← Product Code [OYE](/productcode/OYE) · K232600

# DxFLEX Flow Cytometer; ClearLLab 10C Panels (K232600)

_Beckman Coulter, Inc. · OYE · Nov 22, 2023 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K232600

## Device Facts

- **Applicant:** Beckman Coulter, Inc.
- **Product Code:** [OYE](/productcode/OYE.md)
- **Decision Date:** Nov 22, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K232600 · Nov 22, 2023 | DxFLEX Flow Cytometer; ClearLLab 10C Panels | Beckman Coulter, Inc. | De-identified residual clinical specimens (whole blood, bone marrow, lymph node) from routine clinical practice | Retrospective clinical study used to demonstrate method agreement between the candidate device and the predicate device, and to assess diagnostic accuracy against clinical outcomes. | Retrospective study; Residual clinical specimens; Method comparison; Diagnostic accuracy |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Multi-site retrospective clinical study; Retrospective method comparison study | Patients suspected of having hematopoietic neoplasms (acute leukemia, chronic leukemia, lymphoma, plasma cell neoplasm, MDS, MPN); Sample Size: 512 specimens; Number of Sites: 5 | ClearLLab 10C on Navios EX Flow Cytometer (Predicate) | Qualitative immunophenotype agreement (presence/absence of abnormal cells); Diagnostic accuracy (sensitivity, specificity, PPV, NPV) |

## Indications for Use

The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to ten fluorescent detection channels using three lasers (488 nm, 638 nm, and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument. The ClearLLab 10C Panels are intended for in vitro diagnostic use for qualitative identification of various cell populations by multiparameter immunophenotyping on the Navios, Navios EX and DxFLEX flow cytometers. These reagents are used as an aid in the differential diagnosis of hematologically abnormal patients having, or suspected of having, the following hematopoietic neoplasms: chronic leukemia, acute leukemia, non-Hodgkin lymphoma, myeloma, myelodysplastic syndrome (MDS), and/or myeloproliferative neoplasms (MPN). The reagents can be used with peripheral whole blood (collected in K2EDTA, Acid Citrate Dextrose (ACD) or Heparin), bone marrow (collected in K2EDTA, ACD or Heparin) and lymph node specimens. Interpretation of the results should be confirmed by a pathologist or equivalent professional in conjunction with other clinical and laboratory findings. These reagents provide multiparameter, qualitative results for the surface antigens listed below: - ClearLLab 10C B Cell Tube: Kappa, Lambda, CD10, CD5, CD200, CD34, CD38, CD20, CD19, CD45 - ClearLLab 10C T Cell Tube: TCRγδ, CD4, CD2, CD56, CD5, CD34, CD3, CD8, CD7, CD45 - ClearLLab 10C M1 Cell Tube: CD16, CD7, CD10, CD13, CD64, CD34, CD14, HLA-DR, CD11b, CD45 - ClearLLab 10C M2 Cell Tube: CD15, CD123, CD117, CD13, CD33, CD34, CD38, HLA-DR, CD19, CD45

## Device Story

DxFLEX Flow Cytometer performs qualitative immunophenotyping of cells; uses hydrodynamic focusing to align cells in single file through three lasers (488nm, 638nm, 405nm); scattered/fluorescent light collected by Avalanche Photodiodes (APDs); transduced to electronic pulses. System uses ClearLLab 10C reagent panels for identification of cell populations. Operated by laboratory personnel in clinical settings. Data acquired via CytExpert software; post-acquisition analysis performed manually using Kaluza C software on independent workstation. Output includes plots, statistics, and reports for pathologist review. Assists in differential diagnosis of hematopoietic neoplasms by identifying abnormal cell phenotypes. Benefits include standardized, multiparameter qualitative assessment of hematological specimens.

## Clinical Evidence

Multi-site retrospective clinical study (5 sites, 512 specimens) compared DxFLEX to Navios EX. Qualitative phenotypic agreement was 100% (PPA 100%, NPA 100%). Diagnostic accuracy against clinical outcome showed 70% sensitivity and 94% specificity across all specimen types. Precision/reproducibility studies (repeatability, operator-to-operator, instrument-to-instrument) showed 100% qualitative agreement. Carryover was <1%.

## Technological Characteristics

Flow cytometer with 3 lasers (488nm, 638nm, 405nm) and 10 fluorescence channels using Avalanche Photodiodes (APDs). Features hydrodynamic focusing, automated carousel loader, and digital pulse processing. Connectivity via Ethernet. Software: CytExpert (acquisition) and Kaluza C (analysis). Complies with IEC 61326-1 and 61326-2-6 for EMC.

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- Navios EX Flow Cytometer ([K183592](/device/K183592.md))
- ClearLLab 10C Reagent System ([K183592](/device/K183592.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

INSTRUMENT ONLY

## I Background Information:

A 510(k) Number

K232600

B Applicant

Beckman Coulter, Inc

C Proprietary and Established Names

DxFLEX Flow Cytometer

ClearLLab 10C Panels (B, T, M1, M2)

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  OYE | Class II | 21 CFR 864.5220 - Automated differential cell counter | HE - Hematology  |
|  PWD | Class II | 21 CFR 864.7010 - Flow cytometric test system for hematopoietic neoplasms | HE - Hematology  |

## II Submission/Device Overview:

A Purpose for Submission:

New instrument for previously cleared assay

B Type of Test:

Immunophenotyping, qualitative, flow cytometric instrument

## III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

B Indication(s) for Use:

DxFLEX Flow Cytometer

The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to ten fluorescent detection channels using three lasers (488 nm, 638 nm, and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.

The ClearLLab 10C Panels

The ClearLLab 10C Panels are intended for in vitro diagnostic use for qualitative identification of various cell populations by multiparameter immunophenotyping on the Navios, Navios EX and DxFLEX flow cytometers. These reagents are used as an aid in the differential diagnosis of hematologically abnormal patients having, or suspected of having, the following hematopoietic neoplasms: chronic leukemia, acute leukemia, non-Hodgkin lymphoma, myeloma, myelodysplastic syndrome (MDS), and/or myeloproliferative neoplasms (MPN). The reagents can be used with peripheral whole blood (collected in K2EDTA, Acid Citrate Dextrose (ACD) or Heparin), bone marrow (collected in K2EDTA, ACD or Heparin) and lymph node specimens. Interpretation of the results should be confirmed by a pathologist or equivalent professional in conjunction with other clinical and laboratory findings. These reagents provide multiparameter, qualitative results for the surface antigens listed below:

- ClearLLab 10C B Cell Tube: Kappa, Lambda, CD10, CD5, CD200, CD34, CD38, CD20, CD19, CD45
- ClearLLab 10C T Cell Tube: TCRγδ, CD4, CD2, CD56, CD5, CD34, CD3, CD8, CD7, CD45
- ClearLLab 10C M1 Cell Tube: CD16, CD7, CD10, CD13, CD64, CD34, CD14, HLA-DR, CD11b, CD45
- ClearLLab 10C M2 Cell Tube: CD15, CD123, CD117, CD13, CD33, CD34, CD38, HLA-DR, CD19, CD45

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only
For in vitro diagnostic use.

IV Device/System Characteristics:

A Device Description:

1. Device Description:

The DxFLEX Flow Cytometer uses flow cytometric principles to determine qualitative measurements of biological and physical properties of cells and other particles when the cells pass through the laser beam(s) in single file.

K232600 - Page 2 of 27

{2}

The DxFlex Flow cytometer is being cleared for use with the ClearLLab 10C Reagent System which was cleared on the Navios EX Flow cytometer (K183592). The ClearLLab 10C Reagent System on the DxFLEX Flow Cytometer is comprised of the following:

- DxFLEX Flow Cytometer [3 laser/10 color configurations]
- DxFLEX Daily QC Fluorospheres
- ClearLLab 10C Panels [B, T, M1 and M2]
- ClearLLab Compensation Kit
- ClearLLab Compensation Beads
- ClearLLab Control Cells, normal and abnormal
- IOTest 3 Fixative Solution [Existing] • IOTest 3 Lysing Solution
- Kaluza C Software

The DxFLEX Flow Cytometer includes ten fluorescent detection channels (FL1-FL10) and three laser configurations (blue, red and violet).

The ClearLLab 10C Reagent System on the DxFLEX Flow Cytometer includes the CytExpert for DxFLEX Software (i.e., acquisition software) and the Kaluza C Analysis Software (i.e., analysis software) (cleared in K183592 (version 1.1)).

As with the workflow on the predicate system (K183592), listmode (LMD) data analysis is performed manually using the Kaluza C Analysis Software. This Analysis Software package is supplied separately from the DxFLEX and is installed on an independent computer workstation for off-line analysis of listmode files generated on the flow cytometer with the associated reagents and cytometer system software package, including Control Cell Quality Control (QC) data and sample data analysis. The CytExpert for DxFLEX software is used to run the DxFLEX Daily QC Fluorospheres and Compensation products, Kaluza C Software is a software tool designed to work with *.fcs and *.lmd files generated from flow cytometers. Preset Kaluza C analysis templates for the ClearLLab 10C reagent system will be provided. The total WBC gate is defined in the histogram of SS vs. CD45-KO525-A by the inclusion of all CD45+ events with low, medium, and high Side Scatter after the exclusion of the cells that were compromised and/or aggregated. A subsequent histogram defines specific leukocyte subsets such as lymphocytes, monocytes, granulocytes. The low SS/bright CD45+ population identifies lymphocytes. Applicable markers are then displayed in subsequent histograms gated on lymphocytes. This process is repeated for mid SS/medium CD45+ populations to identify monocytes and for high SS/medium CD45+ to identify granulocytes

As acquisition software, CytExpert.exe enables the user to acquire data from the DxFLEX flow cytometer and allows the user to analyze, display, print and export acquired data. The major features of the software are as follows:

- System Startup
- Instrument Quality Control and Standardization
- Data Acquisition
- Sample Analysis
- Compensation
- Data Review
- Report Generation
- Daily Clean

K232600 - Page 3 of 27

{3}

Kaluza C software is used for post-acquisition analysis and reporting.

2. Principles of Operation:

The DxFLEX flow cytometer aligns cells in suspension by hydrodynamic focusing and passes them through a laser beam, one cell at a time. The scattered and fluoresced light that emanates from these cells as they intersect the laser beam is collected by Avalanche Photodiode detectors and transduced to an electronic pulse and counted by the instrument's computer. This test depends on the ability of a monoclonal or polyclonal antibody to bind to the surface of cells expressing discrete antigenic determinants. Specific cell staining is accomplished by incubating specimens prepared for staining with the appropriate antibody reagent. After sample preparation, the specimens are acquired on the QC controlled flow cytometer with manual gating.

B Instrument Description Information:

1. Instrument Name:

DxFLEX Flow Cytometer

2. Specimen Identification:

Barcode reader or manual entry

3. Specimen Sampling and Handling:

The automated sample loader for the instrument uses a carousel that holds thirty-two 12 x 75-mm test tubes and includes a barcode reader.

4. Calibration:

Fluorescence intensity is calibrated and confirmed using the DxFLEX Daily QC Fluorospheres with lot specific target values.

5. Quality Control:

The DxFLEX QC process verifies important system functions. The system verifies or automatically adjusts the following parameters:

- Hardware configuration
- Laser power of each individual laser
- Laser delays parameters
- Gain settings

Beckman Coulter recommends performing QC on a daily basis.

