Fluidshield* 2 Fog-Free Surgical Mask (62113); Fluidshield* 2 Fog-Free Surgical Mask with Wraparound Visor (62114); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops (62115); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops and WrapAround Visor (62116)
K232812 · Owens & Minor (O&M) Halyard, Inc. · FXX · Jun 14, 2024 · General, Plastic Surgery
Device Facts
Record ID
K232812
Device Name
Fluidshield* 2 Fog-Free Surgical Mask (62113); Fluidshield* 2 Fog-Free Surgical Mask with Wraparound Visor (62114); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops (62115); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops and WrapAround Visor (62116)
Applicant
Owens & Minor (O&M) Halyard, Inc.
Product Code
FXX · General, Plastic Surgery
Decision Date
Jun 14, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Fluidshield* 2 Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Surgical Mask is a single use, disposable device, provided non-sterile. The Fluidshield* 2 Procedure Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Procedure Mask is a single use, disposable device, provided non-sterile.
Device Story
Fluidshield* 2 masks are 4-layer, non-woven, single-use, disposable surgical and procedure masks. Surgical versions feature ties; procedure versions feature earloops and a malleable nosepiece. Some models include a wraparound visor. Masks are worn by healthcare personnel and patients to provide a barrier against microorganisms, body fluids, and particulates. They are used in clinical settings for infection control. The device acts as a physical barrier; no electronic or algorithmic components are involved. Benefits include reduced potential exposure to blood and body fluids for the wearer.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM F2100 (Level 2), ASTM F2101 (Bacterial Filtration Efficiency >98%), ASTM F2299 (Particulate Filtration Efficiency >98%), EN 14683 (Differential Pressure <6.0 mmH2O/cm2), ASTM F1862 (Fluid Resistance 120 mmHg), 16 CFR Part 1610 (Flammability Class 1), and ISO 10993 (Biocompatibility: Cytotoxicity, Sensitization, Irritation).
Technological Characteristics
4-layer non-woven construction: polyester cellulose, spunbond polypropylene, and meltblown polypropylene filter media. Features include aluminum nose wire, polyester lycra earloops or polypropylene/spunlace ties, and optional polyester film visors. Single-use, non-sterile. Complies with ASTM F2100 Level 2, ASTM F2101, ASTM F2299, EN 14683, ASTM F1862, 16 CFR 1610, and ISO 10993 standards.
Indications for Use
Indicated for healthcare personnel and patients to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices. Single-use, disposable, non-sterile.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
KC 200 Fluidshield* Surgical Masks and Procedure Masks and KC 300 Fluidshield* Surgical Masks and Procedure Masks (K111402)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
June 14, 2024
Owens & Minor (O&M) Halyard, Inc. Angela Bunn Sr. Director, Global Regulatory Affairs 9120 Lockwood Blvd Mechanicsville, Virginia 23116
### Re: K232812
Trade/Device Name: Fluidshield* 2 Fog-Free Surgical Mask (62113); Fluidshield* 2 Fog-Free Surgical Mask with Wraparound Visor (62114); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops (62115); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops and WrapAround Visor (62116) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 9, 2024 Received: May 9, 2024
Dear Angela Bunn:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image shows the text "Allan Guan-S" in a large, sans-serif font. The text is black and is set against a white background. There is a faint, light blue watermark in the background that is partially obscured by the text.
For Bifeng Qian, M.D., Ph.D. Assistant Director
{2}------------------------------------------------
DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232812
#### Device Name
Fluidshield * 2 Fog-Free Surgical Mask (62113)
Fluidshield * 2 Fog-Free Surgical Mask with Wraparound Visor (62114)
Fluidshield * 2 Procedure Mask with SO SOFT* Earloops (62115)
Fluidshield * 2 Procedure Mask with SO SOFT* Earloops and WrapAround Visor (62116)
Indications for Use (Describe)
The Fluidshield* 2 Surgical Mask is intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Surgical Mask is a single use, disposable device, provided non-sterile.
