The Vibrant System is an orally administered Capsule that is indicated for the treatment of adults with chronic idiopathic constipation who have not experienced relief of their bowel symptoms by using laxative therapies at the recommended dosage for at least one month.
Device Story
Vibrant System is an orally ingested, transient, drug-free capsule designed to mechanically stimulate the colon to induce peristaltic activity and facilitate defecation. The system consists of a reusable pod and a single-use capsule. The capsule contains a coin vibration motor, electronic card, and batteries. It operates on a programmed vibration cycle (190 cycles/hour; 3s vibration/16s repose) to promote transit. The subject device introduces an adaptive delay mechanism that automatically adjusts stimulation start time if the capsule is ingested at the wrong time. Patients use the system at home; the pod communicates with the capsule via RF and with the cloud via GSM. Healthcare providers use the system to treat chronic idiopathic constipation in adults. The device is expelled naturally during bowel movements. Benefits include mechanical relief of constipation symptoms without pharmacological intervention.
Clinical Evidence
Prospective, randomized, multi-center, double-blinded, placebo-controlled study (n=100) evaluated bi-weekly dosing. Primary endpoint: CSBM1 success rate (increase of ≥1 weekly CSBM for ≥6 of 8 weeks). Results: mITT active arm CSBM1 success rate 55.81% vs. 32.08% placebo (p=0.0161). CSBM2 success rate (increase of ≥2 weekly CSBM) was not statistically significant. Safety: 16% AE incidence in active arm; AEs primarily limited to mild GI vibration/tickling sensations. Study concluded the device is safe and effective for bi-weekly use.
Indicated for adults with chronic idiopathic constipation refractory to laxative therapies for at least one month.
Regulatory Classification
Identification
An orally ingested transient device for constipation is an electric swallowable capsule that naturally passes through the gastrointestinal tract for the treatment of constipation. The Vibrant System is indicated for the treatment of adults with chronic idiopathic constipation who have not experienced relief of their bowel symptoms by using laxative therapies at the recommended dosage for at least one month.
Special Controls
In combination with the general controls of the FD&C Act, an orally ingested transient device for constipation is subject to the following special controls:
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 8, 2023
Vibrant Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103
Re: K232830
Trade/Device Name: Vibrant System Regulation Number: 21 CFR 876.5940 Regulation Name: Orally Ingested Transient Device For Constipation Regulatory Class: Class II Product Code: QTN Dated: September 13, 2023 Received: September 13, 2023
Dear Janice Hogan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephanie Cole -S
for Glenn B. Bell, PhD Division Director DHT3A: Division of Renal, Gastrointestinal. Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality
{2}------------------------------------------------
Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232830
Device Name Vibrant System
Indications for Use (Describe)
The Vibrant System is an orally administered Capsule that is indicated for the treatment of adults with chronic idiopathic constipation who have not experienced relief of their bowel symptoms by using laxative therapies at the recommended dosage for at least one month.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY Vibrant's Vibrant System
## Submitter
Vibrant Ltd. Hakochav P.O. Box 516 Yokneam, Israel 2069206
Phone: +972.46.662379 Facsimile: +972.46.664424 Contact Person: Martha Bezalel, Ph.D.
Date: September 13, 2023
Trade Name : Vibrant System
Regulation Name: Orally Ingested Transient Device For Constipation
Regulation Number: 21 CFR 876.5940
Requlatory Class: II
Product Code: QTN
Predicate Device Vibrant Ltd.'s Vibrant System (K223031)
#### A. Device Description
The Vibrant System is designed to mechanically induce peristaltic activity in the colon, thus aiding in relieving constipated patients. Constipation relief is achieved by the Capsule's stimulation in the colon, consequently inducing peristaltic activity which promotes transit and facilitates defecation. The stimulation protocol is designed to activate at specific times during the day (i.e., afternoon and evening).
The Vibrant System is comprised of two components: a reusable pod and a single use capsule. The Capsule is expelled from the body during the patient's bowel movements.
#### B. Intended Use/Indications for Use
The Vibrant System is an orally administered Capsule that is indicated for the treatment of adults with chronic idiopathic constipation who have not experienced relief of their bowel symptoms by using laxative therapies at the recommended dosage for at least one month.
#### C. Summary of Technological Characteristics
The following changes are being made in this 510(k) with respect to the predicate:
{5}------------------------------------------------
- -An optional change in the frequency of administration. While the labeling in the predicate (K223031) recommends a dosing of 5 times a week, the Vibrant System includes an alternative dosing option of 2 times/week (in addition to the 5 times a week dosing option).