V Substantial Equivalence Information:

K232600 - Page 4 of 27

{4}

K232600 - Page 5 of 27

A Predicate Device Name(s):

ClearLLab 10C Panels (B, T, M1, M2)
Navios Flow Cytometer
Navios EX Flow Cytometer

B Predicate 510(k) Number(s):

K183592

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K232600
(Candidate) | K183592
(Predicate)  |
| --- | --- | --- |
|  Device Trade Name | DxFLEX Flow Cytometer | Navios EX Flow Cytometer  |
|  General Device Characteristic Similarities  |   |   |
|  Intended Use/Indications For Use | The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to ten fluorescent detection channels using three lasers (488 nm, 638 nm, and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument. | The Navios EX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to ten fluorescent detection channels using three lasers (488 nm, 638 nm, and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.  |
|  Sample Preparation | Off-board sample preparation following instructions provided with cleared antibody reagent | Same  |
|  Sample Presentation | Sample prepared in dry reagent tube presented in the same tube | Same  |
|  Resuspension of Prepared Sample Prior to Introduction to System | Prepared sample is vortex mixed | Same  |
|  Sample Introduction | • Automated presentation with Multi-tube Carousel Loader (MCL) from 32 test tube capacity carousel.
• Manual presentation into a tube location on a MCL | Same  |
|  Sample Aspiration Probe | Adjustable (both Beckman Coulter field service personnel and user) | Adjustable (by Beckman Coulter field service personnel)  |

{5}

K232600 - Page 6 of 27
|  Sample Identification | Bar-code reading of carousel position and labeled sample tube. User may also identify samples based on carousel location with a worklist. | Same  |
| --- | --- | --- |
|  Lasers / Driver Boards | Blue (488 nm)
• Diode Pumped Solid State (DPSS), 50mW
• 488 nm laser driver board | Blue (488 nm)
• Diode Pumped Solid State (DPSS), 55mW
• 488 nm laser driver board  |
|   |  Red (638 nm)
• Diode Pumped Solid State (DPSS), 50 mW
• 638 nm laser driver board | Same  |
|   |  Violet (405 nm)
• Diode Pumped Solid State (DPSS), 85 mW
• 405 nm laser driver board | Violet (405 nm)
• Diode Pumped Solid State (DPSS), 80 mW
• 405 nm laser driver board  |
|  Forward Angle Light Scatter | • Forward scatter mask
• Band-pass filter passes only 488nm light
• Diode detector converts light into electrical signals | Same  |
|  Color Separation | Fluorescent signal is separated into desired components with dichroic filters. | Same  |
|  Reagents Used for Compensation and Sample Preparation | ClearLLab Compensation Kit and Compensation Beads will be used with the ClearLLab 10C application. | Same, except without labeling changes to include DxFLEX Flow Cytometer only  |
|  Analysis Software | Kaluza C data analysis software to be used with the ClearLLab 10C application and to be used off-line. Analysis protocols are specific for ClearLLab Reagents on DxFLEX Flow Cytometer | Same, except analysis protocols are instrument specific  |
|  Data Reporting | Report, Plots and Statistics printouts | Same  |
|  Cleaning Cycle Between Samples | Executed with Sheath Fluid ensuring carryover specification is met | Same  |
|  Quality Control Techniques | Daily Instrument Checks Supplied Controls for ClearLLab 10C application | Same  |
|  Cleanse Cycle | Cleaning cycle performed with FlowClean cleaning reagent as part of the daily shutdown process, before and after running samples with vital dyes that stain the tubing, and as part of troubleshooting. (Additional cleaning cycle required after running | Same, except no additional cleaning cycle required after running Daily QC Fluorospheres  |

{6}

K232600 - Page 7 of 27
|   | DxFLEX Daily QC Fluorospheres) |   |
| --- | --- | --- |
|  Process Controls | ClearLLab Control Cells (Normal and Abnormal) except with labeling changes to include DxFLEX Flow Cytometer and assay ranges specific for the DxFLEX Flow Cytometer | ClearLLab Control Cells (Normal and Abnormal)  |
|  Processor | PowerPC | Same  |
|  Workstation Interface | Ethernet | USB, Ethernet  |
|  Data Acquisition | Digital Pulse Processing | Same  |
|  Safety Features | Interlocks and mitigation of hazards via software and hardware controls | Same  |
|  Software Architecture | The acquisition software enables the user to acquire data from the instrument and to analyze, display, print and export acquired data. The embedded software resides in the instrument. It controls the instrument functionality including the multi-carousel loader (MCL) for sample introduction. The embedded software controls the instruments’ lasers, acquisition system and fluidics. The instrument fluidics aspirates the sample, and performs instrument maintenance functions such as startup/shutdown. The embedded software also captures and provides the data to the workstation for processing. | Same  |
|  Software Functionality | Allows Patient and Control data management – storage, review, reporting and system configuration management | Same  |
|  Data Processing | • Region/gates evaluation
• Statistics generation
• Export of results to MS Excel/PDF file

Incoming data stream formatted into per-parameter vectors for improved performance of lin/log data transformations and compensation calculations | • Region/gates evaluation
• Statistics generation
• Export of results to CSV/PDF file

Incoming data stream formatted into per-parameter vectors for improved performance of lin/log data transformations and compensation calculations  |
|  User Interface | Plots and reports | Same  |
|  Report Generator | Panel reports & QC | Same  |

{7}

K232600 - Page 8 of 27
|  Device & Predicate Device(s): | K232600
(Candidate) | K183592
(Predicate)  |
| --- | --- | --- |
|  Device Trade Name | DxFLEX Flow Cytometer | Navios EX Flow Cytometer  |
|  **General Device Characteristic Differences**  |   |   |
|  Forward Angle Light Scatter | Ellipsoidal mirror to focus the light | Single lens collimates the light  |
|  Optics | Aspherical and cylindrical lenses coupled to the flow cell | Spherical and cylindrical lenses coupled to the flow cell  |
|  Flow Cell | • Channel Size: 180 x 430 microns
• Sheath Flow Pressure: 18.8 kPa
• Optics module temperature management hardware: two thermoelectric devices at the top and bottom of the optics module box | • Channel Size: 140 x 460 microns
• Sheath Flow Pressure: 4 psi
• Thin wall to allow for gel coupling distance
• Optics module temperature management hardware: one thermoelectric device at the top and bottom of the optics module box  |
|  Aspiration Pathway | Aspiration pathway uses peristaltic pump for automatic presentation | Aspiration path uses compressor for automated and manual presentation  |
|  Fluorescence Collection | • Collection lens and light path
• Spherical mirror and aspheric Schmidt lens
• Gel coupled to flow cell
• Light passes through optical fiber to WDM area
• Light leaves the fiber and is collimated
• Collimated light is separated by filters into desired bands
APDs (Avalanche Photodiodes) detect light and convert to electrical signals | • Collection lens and light path:
• Spherical mirror and aspheric Schmidt lens
• No gel coupling, NA=1.2
• Light passes through optical fiber to PMT area
• Light leaves the fiber and is collimated
• Collimated light is separated by filters into desired bands
PMTs detect light and convert to electrical signals  |
|  Electronics | 25 MHz sampling | 40 MHz sampling  |
|  Detectors/Colors | ClearLLab 10C application will require the 10 APDs:
• Standard 5 APDs (FITC – PC7 off of 488 nm laser (blue)
• Standard 3 APDs (APC – APC-A750) off of 638 nm laser (red)
• Standard 2 APDs (PB450 – KO525) off of 405 nm laser (violet) | • Standard 5 PMTs (FL1 – FL5) off of 488 nm laser (blue)
• Standard 3 PMTs (FL6 – FL8) off of 638 nm laser (red)
• Standard 2 PMTs (FL9 – FL10) off of 405 nm laser (violet)  |

{8}

K232600 - Page 9 of 27
|  Maximum Parameter Detectors | Twelve (Forward Scatter, Side Scatter and 10 fluorescence channels) | Twelve (Forward Scatter, Side Scatter and 10 fluorescence channels)  |
| --- | --- | --- |
|  Optical Filters | Optical Filters (all band pass):
• FITC 525/40
• PE 585/42
• ECD 610/20
• PC5.5 690/50
• PC7 780/60
• APC-A750 780/60
• APC 660/10
• APC-700 712/25
• PB450 450/45
• KO525 525/40 | Optical Filters: (SP = Short Pass, BP = Band Pass, LP = Long Pass)
• FITC 550 SP - 525 BP
• PE 595 SP- 575/30 BP
• ECD 655 SP - 614/20 BP
• PC5.5 730 SP - 695/30 BP
• PC7 755 LP
• APC 710 SP – 660/20 BP
• APC-A700 750 SP –725/20 BP
• APC-A750 755 LP
• PB 480 SP – 450/50 BP
• KRO 550/40 BP  |
|  Assay Controls and Calibrators | DxFLEX Daily QC Fluorospheres have target values assigned to monitor optical alignment and fluidics and to aid in standardization. | Flow-Check Pro Fluorospheres
Flow-Set Pro Fluorospheres  |
|  Programming Language | C#
C++ | C++  |
|  Real-time Operating System | N/A | VxWorks  |
|  Standardization | Automatic voltage/gain adjustments | Automatic gain calibration using target value  |
|  Control Software | CytExpert for DxFLEX software enables the user to acquire data from the DxFLEX flow cytometer instrument and to analyze, display, print and export acquired fcs file.
CytExpert for DxFLEX controls the instrument functionality including the multi-carousel loader (MCL) for sample introduction.
CytExpert for DxFLEX controls the instruments’ lasers, acquisition system and fluidics. The instrument fluidics aspirates the sample and performs instrument maintenance functions such as startup/shutdown.
CytExpert for DxFLEX also captures and provides the data to the workstation for processing.
Communication protocol | The acquisition software enables the user to acquire data from the Navios EX Flow Cytometer and to analyze, display, print and export acquired listmode data. The embedded software resides in the Navios EX instrument. It controls the instrument functionality including the multi-carousel loader (MCL) for sample introduction.
The embedded software controls the instruments’ lasers, acquisition system and fluidics. The instrument fluidics aspirates the sample and performs instrument maintenance functions such as startup/shutdown.
The embedded software also captures and provides the data to the workstation for processing.
Communication protocol  |
|  Data Processing | Region and Gate processing
Statistics Generation
Export to CSV | Region and Gate processing
Statistics Generation
Export to Excel  |

{9}

VI Standards/Guidance Documents Referenced:

- CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures
- CLSI EP09c 3rd Edition, Measurement Procedure Comparison and Bias Estimation Using Patient Samples
- CLSI EP12 3rd Edition, Evaluation of Qualitative, Binary Output Examination Performance
- CLSI EP43-A2, Clinical Flow Cytometric Analysis of Neoplastic Hematolymphoid Cells
- CLSI H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzers
- IEC 61326-1 Edition 3.0 2020-10, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- IEC 61326-2-6 Edition 3.0 2020-10, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

Precision studies were conducted according to the methodology presented in CLSI EP12-Ed3 and CLSI EP05-A3.

a. Precision (repeatability):

The repeatability of normal and abnormal clinical specimens representing clinical matrices and anticoagulants was evaluated with the ClearLLab 10C Panels on the DxFLEX Flow Cytometer at three sites. Sixty-four specimens across the three specimen types, i.e., whole blood (WB), bone marrow (BM), and lymph node (LN), were evaluated, with two excluded (due to specimen age and operator error). Each site employed a different anticoagulant for specimens, according to their current clinical practice: K2EDTA, acid citrate dextrose (ACD), or heparin.

Unless otherwise specified, a minimum of two specimens per specimen type per panel for each anticoagulant (site) was included. For LN specimens, there was no minimum requirement for the T panel (due to the low incidence of T lineage abnormalities), nor for M1 and M2 (due to the development property of myeloid cells and the pathogenesis of Myeloid abnormalities). Each specimen per panel was analyzed twice each day. There were six sample preparations in the morning and six sample preparations in the afternoon, resulting in a total of 12 replicates per specimen per panel. From 768 initial total replicates, 24 were excluded for reasons of specimen quality or operator error. Both quantitative and qualitative assessment were applied to the precision assessment where the presence or absence of an abnormal phenotype was reported.

K232600 - Page 10 of 27

{10}

All results showed 100% qualitative agreement across all sites, panels, anticoagulants, specimen types, and phenotypes.

