Patient-reported outcomes (PFDI-20 scores); Routine clinical care pessary usage (ring or Gellhorn)
The study compared the Reia pessary against the patients' existing, routine-care pessaries (ring or Gellhorn) to evaluate safety and effectiveness, specifically using PFDI-20 scores to assess symptom bother and quality of life.
Pelvic Organ Prolapse; Pessary; PFDI-20; Comparative effectiveness
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Prospective open label single-arm study; Prospective open-label single-arm study with patients serving as their own control; Follow-up/Duration: 1 month baseline (current pessary) followed by 3 months (Reia pessary)
Patients with pelvic organ prolapse currently using ring or Gellhorn pessaries; Sample Size: 62 subjects treated; Number of Sites: 7
Current routine-care pessary (ring or Gellhorn)
Difference in Pelvic Floor Distress Inventory - 20 (PFDI-20) scores; incidence of serious and related adverse events
Indications for Use
The Reia pessary is indicated for the temporary, non-surgical management of Pelvic Organ Prolapse.
Device Story
The Reia pessary is a single-piece, medical-grade silicone device designed for the non-surgical management of pelvic organ prolapse. It functions by supporting the vaginal wall to prevent descending organs from dropping into the vaginal canal. The device features a collapsible, domed geometry with a stem and loop at the bottom, allowing users to pinch the device to elongate it for easier insertion and removal. It is intended for prescription use, with an initial size fitting performed by a healthcare professional in a clinical setting, followed by patient self-management (insertion/removal) in the home environment. The device is reusable and cleaned by the user with soap and water. By providing mechanical support, the pessary aims to alleviate pelvic floor symptoms and improve quality of life for patients with prolapse.
Clinical Evidence
A prospective, open-label, single-arm study (n=62) compared the Reia pessary to subjects' current ring or Gellhorn pessaries. The primary effectiveness endpoint, measured by the Pelvic Floor Distress Inventory-20 (PFDI-20), showed the Reia pessary was equivalent to current pessaries (p < 0.05). Safety was evaluated by the absence of serious, related adverse events; all related AEs were mild or moderate. The study confirmed the device is safe and effective for its intended use.
Technological Characteristics
Constructed of 100% medical-grade silicone. Single-piece, collapsible, domed geometry with a stem and loop. Mechanical support principle. Reusable; cleaned by user with soap and water. Biocompatibility tested per ISO 10993-1, -3, -5, -6, -10, -11, -17, -18, -23. Bench testing confirmed insertion/removal force, use-life, and hinge strength. No electronic components or software.
Indications for Use
Indicated for temporary, non-surgical management of Pelvic Organ Prolapse in patients requiring vaginal support.
Regulatory Classification
Identification
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse (falling down of uterus), uterine retroposition (backward displacement), or gynecologic hernia.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 21, 2024
Reia, LLC Kaitlin Maier CEO 331 River Road Lyme, New Hampshire 03768
Re: K232677 Trade/Device Name: Reia pessary Regulation Number: 21 CFR 884.3575 Regulation Name: Vaginal Pessary Regulatory Class: II Product Code: HHW Dated: April 18, 2024 Received: April 19, 2024
## Dear Kaitlin Maier:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232677
Device Name Reia pessary
Indications for Use (Describe)
The Reia pessary is indicated for the temporary, non-surgical management of Pelvic Organ Prolapse.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
## Submitter
Reia, LLC 331 River Road Lyme, NH 03768
Phone: (917) 719-0045 Contact person: Kaitlin Maier, CEO
Date Prepared: May 21, 2024
Name of Device Reia pessary
## Common or Usual Name Vaginal pessary
Classification Name Vaginal pessary (21 CFR 884.3575)
Regulatory Class Class II
Product Code HHW (Pessary, vaginal)
Predicate Devices ConTIPI ProVate Vaginal Support (K190277)
The predicate device has not been subject to a design-related recall.
## Device Description
The Reia pessary is a collapsible vaginal pessary for non-surgical treatment of pelvic organ prolapse. The Reia pessary is reusable by a single user and is intended for prescription home use following an in-office size fitting process performed by a healthcare professional using fitting pessaries.
The Reia pessary is placed in the vagina, similar to a tampon or diaphragm, and by supporting the vaginal wall, the pessary prevents the descending organs from dropping into the vaginal canal. Similar to currently available pessaries, the Reia pessary has been evaluated for continuous use; however, its collapsible design enables simpler insertion and removal.
The Reia pessary is made entirely of medical-grade silicone and is a single part. The geometry of the pessary and the flexible nature of the silicone allows the Reia pessary to collapse to facilitate easy, comfortable insertion and removal. In its supportive state, the Reia pessary is domed at the top with a stem at the bottom. The end of the stem has a large loop that allows the user to easily locate, grasp, and pull the pessary out for removal. The expandable and collapsible technology fits the user's anatomy during insertion, use, and removal.
The Reia pessary is designed for self-management and to be easily collapsed for insertion and removal by any user. To insert the pessary, a user (e.g., patient or practitioner) pinches the outside of the pessary to
## 510(k) Summary
Traditional 510(k): Reia pessary
{5}------------------------------------------------
elongate it into a narrower shape, allowing for easier, more comfortable insertion. Once it is released in the vagina, the pessary naturally expands to return to its supportive state. To remove the user pulls on the loop to re-elongate the pessary allowing for easier, more comfortable removal.
The Reia pessary is manually cleaned with soap and water by the user in the home environment or a healthcare facility between periods of use. The Reia fitting pessaries are manually cleaned and sterilized at a healthcare facility between uses.
## Intended Use / Indications for Use
The Reia pessary is indicated for the temporary, non-surgical management of Pelvic Organ Prolapse.
