← Product Code MLY · K232674

# PainFreeze II (K232674)

_Nuance Medical, LLC · MLY · Feb 12, 2024 · PM · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K232674

## Device Facts

- **Applicant:** Nuance Medical, LLC
- **Product Code:** MLY
- **Decision Date:** Feb 12, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** PM
- **Attributes:** Therapeutic

## Intended Use

PainFreeze II Medium Stream and Mist Spray are vapocoolants (skin refrigerants) intended for topical application to skin, intact mucous membranes (oral cavity, nasal passageways and the lips) and minor open wounds. PainFreeze II's skin refrigerant controls pain associated with injections (venipuncture, IV starts, cosmetic procedures), minor surgical (lancing boils, incisions, drainage of small abscesses, and sutures), and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions).

## Device Story

PainFreeze II is a non-flammable vapocoolant aerosol spray; utilizes chemical blend of 60% HFO-1233zd and 40% HFO-1234ze. Device consists of pressurized canister, valve, and actuator; available in 35ml and 115ml volumes. Operates by manual depression of actuator to dispense mist or stream spray onto target site. Immediate evaporation of chemical blend upon skin contact creates cooling sensation; cooling effect intensity modulated by spray distance and droplet size. Used in clinical settings for pain management during injections, minor surgical procedures, or sports injury relief. Healthcare provider applies spray to patient; cooling effect provides temporary analgesia. Benefits include non-invasive pain control for minor procedures.

## Clinical Evidence

No clinical data. Bench testing only. Evaluations included temperature comparison to predicate, biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity) per ISO 10993-1, and stability testing (visual inspection, total delivery, pressure, spray rate, leakage, microbiological) over 24-month shelf-life. All criteria met.

## Technological Characteristics

Chemical blend: 60% HFO-1233zd and 40% HFO-1234ze. Pressurized aerosol canister with valve/actuator. Thermal energy delivery via refrigerant evaporation. Non-flammable. Biocompatibility per ISO 10993-1. Shelf-life: 24 months. Storage: <49°C.

## Predicate Devices

- Pain Freeze ([K162218](/device/K162218.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 12, 2024

Nuance Medical, LLC Neal Hartman Regulatory Affairs/Quality Assurance 5931 Sea Lion Place, Suite 113 Carlsbad, California 92010

Re: K232674

Trade/Device Name: PainFreeze II Regulatory Class: Unclassified Product Code: MLY Dated: August 31, 2023 Received: September 1, 2023

Dear Neal Hartman:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Lauren E. Woodard -S

for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K232674

Device Name PainFreeze II

#### Indications for Use (Describe)

PamFreeze II Medium Stream and Mist Spray are vapocoolants (skin refrigerants) intended for topical op skin. intact mucous membranes (oral cavity, nasal passageways and the lips) and minor open wounds. PainFreeze II's skin refrigerant controls pain associated with injections (venipuncture, IV starts, cosmetic procedures), minor surgical (lancing boils, incisions, drainage of small abscesses, and sutures), and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions).

| Type of Use (Select one or both, as applicable)                                                        |                                                                                                       |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b> </b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;"><b> </b></span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Nuance Medical. The word "NUANCE" is written in large, bold, dark blue letters. Below it, the word "MEDICAL" is written in smaller, orange letters. The logo is framed by thin orange lines on the left and right sides.

## K232674 - 510(K) SUMMARY

#### Submitter Information

| Company Name:      | Nuance Medical, Inc.                                                            |
|--------------------|---------------------------------------------------------------------------------|
| Company Address:   | 5931 Sea Lion Place, Suite 113<br>Carlsbad, CA 92010                            |
| Company Phone:     | (760) 585-9548                                                                  |
| Company Facsimile: | (760) 235-4572                                                                  |
| Contact Person:    | Neal Hartman<br>Regulatory Affairs/Quality Assurance<br>nealenhartman@gmail.com |
| Date:              | January 29, 2024                                                                |

#### Device Identification

| Device Trade Name:      | PainFreeze II                                                                                                         |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Common Name:            | Cold Spray (Skin Refrigerant)                                                                                         |
| Models:                 | #1830 - Mist Stray, 115ml<br>#1831 - Medium Stream, 115ml<br>#1832 - -Mist Spray, 35ml<br>#1833 - Medium Stream, 35ml |
| Classification Name(s): | Vapocoolant Device                                                                                                    |
| Regulation(s):          | Unclassified                                                                                                          |
| Device Class:           | Unclassified                                                                                                          |
| Product Code(s):        | MLY                                                                                                                   |