For repeatability evaluation, pre-defined hematopoietic lineage populations were defined as in the table below:

|  Population definitions for evaluation of multi-site reproducibility of ClearLLab 10C  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Panel | Normal Population | Immunophenotype of Normal Population | Specimen Type | Abnormal Markers  |
|  B | Mature κ⁺
B cells | CD45bright SSlow CD19⁺ CD20⁺
Kappa⁺ CD38dim CD200⁺ CD10⁻
CD5⁻ CD34⁻ | WB | CD5
CD10
CD34  |
|   |   |   |  BM  |   |
|   |   |   |  LN  |   |
|   |  Mature λ⁺
B cells | CD45bright SSlow CD19⁺ CD20⁺
Lambda⁺ CD38dim CD200⁺ CD10⁻
CD5⁻ CD34⁻ | WB  |   |
|   |   |   |  BM  |   |
|   |   |   |  LN  |   |
|  T | Helper T cells | CD45bright SSlow CD2⁺ CD3⁺
CD4⁺ CD8⁻ CD5⁺ CD7⁺ CD56⁻
TCRγδ⁻ CD34⁻ | WB | CD34  |
|   |   |   |  BM  |   |
|   |   |   |  LN  |   |
|   |  Cytotoxic T cells | CD45bright SSlow CD2⁺ CD3⁺
CD8⁺ CD4⁻ CD5⁺ CD7⁺ CD56⁻
TCRγδ⁻ CD34⁻ | WB  |   |
|   |   |   |  BM  |   |
|   |   |   |  LN  |   |
|   |  NK cells | CD45bright SSClow CD56⁺ CD2⁺ CD3⁻
CD4⁻ CD8⁺/⁻ CD5⁻ CD7⁺
TCRγδ⁻ CD34⁻ | WB  |   |
|   |   |   |  BM  |   |
|   |   |   |  LN  |   |
|   |  γδ T cells | CD45bright SSlow CD2⁺ CD3bright
TCRγδ⁺ CD4⁻ CD8⁻ CD5⁺ CD7⁺
CD56⁻ CD34⁻ | WB  |   |
|   |   |   |  BM  |   |
|   |   |   |  LN  |   |
|  M1 | Neutrophils | CD45dim SShigh CD16⁺ CD10⁺
CD13⁺ CD11b⁺ CD7⁻ CD64⁻ CD14⁻
CD34⁻ HLA-DR⁻ | WB | CD7
CD34  |
|   |   |   |  BM  |   |
|   |  Monocytes | CD45med SSmed CD13⁺ CD64⁺
CD14⁺ HLA-DR⁺ CD11b⁺ CD16⁻
CD7⁻
CD10⁻ CD34⁻ | WB  |   |
|   |   |   |  BM  |   |
|  M2 | Neutrophils | CD45dim SShigh CD15⁺ CD13⁺
CD33⁺/⁻ CD123⁻ CD117⁻ CD34⁻
CD19⁻ CD38⁻ HLA-DR⁻ CD34⁻ | WB | CD19
CD34
CD117  |
|   |   |   |  BM  |   |
|   |  Monocytes | CD45med SSmed CD13⁺ CD33⁺
CD38⁺ HLA-DR⁺ CD15⁻ CD123⁻
CD19⁻ CD34⁻ CD117⁻ | WB  |   |
|   |   |   |  BM  |   |
|   |  Basophils | CD45dim CD123bright CD38bright
CD13⁺ CD33⁺ CD15⁻ CD117⁻
CD34⁻ HLA-DR⁻ CD19⁻ | WB  |   |
|   |   |   |  BM  |   |

The results based on qualitative agreement are summarized in the following tables:

K232600 - Page 11 of 27

{11}

|  Qualitative Agreement – by Specimen, by Site  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Site | Panel | Anticoagulant | Specimen Type | Phenotype | N  |
|  1 | B | ACD | BM | Abnormal (+) | 12  |
|  1 | B | ACD | BM | Abnormal (+) | 12  |
|  1 | B | ACD | BM | Normal (-) | 12  |
|  1 | B | n/a | LN | Abnormal (+) | 12  |
|  1 | B | n/a | LN | Abnormal (+) | 12  |
|  1 | B | n/a | LN | Normal (-) | 12  |
|  1 | B | ACD | WB | Abnormal (+) | 12  |
|  1 | B | ACD | WB | Normal (-) | 12  |
|  2 | B | K2EDTA | BM | Abnormal (+) | 12  |
|  2 | B | K2EDTA | BM | Abnormal (+) | 12  |
|  2 | B | K2EDTA | BM | Normal (-) | 12  |
|  2 | B | n/a | LN | Abnormal (+) | 12  |
|  2 | B | n/a | LN | Normal (-) | 12  |
|  2 | B | K2EDTA | WB | Normal (-) | 12  |
|  2 | B | K2EDTA | WB | Abnormal (+) | 12  |
|  3 | B | Heparin | BM | Abnormal (+) | 12  |
|  3 | B | Heparin | BM | Normal (-) | 12  |
|  3 | B | Heparin | WB | Abnormal (+) | 12  |
|  3 | B | Heparin | WB | Abnormal (+) | 12  |
|  3 | B | Heparin | WB | Normal (-) | 12  |
|  3 | B | Heparin | LN | Abnormal (+) | 12  |
|  3 | B | Heparin | LN | Abnormal (+) | 12  |
|  1 | M1 | ACD | BM | Abnormal (+) | 12  |
|  1 | M1 | ACD | BM | Normal (-) | 12  |
|  1 | M1 | ACD | WB | Abnormal (+) | 12  |
|  1 | M1 | ACD | WB | Normal (-) | 12  |
|  2 | M1 | K2EDTA | BM | Normal (-) | 12  |
|  2 | M1 | K2EDTA | BM | Abnormal (+) | 12  |
|  2 | M1 | K2EDTA | WB | Abnormal (+) | 12  |
|  2 | M1 | K2EDTA | WB | Abnormal (+) | 12  |
|  2 | M1 | K2EDTA | WB | Normal (-) | 12  |
|  3 | M1 | Heparin | BM | Abnormal (+) | 12  |
|  3 | M1 | Heparin | BM | Normal (-) | 12  |
|  3 | M1 | Heparin | WB | Abnormal (+) | 12  |
|  3 | M1 | Heparin | WB | Normal (-) | 12  |
|  1 | M2 | ACD | BM | Abnormal (+) | 12  |
|  1 | M2 | ACD | BM | Normal (-) | 12  |
|  1 | M2 | ACD | WB | Abnormal (+) | 12  |
|  1 | M2 | ACD | WB | Normal (-) | 12  |
|  2 | M2 | K2EDTA | BM | Abnormal (+) | 12  |
|  2 | M2 | K2EDTA | BM | Abnormal (+) | 12  |
|  2 | M2 | K2EDTA | BM | Normal (-) | 12  |
|  2 | M2 | K2EDTA | WB | Abnormal (+) | 12  |

K232600 - Page 12 of 27

{12}

|  Qualitative Agreement – by Specimen, by Site  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Site | Panel | Anticoagulant | Specimen Type | Phenotype | N  |
|  3 | M2 | Heparin | BM | Abnormal (+) | 12  |
|  3 | M2 | Heparin | BM | Abnormal (+) | 12  |
|  3 | M2 | Heparin | WB | Normal (-) | 12  |
|  3 | M2 | Heparin | WB | Abnormal (+) | 12  |
|  1 | T | ACD | BM | Abnormal (+) | 12  |
|  1 | T | ACD | BM | Normal (-) | 12  |
|  1 | T | ACD | WB | Normal (-) | 12  |
|  1 | T | ACD | WB | Normal (-) | 12  |
|  2 | T | K2EDTA | BM | Normal (-) | 12  |
|  2 | T | K2EDTA | BM | Normal (-) | 12  |
|  2 | T | K2EDTA | WB | Abnormal (+) | 12  |
|  2 | T | K2EDTA | WB | Normal (-) | 12  |
|  2 | T | n/a | LN | Normal (-) | 12  |
|  2 | T | n/a | LN | Normal (-) | 12  |
|  3 | T | Heparin | BM | Normal (-) | 12  |
|  3 | T | Heparin | BM | Abnormal (+) | 12  |
|  3 | T | Heparin | WB | Normal (-) | 12  |
|  3 | T | Heparin | WB | Normal (-) | 12  |
|  Qualitative Agreement – by Anticoagulant and Specimen Type (Sites and Specimens combined)  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Panel | Anticoagulant | Specimen Type | Phenotype | N  |
|  B | ACD | BM | Normal (-) | 12  |
|  B | ACD | BM | Abnormal (+) | 12  |
|  B | ACD | BM | Both (+/-) | 24  |
|  B | ACD | WB | Normal (-) | 12  |
|  B | ACD | WB | Abnormal (+) | 12  |
|  B | ACD | WB | Both (+/-) | 24  |
|  B | Heparin | BM | Normal (-) | 12  |
|  B | Heparin | BM | Abnormal (+) | 12  |
|  B | Heparin | BM | Both (+/-) | 24  |
|  B | Heparin | WB | Normal (-) | 12  |
|  B | Heparin | WB | Abnormal (+) | 24  |
|  B | Heparin | WB | Both (+/-) | 36  |
|  B | K2EDTA | BM | Normal (-) | 12  |
|  B | K2EDTA | BM | Abnormal (+) | 24  |
|  B | K2EDTA | BM | Both (+/-) | 36  |
|  B | K2EDTA | WB | Normal (-) | 12  |
|  B | K2EDTA | WB | Abnormal (+) | 12  |
|  B | K2EDTA | WB | Both (+/-) | 24  |
|  B | n/a | LN | Normal (-) | 24  |
|  B | n/a | LN | Abnormal (+) | 60  |
|  B | n/a | LN | Both (+/-) | 84  |
|  M1 | ACD | BM | Normal (-) | 12  |

K232600 - Page 13 of 27

{13}

K232600 - Page 14 of 27

|  Qualitative Agreement – by Anticoagulant and Specimen Type
(Sites and Specimens combined)  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Panel | Anticoagulant | Specimen Type | Phenotype | N  |
|  M1 | ACD | BM | Abnormal (+) | 12  |
|  M1 | ACD | BM | Both (+/-) | 24  |
|  M1 | ACD | WB | Normal (-) | 12  |
|  M1 | ACD | WB | Abnormal (+) | 12  |
|  M1 | ACD | WB | Both (+/-) | 24  |
|  M1 | Heparin | BM | Normal (-) | 12  |
|  M1 | Heparin | BM | Abnormal (+) | 12  |
|  M1 | Heparin | BM | Both (+/-) | 24  |
|  M1 | Heparin | WB | Normal (-) | 12  |
|  M1 | Heparin | WB | Abnormal (+) | 12  |
|  M1 | Heparin | WB | Both (+/-) | 24  |
|  M1 | K_{2}EDTA | BM | Normal (-) | 12  |
|  M1 | K_{2}EDTA | BM | Abnormal (+) | 12  |
|  M1 | K_{2}EDTA | BM | Both (+/-) | 24  |
|  M1 | K_{2}EDTA | WB | Normal (-) | 12  |
|  M1 | K_{2}EDTA | WB | Abnormal (+) | 24  |
|  M1 | K_{2}EDTA | WB | Both (+/-) | 36  |
|  M2 | ACD | BM | Normal (-) | 12  |
|  M2 | ACD | BM | Abnormal (+) | 12  |
|  M2 | ACD | BM | Both (+/-) | 24  |
|  M2 | ACD | WB | Normal (-) | 12  |
|  M2 | ACD | WB | Abnormal (+) | 12  |
|  M2 | ACD | WB | Both (+/-) | 24  |
|  M2 | Heparin | BM | Normal (-) | 12  |
|  M2 | Heparin | BM | Abnormal (+) | 12  |
|  M2 | Heparin | BM | Both (+/-) | 24  |
|  M2 | Heparin | WB | Normal (-) | 12  |
|  M2 | Heparin | WB | Abnormal (+) | 12  |
|  M2 | Heparin | WB | Both (+/-) | 24  |
|  M2 | K_{2}EDTA | BM | Normal (-) | 12  |
|  M2 | K_{2}EDTA | BM | Abnormal (+) | 24  |
|  M2 | K_{2}EDTA | BM | Both (+/-) | 36  |
|  M2 | K_{2}EDTA | WB | Normal (-) | 24  |
|  M2 | K_{2}EDTA | WB | Abnormal (+) | 12  |
|  M2 | K_{2}EDTA | WB | Both (+/-) | 36  |
|  T | ACD | BM | Normal (-) | 12  |
|  T | ACD | BM | Abnormal (+) | 12  |
|  T | ACD | BM | Both (+/-) | 24  |
|  T | ACD | WB | Normal (-) | 24  |
|  T | Heparin | BM | Normal (-) | 12  |
|  T | Heparin | BM | Abnormal (+) | 12  |
|  T | Heparin | BM | Both (+/-) | 24  |
|  T | Heparin | WB | Normal (-) | 24  |
|  T | K_{2}EDTA | BM | Normal (-) | 24  |

{14}

|  Qualitative Agreement – by Anticoagulant and Specimen Type (Sites and Specimens combined)  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Panel | Anticoagulant | Specimen Type | Phenotype | N  |
|  T | K2EDTA | WB | Normal (-) | 12  |
|  T | K2EDTA | WB | Abnormal (+) | 12  |
|  T | K2EDTA | WB | Both (+/-) | 24  |
|  T | n/a | LN | Normal (-) | 24  |

