← Product Code [NXC](/productcode/NXC) · K232653

# Blue Sky Plan Software, Blue Sky Bio Aligner Z, and Blue Sky Bio Aligner Z FLX (K232653)

_Blue Sky Bio, LLC · NXC · Apr 3, 2024 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K232653

## Device Facts

- **Applicant:** Blue Sky Bio, LLC
- **Product Code:** [NXC](/productcode/NXC.md)
- **Decision Date:** Apr 3, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Software as a Medical Device

## Indications for Use

Blue Sky Plan Software orthodontic module is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual design of a series of dental casts which may be used for sequential aligner trays and aligners, based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of treatment, and at the end of treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. Blue Sky Bio Aligner Z and Blue Sky Bio Aligner Z FLX are series of clear, lightweight, plastic appliances indicated for the treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental tooth movements, it sequentially positions teeth by way of continuous force. The use of the Blue Sky Plan Software module requires the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

## Device Story

Blue Sky Plan Software is a medical front-end device for orthodontic treatment planning. Clinicians input patient dental data via intraoral scans or physical molds (digitized as .STL files). The software provides tools for model management, inspection, analysis, treatment simulation, and virtual design of dental casts for sequential aligners. The clinician reviews and approves the treatment plan; the dental lab then fabricates the aligners using thermoplastic materials (polyurethane or copolyester/polyurethane composite). The finished aligners are provided to the patient by the clinician, who monitors progress throughout the treatment. The aligners are removable and apply continuous force to move teeth incrementally. The device is used in a clinical setting by trained orthodontic professionals.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993 (Ames Assay, Cytotoxicity, Intracutaneous Reactivity, Oral Mucosa Irritation). Performance characteristics characterized per ASTM D638. Software verification and validation documentation provided.

## Technological Characteristics

Software: Orthodontic planning/design module. Aligners: Clear, lightweight, thermoplastic (polyurethane or copolyester/polyurethane composite). Manufacturing: Thermoforming. Connectivity: Standalone/networked (STL file import). Biocompatibility: ISO 10993 compliant. Performance: ASTM D638.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- iOrtho ([K223518](/device/K223518.md))
- uLab Systems Dental Aligner Kit ([K192596](/device/K192596.md))

## Reference Devices

- Blue Sky Bio Aligner ([K180107](/device/K180107.md))
- Blue Sky Bio Aligner G & Blue Sky Plan Software for Blue Sky Bio Aligner G ([K221845](/device/K221845.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 3, 2024

Blue Sky Bio, LLC % Patsy Trisler Regulatory Consultant Trisler Consulting, dba 306 Turnberry Court Lebanon, Indiana 46052

Re: K232653

Trade/Device Name: Blue Sky Plan Software, Blue Sky Bio Aligner Z, and Blue Sky Bio Aligner Z FLX Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC, PNN Dated: August 30, 2023 Received: March 6, 2024

Dear Patsy Trisler:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Michael E. Adjodha -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

### 510(k) Number (if known)

K232653

#### Device Name

Blue Sky Plan Software, and Blue Sky Bio Aligner Z and Blue Sky Bio Aligner Z FLX

#### Indications for Use (Describe)

Blue Sky Plan Software orthodontic module is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual design of a series of dental casts which may be used for sequential aligners, based on 3D models of the patient's dentition before the start of an orthodonic treatment. It can also be applied during the treatment to inspect and analyze the progress of treatment, and at the end of treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.

Blue Sky Bio Aligner Z and Blue Sky Bio Aligner Z FLX are series of clear, lightweight, plastic appliances indicated for the treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental tooth movements, it sequentially positions teeth by way of continuous force.

The use of the Blue Sky Plan Software module requires the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

| Type of Use ( <i>Select one or both, as applicable</i> )                                                                               |
|----------------------------------------------------------------------------------------------------------------------------------------|
| <span>       Prescription Use (Part 21 CFR 801 Subpart D)   </span> <span>       Over-The-Counter Use (21 CFR 801 Subpart C)   </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary K232653

Submitter Address: 800 Liberty Drive Libertyville, IL 60048

Azickmann@blueskybio.com Email Address:

Contact Person: Dr. Albert Zickmann, VP, Product Development

#### Date Prepared: March 6, 2024

Device Trade Name:

