← Product Code [NKB](/productcode/NKB) · K232607

# 4CIS SARA SPINE SYSTEM, 4CIS VERTU SPINE SYSTEM, 4CIS WILL PEDICLE SCREW SYSTEM, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM (K232607)

_Solco Biomedical Copany India Private Limited · NKB · Feb 27, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K232607

## Device Facts

- **Applicant:** Solco Biomedical Copany India Private Limited
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Feb 27, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are the pedicle screw systems indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-Sl vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative Spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudarthrosis ). aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are intended to used for Minimal Invasive Surgery (MIS).

## Device Story

Top-loading posterior spinal fixation system; consists of pedicle screws, rods, nuts, and transverse links; used by surgeons in clinical settings to stabilize spinal segments and promote fusion; implants inserted into vertebral bodies via posterior approach; provides rigid fixation; components locked into configurations matching patient pathology; supports both open and minimally invasive surgery (MIS); benefits patients by immobilizing unstable segments to facilitate bone graft fusion; removed after solid fusion achieved.

## Clinical Evidence

Bench testing only. Mechanical testing (static/dynamic compression bending, static tension bending, static torsion) conducted per ASTM F1717 on worst-case constructs. Results met performance requirements against predicates. No clinical data provided.

## Technological Characteristics

Top-loading posterior spinal fixation system. Materials: Titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136. Components: pedicle screws, rods, nuts, transverse links. Supplied non-sterile for single use. Mechanical testing performed per ASTM F1717.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- 4CIS SARA Spine System ([K202498](/device/K202498.md))
- 4CIS VERTU Spine System ([K202498](/device/K202498.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

February 27, 2024

Solco Biomedical Copany India Private Limited Darshak Shah Director Survey No.1540, Beside Torrent Pharma, Village-Rajpur Ahmedabad Mehsana Highway Kadi, Gujarat/ Mehasana 382715 India

Re: K232607

Trade/Device Name: 4CIS SARA SPINE SYSTEM, 4CIS VERTU SPINE SYSTEM, 4CIS WILL PEDICLE SCREW SYSTEM, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: January 29, 2024 Received: January 29, 2024

Dear Darshak Shah:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Colin .. O'neill -S

 Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K232607

Device Name 4CIS SARA Spine System, 4CIS VERTU Spine System 4CIS WILL PEDICLE SCREW SYSTEM aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM

Indications for Use (Describe)

The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are the pedicle screw systems indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-Sl vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative Spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudarthrosis ).

aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are intended to used for Minimal Invasive Surgery (MIS).

| Type of Use ( <i>Select one or both, as applicable</i> )                                                          |                                                         |
|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| <div><span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

PAGE NO 1 of 4

Image /page/4/Picture/1 description: The image shows the word "SOLCO." in a bold, sans-serif font. The letters are a dark blue color, and there is a red dot to the right of the "O". The logo is simple and modern.

NO : 1.0 TITLE: 510(k) Summary

## "510(K) SUMMARY"

This summary of 510(k) substantial equivalence information is being submitted in accordance with requirement of 21 CFR 807.92.

|                             |                                                       | SOLCO Biomedical Company India Private Limited    |                       |  |
|-----------------------------|-------------------------------------------------------|---------------------------------------------------|-----------------------|--|
|                             | Survey No.1540, Beside Torrent Pharma, Village-Rajpur |                                                   |                       |  |
| Submitter                   | Ahmedabad Mehsana Highway, Ta-Kadi, Dist-Mehsana      |                                                   |                       |  |
|                             | Gujarat- 382715, INDIA.                               |                                                   |                       |  |
|                             | Tel : +91 851112597, Email: darshak@solco.co.in       |                                                   |                       |  |
|                             | Darshak Shah- Director                                |                                                   |                       |  |
| Contact Person              | SOLCO Biomedical Company India Private Limited        |                                                   |                       |  |
|                             | Phone : +91 98252 06091, Email: darshak@solco.co.in   |                                                   |                       |  |
| Submission Date             | 24 FEB 2024.                                          |                                                   |                       |  |
|                             |                                                       | 4CIS SARA Spine System                            |                       |  |
| Trade /<br>Proprietary name | 4CIS VERTU Spine System                               |                                                   |                       |  |
|                             | 4CIS WILL PEDICLE SCREW SYSTEM                        |                                                   |                       |  |
|                             | aBle SPINAL FIXATION SYSTEM                           |                                                   |                       |  |
|                             | aBle Xt SPINAL FIXATION SYSTEM                        |                                                   |                       |  |
| Classification              |                                                       | Thoracolumbosacral Pedicle screw system           |                       |  |
| Name                        | NKB                                                   |                                                   |                       |  |
| Classification              | Class II                                              |                                                   |                       |  |
| Code                        | 888.3070                                              |                                                   |                       |  |
| Regulatory Class            |                                                       |                                                   |                       |  |
| Regulation                  |                                                       |                                                   |                       |  |
| Number                      |                                                       |                                                   |                       |  |
|                             | 510K                                                  | Trade or Property or                              | Manufacturer          |  |
|                             | Number                                                | model Name                                        | SOLCO Biomedical      |  |
| Primary<br>Predicate Device | K202498                                               | 4CIS SARA Spine System<br>4CIS VERTU Spine System |                       |  |
|                             |                                                       |                                                   | Company India Private |  |
|                             |                                                       |                                                   | Limited               |  |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "SOLCO." in a bold, sans-serif font. The letters are a dark blue color. There is a red dot to the right of the "O".

