The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.
Device Story
Fingertip pulse oximeter; reusable; intended for adult patients in healthcare settings. Device uses probe with LEDs (red/infrared) and photodetector to measure light absorption through finger tissue. Microprocessor calculates SpO2 and pulse rate based on light transmission variations. Models Alpine20B/BA/BH/BHA, Alpine30B/BA/BH/BHA, Alpine20LB/LBH, and Alpine30LB/LBH include Bluetooth for data transmission. Operated by clinicians for spot-checks; output displayed on device screen. Provides real-time oxygen saturation and pulse rate data to assist clinical decision-making. Benefits include non-invasive monitoring of patient oxygenation status.
Clinical Evidence
Clinical study conducted per ISO 80601-2-61:2017 and FDA guidance. 12 healthy adult volunteers (ages 22-37, Fitzpatrick skin tones I-VI, 5 female/7 male) evaluated SpO2 accuracy against arterial CO-Oximetry. Results confirmed 2% accuracy in the 70-100% SpO2 range under stationary conditions.
Indicated for spot-checking oxygen saturation and pulse rate in adult patients in healthcare environments. Not for use under motion or low perfusion conditions.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Shenzhen AOJ Medical Technology Co.,Ltd., AOJ-70A, AOJ-70B, AOJ-70C, AOJ-70D and AOJ-70E Pulse Oximeter (K202173)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
November 21, 2024
Shenzhen Smallsignal Technology Co., Ltd. Qingliang Liu General Manager Room 901-7,Building 1, No.9, Jinxiu Middle Road, Laokeng Community, Longtian Street, Pingshan Shenzhen, Guangdong 518122 China
Re: K232520
Trade/Device Name: Fingertip Pulse Oximeter, models Alpine20, Alpine20A, Alpine20H, Alpine20HA, Alpine20B, Alpine20BA, Alpine20BH, Alpine20BHA, Alpine30, Alpine30A, Alpine30H, Alpine30HA, Alpine30B, Alpine30BA, Alpine30BH, Alpine30BHA, Alpine20L, Alpine20LB, Alpine20LH, Alpine20LBH, Alpine30L, Alpine30LB, Alpine30LH, Alpine30LBH Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: March 2, 2024 Received: October 22, 2024
Dear Qingliang Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K232520
Device Name
Fingertip Pulse Oximeter (Models Alpine20, Alpine20H, Alpine20H, Alpine20HA, Alpine20B, Alpine20BA, Alpine20BH, Alpine30, Alpine30, Alpine30H, Alpine30H, Alpine30HA, Alpine30B, Alpine30BA, Alpine30BH, Alpine20L, Alpine20L, Alpine20LB, Alpine20LH, Alpine20LBH, Alpine30L, Alpine30LB, Alpine30LH, Alpine30LBH)
Indications for Use (Describe)
The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in healthcare environment. And it is not intended to be used under motion or low perfusion scenarios.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### Prepared in accordance with the requirements of 21 CFR Part 807.92
- 1. Submitter: Shenzhen Smallsignal Technology Co., Ltd. Address: Room 40205, Building 1, Huahan Science and Technology Industrial Park, No. 19, Qiyun West Road, Heping Community, Pingshan Street, Pingshan District, Shenzhen, Guangdong Province, China, 518083 TEL: 86 755 23010489
- Contact Person: Qingliang Liu
- Prepare date: October 22, 2024
2. Device name Device Name: Fingertip Pulse Oximeter
and classification Models: Alpine20, Alpine20A, Alpine20H, Alpine20HA, Alpine20BA, Alpine20BA, Alpine20BH, Alpine20BHA, Alpine30, Alpine30A, Alpine30H, Alpine30HA, Alpine30B, Alpine30BA, Alpine30BH, Alpine30BHA, Alpine20L, Alpine20LB, Alpine20LH, Alpine20LBH, Alpine30L, Alpine30LB, Alpine30LH, Alpine30LBH Classification Name: 21 CFR 870.2700 Oximeter
Product code: DQA
Regulatory Class: Class II
- New submission. No prior submission associated with the current submission. 3. Reason for
Submission
Shenzhen AOJ Medical Technology Co.,Ltd., AOJ-70A, AOJ-70B, AOJ-70C, AOJ-70D and 4. Predicate Device(s) AOJ-70E Pulse Oximeter / K202173
- 5. Device The oximeter consists of probe, electronic circuits, and display and plastic enclosures. And one Description side of probe is designed to locate light emitting diodes and a light detector (called a photo-detector). Red and Infrared lights are shone through the tissues from one side of the probe to the other. Then parts of the light emitted absorbed by blood and tissues. The light absorbed by the blood varies with the oxygen saturation of haemoglobin. After that, the photo-detector detects the light volume transmitted through the tissues which depends on blood pulse, Hereafter, the microprocessor calculates a value for the oxygen saturation (SpQ2), The subject device is a reusable device, and need to reprocess as suggested in the user manual after each use. And the device is intended to be used on the finger, and powered by 2*1.5V AAA battery. Bluetooth function is available for Alpine20B, Alpine20BA, Alpine20BH, Alpine20BHA, Alpine30B, Alpine30BA, Alpine30BH, Alpine20LB, Alpine20LB, Alpine20LBH, Alpine30LB and Alpine30LBH only.
