Femoral Block – Safety and Accuracy validation: 183 scans
—
Indications for Use
ScanNav Anatomy Peripheral Nerve Block is indicated to assist qualified healthcare professionals to identify and label the below mentioned anatomy in live ultrasound images in preparation for ultrasound guided regional anesthesia prior to needle insertion for patients 18 years of age or older. The highlighting of structures in the following anatomical regions is supported: - Axillary level brachial plexus - Erector spinae plane - Interscalene level brachial plexus - Popliteal level sciatic nerve - Rectus sheath plane - Sub-sartorial femoral triangle / Adductor canal - Superior trunk of brachial plexus - Supraclavicular level brachial plexus - Longitudinal suprainguinal fascia iliaca plane - Femoral block ScanNav Anatomy Peripheral Nerve Block is an accessory to compatible general-purpose diagnostic ultrasound systems.
Device Story
ScanNav Anatomy Peripheral Nerve Block is a software-as-a-medical-device (SaMD) accessory for general-purpose ultrasound systems. It receives real-time ultrasound video input; processes images using deep learning AI to identify and highlight anatomical structures relevant to regional anesthesia; and displays the highlighted structures on a dedicated panel PC. The primary ultrasound machine display remains unaffected. Used in professional healthcare environments by qualified clinicians, the device assists in anatomical identification prior to needle insertion. By providing real-time visual guidance, it aims to improve the accuracy and safety of nerve block procedures for the patient.
Clinical Evidence
Bench testing only. Validation of the femoral nerve block feature was performed using 183 ultrasound scans. Results showed a 1.1% false positive rate (mis-identification of safety-critical structures), 96.7% accuracy (correct highlighting), and 2.2% false negative rate (non-identification). All software verification tests (unit, integration, fault injection, endurance, system-level) passed without anomalies.
Technological Characteristics
SaMD utilizing machine learning-based AI and image processing algorithms. Hardware includes a panel PC (model IUP10W33M), AC/DC power cord, and 2m shielded HDMI cable; optional external battery. Operates as an accessory to general-purpose diagnostic ultrasound systems. Software-based highlighting displayed on a separate panel PC.
Indications for Use
Indicated for qualified healthcare professionals to identify and label anatomy in live ultrasound images for ultrasound-guided regional anesthesia in patients 18 years or older. Supported regions: Axillary/Interscalene/Supraclavicular brachial plexus, Erector spinae plane, Popliteal sciatic nerve, Rectus sheath plane, Sub-sartorial femoral triangle/Adductor canal, Longitudinal suprainguinal fascia iliaca plane, and Femoral block.
Regulatory Classification
Identification
ScanNav Anatomy Peripheral Nerve Block is a software medical device that assists qualified healthcare professionals in the identification and labeling of anatomical structures in live ultrasound images in preparation for ultrasound-guided regional anesthesia (UGRA) prior to needle insertion for patients 18 years of age or older. It provides real-time interpretation and enhanced visualization by highlighting anatomical landmarks. The device is an accessory to compatible general-purpose diagnostic ultrasound systems and supports the following anatomical regions: Axillary level brachial plexus, Erector spinae plane, Interscalene level brachial plexus, Popliteal level sciatic nerve, Rectus sheath plane, Sub-sartorial femoral triangle/Adductor canal, Superior trunk of brachial plexus, Supraclavicular level brachial plexus, and Longitudinal suprainguinal fascia iliaca plane.
Special Controls
In combination with the general controls of the FD&C Act, the real-time ultrasound anatomy visualization and labeling device for ultrasound guided regional anesthesia is subject to the following special controls:
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
October 6, 2023
Intelligent Ultrasound Limited Bhaskar Chikkanna Head of Regulatory Affairs Floor 6a, Hodge House, 114-116 St Mary Street Cardiff. CF101DY United Kingdom
Re: K232787
Trade/Device Name: ScanNav Anatomy Peripheral Nerve Block Regulation Number: 21 CFR 868.1980 Regulation Name: Real-Time Ultrasound Anatomy Visualization And Labeling Device For Ultrasound Guided Regional Anesthesia Regulatory Class: Class II Product Code: QRG Dated: September 11, 2023 Received: September 11, 2023
Dear Bhaskar Chikkanna:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{3}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232787
Device Name
ScanNav Anatomy Peripheral Nerve Block
#### Indications for Use (Describe)
ScanNav Anatomy Peripheral Nerve Block is indicated to assist qualified healthcare professionals to identify and label the below mentioned anatomy in live ultrasound mages in preparation for ultrasound guided regional anesthesia prior to needle insertion for patients 18 years of age or older.
