K232588 · Knox Medical Diagnostics · BZH · Nov 25, 2024 · Anesthesiology
Device Facts
Record ID
K232588
Device Name
Aluna 2
Applicant
Knox Medical Diagnostics
Product Code
BZH · Anesthesiology
Decision Date
Nov 25, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1860
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Aluna 2 is intended for monitoring FEV1 (Forced exhalation in the first second) and PEF (Peak Expired Flow Rate) for home use. The device is designed for pediatric to adult users. Aluna is not recommended for children under 5 years of age. Additionally, the device may be used by clinicians for in-office monitoring.
Device Story
Aluna 2 is a handheld Lilly-type pneumotachometer for monitoring FEV1 and PEF. Device captures differential pressure across a stainless-steel mesh during patient exhalation; transmits digital data via Bluetooth to a mobile application. Mobile app processes pressure readings to estimate exhaled air flow rate and volume; displays results to user; transmits data to cloud server via API. Used in home or clinical settings by patients or clinicians. Output allows tracking of pulmonary function over time, aiding clinical decision-making for respiratory conditions. Benefits include convenient, remote monitoring of lung function.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including FEV1/PEF measurement accuracy, repeatability, linearity, and frequency response per 2019 ATS standards, plus electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), battery safety (IEC 62133-2), and software verification (IEC 62304).
Technological Characteristics
Lilly-type pneumotachometer; stainless-steel mesh; differential pressure sensor. Dimensions: 90.3mm x 43mm x 115.5mm; Weight: 95g. Power: rechargeable battery. Connectivity: Bluetooth 4.0 LE. Materials: polymers, stainless steel. Non-sterile. Software: mobile app (iOS/Android) and cloud-based API.
Indications for Use
Indicated for monitoring FEV1 and PEF in pediatric (5+ years) to adult patients for home or in-office use. Not for children under 5 years.
Regulatory Classification
Identification
A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.
Predicate Devices
Aluna manufactured by Knox Medical Diagnostics (K193311)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it.
November 25, 2024
Knox Medical Diagnostics % Carol Vierling Senior Principal Ram+ 2790 Mosside Blvd. Suite 800 Monroeville, Pennsylvania 15146
Re: K232588
Trade/Device Name: Aluna 2 Regulation Number: 21 CFR 868.1860 Regulation Name: Peak-Flow Meter For Spirometry Regulatory Class: Class II Product Code: BZH Dated: October 25, 2024 Received: October 25, 2024
Dear Carol Vierling:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232588
Device Name Aluna 2
Indications for Use (Describe)
Aluna 2 is intended for monitoring FEV1 (Forced exhalation in the first second) and PEF (Peak Expired Flow Rate) for home use. The device is designed for pediatric to adult users. Aluna is not recommended for children under 5 years of age. Additionally, the device may be used by clinicians for in-office monitoring.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
escription 303 (Part 21 CFR 1301, Subpart B)
[Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
DATE PREPARED
November 25, 2024
# MANUFACTURER
Knox Medical Diagnostics 650 California Street, FL 7 San Francisco, CA 94108 Telephone: (415) 637-9852 Official contact: Michael Raftery, VP Operations and Quality
## REPRESENTATIVE/CONSULTANT
Allison Komiyama, Ph.D., RAC Pierre Bounaud, Ph.D. ROM+ 2790 Mosside Blvd., Suite 800 Monroeville, PA 15146 Telephone: (412) 816-8253 Email: akomiyama@rqmplus.com, pbounaud(@rqmplus.com
# DEVICE INFORMATION
- Device trade name, or proprietary name: Aluna 2 ●
- Device common name: Peak flow meter for spirometry ●
- Classification: II
- 21 CFR 868.1860: Meter, peak flow, spirometry ●
- Product Code(s): BZH
- Panel: Anesthesiology ●
- Basis for submission: Device modifications ●
# PREDICATE DEVICE IDENTIFICATION
Aluna 2 is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate |
|---------------|------------------------------------------------|----------------------|
| K193311 | Aluna manufactured by Knox Medical Diagnostics | ✓ |
# DEVICE DESCRIPTION
Aluna 2 consists of three main components:
- A small hand-held peak flow meter that captures differential pressure, converts it to ● digital data points, and transmits the data to a mobile device via Bluetooth.
