The QUEX ED and QUEX S device is indicated for use as an Electrophysiological System. The Electrophysiological System is made up of the following Items which are functions of the QUEX ED and S device: 1. Two channel EEG: The QUEX ED and S Device is intended to record and store EEG signals, and to present the EEG signals in visual formats in real time. The visual signals assist trained medical staff to make neurological diagnoses. The EEG device does not provide any diagnoses conclusion about the subject's condition and does not provide any automated alerts of an adverse clinical event. The QUEX ED AND S Device is intended to be used in a professional healthcare facility environment. 2. GSR [galvanic skin response measuring] The QUEX ED and S device is intended for use by trained healthcare professionals, to detect and monitor electrical signals produced by the body skin conductance by using electrodes placed on the skin.
Device Story
System records two-channel EEG and GSR signals; inputs acquired via head harness (EEG) and limb electrodes (GSR). Hardware performs signal gain and A/D conversion; data transmitted via USB to PC/laptop running QUEX Monitor software. Software displays raw waveforms in real-time for clinician review. Used in professional healthcare environments by trained therapists/physicians. Output assists clinical neurological diagnosis; device provides no automated diagnostic conclusions or clinical alerts. Benefits include real-time physiological monitoring to support professional decision-making.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1, IEC 80601-2-26), electromagnetic compatibility (IEC 60601-1-2), EEG essential performance, and GSR resistance measurement accuracy (1k-1MΩ range).
Technological Characteristics
Two-channel EEG (10-100 µV) and GSR (1k-1MΩ) measurement. Hardware connects via USB (5V DC). EEG features: 8-bit ADC, 0-100 Hz LP filter, 50/60 Hz notch filter, >10 MΩ input impedance. Uses Ag/AgCl electrodes. Software (QUEX Monitor) runs on Windows PCs. Complies with IEC 60601-1, IEC 80601-2-26, and IEC 62368.
Indications for Use
Indicated for use by trained healthcare professionals in professional healthcare facilities to record/store/display EEG signals and monitor galvanic skin response (GSR) via skin electrodes. Does not provide diagnostic conclusions or automated alerts.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 11, 2024
OX World Ltd Lilla Strobel QA Advisor Tinodi utca 1-3 A.ep, IV.em, 93 Budapest, Hungary 1095 Hungary
Re: K232779
Trade/Device Name: Quex Ed; Quex S Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMC, GZO Dated: September 11, 2024 Received: September 11, 2024
Dear Lilla Strobel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Image /page/2/Picture/3 description: The image shows the name "Patrick Antkowiak -S" in a simple, sans-serif font. The name is likely associated with a person, possibly an employee or individual affiliated with an organization. To the left of the name is a faded FDA logo.
for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232779
Device Name QUEX ED and S
#### Indications for Use (Describe)
The QUEX ED and QUEX S device is indicated for use as an Electrophysiological System. The Electrophysiological System is made up of the following Items which are functions of the QUEX ED and S device:
1. Two channel EEG:
The QUEX ED and S Device is intended to record and store EEG signals, and to present the EEG signals in visual formats in real time. The visual signals assist trained medical staff to make neurological diagnoses. The EEG device does not provide any diagnoses conclusion about the subject's condition and does not provide any automated alerts of an adverse clinical event. The QUEX ED AND S Device is intended to be used in a professional healthcare facility environment. 2. GSR [galvanic skin response measuring]
The QUEX ED and S device is intended for use by trained healthcare professionals, to detect and monitor electrical signals produced by the body skin conductance by using electrodes placed on the skin.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### Submitter information:
#### Company:
QX WORLD Ltd. Tinódi street 1-3. A building 4. floor 93. door Budapest H-1095, Hungary
#### Contact name:
CEO: Ms. Andreea Taflan OX WORLD LTD phone:+1 (989) 681-1063; mobile:+36 70 618 8508 Tinodi street 1-3, A building 4th floor / 93, Budapest, HUNGARY-1095 Email: andreea@gxworldltd.com
QA Advisor: Ms. Lilla Strobel Phone: +36-30-9251906 Email: strobel.lilla@gmail.com
### Device:
Name of Device: 1) QUEX ED Electrophysiological Device
2) QUEX S Electrophysiological Device
Common or Usual Name: Electrophysiological Device
Classification Name: reduced- montage standard electroencephalograph,
electroencephalograph (21 CFR 882.1400)
Regulatory Class: II Product Code: OMC, GZO
## Predicate device
Manufacturer: Neurosteer Inc. Neurosteer EEG Recorder (K221563)
## Device Description
Device identification:
QUEX ED or QUEX S devices are contains the following components:
- QUEX ED or QUEX S hardware.