The highlighting of structures in the following anatomical regions is supported:
- · Axillary level brachial plexus
- · Erector spinae plane
- · Interscalene level brachial plexus
- · Popliteal level sciatic nerve
- · Rectus sheath plane
- · Sub-sartorial femoral triangle / Adductor canal
- · Superior trunk of brachial plexus
- · Supraclavicular level brachial plexus
- · Longitudinal suprainguinal fascia iliaca plane
- Femoral block
ScanNav Anatomy Peripheral Nerve Block is an accessory to compatible general-purpose diagnostic ultrasound systems.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Intelligent Ultrasound. The logo consists of a blue sphere made up of interconnected dots on the left, and the words "INTELLIGENT ULTRASOUND" in gray on the right. Below the words is the phrase "for smarter scanning" in blue.
# 510(k) Summary K232787
September 27, 2023
# Applicant details:
Applicant Name: Intelligent Ultrasound Limited Applicant Address: Floor 6a, Hodge House, 114-116 St Mary Street, Cardiff, UK Applicant Contact Name: Bhaskar Chikkanna Applicant Contact Number: +442920756534 Applicant Contact Email: bhaskar.chikkanna@intelligentultrasound.com
# Device details:
Device Trade Name: ScanNav Anatomy Peripheral Nerve Block Common Name: Software as a Medical Device Classification: Class II Regulation Number: 21 CFR 868.1980 Product Code: QRG
# Predicate Device details:
Device Trade Name: ScanNav Anatomy Peripheral Nerve Block Predicate #: DEN220024 Product code: QRG
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Intelligent Ultrasound. The logo consists of a blue sphere made up of interconnected circles on the left, and the words "INTELLIGENT ULTRASOUND" in gray on the right. Below the words is the phrase "for smarter scanning" in blue. The logo is clean and modern, and the colors are calming and professional.
#### Device description summary:
ScanNav Anatomy Peripheral Nerve Block is a software (SaMD) which assists qualified healthcare professionals to identify and label relevant anatomical structures in preparation for ultrasound guided regional anesthesia prior to needle insertion for patients 18 years of age or older.
The device receives ultrasound images in real-time from a compatible general-purpose ultrasound machine. It processes these images using deep learning artificial intelligence algorithms and highlights relevant anatomical structures. The ultrasound machine display remains unaffected, and the highlighting is only displayed on a general-purpose panel PC provided with the device.
#### Intended use/Indications for use:
ScanNav Anatomy Peripheral Nerve Block is indicated to assist qualified healthcare professionals to identify and label the below mentioned anatomy in live ultrasound images in preparation for utrasound guided regional anesthesia prior to patients 18 years of age or older.
The highlighting of structures in the following anatomical regions is supported:
- Axillary level brachial plexus
- Erector spinae plane
- Interscalene level brachial plexus
- · Popliteal level sciatic nerve
- Rectus sheath plane
- Sub-sartorial femoral triangle / Adductor canal
- Superior trunk of brachial plexus
- Supraclavicular level brachial plexus
- · Longitudinal suprainguinal fascia iliaca plane
- Femoral block
ScanNav Anatomy Peripheral Nerve Block is an accessory to compatible general-purpose diagnostic ultrasound systems.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Intelligent Ultrasound. The logo consists of a blue sphere made up of interconnected circles on the left. To the right of the sphere are the words "INTELLIGENT ULTRASOUND" in gray, with the word "INTELLIGENT" stacked on top of the word "ULTRASOUND". Below the words "INTELLIGENT ULTRASOUND" is the phrase "for smarter scanning" in a light blue script font.
