Stable-L Standalone Lumbar Interbody System, Stable-C Interbody System

K232530 · Nexus Spine, LLC · OVD · Sep 18, 2023 · Orthopedic

Device Facts

Record IDK232530
Device NameStable-L Standalone Lumbar Interbody System, Stable-C Interbody System
ApplicantNexus Spine, LLC
Product CodeOVD · Orthopedic
Decision DateSep 18, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Stable-L Standalone Lumbar Interbody System is indicated for spinal fusion procedures in skeletally mature patients. Stable-L is designed for use with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are to be used in patients who have had at least six months of non-operative treatment. Stable-L is intended for use in interbody fusions in the lumbar spine from L2 to S1 at one or two adjacent levels in the treatment of symptomatic degenerative disc disease (DDD). The DDD patients may also have up to Grade 1 spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Each interbody fusion device having a lordotic angle 20° or less is intended to be used with the bone screws provided and requires no additional fixation. However, for hyperlordotic devices (> 20° lordosis), due to the increased risk of anterior migration with hyperlordotic implants, the devices should be used with the bone screws provided and supplemental fixation such as posterior fixation. The Nexus Spine Stable-C Interbody System is an anterior cervical interbody fusion system indicated for use in skeletally mature patients with cervical disc disease (DDD) at one level from C2-T1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Stable-C Interbody System is a standalone system intended to be used with the bone anchors provided. The system is intended to be used with autogenous or allogenic bone graft comprised of cancellous, cortical, and/or corticocancellous bone graft to facilitate fusion. The system is to be used in patients who have had six weeks of non-operative treatment.

Device Story

Stable-L and Stable-C are interbody fusion systems consisting of additively manufactured titanium alloy cages and associated bone screws or anchors. Systems are implanted by surgeons during spinal fusion procedures to facilitate bone growth using autogenous or allogenic bone graft. Stable-L is used in the lumbar spine; Stable-C is used in the cervical spine. Devices provide structural support and stabilization of the disc space. Subject submission updates the delivery method from non-sterile to sterile via gamma radiation.

Clinical Evidence

No clinical data. Bench testing limited to sterilization validation, packaging validation, and shelf-life validation to support the change to gamma radiation sterilization.

Technological Characteristics

Interbody cages (Ti-6Al-4V ELI per ASTM F3001) and fixation screws/anchors (Ti-6Al-4V ELI per ASTM F136). Additively manufactured. Provided sterile via gamma radiation.