Results based on quantitative analysis are summarized in the table below:

|  Quantitative Results – by Specimen, by Site  |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Panel | Site | Specimen Type | Anti-coagulant | Population | N | Mean (%) | Repeatability |   | Between-Run |   | Within-Sample  |
|   |   |   |   |   |   |   |  SD | %CV | SD | %CV | SD  |
|  B | 1 | LN | N/A | CD19+CD5+ | 12 | 91.96 | 0.62 | 0.67 | 0.82 | 0.89 | 1.02  |
|  B | 1 | LN | N/A | CD19+CD5+ | 12 | 32.64 | 1.37 | 4.18 | 1.40 | 4.30 | 1.96  |
|  B | 1 | LN | N/A | Mature κ+ B cells | 12 | 66.93 | 1.82 | 2.72 | 0.00 | 0.00 | 1.82  |
|  B | 1 | LN | N/A | Mature λ+ B cells | 12 | 30.19 | 1.23 | 4.07 | 0.64 | 2.13 | 1.39  |
|  B | 1 | BM | ACD | Mature κ+ B cells | 12 | 42.04 | 1.45 | 3.46 | 0.00 | 0.00 | 1.45  |
|  B | 1 | BM | ACD | Mature λ+ B cells | 12 | 27.88 | 1.06 | 3.80 | 0.00 | 0.00 | 1.06  |
|  B | 1 | WB | ACD | Mature κ+ B cells | 12 | 64.03 | 0.61 | 0.95 | 0.00 | 0.00 | 0.61  |
|  B | 1 | WB | ACD | Mature λ+ B cells | 12 | 35.54 | 0.71 | 1.99 | 0.00 | 0.00 | 0.71  |
|  B | 1 | WB | ACD | CD19+CD5+ | 12 | 98.66 | 0.38 | 0.39 | 0.00 | 0.00 | 0.38  |
|  B | 1 | BM | ACD | CD19+CD5+ | 12 | 95.01 | 0.81 | 0.86 | 1.14 | 1.20 | 1.40  |
|  B | 2 | WB | Heparin | CD19+CD5+ | 12 | 98.81 | 0.09 | 0.09 | 0.00 | 0.00 | 0.03  |
|  B | 2 | BM | Heparin | CD19+CD5+ | 12 | 99.37 | 0.03 | 0.04 | 0.00 | 0.00 | 0.02  |
|  B | 2 | WB | Heparin | Mature λ+ B cells | 12 | 99.81 | 0.02 | 0.02 | 0.00 | 0.00 | 2.64  |
|  B | 2 | WB | Heparin | Mature κ+ B cells | 12 | 60.28 | 2.64 | 4.38 | 0.00 | 0.00 | 2.57  |
|  B | 2 | WB | Heparin | Mature λ+ B cells | 12 | 39.49 | 2.57 | 6.52 | 0.00 | 0.00 | 0.79  |
|  B | 2 | BM | Heparin | Mature κ+ B cells | 12 | 41.46 | 0.79 | 1.92 | 0.12 | 0.41 | 0.81  |
|  B | 2 | BM | Heparin | Mature λ+ B cells | 12 | 30.03 | 0.80 | 2.67 | 0.12 | 0.41 | 0.81  |
|  B | 2 | LN | N/A | CD19+CD5+ | 12 | 7.59 | 0.51 | 6.78 | 0.00 | 0.00 | 0.51  |
|  B | 2 | LN | N/A | CD19+CD10+ | 12 | 39.65 | 1.14 | 2.87 | 0.48 | 1.20 | 1.23  |
|  B | 2 | LN | N/A | CD19+CD5+ | 12 | 39.90 | 1.39 | 3.50 | 1.20 | 3.00 | 1.84  |
|  B | 3 | WB | K2EDTA | Mature κ+ B cells | 12 | 71.02 | 0.63 | 0.89 | 0.00 | 0.00 | 0.63  |
|  B | 3 | WB | K2EDTA | Mature λ+ B cells | 12 | 28.83 | 0.63 | 2.19 | 0.00 | 0.00 | 0.63  |
|  B | 3 | BM | K2EDTA | Mature κ+ B cells | 12 | 67.86 | 0.82 | 1.22 | 2.06 | 3.04 | 2.22  |
|  B | 3 | BM | K2EDTA | Mature λ+ B cells | 12 | 31.78 | 0.87 | 2.73 | 2.07 | 6.53 | 2.25  |
|  B | 3 | LN | NA | Mature κ+ B cells | 12 | 48.60 | 1.40 | 2.88 | 0.00 | 0.00 | 1.40  |
|  B | 3 | LN | NA | Mature λ+ B cells | 12 | 50.15 | 1.44 | 2.88 | 0.00 | 0.00 | 1.44  |
|  B | 3 | BM | K2EDTA | CD19+CD5+ | 12 | 99.42 | 0.09 | 0.09 | 0.06 | 0.06 | 0.11  |
|  B | 3 | WB | K2EDTA | CD19+CD10+ | 12 | 98.49 | 0.12 | 0.12 | 0.31 | 0.31 | 0.33  |
|  B | 3 | LN | N/A | CD19+CD5+ | 12 | 78.61 | 1.36 | 1.73 | 0.00 | 0.00 | 1.36  |
|  B | 3 | BM | K2EDTA | CD19+CD5+ | 12 | 99.51 | 0.03 | 0.03 | 0.00 | 0.00 | 0.03  |
|  M1 | 1 | BM | ACD | CD16+Neutrophiles | 12 | 19.39 | 0.58 | 2.99 | 0.74 | 3.82 | 0.94  |
|  M1 | 1 | BM | ACD | CD64+Monocytes | 12 | 7.24 | 0.17 | 2.32 | 0.12 | 1.63 | 0.21  |
|  M1 | 1 | WB | ACD | CD16+Neutrophiles | 12 | 28.62 | 1.57 | 5.49 | 1.15 | 4.02 | 1.95  |
|  M1 | 1 | WB | ACD | CD64+Monocytes | 12 | 2.82 | 0.18 | 6.51 | 0.09 | 3.11 | 0.20  |
|  M1 | 1 | WB | ACD | CD34-CD7+ | 12 | 2.21 | 0.08 | 3.55 | 0.00 | 0.00 | 0.08  |

K232600 - Page 15 of 27

{15}

Quantitative Results – by Specimen, by Site

|  Panel | Site | Specimen Type | Anti-coagulant | Population | N | Mean (%) | Repeatability |   | Between-Run |   | Within-Sample  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |   |   |   |   |  SD | %CV | SD | %CV | SD | %CV  |
|  M1 | 1 | BM | ACD | CD34-CD7+ | 12 | 40.89 | 0.67 | 1.63 | 1.03 | 2.53 | 1.23 | 3.01  |
|  M1 | 2 | BM | Heparin | CD34-CD7+ | 12 | 11.94 | 0.38 | 3.22 | 0.48 | 4.00 | 0.61 | 5.14  |
|  M1 | 2 | WB | Heparin | CD16+Neutrophiles | 12 | 70.20 | 1.31 | 1.87 | 0.60 | 0.85 | 1.44 | 2.05  |
|  M1 | 2 | WB | Heparin | CD64+Monocytes | 12 | 3.72 | 0.16 | 4.18 | 0.16 | 4.33 | 0.22 | 6.02  |
|  M1 | 2 | BM | Heparin | CD16+Neutrophiles | 12 | 31.90 | 1.21 | 3.78 | 0.81 | 2.53 | 1.45 | 4.55  |
|  M1 | 2 | BM | Heparin | CD64+Monocytes | 12 | 37.78 | 1.22 | 3.22 | 1.35 | 3.58 | 1.82 | 4.82  |
|  M1 | 2 | WB | Heparin | CD34-CD7+ | 12 | 36.44 | 0.32 | 0.87 | 0.13 | 0.36 | 0.34 | 0.94  |
|  M1 | 3 | BM | K2EDTA | CD16+Neutrophiles | 12 | 23.01 | 0.24 | 1.02 | 0.11 | 0.46 | 0.26 | 1.12  |
|  M1 | 3 | BM | K2EDTA | CD64+Monocytes | 12 | 4.20 | 0.15 | 3.54 | 0.26 | 6.17 | 0.30 | 7.11  |
|  M1 | 3 | WB | K2EDTA | CD34-CD7+ | 12 | 12.43 | 0.24 | 1.91 | 0.00 | 0.00 | 0.24 | 1.91  |
|  M1 | 3 | BM | K2EDTA | CD16+Neutrophiles | 12 | 64.05 | 0.41 | 0.63 | 0.36 | 0.56 | 0.54 | 0.85  |
|  M1 | 3 | WB | K2EDTA | CD64+Monocytes | 12 | 8.02 | 0.17 | 2.10 | 0.52 | 6.44 | 0.54 | 6.77  |
|  M1 | 3 | WB | K2EDTA | CD34-CD7+ | 12 | 69.35 | 0.36 | 0.52 | 1.42 | 2.04 | 1.46 | 2.11  |
|  M1 | 3 | BM | K2EDTA | CD34-CD7+ | 12 | 12.23 | 0.58 | 4.77 | 0.00 | 0.00 | 0.58 | 4.77  |
|  M2 | 1 | BM | ACD | CD117+ | 12 | 65.10 | 0.73 | 1.12 | 0.51 | 0.78 | 0.89 | 1.37  |
|  M2 | 1 | BM | ACD | CD123+ | 12 | 64.57 | 0.98 | 1.52 | 0.56 | 0.87 | 1.13 | 1.75  |
|  M2 | 1 | BM | ACD | CD15+Neutrophile | 12 | 57.41 | 1.02 | 1.78 | 0.25 | 0.44 | 1.05 | 1.83  |
|  M2 | 1 | BM | ACD | CD33+ Monocytes | 12 | 10.80 | 0.16 | 1.51 | 0.00 | 0.00 | 0.16 | 1.51  |
|  M2 | 1 | BM | ACD | CD15+Neutrophile | 12 | 34.68 | 2.19 | 6.32 | 0.00 | 0.00 | 2.19 | 6.32  |
|  M2 | 1 | BM | ACD | CD33+ Monocytes | 12 | 3.46 | 0.30 | 8.68 | 0.25 | 7.21 | 0.39 | 11.28  |
|  M2 | 1 | BM | ACD | CD117+ | 12 | 9.52 | 0.18 | 1.87 | 0.13 | 1.33 | 0.22 | 2.30  |
|  M2 | 1 | BM | ACD | CD34+CD123+ | 12 | 6.63 | 0.12 | 1.76 | 0.29 | 4.40 | 0.31 | 4.74  |
|  M2 | 2 | BM | Heparin | CD117+ | 12 | 10.79 | 0.26 | 2.38 | 0.55 | 5.12 | 0.61 | 5.65  |
|  M2 | 2 | BM | Heparin | CD34+CD123+ | 12 | 5.83 | 0.27 | 4.57 | 0.30 | 5.18 | 0.40 | 6.91  |
|  M2 | 2 | WB | Heparin | CD15+Neutrophile | 12 | 80.51 | 0.84 | 1.04 | 1.02 | 1.27 | 1.33 | 1.65  |
|  M2 | 2 | WB | Heparin | CD33+ Monocytes | 12 | 2.11 | 0.12 | 5.90 | 0.38 | 17.96 | 0.40 | 18.90*  |
|  M2 | 2 | BM | Heparin | CD15+Neutrophile | 12 | 81.53 | 0.49 | 0.60 | 0.18 | 0.22 | 0.52 | 0.64  |
|  M2 | 2 | BM | Heparin | CD33+ Monocytes | 12 | 4.59 | 0.17 | 3.75 | 0.28 | 6.05 | 0.33 | 7.12  |
|  M2 | 2 | WB | Heparin | CD117+ | 12 | 65.13 | 0.90 | 1.38 | 0.58 | 0.89 | 1.07 | 1.64  |
|  M2 | 2 | WB | Heparin | CD34+CD123+ | 12 | 78.57 | 0.53 | 0.68 | 0.00 | 0.00 | 0.53 | 0.68  |
|  M2 | 3 | BM | K2EDTA | CD15+Neutrophile | 12 | 67.45 | 0.78 | 1.16 | 2.00 | 2.96 | 2.15 | 3.18  |
|  M2 | 3 | BM | K2EDTA | CD33+ Monocytes | 12 | 5.50 | 0.18 | 3.27 | 0.24 | 4.41 | 0.30 | 5.49  |
|  M2 | 3 | WB | K2EDTA | CD117+ | 12 | 14.31 | 0.23 | 1.58 | 0.31 | 2.15 | 0.38 | 2.67  |
|  M2 | 3 | WB | K2EDTA | CD34+CD123+ | 12 | 11.80 | 0.25 | 2.09 | 0.13 | 1.13 | 0.28 | 2.37  |
|  M2 | 3 | BM | K2EDTA | CD34+CD123+ | 12 | 94.61 | 0.28 | 0.29 | 0.00 | 0.00 | 0.28 | 0.29  |
|  M2 | 3 | WB | K2EDTA | CD15+Neutrophile | 12 | 47.82 | 0.41 | 0.86 | 0.25 | 0.52 | 0.48 | 1.00  |
|  M2 | 3 | WB | K2EDTA | CD33+ Monocytes | 12 | 2.25 | 0.19 | 8.56 | 0.70 | 31.09 | 0.73 | 32.24^  |
|  M2 | 3 | BM | K2EDTA | CD117+ | 12 | 70.74 | 0.44 | 0.62 | 1.19 | 1.68 | 1.27 | 1.80  |
|  M2 | 3 | BM | K2EDTA | CD34+CD123+ | 12 | 69.12 | 0.43 | 0.62 | 1.78 | 2.58 | 1.83 | 2.65  |
|  M2 | 3 | WB | K2EDTA | CD15+Neutrophile | 12 | 73.72 | 0.71 | 0.97 | 0.00 | 0.00 | 0.71 | 0.97  |
|  M2 | 3 | WB | K2EDTA | CD33+ Monocytes | 12 | 7.06 | 0.20 | 2.77 | 0.16 | 2.25 | 0.25 | 3.57  |
|  T | 1 | BM | ACD | NK | 12 | 4.94 | 0.29 | 5.97 | 12 | 0.25 | 5.00 | 0.38  |
|  T | 1 | BM | ACD | γδ T cells | 12 | 28.54 | 0.54 | 1.90 | 12 | 0.00 | 0.00 | 0.54  |
|  T | 1 | BM | ACD | T Helper %Gated | 12 | 39.49 | 0.45 | 1.15 | 12 | 0.19 | 0.48 | 0.49  |
|  T | 1 | BM | ACD | Cytotoxic T cells | 12 | 30.52 | 0.50 | 1.65 | 12 | 0.37 | 1.22 | 0.63  |
|  T | 1 | WB | ACD | NK | 12 | 6.36 | 0.10 | 1.52 | 12 | 0.05 | 0.71 | 0.11  |