Blue Sky Plan Software and Blue Sky Bio Aligner Z and Blue Sky Bio Alianer Z FLX

Device Names Requlation Numbers Product Codes Class

| Orthodontic Software | Sequential Aligner |
|----------------------|--------------------|
| 21 CFR 872.5470      | 21 CFR 872.5470    |
| PNN                  | NXC                |
| 2                    | 2                  |

# Predicate Device Information

510(k) #, Device Trade name & Manufacturer Device Name Regulation Number Product Code

### Reference Devices

510(k) #, Device Trade name & Manufacturer Device Name Regulation Number Product Code

Indications for Use:

| Predicate #1 (Primary):                                            | Predicate #2 (Secondary):                                                                           |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| K223518<br>iOrtho,<br>Shanghai EA Medical Instruments<br>Co., Ltd. | K192596,<br>uLab Systems Dental Aligner Kit;<br>uLab Systems, Inc.                                  |
| Orthodontic Software                                               | Aligner, Sequential                                                                                 |
| 21 CFR 872.5470                                                    | 21 CFR 872.5470                                                                                     |
| PNN                                                                | NXC                                                                                                 |
| K180107                                                            | K221845,                                                                                            |
| Blue Sky Bio Aligner; Blue Sky Bio<br>LLC                          | Blue Sky Bio Aligner G & Blue<br>Sky Plan Software for Blue Sky<br>Bio Aligner G; Blue Sky Bio, LLC |
| Aligner, Sequential                                                | Orthodontic Software                                                                                |
| 21 CFR 872.5470                                                    | 21 CFR 872.5470                                                                                     |
| NXC                                                                | PNN, NXC                                                                                            |

Blue Sky Plan Software orthodontic module is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual design of a series of dental casts which may be used for sequential aligner trays and aligners, based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of treatment, and at the end of treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.

Blue Sky Bio Aligner Z and Blue Sky Bio Aligner Z FLX are series of clear, lightweight, plastic appliances indicated for the treatment

{5}------------------------------------------------

of tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental tooth movements, it sequentially positions teeth by way of continuous force.

The use of the Blue Sky Plan Software requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

Device Description: A dental clinician, using a standard personal computer prescribes Summary of the orthodontic appliance based on an assessment of the patient's Technological teeth and determines the course of treatment using the Blue Sky Characteristics. Plan Software. The clinician takes molds of the patient's teeth Material Composition and completes a prescription form and sends the molds to the dental lab, which in turn scans the molds and uploads the .STL and Mechanism of Action files of the molds in Blue Sky Plan software. This digital file is a series of CAD files (.STL) for building models that can be used to fabricate the aligners. Alternatively, the dental clinician may qenerate the digital files by scanning the patient's mouth directly, using an intraoral scanner, and then send the files to the dental lab.

> The Blue Sky Bio Aligners (Z and Z FLX) consist of a series of clear plastic aligner trays which are fabricated with the selected thermoplastic material (either a thin polyurethane sheet or a composite of copolyester and polyurethane). Both materials are commercially used for thermoforming customized removable aligners for the intended use.

> The prescribing physician reviews and approves the model scheme before the aligners are produced. Once approved by the clinician, the dental lab produces the Blue Sky Bio Aligner Z or Z FLX.

> The finished aligners are sent to the dental clinician who then provides them to the patient, confirming fit and form. Over a period of time, additional aliqner trays are provided sequentially to the patient by the clinician to gradually move the target teeth to the designed position. The dental clinician monitors treatment from delivery of the first aligner to the final aligner.

> The trays are held in place by pressure and can be removed by the patient at any time.

#### Device Testing Laboratory Testing

Test data were submitted to validate the processes used for the design and manufacture of the clear customized aligners.

Testing was conducted to characterize device performance characteristics according to ASTM D638.

Testing to verify and validate the software has been included in this 510(k) documentation.

{6}------------------------------------------------

Biocompatibility

ISO 10993 testing, according to Good Laboratory Practices, has been performed to assure the two thermoplastic materials are biocompatible and non-toxic for the oral contact use:

Part 3 (Bacterial Mutagenicity - Ames Assay)

Part 5 (Cytotoxicity Elution - MEM),

Part 10 (Intracutaneous/Intradermal) Reactivity),

Part 10 (Oral Mucosa Irritation)

### Animal | Human Testing

No animal or human testing are required for this product.