## ldentification of Predicate Devices

The System Device is substantially equivalent to the following device:

| Device Name | Classification Name/Regulation   | Product<br>Code | 510(K)<br>Number | Clearance<br>Date |
|-------------|----------------------------------|-----------------|------------------|-------------------|
| Pain Freeze | Vapocoolant Device/ Unclassified | MLY             | K162218          | 11/22/2016        |

{4}------------------------------------------------

Common Name: Cold Spray – 245fa (1.1.1.3.3-Pentafluorpropane) and 134a (1.1.1.2-Tetrafluoroethane)

## Device Description

The Subject Device consists of a chemical blend of 60% HFO-1233zd (Trans-1-chloro-3.3.3- trifluoropropene) and 40% HFO-1234ze (trans- 1.3.3.3-tetrafluoroprop-1- ene) that produces a cooling effect upon contact with the skin, intact mucous membranes, and minor open wounds. The product is delivered in the form of an aerosol in either a mist or stream spray. The device's delivery system controls the amount of chemical blend that is dispensed, the mist spray configuration produces very fine droplets that cools at the points of contact. The stream spray configuration produces a pinpoint stream that contacts the skin surface at a specific single location. Upon contact with the skin, the product begins to evaporate immediately as it contacts the skin. The low evaporation rate of the product's chemical blend is what creates the coldness. The Subject Device is non- flammable and has a lower global warming potential than the predicate and provides equivalent performance.

The Subject Device is topically sprayed onto the skin. It creates a cooling effect on the surface of the application site by the immediate evaporation of the product from the skin surface.

With both product configurations, the evaporation creates the cold sensation. As the distance from the target surface is increased, the dispersion of the droplets in both the mist and stream is increased. Increasing the surface area of contact and decreasing the size of the droplets increases the evaporation rate. The increase in evaporation rate correlates to an increase in the cooling effect.

The chemical blend is incorporated in pressurized canisters, which will be offered in two (2) volumes 35 and 115 ml. The canister consists of a high-pressure canister, valve, and actuator.

#### Indications for Use

PainFreeze II Medium Stream and Mist Spray are vapocoolants (skin refrigerants) intended for topical application to skin, intact mucous membranes (oral cavity, nasal passageways and the lips) and minor open wounds. PainFreeze II's skin refrigerant controls pain associated with iniections (venipuncture, IV starts, cosmetic procedures). minor surgical (lancing boils, incisions, drainage of small abscesses, and sutures), and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions).