K232600 - Page 16 of 27

{16}

|  Quantitative Results – by Specimen, by Site  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Panel | Site | Specimen Type | Anti-coagulant | Population | N | Mean (%) | Repeatability |   | Between-Run |   | Within-Sample  |   |
|   |   |   |   |   |   |   |  SD | %CV | SD | %CV | SD | %CV  |
|  T | 1 | WB | ACD | γδ T cells | 12 | 2.21 | 0.16 | 7.09 | 12 | 0.17 | 7.52 | 0.23  |
|  T | 1 | WB | ACD | T Helper %Gated | 12 | 68.34 | 0.59 | 0.86 | 12 | 0.22 | 0.32 | 0.63  |
|  T | 1 | WB | ACD | Cytotoxic T cells | 12 | 23.80 | 0.49 | 2.06 | 12 | 0.44 | 1.83 | 0.66  |
|  T | 1 | BM | ACD | NK | 12 | 8.78 | 0.43 | 4.87 | 12 | 0.26 | 3.00 | 0.50  |
|  T | 1 | BM | ACD | γδ T cells | 12 | 6.99 | 0.21 | 3.02 | 12 | 0.12 | 1.69 | 0.24  |
|  T | 1 | BM | ACD | T Helper %Gated | 12 | 38.92 | 0.59 | 1.51 | 12 | 0.36 | 0.92 | 0.69  |
|  T | 1 | BM | ACD | Cytotoxic T cells | 12 | 49.85 | 0.64 | 1.28 | 12 | 0.00 | 0.00 | 0.64  |
|  T | 1 | WB | ACD | NK | 12 | 20.43 | 0.48 | 2.34 | 12 | 0.19 | 0.94 | 0.52  |
|  T | 1 | WB | ACD | T Helper %Gated | 12 | 39.15 | 0.48 | 1.23 | 12 | 0.00 | 0.00 | 0.48  |
|  T | 1 | WB | ACD | Cytotoxic T cells | 12 | 55.24 | 0.48 | 0.87 | 12 | 0.49 | 0.89 | 0.69  |
|  T | 2 | WB | Heparin | NK | 12 | 18.60 | 0.83 | 4.46 | 0.68 | 3.64 | 1.07 | 5.76  |
|  T | 2 | WB | Heparin | T Helper %Gated | 12 | 83.71 | 1.34 | 1.60 | 0.98 | 1.17 | 1.66 | 1.98  |
|  T | 2 | WB | Heparin | Cytotoxic T cells | 12 | 11.97 | 1.05 | 8.75 | 0.00 | 0.00 | 1.05 | 8.75  |
|  T | 2 | BM | Heparin | NK | 12 | 12.27 | 0.46 | 3.73 | 0.19 | 1.57 | 0.50 | 4.05  |
|  T | 2 | BM | Heparin | T Helper %Gated | 12 | 58.43 | 1.80 | 3.07 | 0.00 | 0.00 | 1.80 | 3.07  |
|  T | 2 | BM | Heparin | Cytotoxic T cells | 12 | 36.64 | 1.48 | 4.03 | 0.00 | 0.00 | 1.48 | 4.03  |
|  T | 2 | WB | Heparin | NK | 12 | 27.41 | 0.31 | 1.14 | 0.69 | 2.51 | 0.76 | 2.76  |
|  T | 2 | WB | Heparin | γδ T cells | 12 | 3.11 | 0.17 | 5.39 | 0.00 | 0.00 | 0.17 | 5.39  |
|  T | 2 | WB | Heparin | T Helper %Gated | 12 | 50.26 | 0.73 | 1.46 | 0.35 | 0.70 | 0.81 | 1.62  |
|  T | 2 | WB | Heparin | Cytotoxic T cells | 12 | 44.48 | 0.69 | 1.54 | 0.00 | 0.00 | 0.69 | 1.54  |
|  T | 2 | BM | Heparin | NK | 12 | 11.83 | 0.47 | 3.94 | 0.71 | 5.97 | 0.85 | 7.15  |
|  T | 2 | BM | Heparin | γδ T cells | 12 | 2.62 | 0.21 | 7.98 | 0.21 | 8.18 | 0.30 | 11.43  |
|  T | 2 | BM | Heparin | T Helper %Gated | 12 | 14.37 | 0.59 | 4.10 | 0.00 | 0.00 | 0.59 | 4.10  |
|  T | 2 | BM | Heparin | Cytotoxic T cells | 12 | 77.33 | 0.79 | 1.02 | 1.82 | 2.35 | 1.98 | 2.56  |
|  T | 3 | LN | NA | T Helper %Gated | 12 | 73.64 | 0.34 | 0.46 | 0.00 | 0.00 | 0.34 | 0.46  |
|  T | 3 | LN | NA | Cytotoxic T cells | 12 | 22.85 | 0.23 | 0.99 | 0.20 | 0.89 | 0.30 | 1.33  |
|  T | 3 | BM | K2EDTA | NK | 12 | 7.75 | 0.20 | 2.63 | 0.00 | 0.00 | 0.20 | 2.63  |
|  T | 3 | BM | K2EDTA | T Helper %Gated | 12 | 72.87 | 0.49 | 0.68 | 0.00 | 0.00 | 0.49 | 0.68  |
|  T | 3 | BM | K2EDTA | Cytotoxic T cells | 12 | 22.76 | 0.52 | 2.30 | 0.00 | 0.00 | 0.52 | 2.30  |
|  T | 3 | WB | K2EDTA | NK | 12 | 27.78 | 0.91 | 3.28 | 0.26 | 0.95 | 0.95 | 3.42  |
|  T | 3 | WB | K2EDTA | T Helper %Gated | 12 | 70.07 | 0.50 | 0.72 | 0.00 | 0.00 | 0.50 | 0.72  |
|  T | 3 | WB | K2EDTA | Cytotoxic T cells | 12 | 19.93 | 0.80 | 4.04 | 0.00 | 0.00 | 0.80 | 4.04  |
|  T | 3 | LN | NA | γδ T cells | 12 | 1.17 | 0.07 | 6.39 | 0.00 | 0.00 | 0.07 | 6.39  |
|  T | 3 | LN | NA | T Helper %Gated | 12 | 93.44 | 0.16 | 0.17 | 0.00 | 0.00 | 0.16 | 0.17  |
|  T | 3 | LN | NA | Cytotoxic T cells | 12 | 4.50 | 0.11 | 2.54 | 0.09 | 2.06 | 0.15 | 3.27  |
|  T | 3 | WB | K2EDTA | NK | 10 | 5.90 | 0.24 | 4.15 | 0.00 | 0.00 | 0.24 | 4.15  |
|  T | 3 | WB | K2EDTA | γδ T cells | 10 | 1.15 | 0.07 | 6.35 | 0.05 | 4.19 | 0.09 | 7.61  |
|  T | 3 | WB | K2EDTA | T Helper %Gated | 10 | 65.58 | 0.53 | 0.81 | 0.57 | 0.87 | 0.78 | 1.19  |
|  T | 3 | WB | K2EDTA | Cytotoxic T cells | 10 | 31.11 | 0.44 | 1.41 | 0.23 | 0.74 | 0.50 | 1.60  |
|  T | 3 | BM | K2EDTA | NK | 12 | 24.47 | 0.95 | 3.89 | 0.00 | 0.00 | 0.95 | 3.89  |
|  T | 3 | BM | K2EDTA | γδ T cells | 12 | 2.24 | 0.26 | 11.58 | 0.38 | 17.07 | 0.46 | 20.63&  |
|  T | 3 | BM | K2EDTA | T Helper %Gated | 12 | 55.28 | 1.03 | 1.86 | 0.00 | 0.00 | 1.03 | 1.86  |
|  T | 3 | BM | K2EDTA | Cytotoxic T cells | 12 | 38.60 | 0.92 | 2.39 | 0.37 | 0.97 | 1.00 | 2.58  |

* Sample is a whole blood specimen collected in K2EDTA anticoagulant that demonstrated %CV of 17.96% for between-runs and 18.90% for within-sample for CD33 marked Monocytes, as the CD33 marked Monocytes recovery in this specimen was 2.11% of total CD45+ cells. The data analysis indicated that imprecision was

K232600 - Page 17 of 27

{17}

likely due to variations from morning preparation to afternoon preparation within the sample. The high imprecision for this sample has no clinical impact.

Samples is a phenotypically normal WB specimen collected in K2EDTA anticoagulant that demonstrated  $\% \mathrm{CV}$  of  $31.09\%$  for between-runs and  $32.24\%$  for within-sample for CD33 marked Monocytes, as the CD33 marked Monocytes recovery in this specimen was  $2.25\%$  of total  $\mathrm{CD45 + }$  cells. The data analysis indicated that imprecision was likely due to variations from morning preparation to afternoon preparation within the sample. The high imprecision for this sample has no clinical impact.

Sample is a phenotypically normal BM specimen collected in K2EDTA anticoagulant that demonstrated  $\% \mathrm{CV}$  of  $20.63\%$  for the TCRg/d cells, as the TCR g/d cells recovery in this specimen was extremely low with only an average of 73 events collected, representing only  $\sim 0.12\%$  of total CD45+ cells. The high imprecision for this sample has no clinical impact.

# b. Operator-to-operator and instrument-to-instrument imprecision:

The imprecision of between-operator and between-instrument was evaluated by testing four normal and four abnormal clinical specimens with the ClearLLab 10C Panels at one site, with three operators analyzing three replicates on each of two DxFLEX instruments on each of two runs. Each operator performed separate sample preparations in triplicate in both the morning and afternoon. Samples were acquired on each of two instruments, the order of which was randomized. Following acquisition, each operator analyzed listmode datafiles (LMD), using Kaluza C software. A reviewer assessed each LMD analysis as normal or abnormal, and further described the immunophenotype of abnormal results. The study included 288 analyses (three operators/replicate x three replicates/run x two runs/instrument x two instruments = 36 replicates/sample x eight samples).

All results showed  $100\%$  qualitative agreement across Instruments and Operators.