- I QUEX Monitor software
- . USB - USB cable for data communication and power supply.
- EEG electrodes (head electrodes)
- Electrodes for GSR measurements (limbs)
- I Cable harnesses for head and limbs.
The QUEX Monitor software is exclusively used for QUEX ED and QUEX S as part of the system.
The EEG electrodes and the electrodes used for GSR measurements are connected to the QUEX ED or S devices. The device makes the necessary gain on the analog signals and digitalize the signals. The digitalized information is acquired and displayed by the QUEX Monitor software. The QUEX monitor software can run on PCs, notebooks with Windows OS.
{5}------------------------------------------------
Electroencephalography (EEG), measures brain wave activity. The system uses the head harness and electrodes to acquire the signals. The QUEX Devices are measures brain activity on two channels. The EEG wave recording electrodes are connected to the forehead with dry electrodes.
The galvanic skin response (GSR, which falls under the umbrella term of electrodermal activity, or EDA) refers to changes in sweat gland activity.
The GSR recording is done via limb electrodes attached to the wrists and ankles.
### Device characteristics:
### EEG
- Two channel EEG -
- -Measurement on the scalp
- -10 µV to 100 µV signal measurement
#### GSR (Galvanic Skin Response)
- skin conductance change measurement effected by changes by sweat gland activity, -
- measurement with DC current for skin resistance,
- measuring range: 1 Kohm 1 Mohm, -
### Environment:
The device to be used in professional healthcare environment.
### Indications for use
The QUEX ED and S device is indicated for use as an Electrophysiological System. The Electrophysiological System is made up of the following Items which are functions of the OUEX ED and S device:
### 1. Two channel EEG:
The QUEX ED and S Device is intended to record and store EEG signals, and to present the EEG signals in visual formats in real time. The visual signals assist trained medical staff to make neurological diagnoses. The EEG device does not any diagnoses conclusion about the subject's condition and does not provide any automated alerts of an adverse clinical event. The OUEX ED and S Device is intended to be used in a professional healthcare facility environment.
### 2. GSR [galvanic skin response measuring]
The QUEX ED and S device is intended for use by trained healthcare professionals, to detect and monitor electrical signals produced by the body skin conductance by using electrodes placed on the skin.
Intended user:
Professional use (Therapist)
The device itself does not generate any stimulation to the patient. External visual, audible, or other stimulation shall be used based on the trained operator (physician) decision. The QX Word does not take the responsibility of the possible adverse effect of an external stimulation.
{6}------------------------------------------------
| Technical specification of QUEX ED and QUEX S Electrophysiological Device | | | | |
|---------------------------------------------------------------------------|--|--|--|--|
|---------------------------------------------------------------------------|--|--|--|--|
| Electrical characteristics | |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Voltage [power supply] | 5V DC (Port USB) |
| Maximum consumption | 1200 mA |
| Time to change the signal | Max. 300 ms |
| Time to change the channel | Max. 100 ms |
| Rated output current | Max. 1 mA |
| EEG Characteristics | |
| Digital sampling rate per channel | Programable |
| Calibration signals | programable, 0-200 $ \mu V $ |
| Number of channels | 2 |
| EEG channel characteristics | 1. No. of channels: 2<br>2. A/D Conversion: 8 bit ADC<br>3. Max Sampling Rate: 1ms<br>4. Sensitivity: User definable 10 $ \mu V $ /mm, 20 $ \mu V $ /mm, 100 $ \mu V $ /mm<br>5. LP filter: 0 - 100 Hz (8 degree digital filter)<br>6. Notch Filter: 50/60 Hz<br>7. Input impedance: > 10 M ohm<br>8. CMRR: > 100 $ \mu V $<br>9. Noise level : < 5 $ \mu V $ RMS |
| accuracy of signal reproduction; | difference less than $ \pm $ 20 %<br>acc to IEC 80601-2-26, 201.12.1.102 |
| input dynamic range and maximum offset voltage | less than $ \pm $ 10 % acc to IEC 80601-2-26, 201.12.1.103 |
| frequency range and bandwidth | 0,5 Hz - 50 Hz, relative output within<br>71% to 110% of the output at 5Hz |