Indications for Use

Indicated for skeletally mature patients with symptomatic degenerative disc disease (DDD) requiring spinal fusion. Stable-L: Lumbar spine (L2-S1), 1-2 levels, DDD with/without Grade 1 spondylolisthesis, requires 6 months non-operative treatment. Stable-C: Cervical spine (C2-T1), 1 level, requires 6 weeks non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ September 18, 2023 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the full name. The full name of the agency is "U.S. Food & Drug Administration". Nexus Spine, LLC % Christine Scifert Partner MRC Global 9085 E. Mineral Cir., Suite 110 Centennial. Colorado 80112 Re: K232530 Trade/Device Name: Stable-L Standalone Lumbar Interbody System, Stable-C Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD. OVE Dated: August 18, 2023 Received: August 21, 2023 Dear Christine Scifert: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Katherine D. Kavlock -S for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232530 Device Name Stable-L Standalone Lumbar Interbody System ### Indications for Use (Describe) The Stable-L Standalone Lumbar Interbody System is indicated for spinal fusion procedures in skeletally mature patients. Stable-L is designed for use with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are to be used in patients who have had at least six months of non-operative treatment. Stable-L is intended for use in interbody fusions in the lumbar spine from L2 to S1 at one or two adjacent levels in the treatment of symptomatic degenerative disc disease (DDD). The DDD patients may also have up to Grade 1 spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Each interbody fusion device having a lordotic angle 20° or less is intended to be used with the bone screws provided and requires no additional fixation. However, for hyperlordotic devices (> 20° lordosis), due to the increased risk of anterior migration with hyperlordotic implants, the devices should be used with the bone screws provided and supplemental fixation such as posterior fixation. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232530 Device Name Stable-C Interbody System ### Indications for Use (Describe) The Nexus Spine Stable-C Interbody System is an anterior cervical interbody fusion system indicated for use in sketally mature patients with cervical disc disease (DDD) at one level from C2-T1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Stable-C Interbody System is a standalone system intended to be used with the bone anchors provided. The system is intended to be used with autogenous or allogenic bone graft comprised of cancellous, cortical, and/or corticocancellous bone graft to facilitate fusion. The system is to be used in patients who have had six weeks of non-operative treatment. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------| | <div> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary # Stable-L Standalone Lumbar Interbody System and Stable-C Interbody System August 31, 2023 | Company: | Nexus Spine, LLC<br>2825 East Cottonwood Parkway Suite 330<br>Salt Lake City, UT 84121 | |---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Primary/Secondary Contact: | Christine Scifert – Partner<br>MRC Global<br>9085 E. Mineral Cir., Suite 110<br>Centennial, CO 80112<br>Phone: (901) 831-8053<br>Email: christine.scifert@AskMRCGlobal.com | | | Jen McBride - Regulatory Consultant<br>(901) 481-5902<br>Email: jen.mcbride@askmcrglobal.com | | Company Contact: | Jared Crocker<br>Vice President of Quality and Regulatory Affairs<br>Nexus Spine, LLC<br>Phone: (801) 702-8592<br>jared.crocker@nexusortho.com | | Trade Name: | Stable-L Standalone Lumbar Interbody System<br>Stable-C Interbody System | | Common Name:<br>Classification: | Intervertebral Fusion Device With Integrated Fixation, Lumbar<br>Intervertebral Fusion Device With Integrated Fixation, Cervical<br>Class II | | Regulation: | 21 CFR 888.3080 (Intervertebral Fusion Device With Integrated<br>Fixation, Lumbar; Intervertebral Fusion Device With Integrated<br>Fixation, Cervical) | | Panel: | Orthopedic | | Product Code: | OVD, OVE | | Primary Predicate: | Nexus Spine, LLC Stable-L Standalone Lumbar Interbody System -<br>K212498 | {5}------------------------------------------------ # Device Description: The Stable-L Interbody System is a stand-alone lumbar interbody system including an interbody cage additively manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F3001 and associated bone screws manufactured from titanium alloy per ASTM F136. The device is offered in a variety of sizes to accommodate patient anatomy. Stable-L Interbody devices are provided nonsterile and are expected to be sterilized by the end-user prior to use. The Stable-C Interbody System is an anterior cervical interbody device comprised of an interbody cage (lordotic angles of 0 , 6°, and 12°) made from titanium alloy (Ti-6Al-4V) per ASTM F3001 and two fixation anchors made from titanium alloy (Ti-6-Al-4V ELI) per ASTM F136. The device is offered in a variety of sizes to accommodate patient anatomy. The purpose of this Special 510(k) submission is to gain clearance for the subject devices to be provided sterile via gamma radiation. # Indications for Use: The Stable-L Standalone Lumbar Interbody System is indicated for spinal fusion procedures in skeletally mature patients. Stable-L is designed for use with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are to be used in patients who have had at least six months of non-operative treatment. Stable-L is intended for use in interbody fusions in the lumbar spine from L2 to S1 at one or two adjacent levels in the treatment of symptomatic degenerative disc disease (DDD). The DDD patients may also have up to Grade 1 spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Each interbody fusion device having a lordotic angle 20° or less is intended to be used with the bone screws provided and requires no additional fixation. However, for hyperlordotic devices (> 20° lordosis), due to the increased risk of anterior migration with hyperlordotic implants, the devices should be used with the bone screws provided and supplemental fixation such as posterior fixation. The Nexus Spine Stable-C Interbody System is an anterior cervical interbody fusion system indicated for use in skeletally mature patients with cervical disc disease (DDD) at one level from C2-T1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Stable-C Interbody System is a stand alone system intended to be used with the bone anchors provided. The system is intended to be used with autogenous or allogenic bone graft comprised of cancellous, cortical, and/or corticocancellous bone graft to facilitate fusion. The system is to be used in patients who have had six weeks of non-operative treatment. # Substantial Equivalence: The subject Nexus Spine Stable-L Standalone Lumbar Interbody System and Stable-C Interbody System are substantially equivalent to the legally marketed predicate devices: {6}------------------------------------------------ Primary Predicate: - Nexus Spine, LLC, Stable-L Standalone Lumbar Interbody System -K212498 ### Secondary Predicate: - Nexus Spine, LLC, Stable-C Interbody System - K231763 The subject components are identical in indications, sizing and geometry, technological characteristics, and materials to the predicates. The only difference between the subject device and predicate device is that the subject device is provided sterile via gamma radiation and the predicate device is provided nonsterile. Thus, it can be concluded that the subject does not raise new questions about safety and effectiveness. ### Performance Testing: The is no change to the intended use, indication for use, materials, geometry, or dimensions to the Stable-L Stnadalone Lumbar Interbody System or Stable-C Interbody System. There is no mechanical performance testing included in this submission. Sterilization validation, packaging validations, and shelf life validation have all been performed to support the gamma radiation sterilization change. ### Conclusion: The subject device is determined to be substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...