- -The modified Vibrant capsule includes an adaptive delay mechanism. This feature involves the automatic delay in stimulation start time of the capsule.
- -There is also a change to the raw materials used for the capsule shell component of the device.
Performance testing and the clinical study confirm that these changes do not adversely impact performance with respect to safety or effectiveness, and the device continues to meet the same special controls as the predicate.
Table 1, comparing the key features of the subject and predicate devices is provided below.
| | Predicate Vibrant System<br>(K223031) | Subject Vibrant System |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Indications for Use | The Vibrant System is an orally administered Capsule that is indicated for the treatment of adults with chronic idiopathic constipation who have not experienced relief of their bowel symptoms by using laxative therapies at the recommended dosage for at least one month. | Same |
| Population | Adults | Same |
| Components | Pod, capsule, cloud software | Same |
| Pod-capsule Communication | RF | Same |
| Pod- Cloud Communication | GSM | Same |
| Capsule Calibration | During capsule production | Same |
| Capsule Specifications | Length: 24.5 mm<br>Diameter: 11.20 mm<br>Weight: 4.0 g | Same |
| Capsule Components | Coin vibration motor, electronic card, batteries (2 x 1.5v 60mAh), and connector, all encapsulated in a 2-piece shell | Same |
| Capsule External Material | PC Makrolon 2458 | Makrolon 2458 010110 |
# Table 1: Key Features of the Subject and Predicate Devices
{6}------------------------------------------------
| | Predicate Vibrant System<br>(K223031) | Subject Vibrant System |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Vibration Cycle | 190 vibration cycles/ hour; each cycle: 3 sec. vibration and 16 sec. of repose (rest). The duration of each vibration cycle set alternates between 100 minutes (sets 1 and 3), 120 minutes (set 2), and 180 minutes (set 4). | Same |
| Adaptive delay mechanism | None. Device has a set delay mechanism of 14 hours since activation. | If device is taken at the wrong time, the device automatically adjusts the delay, ensuring the stimulation happens at the intended time. |
| Dosage | 5 times weekly | Two dosage options:<br>- 5 times weekly <b>or</b><br>- twice weekly |
#### D. Summary of Nonclinical/Bench Studies
The Vibrant System conforms to the following relevant standards:
- . ASTM D4332-14 - Standard Practice for Conditioning Containers, Packaging Components for Testing
- . ASTM D4169-16 - Standard Practice for Performance Testing of Shipping Containers and Systems
- . ANSI AAMI ISO 10993-1:2018 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- . ANSI AAMI IEC 62304:2006/A1:2016 - Medical device software - Software life cycle processes [Including Amendment 1 (2016)]
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 -Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- ANSI AAMI IEC 60601-1-2:2014 - Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests
- IEC 60601-1-11 Edition 2.0 2015-01 - Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- FCC Part 15, Subpart C, Section 15.209: Unlicensed Wireless Devices Compliance Testing ●
In addition, the performance of the modified Capsule was evaluated with nonclinical bench testing as listed below.
{7}------------------------------------------------
- Mechanical Integrity when placed under loads
- . Seal integrity
- . pH Resistance
- . Dimensional Verification of the capsules
- . Visual Inspection for absence of surface defects and extraneous matter
#### ய் Summary of Clinical Information
A clinical study was conducted, the objective of which was to assess the efficacy and safety of the Vibrant Capsule, administered 2 times a week. This was a prospective, randomized, multi-center, double-blinded, placebo-controlled study. The study enrolled 100 subjects. Study duration per patient included 8 weeks of treatment. Two arms were assessed (at a ratio of 1:1): (a) Vibrant Capsule, and (b) Placebo Capsule.
The study population was comprised of subjects with chronic idiopathic constipation (CIC) who were refractory to existing OTC or Rx treatments or could not tolerate the side effects.
There were two primary efficacy endpoints:
- CSBM1 success rate, defined as an increase of at least one weekly Complete Spontaneous Bowel Movement (CSBM) during at least 6 of the 8 weeks of treatment, and
- CSBM2 success rate, defined as an increase of at least two weekly CSBM during at least 6 of the 8 weeks of treatment.
Secondary efficacy endpoints included:
- · Change from baseline in average straining.
- . Increase of at least one weekly CSBM during at least 5 of the 8 weeks of treatment.
- . Change from baseline in average stool consistency.
