K232593 · Keosan Trading Co. · FWP · Nov 22, 2023 · General, Plastic Surgery
Device Facts
Record ID
K232593
Device Name
FITme Customized Silicone Implant
Applicant
Keosan Trading Co.
Product Code
FWP · General, Plastic Surgery
Decision Date
Nov 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FITme Customized Silicone Implant is intended for augmentation, reconstructive and cosmetic surgery of the facial regions. The FITme Customized Silicone Implant is pre-shaped to the surgeon's specification to meet the needs of a particular patient.
Device Story
Patient-specific silicone implant for facial augmentation/reconstruction; nasal, malar, and chin regions. Input: patient CT scan data. Process: Mimics software converts CT to digital skull model; surgeon fashions solid model or digital STL file; manufacturer creates customized plaster or 3D-printed molds; silicone elastomer cast in molds. Output: non-sterile, patient-specific silicone implant. Used in clinical/surgical settings by surgeons. Implant provides space-occupying contouring to facial features; minimizes intraoperative carving; promotes natural aesthetic lines. Benefits: customized fit to patient anatomy.
Clinical Evidence
Bench testing only. Testing included sterilization validation, biocompatibility (ISO 10993), hardness (ASTM D2240), tensile strength (ASTM D412), elongation (ASTM D412), tear strength (ASTM D624), and 3D printer validation.
Technological Characteristics
Implant-grade silicone elastomer; patient-specific geometry derived from CT scans via 3D printing/molding. Standards: ASTM D2240 (hardness), ASTM D412 (tensile/elongation), ASTM D624 (tear strength). Non-sterile delivery. Software-assisted design.
Indications for Use
Indicated for patients requiring augmentation, reconstructive, or cosmetic surgery of facial regions (nasal, malar cheek, and chin contours).
Regulatory Classification
Identification
A chin prosthesis is a silicone rubber solid device intended to be implanted to augment or reconstruct the chin.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2023
KEOSAN TRADING CO. % Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, CA 92612
Re: K232593
Trade/Device Name: FITme Customized Silicone Implant Regulation Number: 21 CFR 878.3550 Regulation Name: Chin Prosthesis Regulatory Class: Class II Product Code: FWP, KKY Received: August 26, 2023
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Tek N. Tek N. Lamichhane -S Lamichhane -S Date: 2023.11.22 08:44:36 -05'00' Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232539
Device Name FITme Customized Silicone Implant
### Indications for Use (Describe)
The FITme Customized Silicone Implant is intended for augmentation, reconstructive and cosmetic surgery of the facial regions.
The FITme Customized Silicone Implant is pre-shaped to the surgeon's specification to meet the needs of a particular patient.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
K232593
Date Prepared: Nov 20, 2023
#### Applicant / Submitter: 1.
KEOSAN TRADING CO. #805, #806, Human Teco B/D, 57, Achasan-ro 17-gil Seongdong-gu, Seoul, Republic of Korea Tel : +82-2-466-1188
#### 2. Submission Correspondent:
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine CA 92620 Fax: 714-409-3357 Phone: 714-202-5789 Email: juhee.c(@lkconsultinggroup.com
#### 3. Device Information:
| Proprietary Name: | FITme Customized Silicone Implant |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Common Name: | Silicone Implant |
| Device Class: | II |
| Primary Product Code &<br>Regulation: | FWP<br>(21CFR§878.3550 - Prosthesis, Chin, Internal) |
| Secondary Product Code<br>& Regulation: | KKY<br>(21CFR§878.3500 - Material, Polytetrafluoroethylene<br>Vitreous Carbon, For Maxillofacial Reconstruction) |
#### 4. Predicate Device:
Customized Contour Implant (K191130) by Implantech Associates Inc. (21CFR§878.3550, Product Code: FWP, 21CFR§878.3500, Product Code: KKY)
{6}------------------------------------------------
#### 5. Device Description:
The FITme Customized Silicone Implant is a patient-specific medical device intended for augmentation, reconstructive and cosmetic surgery of facial region, specifically the nasal contour, the malar cheek contour, and chin contour. The device is a single use implant intended for long term implantation as a space occupying device to form a contoured feature. The customized implant, FITme Customized Silicone Implant, is made of implant grade silicone elastomer, in a range of durometers as specified by the surgeon.