DOC NO: K202498/Summary NO : 1.0

PAGE NO 2 of 4

TITLE: 510(k) Summary

|                                                                                    | The Spinal Fixation System is a top-loading posterior spinal fixation system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description of<br>Device                                                           | which consists of pedicle screws, rods, and nuts, transverse (cross) link                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                                    | and associated instruments. Rigid fixation is provided by pedicle screws                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                                    | inserted into the vertebral body through pedicle of the lumbar spine via                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                                    | posterior approach. This system will allow surgeons to build a spinal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                    | implant construct to stabilize and promote spinal fusion through open                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                    | surgery. Implant components can be rigidly locked into a variety of different                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                                    | configurations to suit the individual pathology and anatomical conditions of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                    | the mature patient. The implant components are supplied non-sterile single                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                    | use and are fabricated from titanium alloy (Ti-6A1-4V ELI) and Cobalt                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                    | Chromium ally that conforms to ASTM F136 and ASTM F1537 respectively.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                    | Also, Specialized instruments are available for the application                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                    | and removal of the Spinal Fixation System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Indication for Use                                                                 | The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                    | FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                                    | PEDICLE SCREW SYSTEM are the pedicle screw systems indicated for                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                                    | the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-SI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                                                                    | vertebra in skeletally mature patients receiving fusion by autogenous bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                    | graft having implants attached to the lumbar and sacral spine (L3 to                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                    | sacrum) with removal of the implants after the attainment of a solid fusion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                    | The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                    | FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                                    | PEDICLE SCREW SYSTEM are intended to provide immobilization and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                    | stabilization of spinal segments in skeletally mature patients as an adjunct                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                    | to fusion in the treatment of the following acute and chronic instabilities or                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| TITLE: 510(k) Summary                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                    | deformities of the thoracic, lumbar and sacral spine: degenerative<br>Spondylolisthesis with objective evidence of neurological impairment,<br>fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous<br>fusion (pseudarthrosis ).<br>aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW<br>SYSTEM are intended to used for Minimal Invasive Surgery (MIS).                                                                                                                                                                                                                                                                                                                                                                                                               |
| Comparison of<br>Technological<br>Characteristics<br>with the<br>Predicate Devices | The 4CIS WILL PEDICLE SCREW SYSTEM, aBle Xt SPINAL FIXATION<br>SYSTEM & aBle SPINAL FIXATION SYSTEM have similar indications as<br>compare to cleared 4CIS SARA Spine System and 4CIS VERTU Spine<br>System. The system is composed of the same material as the predicate-<br>K202498 devices conforming to recognized industry standards for<br>permanent implants (titanium alloy (Ti-6A1-4V ELI) that conforms to ASTM<br>F136) and surgical orthopedic instruments. aBle SPINAL FIXATION<br>SYSTEM is modified version of 4CIS SARA Spine System and 4CIS<br>VERTU Spine System which does not raise any concern of safety and<br>performance. 4CIS WILL PEDICLE SCREW SYSTEM is intended to used<br>for Minimal Invasive Surgery (MIS). Also they are provided non-sterile for<br>single use only. |
| Performance Data                                                                   | Mechanical testing (static and dynamic compression bending, static<br>tension bending, static torsion) is conducted in accordance with ASTM<br>F1717 on the worst case construct. Static Compression Bending,<br>Dynamic Compression Bending, Static Torsion, and Static Tension tests<br>results meet performance requirement against predicates.                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Biocompatibility                                                                   | The subject implants are permanent implants (> 30 days) and are classified<br>as body contacting devices according to FDA's Draft Guidance for Industry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                    | and FDA Staff "Use of International Standard ISO-10993, Biological                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                                                                    | Evaluation of Medical Devices Part 1: Evaluation and Testing.” The subject                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                    | implants are manufactured from identical materials as the predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                    | devices, in accordance with the ASTM F136 – Standard Specification for                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                                                                    | Wrought Titanium – 6Aluminum – 4Vanadium ELI (Extra-Low-Interstitial)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                    | Alloy for Surgical Implants The materials used for manufacturing the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                    | subject device have a long history of safe and effective use identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                                                                    | to predicate devices and biocompatibility testing is not required.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Conclusion                                                                         | The overall technology characteristics, material of construction , design                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                                    | characteristics and performance data lead to the conclusion that our Spine                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                    | Systems (4CIS SARA Spine System, 4CIS VERTU Spine System, aBle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                                                    | SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                    | 4CIS WILL PEDICLE SCREW SYSTEM) are substantially equivalent to our                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                    | FDA previously cleared devices of K202498 for intended use, material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                    | composition, principles of operation, and overall design.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for SOLCO. The text "SOLCO" is written in a bold, dark blue font. There is a red dot to the right of the text.

DOC NO: K202498/Summary NO : 1.0

TITLE: 510(k) Summary

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the word "SOLCO." in a bold, sans-serif font. The letters are a dark blue color. There is a red dot to the right of the word.

DOC NO: K202498/Summary NO : 1.0

PAGE NO 4 of 4

TITLE: 510(k) Summary

---

**Source:** [https://fda.innolitics.com/device/K232607](https://fda.innolitics.com/device/K232607)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