- . Increase of at least two weekly CSBM during at least 5 of the 8 weeks of treatment.
- . Change from baseline in average bloating.
Selected additional efficacy endpoints included increase of at least one weekly Spontaneous Bowel Movement during at least 6 out of 8 weeks of treatment and change in quality of life.
Safety endpoints include all adverse events related and unrelated to the study treatment.
Analysis sets: The Intend-to-Treat (ITT) analysis set consisted of all 100 subjects who were enrolled in the study and randomized. The Modified Intend-to-Treat (mlTT) analysis set consisted of all 96 randomized subjects who met the inclusion criteria of the protocol.
{8}------------------------------------------------
Demographics: 100% (100/100) of the subjects were enrolled in US sites. 75% (75/100) of the participants were female, consistent with the expected gender distribution in the indicated population. The study participants were between 22.6 and 68.5 years old, with a mean age of 43.3 years (SD=12.27). The ethnic distribution was: Caucasian - 22%; Hispanic or Latino - 49%; Black or African American - 15%; Native American or American Indian - 2%; Asian/ Pacific Islander - 11%; Other -1.0%.
No difference in gender or ethnicity between the treatment groups was considered to be statistically significant.
Primary Efficacy Results: Table 2 presents the CSBM2 success rates by study arm and analysis set. In the mITT analysis set, the CSBM1 success rate was higher in the active arm than in the placebo arm by 23.7%. The CSBM2 success rate was higher in the active arm than in the placebo arm, but the difference was not statistically significant. These results are repeated in the ITT analysis set as well.
| | | Active | | Placebo | | |
|------|-----------------------------------------------|-------------------|--------------------------|-------------------|--------------------------|------------------|
| | | % (n/N) | 95% CI (Wilson<br>Score) | % (n/N) | 95% CI (Wilson<br>Score) | p-value<br>(CMH) |
| mITT | Imp. by at least 1 CSBM<br>- 6 out of 8 weeks | 55.81%<br>(24/43) | [41.11%;69.57%] | 32.08%<br>(17/53) | [21.09%;45.48%] | 0.0161 |
| | Imp. by at least 2 CSBM<br>- 6 out of 8 weeks | 30.23%<br>(13/43) | [18.60%;45.11%] | 22.64%<br>(12/53) | [13.45%;35.53%] | 0.5099 |
| ITT | Imp. by at least 1 CSBM<br>- 6 out of 8 weeks | 54.55%<br>(24/44) | [40.07%;68.29%] | 30.36%<br>(17/56) | [19.90%;43.34%] | 0.0165 |
| | Imp. by at least 2 CSBM<br>- 6 out of 8 weeks | 29.55%<br>(13/44) | [18.16%;44.22%] | 21.43%<br>(12/56) | [12.71%;33.82%] | 0.5091 |
# Table 2: CSBM1 and CSBM2 Success Rates
Secondary and Additional Endpoints: While secondary endpoints, including quality of life, straining during bowel movements and stool consistency, were numerically higher in the Vibrant System group, the difference between the groups on these secondary assessments was not statistically significant.
Safety Analysis: A total of 42 Adverse Events (AE) were reported during the study, incidence of 16.00% and 19.64% in the active and placebo arm respectively. Eight AE's were considered possibly or related to the study treatment or procedure. Adverse events considered to the use of the device were mostly limited to vibration/tickling sensations in the GI tract (1 case of mild headache was considered "possibly related"). The incidence of AE is decreased when compared to the clinical study presented in DEN210052 in which the device was taken 5 times/week. The decreased incidence in AE in the biweekly dosing study is expected as the dosing is reduced.
Overall, the study was deemed successful as the CSBM1 success rate was statistically significantly higher in the active arm than in the placebo arm.
{9}------------------------------------------------
# Conclusion
Substantial equivalence of the subject Vibrant System compared to the predicate Vibrant System (K223031) was demonstrated by nonclinical and clinical performance testing provided in this 510(k) premarket notification. The same nonclinical testing was used to support the subject Vibrant System as was used for the predicate device. Similarly, although the modified system includes a labeling change that allows an optional change in dosing frequency, i.e. twice weekly administration, clinical testing demonstrates the device is as safe and effective when administered with this frequency as compared to the predicate. In addition, the subject Vibrant System continues to comply with the special controls set forth by FDA for this device type.
In conclusion, the modified Vibrant System is substantially equivalent to the currently marketed predicate device, Vibrant System (K223031), in terms of indications for use, technological characteristics, and safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.