The patient's own medical imaging (e.g., Computed Tomography (CT) scan) is translated into a digital model of the patient's skull using the software, Mimics by Materialise. At the recommendation of the surgeon, the Mimics is used either to 3D-print a skull model for the surgeon to fashion a solid implant model with commercially available plaster or to create a digital implant model as an STL file.
KEOSAN TRADING Co. manufactures the customized molds from the solid implant model provided by the surgeon, or the digital implant model provided by KEOSAN TRADING Co., using plaster molds or 3D-printed molds as appropriate. The FITme Customized Silicone Implant is manufactured from the customized molds and provided to the surgeon, non-sterile.
#### 6. Indications for Use:
The FITme Customized Silicone Implant is intended for augmentation, reconstructive and cosmetic surgery of the facial regions.
The FITme Customized Silicone Implant is pre-shaped to the surgeon's specification to meet the needs of a particular patient.
{7}------------------------------------------------
#### 7. Substantial Equivalence
## 7.1. Comparison Chart
| | Subject Device | Predicate Device |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | FITme Customized Silicone Implant | Customized Contour Implant |
| Manufacturer | KEOSAN TRADING CO. | Implantech Associates Inc. |
| 510k # | K232539 | K191130 |
| Product Code | FWP, KKY | FWP, KKY |
| Indications for<br>Use | The FITme Customized Silicone Implant<br>is intended for augmentation,<br>reconstructive and cosmetic surgery of<br>the facial regions.<br>The FITme Customized Silicone Implant<br>is pre-shaped to the surgeon's<br>specification to meet the needs of a<br>particular patient. | The Customized Contour Implant is intended<br>for augmentation, reconstructive and<br>cosmetic surgery of the facial regions.<br>The Customized Contour Implant is pre-<br>shaped to the surgeon's specification to meet<br>the needs of a particular patient. |
| Material | Silicone elastomer | Silicone elastomer |
| Design Features | | Chin Implant<br>Malar Implant<br>Nasal Implant |
| Principle of<br>Operation | It is made of implant grade silicone<br>elastomer, and the design characteristic<br>minimizes carving and promotes natural<br>line. | It is made of implant grade silicone<br>elastomer, and the design characteristic<br>minimizes carving and promotes natural line. |
| Technological<br>Characteristics | Use of 3D printing technology to create<br>patient-specific molds and implant<br>models derived from CT scans of a<br>patient | Use of 3D printing technology to create<br>patient-specific molds and implant models<br>derived from CT scans of a patient |
| Additive<br>Manufacturing<br>Process | Surgeon's specification-> Input (solid<br>model or digital file) -> Mold Creation<br>(plaster mold or 3D printed mold) -><br>Customized Implant | Surgeon's specification-> Input (solid model<br>or digital file) -> Mold Creation (plaster<br>mold or 3D printed mold) -> Customized<br>Implant |
{8}------------------------------------------------
## 7.2. Substantial Equivalence Discussion
The FITme Customized Silicone Implant is substantially equivalent in material, design, function, and additive manufacturing process to predicate devices. All products have identical intended uses and are offered in similar shapes and sizes. The size might be different between the devices, however, since the physician can carve or cut the material to fit each patient so that this difference does not raise an issue of safety and performance. We have performed the performance and biocompatibility tests and the test results support that the subject device is substantially equivalent to the predicate device in the market.
#### 8. Performance Data (Non-Clinical):
The following tests were performed on the subject device and the test results support that the subject device is substantially equivalent to the predicate devices.
- . Sterilization Validation
- . Biocompatibility Tests in accordance with ISO 10993
- " Hardness Test in accordance with ASTM D2240
- . Tensile Strength Test in accordance with ASTM D412
- Elongation in accordance with ASTM D412 "
- Tear Strength Test in accordance with ASTM D624
- . 3D Printer Validation
#### Conclusion: 9.
Based on the information submitted herein, KEOSAN TRADING CO. concludes that the FITme Customized Silicone Implant is substantially equivalent to the predicate device, Customized Contour Implant (K191130).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.