The Fluidshield* 2 Procedure Mask is intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Procedure Mask is a single use, disposable device, provided non-sterile.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
x Dver-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Owens & Minor logo, which consists of the letters "OM" in a stylized font, with the words "Owens & Minor" to the right of the letters. The letters "OM" are in a dark red color, while the words "Owens & Minor" are in a light gray color. Below the logo, the text "510(k) Summary K232812" is written in a bold font.
| Date Summary<br>was Prepared | June 14, 2024 |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter | O & M Halyard, Inc.<br>1 Edison Drive<br>Alpharetta, GA 30005 |
| Primary Contact for<br>this 510(k) Submission | Angela L. Bunn, RAC<br>Tel: 470-347-7147<br>Email: angela.bunn@owens-minor.com |
| Marketed Common<br>Name | Surgical Mask and Procedure Mask |
| Device Submission Trade<br>Name and Description | Fluidshield * 2 Fog-Free Surgical Mask (62113)<br>Fluidshield * 2 Fog-Free Surgical Mask with Wraparound Visor<br>(62114)<br>Fluidshield * 2 Procedure Mask with SO SOFT* Earloops,<br>(62115)<br>Fluidshield* 2 Procedure Mask with SO SOFT* Earloops and<br>WrapAround Visor (62116) |
| Device Common Name | Surgical Mask and Procedure Mask |
| Device Product Code<br>and Classification Name | FXX Class II, 21 CFR §878.4040 Surgical Apparel |
| Predicate Device | KC 200 Fluidshield* Surgical Masks and Procedure Masks and<br>KC 300 Fluidshield* Surgical Masks and Procedure Masks<br>(K111402) |
| Subject Device<br>Description | The subject device(s) are ASTM Level 2 Surgical and<br>Procedure Mask(s). Fluidshield* 2 Surgical Mask(s),<br>Fluidshield* 2 Procedure Mask(s) family of products are a four-<br>layer mask, constructed of well-known non-woven |
| | materials used in facial protection. The Surgical Masks are<br>provided with ties while the Procedure Masks are provided with<br>earloops and a malleable nosepiece.<br><br>The Fluidshield* 2 Surgical Mask(s), Fluidshield* 2 Procedure<br>Mask(s) family will be provided in designs with and without a<br>protective visor. The Fluidshield* 2 Surgical Mask(s),<br>Fluidshield* 2 Procedure Mask(s) family are single use,<br>disposable devices, provided non-sterile. |
| Indications for Use | The Fluidshield* 2 Surgical Mask is intended to be worn to protect<br>both the patient and healthcare personnel from transfer of<br>microorganisms, body fluids, and particulate material. These<br>face masks are intended for use in infection control practices to<br>reduce the potential exposure of the wearer to blood and body<br>fluids. The Fluidshield* 2 Surgical Mask is a single use,<br>disposable device, provided non-sterile.<br><br>The Fluidshield* 2 Procedure Mask is intended to be worn to<br>protect both the patient and healthcare personnel from transfer of<br>microorganisms, body fluids, and particulate material. These<br>face masks are intended for use in infection control practices to<br>reduce the potential exposure of the wearer to blood and body<br>fluids. The Fluidshield* 2 Procedure Mask is a single use,<br>disposable device, provided non-sterile. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Owens & Minor. The logo features a stylized "OM" in a dark red color, with the words "Owens & Minor" in a gray serif font to the right of the symbol. The logo is clean and professional, suggesting a well-established company.
| Technological Characteristics Comparison Table | | | |
|------------------------------------------------|-----------------|------------------------------------------------------------------|------------|
| | Subject Device | Predicate Device<br>(K111402)<br>KC 300 and KC 200<br>Face Masks | Comparison |
| FDA Product Code | FXX | FXX | Same |
| FDA Classification | Class II | Class II | Same |
| Regulation Number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Common Name | Surgical Mask | Surgical Mask | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Owens & Minor, a healthcare logistics company. The logo consists of a stylized "OM" in a dark red color, with the words "Owens & Minor" in a light gray color to the right of the "OM" symbol. The "OM" symbol is made up of several horizontal lines, creating a modern and abstract design. The overall design is clean and professional, reflecting the company's focus on healthcare and logistics.
