← Product Code [DPS](/productcode/DPS) · K232823

# MCG-S (AM1000) (K232823)

_Amcg Co., Ltd. · DPS · Apr 5, 2024 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K232823

## Device Facts

- **Applicant:** Amcg Co., Ltd.
- **Product Code:** [DPS](/productcode/DPS.md)
- **Decision Date:** Apr 5, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

AM1000 is intended for use as a tool which non-invasively measures and displays the magnetic signals produced by the electric currents of the heart.

## Device Story

MCG-S (AM1000) is a magnetocardiograph system for adult cardiology. It records biomagnetic signals (X- and Y-components) generated by cardiac ionic currents. The system utilizes a Dewar containing 96-channel SQUID (Superconducting Quantum Interference Device) sensors, a gantry, electronics, and software. It employs the DROS (Double Relaxation Oscillation SQUID) method to enhance measurement sensitivity. Sensors are housed in a vacuum-insulated Dewar to minimize mechanical vibration and provide noise shielding. The device is operated in a shielded room. Data is transmitted via noise-free optical fiber to a workstation running Windows 10. The software processes and analyzes the magnetic signals, which are displayed for clinical review. By providing non-invasive, radiation-free mapping of cardiac magnetic fields, the device assists clinicians in evaluating cardiac electrical activity.

## Clinical Evidence

No clinical data provided. Evidence consists of bench testing, including electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2), software verification/validation, and performance testing (sensitivity, boil-off rate, dimensional accuracy, sensor operation, heart positioning, heart rate measurement, and sensor calibration).

## Technological Characteristics

System uses 96-channel SQUID sensors (DROS method) for X and Y magnetic field components. Requires liquid helium cooling (13L volume) and a shielded room. Connectivity via optical fiber. Software runs on Windows 10. Dimensions/specs: 1-10 kHz sampling rate, sensitivity < 10 fTrms/√Hz at 100 Hz, 50 mm baseline. Complies with IEC 60601-1 and IEC 60601-1-2.

## Regulatory Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

## Predicate Devices

- CS-MAGII ([K121825](/device/K121825.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

AMCG Co., Ltd. Soyeon Kim Senior Regulatory Affairs F8, 156, Seochojungang-ro, Seocho-gu Seoul, 06605 Korea, South

Re: K232823

Trade/Device Name: Mcg-s (am1000) Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: February 26, 2024 Received: February 26, 2024

Dear Soyeon Kim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Kimberly

For: Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices

{2}------------------------------------------------

Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K232823

Device Name

MCG-S (AM1000)

Indications for Use (Describe)

AM1000 is intended for use as a tool which non-invasively measures and displays the magnetic signals produced by the electric currents of the heart.

Type of Use (Select one or both, as applicable)

> | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

This 510(k) Summary is being submitted in according with requirement of 21 CFR Part 807.92.

| Submitter's Name             | AMCG Co., Ltd.                                                    |
|------------------------------|-------------------------------------------------------------------|
| Submitter's Address          | F8, 156, Seochojungang-ro, Seocho-gu, Seoul,<br>Republic of Korea |
| Submitter's Telephone number | +82-2-598-8001                                                    |
| Contact Person               | Soyeon Kim<br>AMCG Co., Ltd.<br>+82-2-598-8001                    |
| Date Summary was Prepared    | September 08, 2023                                                |
| Trade or Proprietary Name    | MCG-S (AM-1000)                                                   |
| Common or Usual Name         | Magnetocardiograph                                                |
| Classification               | Class II per 21 CFR §870.2340                                     |
| Product Code                 | DPS                                                               |
| Classification Panel         | Cardiovascular                                                    |

# 1. Device Description

AMCG's MCG-S will be used for purpose of measurement and analysis in adult cardiology. Magnetocardiography (MCG) is a non-invasive, non-contact, radiation-free, multichannel body surface mapping technique to record biomagnetic signals of X- and Y-components from the heart generated by the same ionic currents underling the electrocardiogram.

This system consists of Dewar with 96-channel SQUIDs, gantry, electronics, power supply device, and software. 48 of these SQUID sensors are used for detecting the X-components, and the remaining 48 sensors are for the Y-components of the magnetic field. The device does not measure and/or display the magnetic field in the Z-component.

The system increases the output signal of the SQUID by applying the DROS (Double Relaxation Oscillation SQUID) method to increase the measurement sensitivity of the cardiac magnetic field signal, and a wide range of channel area is applied to measure the magnetic field area throughout the heart within a short time. In addition, in order to minimize the mechanical vibration of the sensor, the sensor is installed in a vacuum area, and Dewar, which has excellent noise shielding and insulation performance, is applied.

The system is controlled through noise-free optical fiber signals and can control, measure, process and analyze signals through software.