| Table 10.1: Comparison Table - Subject & Predicate Devices |                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                           |  |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Comparison<br>Feature                                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similarities/<br>Differences                                                                                                                                              |  |
| Device name                                                | PainFreeze II                                                                                                                                                                                                                                                                                                                                                                                                                                  | PainFreeze                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                           |  |
| Manufacturer                                               | Nuance Medical, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                           | Nuance Medical, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                           |  |
| Indications for Use                                        | PainFreeze II Medium Stream<br>and Mist Spray are<br>vapocoolants (skin<br>refrigerants) intended for<br>topical application to skin,                                                                                                                                                                                                                                                                                                          | Pain Freeze Mist Spray and<br>Medium Stream Spray are<br>vapocoolants (skin refrigerants)<br>intended for topical application<br>to skin, intact mucous                                                                                                                                                                                                                                                                                                                                                                                                             | Same with the exception<br>of removing the<br>statement "Pain Freeze<br>Medium Stream Spray<br>is also intended for the                                                   |  |
| Table 10.1: Comparison Table – Subject & Predicate Devices |                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                           |  |
| Comparison<br>Feature                                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similarities/<br>Differences                                                                                                                                              |  |
|                                                            | intact mucous membranes<br>(oral cavity, nasal<br>passageways, and the lips)<br>and minor open wounds.<br>PainFreeze II's skin refrigerant<br>controls pain associated with<br>injections (venipuncture, IV<br>starts, cosmetic procedures),<br>minor surgical (lancing boils,<br>incisions, drainage of small<br>abscesses, and sutures), and<br>the temporary relief of minor<br>sports injuries (sprains,<br>bruising, cuts and abrasions). | membranes (oral cavity, nasal<br>passageways, and the lips) and<br>minor open wounds. Pain<br>Freeze controls pain associated<br>with injections (venipuncture, IV<br>starts, cosmetic procedures),<br>minor surgical procedures (such<br>as lancing boils, incisions,<br>drainage of small abscesses<br>and sutures) and the temporary<br>relief of minor sports injuries<br>(sprains, bruising, cuts and<br>abrasions). Pain Freeze<br>Medium Stream Spray is also<br>intended for the management of<br>myofascial pain, restricted<br>motion and muscle tension. | management of<br>myofascial pain,<br>restricted motion and<br>muscle tension." This<br>does not impact the<br>safety and efficacy of<br>the device.                       |  |
| Product<br>Configurations                                  | Mist spray<br>Medium stream spray                                                                                                                                                                                                                                                                                                                                                                                                              | Mist spray<br>Medium stream spray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                      |  |
| Chemical<br>Composition                                    | HFO-1233zd (Trans-1-chloro-<br>3,3,3- trifluoropropene) 60%<br>and HFO-1234ze (trans-<br>1,3,3,3-tetrafluoroprop-1- ene)<br>40%                                                                                                                                                                                                                                                                                                                | 1,1,1,3,3-Pentafluoropropane<br>(HFC-245fa) 95% and 1,1,1,2-<br>Tetrafluoroethane (HFC-134a)<br>5%                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | The formation of the<br>device is different.<br>Information incorporated<br>in the submissions<br>demonstrates the<br>formulation is safe and<br>effective.               |  |
| Container Size                                             | 35ml, 115ml                                                                                                                                                                                                                                                                                                                                                                                                                                    | 103.5ml                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Volume offered differs,<br>however there is no<br>impact on safety or<br>efficacy                                                                                         |  |
| Target Populations                                         | Adults                                                                                                                                                                                                                                                                                                                                                                                                                                         | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Safety and efficacy are not<br>impacted.                                                                                                                                  |  |
| Patient Contact                                            | Skin                                                                                                                                                                                                                                                                                                                                                                                                                                           | Skin                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                      |  |
| Energy Delivery                                            | Thermal energy via refrigerant<br>spray                                                                                                                                                                                                                                                                                                                                                                                                        | Thermal energy via refrigerant<br>spray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                      |  |
| Vapocoolant<br>Discharge Method                            | Handheld spray containers.<br>Button on actuator is pressed.                                                                                                                                                                                                                                                                                                                                                                                   | Depress the Actuator Button to<br>release the vapocoolant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                      |  |
| Boiling Point                                              | $-3.0±3°C$                                                                                                                                                                                                                                                                                                                                                                                                                                     | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | A comparison<br>evaluation was<br>conducted, which the<br>temperature exiting the<br>canister is similar to the<br>predicate. Safety and<br>efficacy are not<br>impacted. |  |
| Mechanism of<br>Action & Operating<br>Principle            | Cooling by application of<br>refrigerant; Dispensing valve<br>on canister is depressed to<br>release the refrigerant.                                                                                                                                                                                                                                                                                                                          | Depress the Actuator Button to<br>release the vapocoolant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                      |  |
| Flammability                                               | Non-flammable                                                                                                                                                                                                                                                                                                                                                                                                                                  | Non-flammable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                      |  |
| Mechanical Safety                                          | Mechanism has positive shut-<br>off release                                                                                                                                                                                                                                                                                                                                                                                                    | Mechanism has positive shut-off<br>release                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                      |  |
| Shelf-Life                                                 | 24 months                                                                                                                                                                                                                                                                                                                                                                                                                                      | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Shelf-life may differ;<br>however, stability testing<br>is conducted, which<br>demonstrates no impact<br>to safety and efficacy.                                          |  |
| Storage &<br>Transportation                                | Not to exceed 49°C                                                                                                                                                                                                                                                                                                                                                                                                                             | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Storage conditions may<br>vary; however,<br>restrictions are<br>indicated on the product<br>labeling.                                                                     |  |

#### Summary of Technological Characteristics

{5}------------------------------------------------

{6}------------------------------------------------

#### Non-Clinical Testing

Assessments were performed that includes the following:

- Temperature Comparison - Evaluation demonstrated the temperature exiting from the canister is similar to the Predicate.
- Biocompatibility Evaluations were conducted per ISO 10993-1 Biological . evaluation of medical devices - Part 1: Evaluation and testing within a risk management process for devices in contact with compromised skin and a limited exposure (<24 hours). Evaluations included:
	- Cytotoxicity O
	- Sensitization O
	- Irritation O
	- o Acute Systemic Toxicity
	- Material Mediated Pyrogenicity o

All test criteria were successfully met for each evaluation.

- Stability The following evaluations were conducted at various assessment ● intervals up to the stated shelf-life limits:
	- o Visual Inspection
	- Total Delivery O
	- Pressure O
	- Spray Rate o
	- o Leakage/Weight Loss
	- o Microbiological

Test criteria were met at each assessment interval, which demonstrated the Subject Device's stability.

Results of the evaluations demonstrate that the Subject Device met the safety and performance requirements as it relates to its indication for use and the Predicate.

#### Clinical Testing

Not Applicable

#### Conclusions

The results of the evaluation demonstrate that the Subject Device is substantially equivalent to the Predicate.

---

**Source:** [https://fda.innolitics.com/device/K232674](https://fda.innolitics.com/device/K232674)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