The number of samples tested is depicted in the tables below.

|  Qualitative Agreement across Instruments and Operators  |   |   |
| --- | --- | --- |
|  Panel | Phenotype | N  |
|  B | Normal (-) | 36  |
|  B | Abnormal (+) | 72  |
|  M1 | Normal (-) | 36  |
|  M1 | Abnormal (+) | 36  |
|  M2 | Normal (-) | 36  |
|  M2 | Abnormal (+) | 36  |
|  T | Normal (-) | 36  |

Results based on quantitative analysis are summarized in the table below.

|  Quantitative Results – by Instrument, by Operator  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Panel | Population | N | Mean (%) | Repeatability |   | Between-Run |   | Between-Instruments |   | Between-Operators |   | Within-Specimen  |
|   |   |   |   |  SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD  |
|  B | CD19+CD5+ | 36 | 91.55 | 0.67 | 0.73 | 0.40 | 0.44 | 0.64 | 0.70 | 0.00 | 0.00 | 1.01  |
|  B | Mature κ+ B cells | 36 | 90.63 | 0.68 | 0.75 | 0.17 | 0.19 | 0.00 | 0.00 | 0.00 | 0.00 | 0.70  |
|  B | Mature λ+ B cells | 36 | 8.84 | 0.58 | 6.56 | 0.19 | 2.20 | 0.00 | 0.00 | 0.00 | 0.00 | 0.61  |
|  B | Mature κ+ B cells | 36 | 63.62 | 1.99 | 3.12 | 0.00 | 0.00 | 0.00 | 0.00 | 0.41 | 0.65 | 2.03  |

K232600 - Page 18 of 27

{18}

|  Quantitative Results – by Instrument, by Operator  |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Panel | Population | N | Mean (%) | Repeatability |   | Between-Run |   | Between-Instruments |   | Between-Operators |   | Within-Specimen  |   |
|   |   |   |   |  SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  B | Mature λ+ B cells | 36 | 35.75 | 1.87 | 5.22 | 0.26 | 0.73 | 0.00 | 0.00 | 0.77 | 2.16 | 2.04 | 5.70  |
|  M1 | CD34-CD7+ | 36 | 6.22 | 0.19 | 3.01 | 0.09 | 1.39 | 0.00 | 0.00 | 0.26 | 4.11 | 0.33 | 5.28  |
|  M1 | CD16+Neutrophiles | 36 | 57.46 | 2.66 | 4.63 | 4.11 | 7.14 | 0.00 | 0.00 | 3.61 | 6.29 | 6.08 | 10.59  |
|  M1 | CD64+Monocytes | 36 | 6.91 | 0.47 | 6.86 | 0.72 | 10.37 | 0.00 | 0.00 | 0.00 | 0.00 | 0.86 | 12.43  |
|  M2 | CD15+Neutrophile | 36 | 62.79 | 0.39 | 0.62 | 0.89 | 1.42 | 0.00 | 0.00 | 0.15 | 0.24 | 0.98 | 1.57  |
|  M2 | CD33+ Monocytes | 36 | 4.79 | 0.09 | 1.92 | 0.07 | 1.54 | 0.02 | 0.41 | 0.00 | 0.00 | 0.12 | 2.49  |
|  M2 | CD117+ | 36 | 7.06 | 0.28 | 4.03 | 0.32 | 4.47 | 0.67 | 9.55 | 0.00 | 0.00 | 0.80 | 11.29  |
|  M2 | CD34+CD123+ | 36 | 2.85 | 0.10 | 3.41 | 0.16 | 5.75 | 0.00 | 0.00 | 0.03 | 0.99 | 0.19 | 6.76  |
|  T | NK cells | 36 | 17.65 | 0.46 | 2.62 | 0.36 | 2.03 | 0.14 | 0.80 | 0.19 | 1.07 | 0.63 | 3.57  |
|  T | γδ T cells | 36 | 2.20 | 0.17 | 7.95 | 0.00 | 0.00 | 0.04 | 1.68 | 0.00 | 0.00 | 0.18 | 8.12  |
|  T | Helper T cells | 36 | 80.62 | 0.55 | 0.68 | 0.00 | 0.00 | 0.00 | 0.00 | 0.26 | 0.32 | 0.61 | 0.75  |
|  T | Cytotoxic T cells | 36 | 15.43 | 0.44 | 2.88 | 0.00 | 0.00 | 0.15 | 0.98 | 0.25 | 1.62 | 0.53 | 3.45  |

c. Site-to-Site Reproducibility with Control Material:

The reproducibility of the ClearLLab 10C Reagent System on DxFLEX Flow Cytometer was evaluated utilizing one lot each of control cell materials: ClearLLab Normal Control Cells and ClearLLab Abnormal Control Cells. These control materials express all surface markers comprising the ClearLLab 10C panels. For multi-site testing, control materials allow samples to be shared between sites in a manner not available for native samples. For each panel, the two controls were tested on three instruments, across three total sites using one ClearLLab 10C reagent lot. At Site 1, testing was performed for 20 days × two runs/day × two replicates/run × three sites × one instrument/site. At sites 2 and 3, testing was performed for five days × two runs/day × three replicates/run × three sites × one instrument/site. Across all three sites, a total of 1133 analyses were generated; 13 were excluded due to operator error. The results are summarized in the tables below.

|  Normal Control Cells – across sites  |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Cell Population | Panel | N | Mean (%) | Repeatability |   | Between-Run |   | Between-Day |   | Between-Site |   | Reproducibility  |   |
|   |   |   |   |  SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  WBC | B | 140 | 97.76 | 0.43 | 0.44 | 1.39 | 1.46 | 4.14 | 4.35 | 1.59 | 1.67 | 4.68 | 4.91  |
|  lymphocyte | B | 140 | 30.70 | 0.76 | 2.47 | 1.32 | 1.35 | 2.87 | 2.94 | 1.93 | 1.97 | 3.72 | 3.81  |
|  monocyte | B | 140 | 9.29 | 0.39 | 4.17 | 0.60 | 1.96 | 0.54 | 1.76 | 0.69 | 2.26 | 1.31 | 4.26  |
|  granulocyte | B | 140 | 57.41 | 0.93 | 1.62 | 0.18 | 1.92 | 0.33 | 3.54 | 1.20 | 12.88 | 1.31 | 14.12  |
|  CD5+ | B | 140 | 74.62 | 0.55 | 0.74 | 0.82 | 1.43 | 0.63 | 1.09 | 2.11 | 3.68 | 2.53 | 4.40  |
|  CD10+ | B | 140 | 57.00 | 0.91 | 1.60 | 0.25 | 0.34 | 0.41 | 0.55 | 0.32 | 0.42 | 0.80 | 1.07  |
|  CD19+CD20+ | B | 140 | 10.09 | 0.28 | 2.77 | 0.80 | 1.41 | 0.57 | 1.00 | 2.04 | 3.59 | 2.44 | 4.29  |
|  CD38+ | B | 140 | 8.83 | 0.36 | 4.03 | 0.14 | 1.41 | 0.17 | 1.71 | 1.06 | 10.48 | 1.12 | 11.06  |
|  CD200+ | B | 140 | 9.22 | 0.25 | 2.76 | 0.25 | 2.84 | 0.16 | 1.87 | 1.42 | 16.09 | 1.49 | 16.93*  |
|  Kappa+ | B | 140 | 59.62 | 1.32 | 2.22 | 0.16 | 1.75 | 0.14 | 1.53 | 0.94 | 10.15 | 0.99 | 10.78  |
|  Lambda+ | B | 140 | 40.34 | 1.33 | 3.30 | 0.19 | 0.32 | 0.44 | 0.73 | 0.56 | 0.94 | 1.51 | 2.54  |
|  WBC | M1 | 140 | 97.80 | 0.23 | 0.24 | 1.31 | 1.34 | 2.77 | 2.84 | 1.91 | 1.95 | 3.62 | 3.70  |
|  lymphocyte | M1 | 140 | 30.53 | 0.73 | 2.38 | 0.62 | 2.02 | 0.68 | 2.24 | 1.95 | 6.37 | 2.27 | 7.44  |
|  monocyte | M1 | 140 | 9.10 | 0.39 | 4.31 | 0.27 | 2.95 | 0.27 | 2.96 | 0.51 | 5.61 | 0.75 | 8.22  |

K232600 - Page 19 of 27

{19}

|  Normal Control Cells – across sites  |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Cell Population | Panel | N | Mean (%) | Repeatability |   | Between-Run |   | Between-Day |   | Between-Site |   | Reproducibility  |   |
|   |   |   |   |  SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  granulocyte | M1 | 140 | 57.91 | 0.90 | 1.56 | 0.76 | 1.32 | 0.70 | 1.21 | 2.88 | 4.97 | 3.19 | 5.51  |
|  CD10+ | M1 | 140 | 57.46 | 0.90 | 1.56 | 0.83 | 1.45 | 0.71 | 1.24 | 2.92 | 5.09 | 3.25 | 5.65  |
|  CD7+ | M1 | 140 | 80.54 | 0.51 | 0.63 | 0.21 | 0.26 | 0.82 | 1.01 | 4.83 | 6.00 | 4.93 | 6.12  |
|  CD14+CD11b+ | M1 | 140 | 7.89 | 0.32 | 4.08 | 0.25 | 3.13 | 0.00 | 0.00 | 1.02 | 12.90 | 1.10 | 13.89  |
|  CD13+ | M1 | 140 | 57.54 | 0.93 | 1.62 | 0.83 | 1.44 | 0.71 | 1.24 | 2.81 | 4.88 | 3.16 | 5.48  |
|  CD16+ | M1 | 140 | 54.36 | 0.86 | 1.59 | 0.75 | 1.38 | 0.83 | 1.52 | 2.86 | 5.26 | 3.19 | 5.87  |
|  CD14+CD64+ | M1 | 140 | 7.80 | 0.32 | 4.13 | 0.26 | 3.38 | 0.10 | 1.33 | 0.98 | 12.63 | 1.07 | 13.77  |
|  HLA-DR+ | M1 | 140 | 8.44 | 0.33 | 3.93 | 0.27 | 3.15 | 0.22 | 2.61 | 0.42 | 4.98 | 0.64 | 7.55  |
|  WBC | M2 | 140 | 97.90 | 0.38 | 0.38 | 0.67 | 0.68 | 2.50 | 2.55 | 1.87 | 1.91 | 3.21 | 3.28  |
|  lymphocyte | M2 | 140 | 30.38 | 0.85 | 2.79 | 0.82 | 2.71 | 0.67 | 2.21 | 1.98 | 6.53 | 2.40 | 7.91  |
|  monocyte | M2 | 140 | 9.12 | 0.48 | 5.23 | 0.00 | 0.00 | 0.39 | 4.32 | 0.74 | 8.11 | 0.96 | 10.57  |
|  granulocyte | M2 | 140 | 58.06 | 1.22 | 2.10 | 0.98 | 1.70 | 0.86 | 1.47 | 2.79 | 4.81 | 3.32 | 5.71  |
|  CD38+ | M2 | 140 | 8.69 | 0.42 | 4.84 | 0.13 | 1.50 | 0.38 | 4.40 | 0.83 | 9.59 | 1.02 | 11.71  |
|  CD13+ | M2 | 140 | 57.57 | 1.24 | 2.15 | 1.07 | 1.85 | 0.87 | 1.50 | 2.47 | 4.29 | 3.08 | 5.36  |
|  HLA-DR+ | M2 | 140 | 8.43 | 0.44 | 5.25 | 0.00 | 0.00 | 0.28 | 3.37 | 0.58 | 6.92 | 0.79 | 9.32  |
|  CD15+ | M2 | 140 | 57.44 | 1.24 | 2.17 | 1.03 | 1.80 | 0.94 | 1.64 | 2.71 | 4.72 | 3.29 | 5.74  |
|  CD19+ | M2 | 140 | 9.82 | 0.32 | 3.23 | 0.21 | 2.12 | 0.33 | 3.33 | 0.50 | 5.14 | 0.71 | 7.24  |
|  CD33+ | M2 | 140 | 8.90 | 0.49 | 5.56 | 0.20 | 2.24 | 0.47 | 5.29 | 1.31 | 14.75 | 1.49 | 16.78*  |
|  WBC | T | 140 | 97.72 | 0.50 | 0.51 | 0.95 | 0.97 | 2.88 | 2.95 | 1.95 | 2.00 | 3.64 | 3.72  |
|  lymphocyte | T | 140 | 30.59 | 0.66 | 2.17 | 0.59 | 1.94 | 0.63 | 2.04 | 1.41 | 4.59 | 1.78 | 5.81  |
|  monocyte | T | 140 | 9.17 | 0.29 | 3.18 | 0.37 | 4.04 | 0.27 | 2.99 | 1.09 | 11.87 | 1.22 | 13.28  |
|  granulocyte | T | 140 | 57.56 | 0.81 | 1.40 | 0.82 | 1.43 | 0.77 | 1.33 | 2.58 | 4.49 | 2.93 | 5.09  |
|  CD5+ | T | 140 | 75.54 | 0.46 | 0.61 | 0.40 | 0.53 | 0.49 | 0.65 | 0.60 | 0.80 | 0.99 | 1.31  |
|  CD2+ | T | 140 | 84.68 | 0.37 | 0.44 | 0.26 | 0.31 | 0.28 | 0.33 | 1.04 | 1.23 | 1.17 | 1.38  |
|  CD3+ | T | 140 | 76.05 | 0.43 | 0.57 | 0.29 | 0.39 | 0.42 | 0.56 | 0.49 | 0.64 | 0.83 | 1.10  |
|  CD4+CD8+ | T | 140 | 65.08 | 0.68 | 1.05 | 0.38 | 0.59 | 0.00 | 0.00 | 1.12 | 1.72 | 1.36 | 2.10  |
|  CD4-CD8+ | T | 140 | 29.06 | 0.71 | 2.44 | 0.53 | 1.84 | 0.00 | 0.00 | 2.32 | 7.98 | 2.48 | 8.54  |
|  CD7+ | T | 140 | 79.47 | 0.46 | 0.58 | 0.70 | 0.88 | 1.04 | 1.31 | 3.15 | 3.97 | 3.43 | 4.31  |
|  CD3-CD56+ | T | 140 | 11.33 | 0.31 | 2.70 | 0.24 | 2.08 | 0.35 | 3.09 | 1.84 | 16.28 | 1.92 | 16.91*  |
|  TCRγδ+ | T | 140 | 2.77 | 0.13 | 4.59 | 0.00 | 0.00 | 0.07 | 2.63 | 0.15 | 5.48 | 0.21 | 7.62  |
|  Abnormal Control Cells – across sites  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Cell Population | Panel | N | Mean (%) | Repeatability |   | Between-Run |   | Between-Days |   | Between-Site |   | Reproducibility  |
|   |   |   |   |  SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD  |
|  WBC | B | 140 | 97.92 | 0.61 | 0.62 | 0.58 | 0.59 | 2.49 | 2.54 | 1.73 | 1.76 | 3.14  |
|  lymphocyte | B | 140 | 30.34 | 0.88 | 2.89 | 0.70 | 2.31 | 0.27 | 0.88 | 1.30 | 4.27 | 1.73  |
|  monocyte | B | 140 | 9.09 | 0.35 | 3.85 | 0.19 | 2.04 | 0.25 | 2.72 | 0.87 | 9.56 | 0.99  |
|  granulocyte | B | 140 | 57.80 | 1.15 | 1.98 | 0.96 | 1.66 | 0.38 | 0.66 | 2.45 | 4.24 | 2.90  |
|  CD5+ | B | 140 | 74.89 | 0.48 | 0.64 | 0.36 | 0.48 | 0.41 | 0.55 | 0.16 | 0.21 | 0.75  |
|  CD10+ | B | 140 | 56.99 | 1.19 | 2.09 | 0.89 | 1.57 | 0.37 | 0.66 | 2.71 | 4.75 | 3.11  |
|  CD19+CD20+ | B | 140 | 10.05 | 0.32 | 3.20 | 0.16 | 1.61 | 0.00 | 0.00 | 1.17 | 11.62 | 1.22  |
|  CD34+ | B | 140 | 7.99 | 0.26 | 3.23 | 0.20 | 2.55 | 0.35 | 4.43 | 0.50 | 6.25 | 0.69  |
|  CD38+ | B | 140 | 8.65 | 0.31 | 3.56 | 0.28 | 3.25 | 0.29 | 3.33 | 1.15 | 13.29 | 1.26  |