| Device Trade Name | Fluidshield * 2 Fog-Free<br>Surgical Mask (62113) | KC200 Procedure Mask | Similar |
|----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| | Fluidshield * 2 Fog-Free<br>Surgical Mask with<br>Wraparound Visor (62114) | KC200 Procedure Mask,<br>fog-free with visor | |
| | Fluidshield * 2 Procedure<br>Mask with SO SOFT*<br>Earloops, (62115) | KC200 Surgical Mask,<br>fog-free | |
| | Fluidshield* 2 Procedure<br>Mask with SO SOFT*<br>Earloops and WrapAround<br>Visor (62116) | KC200 Surgical Mask,<br>fog-free with visor | |
| | | KC300 Procedure Mask,<br>Fluidshield*, fog-free | |
| | | KC300 Procedure Mask,<br>Fluidshield*, fog-free with<br>visor | |
| | | KC300 Surgical Mask,<br>Fluidshield*, fog-free | |
| | | KC300 Surgical Mask,<br>Fluidshield*, fog-free with<br>visor | |
| Intended Use | The Fluidshield* 2 Surgical<br>Mask is intended to be worn<br>to protect both the patient<br>and healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential<br>exposure of the wearer to<br>blood and body fluids. The<br>Fluidshield* 2 Surgical<br>Mask is a single use, | The Kimberly-Clark,<br>KC200 and KC300 Face<br>Mask(s) are intended to.be<br>worn to protect both the<br>patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices<br>to reduce the potential<br>exposure of the wearer to<br>blood and body fluids. The<br>Kimberly-Clark, KC200 | Same<br>This<br>submission<br>covers the KC<br>300 and KC<br>200 Blue<br>Surgical or<br>Procedure<br>Masks. |
| | device, provided non-sterile.<br>The Fluidshield* 2<br>Procedure Mask is intended<br>to be worn to protect both<br>the patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential<br>exposure of the wearer to<br>blood and body fluids. The<br>Fluidshield* 2 Procedure<br>Mask is a single use,<br>disposable device, provided<br>non-sterile. | and KC300 face mask(s) is<br>a single use, disposable<br>device(s), provided non-<br>sterile. | |
| Materials of Construction for Surgical Mask Family (Codes 62113 and 62114) | | | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of the letters "OM" in a stylized font, with the "O" made up of horizontal lines. To the right of the letters is the text "Owens & Minor" in a serif font. The "OM" is in a dark red color, while the "Owens & Minor" text is in gray.
| Structure | Material of Layer for<br>K232812 | Material cleared under<br>K111402 | Same or Different |
|--------------------------------------|------------------------------------------|----------------------------------------------------|-------------------|
| Code 62113 Materials of Construction | | | |
| Outer Layer | Polyester Cellulose | Spunbond polypropylene<br>(or) Polyester Cellulose | Same |
| 2nd Layer | Spunbond polypropylene | Spunbond polypropylene | Same |
| Media Filter | Meltblown polypropylene | Meltblown polypropylene | Same |
| Inner Layer | Polyester Cellulose | Polyester Cellulose | Same |
| Top Binding | Spunbond polypropylene | Spunbond polypropylene,<br>Polyester spunlace | Same |
| Lower Binding | Spunbond polypropylene | Spunbond polypropylene,<br>Polyester spunlace | Same |
| Ties (Surgical) | Spunbond polypropylene | Polyester spunlace | Different |
| Foam | Polyester/Polyester<br>Spunlace laminate | Polyester/Polyester<br>Spunlace laminate | Same |
| Foam | Polyester | Polyester | Same |
| Nose Wire | Aluminum | Aluminum | Same |
| Code 62114 Materials of Construction | | | |
| Outer Layer | Polyester Cellulose | Spunbond polypropylene<br>(or) Polyester Cellulose | Same |
| 2nd Layer | Spunbond polypropylene | Spunbond polypropylene | Same |
| Media Filter | Meltblown polypropylene | Meltblown polypropylene | Same |
| Inner Layer | Polyester Cellulose | Polyester Cellulose | Same |
| Top Binding | Spunbond polypropylene | Spunbond polypropylene<br>(or) Polyester spunlace | Same |
| Lower Binding | Polyester spunlace | Spunbond polypropylene<br>(or) Polyester spunlace | Same |
| Ties (Surgical) | Polyester spunlace | Polyester spunlace | Same |
| Foam | Polyester | Polyester/Polyester | Same |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of the letters "OM" in a stylized font, with the word "Owens" above the word "Minor" to the right of the letters. The letters "OM" are in a dark red color, while the words "Owens" and "Minor" are in a gray color.