| a description of waveform displays | EEG signals are displayed |
| Skin resitance | |
| Galvanic Skin Response data acquisition | Measurement Range: 1k - 1MΩ +/- 10% |
| Output Voltage | 0 - 4 V DC |
# Comparison with predicate device
The QUEX ED and QUEX S EEG and GSR recorder and the predicate devices have the same intended use and very similar technological features, including software intended for recording and viewing EEG signals. The differences between the Neurosteer EEG Recorder and QUEX S devices raise no new issues of safety or effectiveness. Performance testing confirms that the device functions as intended, supporting substantial equivalence. The characteristics of the subject and predicate device are summarized in the following table:
{7}------------------------------------------------
| Attribute | Predicate Device:<br>Neurosteer EEG<br>Recorder (K221563) | QUEX ED and QUEX S<br>EEG and GSR devcice | Comparison |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Device class and<br>regulation | Class II per 882.1400 | Class II per 882.1400 | Same |
| Product codes | OMC | OMC, GZO | Same. |
| Indications for use | The Neurosteer EEG<br>Recorder is intended to<br>record and store EEG signals,<br>and to present the EEG<br>signals in visual formats in<br>real time. The visual signals<br>assist trained medical staff to<br>make neurological<br>diagnoses. The EEG Recorder<br>does not provide any<br>diagnostic conclusion about<br>the subject's condition and<br>does not provide any<br>automated alerts of an<br>adverse clinical event. The<br>EEG Recorder is intended to<br>be used in a professional<br>healthcare facility<br>environment. | 1. Two channel EEG:<br>The QUEX ED Device is<br>intended to record and store<br>EEG signals, and to present the<br>EEG signals in visual formats in<br>real time. The visual signals<br>assist trained medical staff to<br>make neurological diagnoses.<br>The EEG device does not any<br>diagnoses conclusion about the<br>subject's condition and does<br>not provide any automated<br>alerts of an adverse clinical<br>event. The QUEX ED Device is<br>intended to be used in a<br>professional healthcare facility<br>environment.<br><br>2. GSR [galvanic skin response<br>measuring resistance]<br><br>The QUEX ED device is<br>intended for use by trained<br>healthcare professionals, to<br>use of sensors or electrodes<br>placed on the skin to detect and<br>monitor electrical signals<br>produced by the body skin<br>conductance. | Substantially<br>equivalent |
| Classification | Review Panel Neurology<br>Product Code OMC<br>Regulation Number 882.1400<br>Device Class 2<br>Submission Type 510(k) | | |
| | | Device Galvanic Skin<br>Response Measurement<br>Review Panel Neurology<br>Product Code GZO<br>Regulation Number 882.1540<br>Device Class 2<br>Submission Type 510(K)<br>Exempt | |
| System<br>components | EEG electrodes with<br>conductive hydrogel<br>EEG sensor<br>Portable EEG monitor to<br>display EEG data in real time<br>in visual format and transfer<br>EEG recording files to cloud<br>storage platform via<br>internet connection Cloud<br>storage platform<br>Browser-based EEG viewing<br>software | EEG electrodes with cables and<br>securing strips.<br>Main box for acquiring and<br>digitalizing the analog<br>physiological signals as EEG<br>and skin resistance for GSR.<br>Application-based EEG<br>viewing software running on<br>desktop and laptops | Substantially<br>equivalent |
| Measurement of<br>physiological<br>data | EEG recording only | EEG and GSR recording | Substantially<br>equivalent<br>GSR function is<br>510(k) exempt |
| Channels | 1 | 2 | Substantially<br>equivalent |
| EEG presentation<br>format(s) | Visual only (raw EEG<br>waveform, spectrogram,<br>metrics and bar graph views) | Visual only (raw EEF<br>waveform) | Substantially<br>equivalent |
| Data transfer<br>method(s) | Bluetooth 2.4 GHz and<br>internet connection (ethernet<br>or Wi-Fi) | USB cable, | No additional risk<br>to patient, more<br>reliable<br>connection. |
| Data file format | Edf | Reusable | Same |
| Type of use | Reusable | Reusable | Same |
| Power source | Lithium polymer battery –<br>rechargeable with 100-240 V<br>AC power adapter (device<br>does not work when<br>connected to AC to recharge) | USB connected to a computer,<br>5V DC, to be used IEC 62368<br>compliance computer. | Same level of<br>electrical safety.<br>No risk from the<br>lithium battery in<br>the device |