#### 6. Indications for The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in healthcare environments. And it is Use not intended to be used under motion or low perfusion scenarios.
{5}------------------------------------------------
### 7. Predicate Device Comparison
Please refer to following table to find differences between the subject device and predicate device.
| ITEM | Proposed Device<br>Alpine20 series Pulse<br>Oximeter | Predicate Device<br>AOJ-70A, AOJ-70B, AOJ-70C,<br>AOJ-70D and AOJ-70E | Comparison<br>Result |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacture | Shenzhen Smallsignal<br>Technology Co., Ltd. | Shenzhen AOJ Medical<br>Technology Co., Ltd. | --- |
| Indications for<br>Use | The pulse oximeter is a<br>reusable device and intended<br>for spot-checking of oxygen<br>saturation and pulse rate for<br>use with the finger of adult<br>patients in healthcare<br>environments. And it is not<br>intended to be used under<br>motion or low perfusion<br>scenarios. | The pulse oximeter is a reusable<br>device and intended for<br>spot-checking of oxygen<br>saturation and pulse rate for use<br>with the finger of adult patients<br>in healthcare environments. And<br>it is not intended to be used<br>under motion or low perfusion<br>scenarios. | Same |
| Operational Specifications | | | |
| Intended patient<br>population | Adult | Adult | Same |
| Intended<br>application site | Finger | Finger | Same |
| use under<br>motion and low<br>perfusion<br>conditions | No | No | Same |
| Measurement<br>Principles | 2-wavelength Relative<br>Optical Absorption | 2-wavelength Relative<br>Optical Absorption | Same |
| Signal Detection<br>Method | Photodetector | Photodetector | Same |
| SpO2 Range | 35~100% | 0~100% | Different¹ |
| SpO₂ Resolution | 1% | 1% | Same |
| SpO2 Accuracy | 70~100%: ±2%<br>0% to 69%: unspecified | 70~100%: ±2%<br>0% to 69%: unspecified | Same |
| Pulse Rate Range | 25 bpm ~ 250 bpm | 25 bpm ~ 250 bpm | Same |
| Pulse Rate<br>Accuracy | ±2 bpm | ±2 bpm | Same |
| Pulse<br>Rate Resolution | 1 bpm | 1 bpm | Same |
| Shipped Sterile | No | No | Same |
| Storage and<br>Transport<br>Environment | Temperature: -10 °C - 40 °C<br>Humidity: 15%-93%<br>Atmospheric pressure:<br>50 kPa-106 kPa | Temperature: -20 °C - 50 °C<br>Humidity (non-condensing):<br>10%-95%<br>Atmospheric pressure: 70<br>kPa-106 kPa | Different 2 |
| Operating<br>Environment | Temperature: 5 °C - 40 °C<br>Humidity: 15%-80%<br>Atmospheric pressure: 86<br>kPa-106 kPa | Temperature: 10 °C - 40 °C<br>Humidity (non-condensing):<br>15%-95%<br>Atmospheric pressure: 70<br>kPa-106 kPa | |
| Compliance Standards | | | |
| Bio-compatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-23 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Different 3 |
| Electrical Safety | IEC 60601-1<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-11 | Same |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | |
| Performance | ISO 80601-2-61 | ISO 80601-2-61 | |
| Physical Specifications | | | |
| Dimension<br>(Width*Height*<br>Depth) | Alpine20/ 20L series:<br>61mm×37mm×33mm<br>Alpine30/ 30L series:<br>61mm×34mm×32mm | 57mm×30mm×30 mm | Different 4 |
| Data Transmission | | | |
| Data<br>transmission | Bluetooth is available for<br>Alpine20B, Alpine20BA,<br>Alpine20BH, Alpine20BHA,<br>Alpine30B, Alpine30BA,<br>Alpine30BH, Alpine30BHA,<br>Alpine20LB, Alpine20LBH,<br>Alpine30LB and Alpine30LBH | No available | Different 5 |
Table 1 Comparison between the predicate and the subject device
{6}------------------------------------------------
#### Justification for the differences:
- 1) Different SpO2 Range
The measuring range of the predicate is 0~100% , while that of the subject is 35~100%, and such specification is verified per the international standard ISO 80601-2-61, same as the standard complied with of the predicate, so the different range will be acceptable for the subject device.