K232600 - Page 20 of 27

{20}

|  Abnormal Control Cells – across sites  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Cell Population | Panel | N | Mean (%) | Repeatability |   | Between-Run |   | Between-Days |   | Between-Site |   | Reproducibility  |
|   |   |   |   |  SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD  |
|  CD200+ | B | 140 | 9.16 | 0.31 | 3.42 | 0.16 | 1.72 | 0.09 | 0.99 | 1.06 | 11.59 | 1.12  |
|  Kappa+ | B | 140 | 59.51 | 1.38 | 2.32 | 0.47 | 0.80 | 0.54 | 0.91 | 0.39 | 0.65 | 1.61  |
|  Lambda+ | B | 140 | 40.45 | 1.37 | 3.38 | 0.48 | 1.19 | 0.56 | 1.39 | 0.23 | 0.56 | 1.57  |
|  WBC | M1 | 140 | 97.88 | 0.95 | 0.98 | 0.71 | 0.72 | 2.38 | 2.43 | 1.88 | 1.92 | 3.26  |
|  lymphocyte | M1 | 140 | 30.12 | 1.04 | 3.44 | 0.75 | 2.49 | 0.42 | 1.39 | 1.97 | 6.53 | 2.38  |
|  monocyte | M1 | 140 | 8.89 | 0.30 | 3.37 | 0.17 | 1.86 | 0.37 | 4.20 | 0.55 | 6.19 | 0.75  |
|  granulocyte | M1 | 140 | 58.38 | 1.27 | 2.18 | 0.94 | 1.61 | 0.63 | 1.08 | 2.98 | 5.11 | 3.43  |
|  CD10+ | M1 | 140 | 57.62 | 1.29 | 2.24 | 1.00 | 1.74 | 0.72 | 1.25 | 3.20 | 5.55 | 3.67  |
|  CD34+ | M1 | 140 | 8.04 | 0.28 | 3.51 | 0.23 | 2.92 | 0.28 | 3.44 | 0.71 | 8.84 | 0.85  |
|  CD7+ | M1 | 140 | 80.95 | 0.44 | 0.55 | 0.34 | 0.42 | 0.52 | 0.64 | 4.00 | 4.94 | 4.07  |
|  CD14+CD11b+ | M1 | 140 | 7.77 | 0.36 | 4.61 | 0.20 | 2.54 | 0.20 | 2.58 | 0.94 | 12.04 | 1.04  |
|  CD13+ | M1 | 140 | 58.14 | 1.31 | 2.25 | 0.90 | 1.54 | 0.80 | 1.38 | 2.85 | 4.91 | 3.36  |
|  CD16+ | M1 | 140 | 54.33 | 1.22 | 2.25 | 0.96 | 1.76 | 0.65 | 1.20 | 3.19 | 5.87 | 3.61  |
|  CD14+CD64+ | M1 | 140 | 7.70 | 0.35 | 4.58 | 0.21 | 2.72 | 0.23 | 3.00 | 0.89 | 11.51 | 1.00  |
|  HLA-DR+ | M1 | 140 | 8.45 | 0.35 | 4.08 | 0.20 | 2.36 | 0.39 | 4.67 | 0.37 | 4.40 | 0.67  |
|  WBC | M2 | 140 | 98.03 | 0.65 | 0.66 | 0.30 | 0.31 | 2.19 | 2.23 | 1.74 | 1.78 | 2.89  |
|  lymphocyte | M2 | 140 | 29.97 | 0.83 | 2.76 | 0.83 | 2.76 | 0.63 | 2.09 | 2.48 | 8.26 | 2.81  |
|  monocyte | M2 | 140 | 8.78 | 0.35 | 3.97 | 0.14 | 1.65 | 0.25 | 2.82 | 0.64 | 7.32 | 0.79  |
|  granulocyte | M2 | 140 | 58.61 | 1.03 | 1.76 | 0.98 | 1.67 | 0.81 | 1.38 | 3.44 | 5.87 | 3.81  |
|  CD34+ | M2 | 140 | 7.96 | 0.31 | 3.85 | 0.27 | 3.34 | 0.32 | 4.07 | 0.63 | 7.93 | 0.82  |
|  CD38++ | M2 | 140 | 8.31 | 0.32 | 3.81 | 0.25 | 3.03 | 0.23 | 2.81 | 0.85 | 10.28 | 0.97  |
|  CD13+ | M2 | 140 | 58.01 | 1.05 | 1.80 | 0.90 | 1.55 | 0.86 | 1.48 | 3.78 | 6.52 | 4.11  |
|  HLA-DR+ | M2 | 140 | 8.35 | 0.34 | 4.02 | 0.21 | 2.51 | 0.23 | 2.76 | 0.60 | 7.21 | 0.76  |
|  CD15 | M2 | 140 | 57.80 | 1.04 | 1.80 | 1.05 | 1.82 | 0.82 | 1.42 | 3.62 | 6.26 | 3.99  |
|  CD19+ | M2 | 140 | 9.69 | 0.33 | 3.45 | 0.06 | 0.63 | 0.18 | 1.88 | 0.65 | 6.74 | 0.76  |
|  CD33+ | M2 | 140 | 8.62 | 0.36 | 4.19 | 0.19 | 2.26 | 0.25 | 2.94 | 1.11 | 12.88 | 1.21  |
|  CD117+ | M2 | 140 | 8.06 | 0.31 | 3.84 | 0.25 | 3.14 | 0.34 | 4.23 | 0.48 | 5.90 | 0.71  |
|  CD123+ | M2 | 140 | 8.16 | 0.31 | 3.79 | 0.28 | 3.46 | 0.34 | 4.12 | 0.47 | 5.81 | 0.72  |
|  WBC | T | 140 | 97.83 | 1.22 | 1.25 | 0.62 | 0.64 | 2.31 | 2.36 | 1.90 | 1.94 | 3.29  |
|  lymphocyte | T | 140 | 30.34 | 0.86 | 2.84 | 1.10 | 3.62 | 0.41 | 1.34 | 2.02 | 6.65 | 2.49  |
|  monocyte | T | 140 | 8.91 | 0.34 | 3.84 | 0.33 | 3.66 | 0.29 | 3.27 | 0.98 | 10.95 | 1.12  |
|  granulocyte | T | 140 | 58.08 | 1.09 | 1.88 | 1.38 | 2.38 | 0.51 | 0.88 | 3.25 | 5.59 | 3.73  |
|  CD5+ | T | 140 | 75.50 | 0.55 | 0.73 | 0.19 | 0.25 | 0.58 | 0.76 | 0.94 | 1.25 | 1.25  |
|  CD34+ | T | 140 | 8.15 | 0.44 | 5.46 | 0.00 | 0.00 | 0.00 | 0.00 | 0.50 | 6.11 | 0.67  |
|  CD2+e | T | 140 | 84.61 | 0.48 | 0.57 | 0.13 | 0.15 | 0.48 | 0.56 | 1.47 | 1.73 | 1.62  |
|  CD3+ | T | 140 | 76.08 | 0.51 | 0.67 | 0.29 | 0.38 | 0.44 | 0.58 | 0.33 | 0.43 | 0.81  |
|  CD4+CD8- | T | 140 | 65.11 | 0.70 | 1.08 | 0.67 | 1.03 | 0.00 | 0.00 | 1.22 | 1.87 | 1.56  |
|  CD4-CD8+ | T | 140 | 29.10 | 0.63 | 2.15 | 0.67 | 2.31 | 0.00 | 0.00 | 2.39 | 8.20 | 2.56  |
|  CD7+ | T | 140 | 79.02 | 0.50 | 0.63 | 0.55 | 0.69 | 0.77 | 0.98 | 3.92 | 4.96 | 4.06  |
|  CD3-CD56+ | T | 140 | 11.16 | 0.33 | 2.95 | 0.18 | 1.57 | 0.22 | 1.93 | 1.81 | 16.17 | 1.86  |
|  TCRγδ+ | T | 140 | 2.79 | 0.11 | 4.09 | 0.09 | 3.22 | 0.04 | 1.44 | 0.15 | 5.30 | 0.21  |

* %CV of reproductivity for the sample is &gt;15%. As the repeatability of individual CD measurements includes unique variability sources, sources of variability include the frequency of cells with a particular phenotype, the level of the marker expression on the specific population being tested (in relation to the expression on other populations that

K232600 - Page 21 of 27

{21}

are present), as well as the fluorescence intensity of the fluorochrome (quantum efficiency) of the specific conjugate used for staining. These differences in cell number, target analyte level, distribution on non-target populations and fluorochrome intensity affect the ability of the instrument/cytometrist to distinguish specific populations. These additional sources of variability contribute to the increased imprecision in several of the phenotypes tested.

d. ClearLLab 10C Reagents Lot-to-Lot Reproducibility:

Refer to K183592

2. Linearity (Fluorescence Detection):

The study was performed to evaluate the linearity and dynamic range of the DxFLEX Flow Cytometers' acquisition systems, from the single-photon detection APD transfer to the electronics, independent of the application and gating methodology using SpheroTECH Rainbow particles (RCP-30-5A) analyzed at six different gain settings covering the dynamic range of the DxFLEX Flow Cytometer. The study was conducted on three DxFLEX Flow Cytometers. The results indicated the constancy (&lt;5% deviation) of the ratio of adjacent bead fluorescence, across a series of gain settings. Representative data from one DxFLEX Flow Cytometer is presented in the following table.

|  Gain Setting | Median-A Ratio of P7/P6  |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  FITC | PE | ECD | PC5.5 | PC7 | APC | APC700 | APC750 | PB450 | KO525  |
|  100 | 2.88 | 2.51 | 3.02 | 3.19 | 3.28 | 2.82 | 2.95 | 2.97 | 2.79 | 2.96  |
|  500 | 2.88 | 2.51 | 3.04 | 3.21 | 3.20 | 2.82 | 2.95 | 2.94 | 2.81 | 2.98  |
|  1000 | 2.86 | 2.52 | 3.03 | 3.19 | 3.16 | 2.81 | 2.94 | 2.94 | 2.79 | 2.96  |
|  1500 | 2.87 | 2.52 | 3.03 | 3.19 | 3.19 | 2.82 | 2.93 | 2.92 | 2.80 | 2.97  |
|  2000 | 2.88 | 2.51 | 3.03 | 3.20 | 3.20 | 2.83 | 2.93 | 2.94 | 2.81 | 2.98  |
|  3000 | 2.88 | 2.51 | 3.03 | 3.19 | 3.19 | 2.81 | 2.95 | 2.96 | 2.78 | 2.96  |
|  Lower Limit | 2.73 | 2.39 | 2.88 | 3.04 | 3.05 | 2.68 | 2.79 | 2.80 | 2.66 | 2.82  |
|  Upper Limit | 3.02 | 2.64 | 3.18 | 3.36 | 3.37 | 2.96 | 3.09 | 3.09 | 2.94 | 3.12  |