| Charging | Micro-USB charging cable; if<br>connected to a computer, all<br>EEG acquisition functions are<br>automatically disabled | No charging is necessary, to be<br>used charged laptops or<br>computer connected to mains<br>supply. | Substantially<br>equivalent,<br>no risk from the<br>battery<br>exhausting. |
| Electrode type | Passive silver/silver-chloride<br>(Ag/AgCl) | | Same |
| Number / locations of<br>electrodes | 3 (Locations: Fp1, Fpz, Fp2) | 5 (Fp1:F7; Fp2:F8; Fpz) | Substantially<br>equivalent |
| Conductive<br>electrolyte gel | Semi-solid hydrogel is<br>included in a layer integrated<br>into each electrode assembly. | | Substantially<br>equivalent |
| Patient contact | Patient forehead (intact skin) | Patient forehead (intact skin) | Same |
| Securing method | Integrated adhesive layer | EEG strip | Substantially<br>equivalent |
| Available sizes and<br>dimensions | One size (10.5 cm) | One size | Substantially<br>equivalent |
| Type of use | Single use, non-sterile,<br>disposable | Reusable electrodes, single use<br>electrodes can be used | Substantially<br>equivalent |
| Connector | Integrated single-cable<br>connector to connect to an<br>EEG recording device | The electrodes connected via<br>connector fit for device inlets. | Same |
| Compatibility | Compatible with the<br>Neurosteer EEG Recorder<br>only | The device can be used with<br>own software only (QUEX<br>Monitor) | Substantially<br>equivalent |
{8}------------------------------------------------
{9}------------------------------------------------
# Non-clinical tests performed:
| Test | Test Method Summary | Results |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Bench tests:<br>Electrical Safety | The QUEX ED and QUEX S devices are tested<br>according to IEC 60601-1 ed 3.1 and EN/IEC<br>60601-2-26 standards. Retesting was<br>performed acc.to the most current EEG<br>Standard: IEC 80601-2-26. the tests were<br>performed by national accredited testing<br>laboratories. | All samples passed the acceptance<br>criteria. The subject device is as safe as<br>the predicates with respect to<br>electrical safety. |
| Bench tests:<br>Electromagnetic<br>Compatibility<br>(EMC) | IEC 60601-1-2:2014+A1:2020 (EN 60601-1-<br>2:2015+A1:2021): Medical electrical<br>equipment - Part 1-2: General requirements<br>for basic safety and essential performance -<br>Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests<br>IEC TS 60601-4-2:2024: Medical electrical<br>equipment - Part 4-2: Guidance and<br>interpretation - Electromagnetic immunity: | All samples passed the acceptance<br>criteria for each test. The subject<br>device is as safe as the predicates with<br>respect to EMC. |
{10}------------------------------------------------
| | performance of medical electrical equipment<br>and medical electrical systems | |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| | The device to be used in professional<br>healthcare environment. | |
| Bench tests:<br>EEG Essential<br>Performance | The essential performance of the QUED ED<br>and S devices is accessed according to the<br>EN/IEC 60601-2-26 and EN/IEC 80601-2-<br>26. | All samples passed the acceptance<br>criteria. The subject device is as<br>effective as the predicates with<br>respect to EEG performance. |
| Bench tests:<br>GSR measurments | The device is tested in accordance a test<br>plan for evaluation of the resistance<br>measurement within the specified limits<br>1Kohm-1Mohm. | The test sample is passed the tests. |
| Biocompatibility | 510(k) approved electrodes and cables are<br>used. | |
| EEG electrode<br>performance | 510(k) approved electrodes and cables are<br>used. | |
### Conclusions
The QUEX ED and QUEX S has the same intended use and similar indications for use, technological characteristics, and principles of operation as the predicate devices. The minor technological differences between the QUEX ED and QUEX S and its predicate devices raise no new or different questions of safety or effectiveness. Performance data demonstrate that the QUEX ED and QUEX S r is as safe and effective as the predicate devices. Thus, the QUEX ED and QUEX S are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.