- A mobile application that collects and transmits the differential pressure readings from ● the spirometry device and processes it to estimate the exhaled air flow rate and volume.
- An Application Programming Interface (API) to facilitate communication with a cloud ● server used for data storage.
{5}------------------------------------------------
# INDICATIONS FOR USE
Aluna 2 is intended for monitoring FEV1 (Forced exhalation in the first second) and PEF (Peak Expired Flow Rate) for home use. The device is designed for pediatric to adult users. Aluna is not recommended for children under 5 years of age. Additionally, the device may be used by clinicians for in-office monitoring.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Modifications are proposed to Aluna, the first generation device cleared under K193311. Modifications include minor cosmetic changes to hardware, additional capabilities to the software and changes to non-patient contacting materials.
The subject device has the same design and dimensions and uses similar materials as the first generation Aluna. The subject device has the same intended use and similar technological characteristics to the device cleared in K193311. Any difference in technological characteristics have undergone testing to ensure the device is as safe and effective as the predicate.
Substantial equivalence comparison between the subject device and the device cleared in K193311 is summarized in the table below.
{6}------------------------------------------------
| | Subject Device | Predicate Device | Comparison |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| | Aluna 2 | Aluna | |
| | (K232588) | (K193311) | |
| Indications<br>for Use | Aluna 2 is intended for<br>monitoring FEV1 (Forced<br>exhalation in the first<br>second) and PEF (Peak<br>Expired Flow Rate) for<br>home use. The device is<br>designed for pediatric to<br>adult users. Aluna is not<br>recommended for children<br>under 5 years of age.<br>Additionally, the device<br>may be used by clinicians<br>for in-office monitoring. | Aluna is intended for<br>monitoring FEV1 (Forced<br>exhalation in the first<br>second) and PEF (Peak<br>Expired Flow Rate) for<br>home use. The device is<br>designed for children 5<br>years of age or older,<br>adolescent and adult<br>subjects. Additionally, the<br>device may be used by<br>clinicians for in-office<br>monitoring. | Similar; Replacing age<br>range of intended users<br>with more concise<br>language with the same<br>meaning |
| Product Code<br>/ Regulation<br>Number | BZH / 21 CFR 868.1860 | BZH / 21 CFR 868.1860 | Identical |
| Regulation | Peak-flow meter for | Peak-flow meter for | Identical |
| Description | spirometry | spirometry | |
| Rx / OTC | Rx or OTC | Rx or OTC | Identical |
| Mode of<br>operation | Aluna 2 is a Lilly type<br>pneumotachometer that<br>uses a differential pressure<br>sensor to measure the<br>pressure difference across<br>the stainless-steel mesh.<br>The resulting pressure<br>drop is used to calculate<br>FEV1 and PEF. | Aluna is a Lilly type<br>pneumotachometer that<br>uses a differential pressure<br>sensor to measure the<br>pressure difference across<br>the stainless-steel mesh.<br>The resulting pressure<br>drop is used to calculate<br>FEV1 and PEF. | Identical |
| Dimensions | 90.3 mm x 43 mm x 115.5<br>mm | 90.3 mm x 43 mm x 115.5<br>mm | Identical |
| Weight | 95 g | 95 g | Identical |
| FEV1 Max | 8 L | 8 L | Identical |
| FEV1 | ±2.5% or 0.05 L | ±3% or 0.05 L whichever | Similar; Tightened |
| accuracy | whichever is greater | is greater | accuracy limits as required<br>by 2019 ATS Update v.<br>2005 ATS Guidelines |
| FEV1<br>linearity | <2.5% | Unspecified | Similar; Stating new<br>parameter and limit<br>required by 2019 ATS<br>Update that was not<br>required in 2005 ATS<br>Guidelines |
| | Subject Device | Predicate Device | Comparison |
| | Aluna 2<br>(K232588) | Aluna<br>(K193311) | |
| FEV1<br>repeatability | ±2.5% or 0.05 L<br>whichever is greater | Unspecified | Similar; Stating new<br>parameter and limit<br>required by 2019 ATS<br>Update that was not<br>required in 2005 ATS<br>Guidelines |
| FEV1<br>impedance | ≤0.15 kPa/(l/s) | Unspecified | Similar; Stating new<br>parameter and limit<br>required by 2019 ATS<br>Update that was not<br>required in 2005 ATS<br>Guidelines |
| PEF max | 14 L/s | 14 L/s | Identical |
| PEF<br>accuracy | ±10% or 0.17 L/s<br>whichever is greater | ±10% or 0.3 L/s<br>whichever is greater | Similar; Tightened<br>accuracy limits as required<br>by 2019 ATS Guidelines<br>v. 2005 ATS Guidelines |