3. Analytical Specificity/Interference:

a. Interfering Substances:

Not applicable. The test system requires washing with phosphate buffered saline (PBS) before staining with reagents to remove residual plasma and interfering substances.

b. Carryover:

Specimen Carryover

Specimen carryover was evaluated by following the protocol from CLSI H26-A2 on three DxFLEX Flow Cytometers. Samples with high concentration (high target value; HTv)

K232600 - Page 22 of 27

{22}

and low concentration (low target value; LTv) were run in multiple replicates on the DxFLEX Flow Cytometer. CD34 marker analysis was performed using a CD34+ clinical specimen. Three high concentration samples (HTv1-HTv2-HTv3) targeting to 20,000 cells /μL white blood cells (WBC) were prepared from one specimen containing a high level of white blood cells Low concentration samples (LTv1-LTv2-LTv3) counts were prepared by staining a diluted (1:10,000) specimen with a blank Red Cell Pool (a red blood cell pool which has been depleted of white blood cells diluted 4:1 with buffer). This resulted in a specimen composition similar to whole blood regarding the red blood cell concentration with residual concentrations of WBCs. On each of three DxFLEX Flow Cytometers, three HTv samples (HTv1, HTv2, HTv3) were analyzed consecutively followed immediately by the three LTv samples (LTv1, LTv2, LTv3) using the four ClearLLab 10C Panels. A total of 54 specimen carryover runs were generated for each panel. The % carryover, which is calculated as % Carryover = (LTV1-LTV3)/average (HTv1, HTv2, HTv3), demonstrated to be &lt; 1%, for each of the 37 phenotypes measured across the four ClearLLab 10C tubes.

## Reagent Carryover

Reagent carryover was evaluated by following the protocol from CLSI H26-A2 on three DxFLEX Flow Cytometers. Samples with high concentration (high target value; HTv) and low concentration (low target value; LTv) were run in multiple replicates on the DxFLEX flow cytometer. Clinical bone marrow specimens stained with all four ClearLLab 10C Panels were used as the high concentration samples. Low concentration samples (LTv1, LTv2, LTv3) were prepared from the same clinical bone marrow specimens as in the high concentration samples, that were processed in triplicate using antibody dilution buffer (RD1), in place of the ClearLLab 10C Panels, to produce an unstained sample. On each of three DxFLEX Flow Cytometers and three HTv samples (HTv1, HTv2, HTv3) were analyzed consecutively followed immediately by the three LTv samples (LTv1, LTv2, LTv3). A total of 54 specimen carryover runs were generated for each panel. The % carryover, which is calculated as % Carryover = (LTV1-LTV3)/average (HTv1, HTv2, HTv3), demonstrated to be &lt; 1%, for each of the 37 phenotypes measured across the four ClearLLab 10C tubes.

## 4. Accuracy (Instrument):

### Method Comparison with Predicate Device:

A method comparison study was performed to demonstrate qualitative agreement for the presence of abnormal cells between ClearLLab 10C Reagent System on the DxFLEX Flow Cytometer vs Navios EX Flow Cytometer. The study was based on the data generated from a multi-site retrospective clinical study conducted at five sites in which the test results from the ClearLLab 10C Reagents used to detect the presence or absence of an abnormal phenotype were compared to the clinical outcome of "malignant" or "non-malignant" based on the clinical sites' final patient diagnosis.

Testing was conducted on the DxFLEX and Navios EX flow cytometers utilizing de-identified residual specimens, covering both hematologically abnormal malignant patients and hematologically abnormal non-malignant patients in line with the intended use of the ClearLLab 10C panels. The study included specimens of whole blood (WB) and bone

K232600 - Page 23 of 27

{23}

marrow (BM) collected in anticoagulant K2EDTA, Acid Citrate Dextrose (ACD) or Heparin, as well as specimens of lymph node (LN), which overall reflect the three major specimen types distributed in leukemia and lymphoma (L&amp;L) diseases and clinical indications encountered in the L&amp;L population. Patients with abnormal B-cell, T-cell, Natural Killer (NK) cell, and myeloid cell populations were tested. Specimen types tested reflected the distribution of the diseases and/or clinical indications encountered in the population of patients that are routinely encountered in evaluations of patients suspected of having hematopoietic neoplasia by flow cytometry. The study design was purposive, and specimens were enrolled consecutively.

Samples were prepared with all four ClearLLab 10C Panels, except LN where only B and T tubes were required. The same prepared tube was randomized for acquisition once on the Navios EX and once on the DxFLEX flow cytometer. Immunophenotyping was performed with provided analysis protocols using Kaluza C software. The phenotype assessments of the data generated from DxFLEX-10C system and Navios EX-10C system were independently assessed from each other. If clinical diagnosis was provided in the subject's case report form, the Principal Investigator at each site was blinded to such diagnosis to further assure the review was exclusively phenotype based. The assessment for presence or absence of an abnormal phenotype was based on Bethesda and WHO guidelines as follows:

- A 3% (in total leukocytes) clinical cutoff will be applied to BM blast population which is normally undetected in peripheral WB (&lt;1%) or LN (&lt;1%)
- 1% (in total leukocytes) clinical cutoff will be applied to all the other abnormal populations detected in BM, WB, or LN specimens

Of the 550 samples enrolled, 38 were excluded due to failing the inclusion criteria, preanalytical requirements or quality control or due to instrument, process or operator errors. A total of 512 specimens were qualified and included in the final analysis for evaluating method agreement per CLSI EP12-Ed3. The distribution of specimens is presented in the following tables.

|  Distribution of Subjects - by Gender and Age  |   |   |   |
| --- | --- | --- | --- |
|  Gender | Male |   | 268  |
|   |  Female |   | 239  |
|   |  Not Provided |   | 5  |
|  Age | Infant
(1 mon-2 yrs) | Mean | 1  |
|   |   |  4 months  |   |
|   |  Child
(>2-12 yrs) | Mean | 12  |
|   |   |  6.3 yrs  |   |
|   |  Adolescent
(>12-21 yrs) | Mean | 14  |
|   |   |  17.6 yrs  |   |
|   |  Adult
(>21-65 yrs) | Mean | 230  |
|   |   |  50.6 yrs  |   |
|   |  Geriatric
(>65-89 yrs) | Mean | 250  |
|   |   |  75.6 yrs  |   |
|   |  >89 yrs | Mean | 5  |
|   |   |  N/A  |   |
|  Total |   | 512  |   |

K232600 - Page 24 of 27

{24}

|  Distribution of Subjects - by Category  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   | Abnormal Hematology No Malignancy | Abnormal Hematology- Malignancy |   |   |   |   | Total  |
|   |   |  Acute Leukemia | Chronic Leukemia | Lymphoma | Plasma Cell Neoplasm | Others: MDS MPN  |   |
|  Site 1 | 46 | 12 | 14 | 26 | 7 | 10 | 115  |
|  Site 2 | 41 | 10 | 12 | 14 | 0 | 8 | 85  |
|  Site 3 | 60 | 14 | 30 | 18 | 5 | 2 | 129  |
|  Site 4 | 21 | 2 | 3 | 26 | 11 | 3 | 66  |
|  Site 5 | 28 | 23 | 18 | 29 | 8 | 11 | 117  |
|  Total | 196 | 61 | 77 | 113 | 31 | 34 | 512  |
|  Distribution of Subjects - by Specimen Type  |   |
| --- | --- |
|  WB | 291  |
|  BM | 167  |
|  LN | 54  |
|  Total | 512  |

The qualitative immunophenotype agreement (presence or absence of an abnormal population) was evaluated by comparing the assessment from the ClearLLab 10C on DxFLEX Flow Cytometer (Test Method) to that from the ClearLLab 10C on the Navios EX Flow Cytometer (Predicate Method). % The results are shown in the following tables:

|  Method Agreement – All Specimens Combined  |   |   |   |
| --- | --- | --- | --- |
|   | ClearLLab 10C on Navios EX  |   |   |
|  ClearLLab 10C on DxFLEX | Presence of Abnormal cells (+) | Absence of Abnormal cells (-) | Total  |
|  Presence of Abnormal cells (+) | 231 | 0 | 231  |
|  Absence of Abnormal cells (-) | 0 | 281 | 281  |
|  Total | 246 | 281 | 512  |

PPA: 100% (95% CI: 98.46% – 100%)
NPA: 100% (95% CI: 99.28% – 100%)

|  Method Agreement – Whole Blood Specimens Only  |   |   |   |
| --- | --- | --- | --- |
|   | ClearLLab 10C on Navios EX  |   |   |
|  ClearLLab 10C on DxFLEX | Presence of Abnormal cells (+) | Absence of Abnormal cells (-) | Total  |
|  Presence of Abnormal cells (+) | 139 | 0 | 139  |
|  Absence of Abnormal cells (-) | 0 | 152 | 152  |
|  Total | 139 | 152 | 291  |

PPA: 100% (95% CI: 97.47% – 100%)
NPA: 100% (95% CI: 97.54% – 100%)

K232600 - Page 25 of 27

{25}

Clinical performance (sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV)) of the ClearLLab 10C Reagent on DxFLEX Flow Cytometer was also assessed by measuring the agreement of the flow cytometric results with the clinical diagnosis. The results of this evaluation are presented below.

|  Diagnostic Accuracy – All Specimen Types Combined  |   |   |   |
| --- | --- | --- | --- |
|   | Clinical Outcome  |   |   |
|  ClearLLab 10C on DxFLEX | Malignant | Non-Malignant | Total  |
|  Malignant (+) | 220 | 11 | 231  |
|  Non-Malignant (-) | 96 | 185 | 281  |
|  Total | 316 | 196 | 512  |

Sensitivity: 70% (95% CI: 64% - 74%)
Specificity: 94% (95% CI: 90% - 97%)
PPV: 95% (95% CI: 92% - 97%)
NPV: 66% (95% CI: 60% - 71%)

|  Diagnostic Accuracy – Whole Blood Only  |   |   |   |
| --- | --- | --- | --- |
|   | Clinical Outcome  |   |   |
|  ClearLLab 10C on DxFLEX | Malignant | Non-Malignant | Total  |
|  Malignant (+) | 133 | 6 | 139  |
|  Non-Malignant (-) | 27 | 125 | 152  |
|  Total | 160 | 131 | 291  |

Sensitivity: 83% (95% CI: 77% - 88%)
Specificity: 95% (95% CI: 90% - 98%)
PPV: 96% (95% CI: 91% - 98%)
NPV: 82% (95% CI: 75% - 88%)

The clinical performance was also assessed with the predicate device. The results demonstrated 100% phenotypic agreement between the candidate and predicate devices. For both methods, when compared to clinical diagnostic outcome, there were a total of 107 discordant cases including 11 non-concordant to non-malignancy (false positive) and 96 non-concordant to malignancy (false negative) cases. The false negative was categorized as being due to specimen type limitation, flow cytometry limitation, or ClearLLab reagents limitation. These limitations occur when the phenotypically abnormal population is not found in a particular specimen, the clinical condition is not typically detected through flow cytometry, or the reagents did not have all the antibodies required to detect a particular clinical condition, respectively. False positive cases were primarily due to increases in populations that could represent reactive responses to viral or bacterial infection or autoimmunity, where lymphoid proliferation is common.

## B Other Supportive Instrument Performance Characteristics Data:

The detection capability of the DxFLEX flow cytometer with the ClearLLab 10C Reagent, was assessed for each reagent tube using one normal specimen (Blank) that was spiked with samples having abnormal populations (CLL, T-ALL, Myelomonocytic Leukemia and AML), to target 0.5%, 1% and 2% of the abnormal population, each in five replicates. The results demonstrated

K232600 - Page 26 of 27

{26}

that 100% of the replicates showed the correct populations, down to a level of 0.5% and verified that the Limit of Detection of 1% was confirmed on the new instrument.

## VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K232600 - Page 27 of 27

---

**Source:** [https://fda.innolitics.com/device/K232600](https://fda.innolitics.com/device/K232600)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