{5}------------------------------------------------

# 2. Indications for Use

AM1000 is intended for use as a tool which non-invasively measures and displays the magnetic signals produced by the electric currents of the heart.

{6}------------------------------------------------

## 3. Technological Characteristics

MCG-S(AM1000) is substantially equivalent in terms of indications for use, technical characteristics, and performance to predicate device (K121825) .

| K number | Proprietary Name | Manufacturer             | Predicate Type   |
|----------|------------------|--------------------------|------------------|
| K121825  | CS-MAGII         | BIOMAGNETIK-PARK<br>GMBH | Predicate device |

Similar or different technical characteristics exist to MCG-S (AM1000), a predicate device.

The following technological differences exist between the subject and predicate device:

| Elements of<br>Comparison                    | MCG-S (AM1000)<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                        | CS-MAGII<br>(K121825)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar |         |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------|
| Regulation<br>Number No.                     | 21CFR 870.2340                                                                                                                                                                                                                                                                                                                                                                                                            | 21CFR 870.2340                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |         |         |
| Product Code                                 | DPS                                                                                                                                                                                                                                                                                                                                                                                                                       | DPS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |         |         |
| Indications<br>for use                       | AM1000 is intended for<br>use as a tool which non-<br>invasively measures and<br>displays the magnetic<br>signals produced by the<br>electric currents of the<br>heart.                                                                                                                                                                                                                                                   | The BMP MCG CS-<br>MAGII<br>Magnetocardiograph is<br>intended for use as tool<br>which non-invasively<br>measures and displays<br>the magnetic signals<br>produced by the electric<br>currents in the heart.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |         |         |
|                                              | Type of use                                                                                                                                                                                                                                                                                                                                                                                                               | Prescription Use only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Prescription Use only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | X       |         |
|                                              | Description                                                                                                                                                                                                                                                                                                                                                                                                               | AMCG's MCG-S will<br>be used for purpose of<br>measurement and<br>analysis in adult<br>cardiology.<br>Magnetocardiography<br>(MCG) is a non-<br>invasive, non-contact,<br>radiation-free                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | The BMP MCG CS-<br>MAGII system will be<br>used for diagnostic<br>purposes in adult<br>cardiology.<br>Magnetocardiography<br>(MCG) is a non-<br>invasive, non-contact,<br>radiation-free                                                                                                                                                                                                                                                                                                                                                                                                             |         | X       |
| Elements of<br>Comparison                    |                                                                                                                                                                                                                                                                                                                                                                                                                           | MCG-S (AM1000)<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | CS-MAGII<br>(K121825)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same    | Similar |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                                           | multichannel body<br>surface mapping<br>technique to record<br>biomagnetic signals of<br>X- and Y-components<br>from the heart generated<br>by the same ionic<br>currents underling the<br>electrocardiogram.<br>This system consists of<br>Dewar with 96-channel<br>SQUIDs, gantry,<br>electronics, power<br>supply device, and<br>software. 48 of these<br>SQUID sensors are used<br>for detecting the X-<br>components, and the<br>remaining 48 sensors<br>are for the Y-<br>components of the<br>magnetic field. The<br>device does not measure<br>and/or display the<br>magnetic field in the Z-<br>component.<br>The system increases<br>the output signal of the<br>SQUID by applying the<br>DROS (Double<br>Relaxation Oscillation<br>SQUID) method to<br>increase the<br>measurement sensitivity<br>of the cardiac magnetic<br>field signal, and a wide<br>range of channel area is<br>applied to measure the<br>magnetic field area<br>throughout the heart | multichannel body<br>surface mapping<br>technique to record<br>biomagnetic signals<br>from the heart generated<br>by the same ionic<br>currents underling the<br>electrocardiogram.<br>Compared to<br>electrocardiography<br>(ECG), MCG has<br>similar morphological<br>features such as T-, P-,<br>and Q-waves, and the<br>QRS complex. The<br>advantages of MCGs<br>over traditional EGGs<br>are increased sensitivity<br>to weak signals, lack of<br>distortion from<br>conductivity in body<br>tissues, and presentation<br>of direct current (DC)<br>component signals and<br>primary currents. |         |         |
| Elements of<br>Comparison                    |                                                                                                                                                                                                                                                                                                                                                                                                                           | MCG-S (AM1000)<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | CS-MAGII<br>(K121825)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same    | Similar |
|                                              | within a short time. In<br>addition, in order to<br>minimize the<br>mechanical vibration of<br>the sensor, the sensor is<br>installed in a vacuum<br>area, and Dewar, which<br>has excellent noise<br>shielding and insulation<br>performance, is applied.<br>The system is controlled<br>through noise-free<br>optical fiber signals and<br>can control, measure,<br>process and analyze<br>signals through<br>software. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |         |         |
| Target<br>population                         | Adult                                                                                                                                                                                                                                                                                                                                                                                                                     | Adult                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |         |         |
| Configurations                               | Dewar, Gantry,<br>Electronics(Electric<br>module, DC Power filter<br>module, Power supply<br>device), Software                                                                                                                                                                                                                                                                                                            | SQUID gradiometer,<br>Insert, Dewar,<br>Electronics, Software,<br>Gantry and Bed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | X       |         |
| Sensor Type                                  | SQUID(DROS)                                                                                                                                                                                                                                                                                                                                                                                                               | SQUID(DROS)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |         |         |
| Shielded<br>Room                             | Required                                                                                                                                                                                                                                                                                                                                                                                                                  | Required                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |         |         |
| Cryogen Used                                 | Liquid Helium                                                                                                                                                                                                                                                                                                                                                                                                             | Liquid Helium                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |         |         |
| Helium<br>reliquefied<br>system              | Not applicable                                                                                                                                                                                                                                                                                                                                                                                                            | Not applicable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |         |         |
| Operating<br>System                          | Windows 10                                                                                                                                                                                                                                                                                                                                                                                                                | Windows                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |         |         |
| Number of<br>SQUID<br>channels               | 96 channels                                                                                                                                                                                                                                                                                                                                                                                                               | 64 channels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | X       |         |
| SQUID<br>gradiometer<br>sensitivity          | less than 10 fTrms/√Hz<br>(at 100 Hz)                                                                                                                                                                                                                                                                                                                                                                                     | 3.5 fTrms/√Hz (at 100<br>Hz)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | X       |         |
| Sampling rate                                | 1-10 kHz                                                                                                                                                                                                                                                                                                                                                                                                                  | 1 kHz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | X       |         |
| Elements of<br>Comparison                    | MCG-S (AM1000)<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                        | CS-MAGII<br>(K121825)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar |         |
| Liquid He<br>volume                          | 13 L                                                                                                                                                                                                                                                                                                                                                                                                                      | 40 L                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | X       |         |
| Liquid He<br>Boil-off and<br>refill interval | about 4 days                                                                                                                                                                                                                                                                                                                                                                                                              | 7 days                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | X       |         |
| Liquid He<br>Boil-off Rate                   | less than 3.5 L/day                                                                                                                                                                                                                                                                                                                                                                                                       | Average 4.41 L/day                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | X       |         |
| Baseline                                     | 50 mm                                                                                                                                                                                                                                                                                                                                                                                                                     | 70 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | X       |         |
| Sensing Coil<br>(Detecting<br>Coil)          | 1st planar gradiometer<br>(X- and Y-component)                                                                                                                                                                                                                                                                                                                                                                            | 1st axial gradiometer<br>(Z-component)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | X       |         |