# Substantial Equivalence
| Attributes | Predicate (DEN220024) | Modified device | Comments |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Intended use | ScanNav Anatomy Peripheral Nerve Block is<br>intended to assist in the identification and<br>labelling of anatomical structures in live<br>ultrasound images. | ScanNav Anatomy Peripheral Nerve Block is<br>intended to assist in the identification and<br>labelling of anatomical structures in live<br>ultrasound images. | No change |
| Indications for use | ScanNav Anatomy Peripheral Nerve Block is<br>indicated to assist qualified healthcare<br>professionals to identify and label the below<br>mentioned anatomy in live ultrasound images in<br>preparation for ultrasound guided regional<br>anesthesia prior to needle insertion for patients<br>18 years of age or older.<br>The highlighting of<br>structures in the following anatomical regions is<br>supported:<br>• Axillary level brachial plexus<br>• Erector spinae plane<br>• Interscalene level brachial plexus<br>• Popliteal level sciatic nerve<br>• Rectus sheath plane<br>• Sub-sartorial femoral triangle / Adductor canal<br>• Superior trunk of brachial plexus | ScanNav Anatomy Peripheral Nerve Block is<br>indicated to assist qualified healthcare<br>professionals to identify and label the below<br>mentioned anatomy in live ultrasound images in<br>preparation for ultrasound guided regional<br>anesthesia prior to needle insertion for patients<br>18 years of age or older.<br>The highlighting of<br>structures in the following anatomical regions is<br>supported:<br>• Axillary level brachial plexus<br>• Erector spinae plane<br>• Interscalene level brachial plexus<br>• Popliteal level sciatic nerve<br>• Rectus sheath plane<br>• Sub-sartorial femoral triangle / Adductor canal<br>• Superior trunk of brachial plexus | Addition of Femoral block<br>to the list of supported<br>ultrasound guided<br>regional anesthesia<br>procedures |
| | | | |
| | • Supraclavicular level brachial plexus | • Supraclavicular level brachial plexus | |
| | • Longitudinal suprainguinal fascia iliaca plane | • Longitudinal suprainguinal fascia iliaca plane | |
| | | • Femoral | |
| | ScanNav Anatomy Peripheral Nerve Block is an<br>accessory to compatible general-purpose<br>diagnostic ultrasound systems. | ScanNav Anatomy Peripheral Nerve Block is an<br>accessory to compatible general-purpose<br>diagnostic ultrasound systems. | |
| Intended users | Qualified healthcare professionals | Qualified healthcare professionals | No change |
| Intended use<br>environment | Professional healthcare environment | Professional healthcare environment | No change |
| Technology/Operating<br>principle | Machine learning based artificial intelligence<br>in combination with image processing<br>algorithms | Machine learning based artificial intelligence<br>in combination with image processing<br>algorithms | No change |
| Device hardware (panel<br>PC and cables) | Panel PC model IUP10W33M<br>Ac/Dc power cord<br>2m shielded HDMI cable | Panel PC model IUP10W33M<br>Ac/Dc power cord<br>2m shielded HDMI cable | No change |
| | Optional: external battery unit with power<br>supply | Optional: external battery unit with power<br>supply | |
| Labelling (incl.<br>instructions for use) | Instructions for use v1.0<br><br>Device labelling | Instructions for use v2.0<br><br>Device Labelling | Labelling updated to<br>reflect the changes to the<br>indications for use with<br>new nerve block. |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Intelligent Ultrasound. The logo features a blue sphere made up of interconnected circles on the left. To the right of the sphere is the text "INTELLIGENT ULTRASOUND" in gray, with the word "INTELLIGENT" stacked on top of "ULTRASOUND". Below the text is the phrase "for smarter scanning" in a light blue, script-like font.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Intelligent Ultrasound. The logo consists of a blue sphere made up of interconnected dots on the left side. To the right of the sphere, the words "INTELLIGENT ULTRASOUND" are written in gray, with the word "INTELLIGENT" above the word "ULTRASOUND". Below the words, the phrase "for smarter scanning" is written in blue.
# Summary of Clinical/Non-Clinical tests:
The device change subject to this 510(k) has undersy and safety validation tests. The non-clinical test results show that the device has met the predefined acceptance criteria and has passed all tests without any unresolved anomalies. The actray validation tests carried out using the established test protocols hat the addition of femoral neve block has met the established acceptance criteria and the device continues to provide reasonable of safety and effectiveness. No new risks have been identified with this device change. The results are summarized in the table below.
| Verification/Validation<br>Methods | Acceptance Criteria | Summary of<br>results |
|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Femoral Block – Safety and Accuracy validation.<br><br>Established protocol and acceptance criteria same as that<br>used for the predicate device. | FP rate: Mis-identification rate of safety critical anatomical<br>structures in the indicated procedures is less than 5%. | 1.1% (2 out of<br>183 scans) |
| | Accuracy (TP+TN) rate: Correct highlighting of safety critical<br>anatomical structures in the indicated procedures at least<br>80% of the time. | 96.7% (177 out<br>of 183 scans) |
| | FN rate: non-identification rate of safety critical anatomical<br>structures in the indicated procedures is less than 15%. | 2.2% (4 out of<br>183 scans) |
| Software tests<br><br>Established test cases and protocol same as that used for<br>the predicate device. | Unit test<br>Integration test<br>Fault Injection test<br>Endurance test<br>Use case test<br>Software system level test<br>Requirements traceability test | All software tests have<br>been successfully<br>completed without any<br>anomalies |
#### Conclusion:
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Intelligent Ultrasound. The logo consists of a blue sphere made up of interconnected circles on the left, and the words "INTELLIGENT ULTRASOUND" in gray on the right. Below the words is the phrase "for smarter scanning" in blue.
The subject device and the predicate device have intended use and the technological different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.