- 2) Different operating, storage and transport environment
Minor difference to operation, storage/transport environment for the subject device, but the system has been proved to be safe and effective since the safety testing was conducted under the suggested environment; Moreover, environment testing data shows the device can work as declared under the suggested conditions. So those changes will not cause any safety and effectiveness problem.
- 3) Different Biocompatibility test standard
{7}------------------------------------------------
The endpoints for the biocompatibility are the same, which include cytoxicity, skin irritation and sensitization, the difference happened due to the irritation was taken from ISO 10993-10 to ISO 10993-23 without no change for the parts applied in the subject device, both of which are used for evaluation for test of irritation. This difference will not raise any safety and effectiveness questions.
- 4) Different Physical Specifications
The subject and the predicate are of different size but proximity. Moreover, such engineering design has been verified during the international standards, so such minor different will not raise any safety and effectiveness questions.
- 5) Different data transmission function
Some models of the subject devices have Bluetooth data transmission function, and the wireless function has been verified to assure its data transmission and cybersecurity. No new safety and effectiveness questions raised.
As seen in the comparison tables, the subject and predicate devices have same design principle, similar design features and performance specifications. The different technological characteristics between the subject and predicate devices will not raise different questions of safety or effectiveness as demonstrated in the non-clinical and clinical evidence.
#### 8. Performance Testing
Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.
#### Non-Clinical Testing:
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility testing
The biocompatibility evaluation for the Pulse Oximeter was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the testing included the following tests, results of which demonstrate the biocompatibility of the subject device:
- Cytotoxicity
- Skin Sensitization
- Skin Irritation
#### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the IEC 60601-1: 2005+A1: 2020 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2015+A1:2020 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests standard for EMC.
#### Bench Testing
Bench testing was conducted and the results show that the subject device complies with the ISO 80601-2-61: 2017 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and
{8}------------------------------------------------
essential performance of Pulse Oximeter Equipment standard. And Pulse Rate Accuracy meets the requirements defined in ISO 80601-2-61, Clause 201.12.1.104.
#### Software Verification and Validation Testing
Software documentation including verification & validation was provided in accordance with FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern.
#### Clinical data
Clinical testing is conducted per Annex EE Guideline for evaluating and documenting SpO2ACCURACY in human subjects of ISO 80601-2-61:2017 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment and the FDA Guidance Document for Pulse Oximeters. The subject device of this study was to evaluate the Sp02 accuracy performance of the subject device during stationary (non motion) conditions over a wide range of arterial blood oxygen saturation levels as compared to arterial blood CO-Oximetry. There were 12 adult healthy volunteers to validated the SpO2 accuracy of the pulse oximeter, aged 22-37 years old, with skin tones varying from Fitzpatrick I-VI,including five females and seven males. The measure results showed that the subject device perform as claimed during the measurement range of 70-100% with 2% accuracy.
#### 9. Conclusion
Verification and validation testing was conducted on the subject device Fingertip Pulse Oximeter and all testing passed pre-specified criteria. The subject device and the predicate device have the same intended use and the differences in technological features do not raise different questions of safety and effectiveness. This premarket notification submission demonstrates that the subject device is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.