| PEF<br>linearity | ≤5% | Unspecified | Similar; Stating new<br>parameter and limit<br>required by 2019 ATS<br>Update that was not<br>required in 2005 ATS<br>Guidelines |
| PEF<br>resistance to<br>flow | ≤0.36 kPa/l/s | Unspecified | Similar; Stating new<br>parameter and limit<br>required by 2019 ATS<br>Update that was not<br>required in 2005 ATS<br>Guidelines |
| PEF<br>frequency<br>response | ≤0.25 l/s), or 12 %,<br>whichever is greater | Unspecified | Similar; Stating new<br>parameter and limit<br>required by 2019 ATS<br>Update that was not<br>required in 2005 ATS<br>Guidelines |
| Materials | Polymers, stainless steel | Polymers, stainless steel | Identical |
| Transmission<br>Method | Bluetooth 4.0 LE | Bluetooth 4.0 LE | Identical |
| Power Source | Rechargeable battery | Rechargeable battery | Identical |
| Smartphone<br>compatibility | iPhone (iOS 15 or later)<br>Android (8 or later) | iPhone (iOS 11 or later) | Similar; Updated to<br>include Android<br>application and to include<br>newer versions of iOS |
| Sterilization | Non-sterile | Non-sterile | Identical |
| | Subject Device | Predicate Device | Comparison |
| | Aluna 2<br>(K232588) | Aluna<br>(K193311) | |
| App<br>functions | • Primary user interface<br>• Data acquisition and<br>processing<br>• Data display<br>• Data transmission to<br>server | • Primary user interface<br>• Data acquisition and<br>processing<br>• Data display<br>• Data transmission to<br>server | Identical |
| Game<br>function | Yes or No depending on<br>user preference | Yes | Similar; Updated to allow<br>the game function to be<br>switched off depending on<br>user preference. |
| Data Storage | Cloud server | Cloud server | Identical |
| Data sharing | Yes, via email | Yes, via email | Identical |
{7}------------------------------------------------
{8}------------------------------------------------
## SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate safety based on current industry standards:
- . Electrical Safety
- IEC 60601-1 Medical electrical equipment Part 1: General requirements for O basic safety and essential performance
- IEC 60601-1-6 Medical electrical equipment Part 1-6: General requirements for o basic safety and essential performance - Collateral standard: Usability
- IEC 60601-1-11 General requirements for basic safety and essential performance о — Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- Electromagnetic Compatibility .
- 0 IEC 60601-1-2 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- Battery
- IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid O electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
{9}------------------------------------------------
- . FEV1/PEF measurement accuracy, repeatability, linearity, and Frequency Response per the American Thoracic Society Technical Statement on "Standardization of Spirometry 2019 Update" to meet the requirements of:
- o ISO 26782:2009 Anaesthetic and respiratory equipment Spirometers intended for the measurement of time forced expired volumes in humans
- ISO 23747:2015 Anaesthetic and respiratorv equipment Peak expiratory flow O meters for the assessment of pulmonary function in spontaneously breathing humans
- . Software Verification
- o IEC 62304 Medical device software Software life cycle processes
- Biocompatibility
- o FDA Guidance: Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (issued September 8, 2023) Note: The patient contacting materials are identical to the predicate device. Biocompatibility testing was leveraged from the predicate device to support substantial equivalence.
- . Cybersecurity
- o FDA Guidance: Cybersecurity in Medical Devices: Oualitv System Considerations and Content of Premarket Submissions (issued September 27, 2023)
The results of these tests indicate that Aluna 2 is substantially equivalent to the predicate device.
## SUMMARY OF CLINICAL TESTING
Clinical testing was not required.
## CONCLUSION
Based on the testing performed, including software testing, electrical safety and EMC testing, battery testing and non-clinical performance bench testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. Aluna 2 has the same design and dimensions and uses similar materials as the first generation Aluna. The subject device has the same intended use and similar technological characteristics as the predicate device. Aluna 2 is assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.