Comparison to Predicate and Reference Devices:

The similarities in intended use and technological characteristics comparing the subject Blue Sky Plan Software and Blue Sky Bio Z and Z FLX aligners to the Predicates follow:

- The software has the same intended use as the Primary ● Predicate and functions similarly as noted.
- The aligners are clear plastic, customized sequential ● aligners with the same intended use and mechanism of action as the Secondary Predicate and Reference K180107:
	- They are intended for realignment of teeth during . orthodontic treatment of malocclusions by way of continuous gentle force.
	- They use the same thermoplastic materials as the ● Secondary Predicate.
	- . They are thermoformed using similar processes as the Secondary Predicate and using the same processes as the Reference Device (K180107).

The subject device software and the Primary Predicate software have proprietary differences, but they have the same intended use and function equivalently. Furthermore, both Reference devices use the same software.

The only differences between the subject device aligners and the Secondary Predicate are the minor proprietary differences in the processes; those processes are standard in this field for thermoforming/manufacturing clear aligners. Furthermore they are the same processes used for Reference aligner K180107.

Details follow on the Substantial Equivalence Comparison Tables (Table A for the PNN code; and Table B for NXC code).

Based on the documentation presented in the 510(k), as Substantial summarized herein, it can be concluded that these clear aligner Equivalence systems are substantially equivalent to the Predicate devices. Conclusion:

{7}------------------------------------------------

### Substantial Equivalence Comparison Tables

| Trade Name:                                                                                                                                                                                                                                              | Subject Device:<br>Blue Sky Plan Software, &                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Primary Predicate:<br>iOrtho                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison<br>to Predicate                                         | Trade Name:                                              | Subject Device:<br>Blue Sky Plan Software, &<br>Blue Sky Bio Aligner Z and<br>Blue Sky Bio Aligner Z FLX                                                                              | Secondary Predicate:<br>uLab Systems Dental<br>Aligner Kit                                                                                                                               | Comparison<br>to Predicate                             |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
|                                                                                                                                                                                                                                                          | Blue Sky Bio Aligner Z and<br>Blue Sky Bio Aligner Z FLX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                    | 510(k) #                                                 | K232653                                                                                                                                                                               | K192596                                                                                                                                                                                  | n/a                                                    |
| 510(k) #                                                                                                                                                                                                                                                 | K232653                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | K223518                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | n/a                                                                | Manufacturer                                             | Blue Sky Bio, LLC                                                                                                                                                                     | uLab Systems, Inc.                                                                                                                                                                       | n/a                                                    |
| Manufacturer                                                                                                                                                                                                                                             | Blue Sky Bio, LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Shanghai EA Medical<br>Instruments Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | n/a                                                                | Device Name<br>21 CFR<br>Product Code<br>Class           | Sequential Aligner &<br>Orthodontic Software<br>872.5470<br>PNN & NXC<br>2                                                                                                            | Sequential Aligner<br>872.5470<br>NXC<br>2                                                                                                                                               | Same                                                   |
| Device Name<br>21 CFR<br>Product Code<br>Class                                                                                                                                                                                                           | Orthodontic Software &<br>Sequential Aligner<br>872.5470<br>PNN & NXC<br>2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Orthodontic Software<br>872.5470<br>PNN<br>2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                               | Intended Use                                             | Same as Table A above                                                                                                                                                                 | The uLab Systems Dental<br>Aligner is indicated for the<br>alignment of permanent teeth<br>during orthodontic treatment<br>of malocclusions by way of<br>continuous gentle forces.       | Same aligner<br>Intended Use<br>as both<br>Predicates. |
| Intended Use                                                                                                                                                                                                                                             | Blue Sky Plan Software<br>orthodontic module is intended<br>for use as a medical front-end<br>device providing tools for<br>management of orthodontic<br>models, systematic inspection,<br>detailed analysis, treatment<br>simulation and virtual design<br>of a series of dental casts<br>which may be used for<br>sequential aligner trays and<br>aligners, based on 3D models<br>of the patient's dentition before<br>the start of an orthodontic<br>treatment.  It can also be<br>applied during the treatment to<br>inspect and analyze the<br>progress of treatment, and at<br>the end of treatment to<br>evaluate if the outcome is<br>consistent with the<br>planned/desired treatment<br>objectives.<br>Blue Sky Bio Aligner Z and<br>Blue Sky Bio Aligner Z FLX<br>are series of clear, lightweight,<br>plastic appliances indicated for<br>the treatment of tooth<br>malocclusions in patients with<br>permanent dentition (i.e. all<br>second molars).  Utilizing a<br>series of incremental tooth<br>movements, it sequentially<br>positions teeth by way of<br>continuous force. | iOrtho is intended for use as a<br>medical front-end device<br>providing tools for management<br>of orthodontic cases,<br>systematic inspection, detailed<br>analysis, treatment simulation<br>and virtual appliance design<br>options (Export of Models,<br>Indirect Bonding Transfer<br>Media, Sequential aligners)<br>based on 3D models of the<br>patient's dentition before the<br>start of an orthodontic<br>treatment. It can also be<br>applied during the treatment to<br>inspect and analyze the<br>progress of the treatment. It can<br>be used at the end of the<br>treatment to evaluate if the<br>outcome is consistent with the<br>planned/desired treatment<br>objectives.<br>The use of iOrtho requires the<br>user to have the necessary<br>training and domain knowledge<br>in the practice of orthodontics,<br>as well as to have received a<br>dedicated training in the use of<br>the software. | Same<br>software<br>Intended Use<br>as the<br>Primary<br>Predicate | Mechanism of<br>Action of<br>Sequential<br>Aligner Trays | Orthodontic tooth movements<br>through forces applied by the<br>appliance to the dentition as<br>each tooth follows the<br>programmed displacement<br>based on doctor's prescription. | Orthodontic tooth movements<br>through forces applied by the<br>appliance to the dentition as<br>each tooth follows the<br>programmed displacement<br>based on doctor's<br>prescription. | SE                                                     |
| The use of the Blue Sky Plan<br>Software module requires the<br>user to have the necessary<br>training and domain<br>knowledge in the practice of<br>orthodontics, as well as to<br>have received a dedicated<br>training in the use of the<br>software. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Treatment<br>planning                                              | Use of orthodontic software<br>program                   | Use of orthodontic software<br>program                                                                                                                                                | SE                                                                                                                                                                                       |                                                        |
| Supported<br>Anatomic Area                                                                                                                                                                                                                               | Maxilla/Mandible                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Maxilla/Mandible                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | SE                                                                 | Method of Use                                            | Each customized aligner is<br>worn as prescribed for 20-22<br>hours per day                                                                                                           | Each customized aligner is<br>worn as prescribed for 20-22<br>hours per day                                                                                                              | SE                                                     |
| Mechanism of<br>Action                                                                                                                                                                                                                                   | Medical front-end software as<br>described in Intended Use<br>statement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Medical front-end software as<br>described in Intended Use<br>statement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | SE.                                                                | Material and<br>Form                                     | Sheet of thin thermoformed<br>polyurethane or copolyester<br>and polyurethane composite                                                                                               | Sheet of thin thermoformed<br>polyurethane or copolyester<br>and polyurethane composite                                                                                                  | SE to<br>Secondary<br>Predicate                        |
| Software<br>Characteristics                                                                                                                                                                                                                              | • Manages patient & case base<br>data<br>• Collects study material<br>• Segments study material<br>• Aligns study material<br>• Measures study material<br>• Measures point selection<br>• Analyzes study material<br>• Prepares treatment<br>simulations and options<br>• Designs virtual appliances<br>• Surface scans for intraoral<br>scanners<br>• Surface scans from STL files                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | • Manages patient & case base<br>data<br>• Collects study material<br>• Segments study material<br>• Aligns study material<br>• Measures study material<br>• Measures point selection<br>• Analyzes study material<br>• Prepares treatment simulations<br>and options<br>• Designs virtual appliances<br>• Surface scans for intraoral<br>scanners<br>• Surface scans from STL files                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | SE                                                                 | Biocompatibility                                         | Materials shown to be<br>biocompatible and non-toxic.                                                                                                                                 | Materials shown to be<br>biocompatible and non-toxic.                                                                                                                                    | SE to<br>Secondary<br>Predicate                        |
| Rx or OTC                                                                                                                                                                                                                                                | Rx only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Rx only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | SE                                                                 | Rx or OTC                                                | Rx only                                                                                                                                                                               | Rx only                                                                                                                                                                                  | SE                                                     |

### Table A – Comparison of Orthodontic Software

{8}------------------------------------------------

{9}------------------------------------------------

### Table B: Comparison of Sequential Aligners

---

**Source:** [https://fda.innolitics.com/device/K232653](https://fda.innolitics.com/device/K232653)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