# Table 2. Comparison Table to Predicate Device

{7}------------------------------------------------

{8}------------------------------------------------

Confidential

{9}------------------------------------------------

Therefore, the differences between the subject device MCG-S(AM1000) and the predicate device do not raise new questions of the safety and effectiveness.

{10}------------------------------------------------

#### 4. Performance Data

MCG-S(AM1000) has been extensively tested in Electromagnetic Compatibility and Electrical Safety, and bench testing.

#### 1) Electromagnetic Compatibility and Electrical Safety

The electromagnetic and electrical safety of MCG-S(AM1000) was tested according to IEC 60601-1 and IEC 60601-1-2 to guarantee that the device meets electromagnetic compatibility and electrical safety requirements.

## 2) Software

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

#### 3) Performance

Performance tests were conducted to evaluate the basic performance. Following tests were performed and met the internal criteria.

- . Sensitivity Test
- L-He Boil-off Rate Test
- . Dimensional Test
- Sensor Operation Test
- Accuracy Test
- . Heart Positioning Test
- Heart Rate Measurement Test
- Sensor Calibration Test

## 5. Summary of Substantial Equivalence

Based on the comparison, we have demonstrated that MCG-S (AM1000) has been shown to be as safe and effective for the proposed Indications for Use as the legally marketed predicate device. Therefore, we conclude that the proposed MCG-S (AM1000) is substantially equivalent to those predicate device.

{11}------------------------------------------------

#### 6. Conclusion

The results of safety and performance tests demonstrate the substantial equivalence of MCG-S (AM1000). There are some differences with the predicate device, however, the differences between the subject device MCG-S (AM1000) and the predicate device do not raise new questions of the safety and effectiveness. Therefore, MCG-S (AM1000) is substantially equivalent to the predicate device and does not raise any new issues regarding safety or efficacy when compared to its predicate device.

---

**Source:** [https://fda.innolitics.com/device/K232823](https://fda.innolitics.com/device/K232823)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