| Visor | Polyester Film | Spunlace laminate | | | Same |
|--------------------------------------------------------------------------|-----------------------------------------------|----------------------------------------------------|--|--|-------------------|
| Nose Wire | Aluminum | Aluminum | | | Same |
| Materials of Construction for Procedure Mask Family (Codes 62115, 62116) | | | | | |
| Structure | Material of Layer for<br>K232812 | Material cleared under<br>K111402 | | | Same or Different |
| | Code 62115 Materials of Construction | | | | |
| Outer Layer | Polyester Cellulose | Spunbond polypropylene<br>(or) Polyester Cellulose | | | Same |
| 2nd Layer | Spunbond polypropylene | Spunbond polypropylene | | | Same |
| Media Filter | Meltblown polypropylene | Meltblown polypropylene | | | Same |
| Inner Layer | Polyester Cellulose | Polyester Cellulose | | | Same |
| Top Binding | Polyester spunlace | Polyester spunlace | | | Same |
| Lower Binding | Polyester spunlace | Polyester spunlace | | | Same |
| Earloops (Procedure) | Polyester lycra | Polyester lycra | | | Same |
| Nose Wire | Aluminum | Aluminum | | | Same |
| | Code 62116 Materials of Construction | | | | |
| Outer Layer | Polyester Cellulose | Spunbond polypropylene<br>(or) Polyester Cellulose | | | Same |
| 2nd Layer | Spunbond polypropylene | Spunbond polypropylene | | | Same |
| Media Filter | Meltblown polypropylene | Meltblown polypropylene | | | Same |
| Inner Layer | Polyester Cellulose | Polyester Cellulose | | | Same |
| Top Binding | Polypropylene spunbond | Polyester spunlace | | | Different |
| Lower Binding | Polyester spunlace | Polyester spunlace | | | Same |
| Earloops (Procedure) | Polyester lycra | Polyester lycra | | | Same |
| Foam | Polyester foam/polyester<br>spunlace laminate | Polyester foam/polyester<br>spunlace laminate | | | Same |
| Visor | Polyester Film | Polyester Film | | | Same |
| Nose Wire | Aluminum | Aluminum | | | Same |
| | Design Attributes for Surgical Masks | | | | |
| # of Layers | 4-layer | 4-layer | | | Same |
| Color | Blue with Orange Print | Blue with Orange Print | | | Same |
| Ties (Surgical) | Yes | Yes | | | Same |
| Visor | Some designs contain a<br>visor | Some designs contain a<br>visor | | | Same |
| Nose Wire | Enclosed | Enclosed | | | Same |
| | Design Attributes for Procedure Masks | | | | |
| # of Layers | 4-layer | 4-layer | | | Same |
| Color | Blue | Blue | | | Same |
| Earloops (Procedure) | Yes | Yes | | | Same |
| Visor | Some designs contain a<br>visor | Some designs contain a<br>visor | | | Same |
| Performance Data/Product Claims | | | | | |
| Fluid Resistance Level | 2 | 2 and 3 | | | Same |
| EN 14683 | Yes | No | | | Different |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Owens & Minor logo above a table. The table has two rows and two columns. The first row indicates "Single Use Device" and "Yes", while the second row indicates "Sterility" and "Non-Sterile".
| Single Use Device | Yes | Yes | Same |
|-------------------|-------------|-------------|------|
| Sterility | Non-Sterile | Non-Sterile | Same |
## Non-Clinical Performance Testing Standards
Non-clinical tests were conducted to verify that the proposed device(s) met all design specifications as was same/similar to the predicate device(s). The test results demonstrated that the proposed device complies with the following standards:
| Purpose | Test | Acceptance Criteria | Result |
|-----------------------------------|------------------|---------------------|--------|
| Level 2 - Face Mask Performance | ASTM F2100 | ASTM Level 2 | Pass |
| Bacterial Filtration Efficiency | ASTM F2101 | >98% | Pass |
| Particulate Filtration Efficiency | ASTM F2299 | >98% | Pass |
| Differential Pressure | EN 14683 | <6.0 mmH2O/cm2 | Pass |
| Fluid Resistance | ASTM F1862 | 120 mmHg | Pass |
| Flammability | 16 CFR Part 1610 | Class 1 | Pass |
| Biocompatibility | ISO 10993 | | |
| Cytotoxicity | ISO 10993-5 | Non-cytotoxic | Pass |
| Sensitization | ISO 10993-10 | Non-sensitizing | Pass |
| Irritation | ISO 10993-10 | Non-irritant | Pass |
## Summary of Clinical Performance Testing:
Not applicable
## Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the Fluidshield* 2 Surgical Mask(s), and Fluidshield* 2 Procedure Mask(s) are as safe, as effective and perform as well as or better than the legally marketed KC200 and KC300 Face Masks (K111402).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.