pump with a waste collection element. Any differences between the subject and predicate device were evaluated through comparative performance evaluation and design verification and validation testing which demonstrated device performance and safety. Comparative tests included Peak maximum vacuum pressure (inHg). Maximum vacuum values were obtained after stabilization of the pump was recorded. Comparative testing yielded substantially similar results.
# Discussion
Based on the comparison presented in our submission, the Quiver Aspiration Pump shares the same indications for use as the Penumbra Pump MAX, as both are intended for general suction use in hospitals or clinics.
Additionally, The Quiver Aspiration Pump and Penumbra Pump MAX also share the same operating principal and have similar technological characteristics and mechanisms of action as
{6}------------------------------------------------
Quiver Aspiration Pump Traditional 510(k)
both serve as the source of vacuum for aspiration of materials. Both systems are provided with sterile aspiration tubing and inline flow control switch. Performance testing was completed to demonstrate safety and effectiveness of the subject device, supporting the company's substantial equivalency claim.
#### PERFORMANCE DATA 7.
ANOXIA MEDICAL. INC.
To demonstrate the substantial equivalence of the Ouiver Aspiration Pump to the selected predicate device, the performance and technological characteristics were evaluated by completion of the following tests:
- Visual Inspection Test
- Leak and Vacuum Test (Air) ●
- Max Vacuum Test (Water) ●
- Simulated Use Test ●
- Tensile Strength Test
- System Comparative Test
- Packaging and Shelf-Life Test
- . Electrical Safety and EMC Testing
#### CONCLUSION 8.
The Ouiver Aspiration Pump has the same/similar intended use, technological characteristics, and principles of operation as the predicate. Any differences between the subject and predicate devices were evaluated through design verification and validation testing and comparative testing which demonstrated device performance and safety, and do not raise any new questions of safety and effectiveness.
Based on the information submitted in this 510(k), the Quiver Aspiration Pump is substantially equivalent to the currently marketed predicate Penumbra Pump Max System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.