## Comparison of Technological Characteristics with the Predicate Device
The Reia pessary is a vaginal pessary that is used to provide support to prolapsed organs, in a manner similar to the predicate, ProVate Vaginal Support (K190277). The devices have the same intended use and similar principles of operation and technological characteristics. Bench tests have been conducted and demonstrated the biocompatibility and performance properties of both devices. Both the Reia pessary and the predicate pessary support the vaginal wall, collapse in shape for insertion and removal, are fit by practitioners, can be inserted and removed by the user, and come in multiple sizes.
The differences between the subject and predicate devices are:
- . The subject device is a single piece construction and made of 100% medical grade silicone while the predicate device has multiple parts and is made of multiple materials. However, the construction of the subject device is similar to other commonly marketed pessaries that are also a single piece and 100% silicone.
- The subject device is reusable with cleaning instructions while the predicate device is single use for . up to 7 days. However, the subject device reprocessing is the same as other currently marketed pessaries. To support reuse, the subject device has been tested for and will be indicated for 1 year of repeated use by a single user.
- The subject device is intended to be inserted and removed by either a healthcare practitioner or the . user in the home while the predicate device is only intended to be inserted by the user in the home. Due to the disposable nature and short use life of the predical for practical for practitioners to insert and remove the predicate device for each use. The subject device is reusable and can be managed by the user at home or by the healthcare practitioner. This dual management of insertion and removal is also similar to commonly marketed vaginal pessaries.
These identified differences were assessed through a combination of biocompatibility, benchtop performance, and clinical performance tests. The testing demonstrated that the identified differences did not impact the subject device performance, safety, or efficacy compared with the predicate device.
#### Performance Data
Testing was performed to evaluate the properties and performance of the device and to confirm the device meets its specifications and intended use requirements. Specifically, the following testing was conducted successfully:
## Bench Testing
Bench testing was performed to confirm that the Reia pessary met key specifications related to insertion and removal force, use-life, and hinge strength. The pessary was also tested to ensure it did not affect vaginal microflora, enhance the growth of Staphylococcus aureus, nor increase the production of Toxic Shock Syndrome Toxin-1. Additional bench testing was performed to confirm the fitting pessary maintains performance properties following repeated sterilization. All tests passed.
## 510(k) Summary
{6}------------------------------------------------
## Biocompatibility Testing
Biocompatibility testing and chemical characterization of the device according to ISO 10993-1 and applicable FDA guidance documents were conducted on new devices to confirm the biocompatibility requirements. All tests passed. The testing included the following:
- ISO 10993-5 Cytotoxicity ●
- ISO 10993-10 Sensitization ●
- ISO 10993-23 Irritation ●
- ISO 10993-3 Genotoxicity ●
- ISO 10993-6 Implantation .
- ISO 10993-11 Systemic Toxicity
- ISO 10993-18 Extractable/Leachable Evaluation with ISO 10993-17 Risk Assessment ●
Additional biocompatibility testing was conducted on repeatedly sterilized devices, simulating typical fitting pessary usage, to confirm sterilization of Reia's fitting pessary does not affect biocompatibility properties. The testing included the following:
- ISO 10993-5 Cytotoxicity ●
- ISO 10993-10 Sensitization
- ISO 10993-10 Irritation ●
## Reprocessing Testing
Reprocessing testing was conducted to confirm the fitting pessary could be adequately sterilized between uses with different patients. Reia's pessaries are not packaged or shipped sterile.
## Clinical Testing
## Study Overview
A prospective open label single-arm study with patients serving as their own control was conducted to evaluate safety and effectiveness of the Reia pessary for the non-surgical management of pelvic organ prolapse. The study was performed at seven sites in the United States. Following an approximately 1 month period using their own pessary (ring or Gellhorn), subjects were fit with a Reia pessary, which they used for approximately 3 months. The Reia pessary was used in the home environment and the protocol mirrored the routine clinical care for the management of pessaries. Surveys were initially conducted to assess baseline effectiveness with the current (ring or Gellhorn) pessary and again at the end of the study with the Reia pessary. Vaginal examinations were conducted during each visit to the study clinic.
78 subjects initially enrolled and 62 subjects (50 ring users and 12 Gellhorn users) were treated with the Reia pessary. All subjects who were treated were included in the effectiveness primary endpoint analyses. The study met both primary endpoints.
## Safetv
The primary safety endpoint was the absence of serious and related adverse events (AEs) that could put the general population at risk. All related AEs were mild or moderate as defined by the Common Terminology Criteria for Adverse Events (Version 5.0). There were 135 AEs that were new and/or worsening with the Reia pessary, 70 AEs that were persisting and unchanged across use of the current and Reia pessary, and 81 AEs existing with the subjects' current (ring or Gellhorn) pessary and improving with the Reia pessary. The use of the Reia pessary did not result in any serious and related adverse events and all related events were associated with typical pessary use.
## Effectiveness
The primary effectiveness endpoint was difference in Pelvic Floor Distress Inventory - 20 (PFDI-20) scores between current (ring or Gellhorn) and Reia pessaries. The PFDI-20 is a validated instrument to assess the presence and bother of symptoms that pelvic floor disorders have on health-related quality of life and widely
{7}------------------------------------------------
accepted as a standard subjective outcome for efficacy in urogynecology literature. The study demonstrated that the Reia pessary is equivalent to current (ring or Gellhorn) pessaries in terms of treatment of the presence and bother of pelvic floor symptoms as evaluated by the PFDI-20 (p < 0.05).
## Conclusions
The performance data demonstrates that the Reia pessary is as safe and effective as the predicate device to support a substantial equivalence determination.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.