The Medline ReNewal Reprocessed LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surqical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include, but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis. The instrument is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, parotidectorny, and tonsillectomy) for ligation and division of vessels. Ivmphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands. The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
Device Story
Sterile, single-use, handheld bipolar vessel sealing device; used with compatible electrosurgical generators. Input: RF energy from generator; user-applied mechanical pressure via hemostat-style handle. Transformation: RF energy delivered to tissue clamped between jaws for vessel/tissue ligation and division. Output: Sealed and divided tissue. Used in open surgical procedures (general, urologic, thoracic, plastic, ENT); operated by surgeons. Device allows blunt dissection and tissue grasping. Benefits: Enables efficient vessel ligation and division in open surgery; reduces need for multiple instruments. Reprocessed version of predicate; lacks original nano-coating on jaws.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical testing: functional performance (simulated use, artificial soiling), visual inspection, device functionality, in vivo acute and chronic animal studies, electrical safety, cleaning validation, biocompatibility, sterilization validation, packaging/shelf-life validation, bioburden testing, and ethylene oxide/ethylene chlorohydrin residuals testing.
Technological Characteristics
Bipolar electrosurgical instrument; hemostat-style handle with single activation button. Energy source: RF energy via compatible electrosurgical generator. Materials: Reprocessed components from predicate. Sterilization: Ethylene oxide. Connectivity: Wired connection to generator. No software/algorithm.
Indications for Use
Indicated for open surgical procedures in adults requiring ligation/division of vessels (up to 7mm), tissue bundles, and lymphatics. Specialties: general, urologic, thoracic, plastic, reconstructive, and ENT (thyroidectomy, radical neck dissection, parotidectomy, tonsillectomy). Contraindicated for tubal sterilization/coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Underneath that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 29, 2024
Surgical Instrument Service and Savings Inc. Stephanie Boyle Mays Senior Regulatory Affaiirs Specialst, QA/RA (dba Medline ReNewal) 1500 NE Hemlock Ave. Redmond, Oregon 97756
Re: K232511
Trade/Device Name: Medline ReNewal Reprocessed LigaSure Exact Dissector, Nano-coated (LF2019) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NUJ, GEI Dated: August 17, 2023 Received: August 18, 2023
Dear Stephanie Boyle Mays:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.04.29 Mark Trumbore, Ph.D. Assistant Director, THT4A1: Robotically-Assisted Surgical Devices Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality
{3}------------------------------------------------
## Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
Submission Number (if known)
232511
Device Name
Medline ReNewal Remanufactured LigaSure Exact Dissector, without Nano-coating (LF2019)
#### Indications for Use (Describe)
The Medline ReNewal Reprocessed LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surqical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include, but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis. The instrument is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, parotidectorny, and tonsillectomy) for ligation and division of vessels. Ivmphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands.
The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline Renewal logo. On the left is the Medline logo, which is a blue square with a white cross inside. To the right of the Medline logo is the word "Renewal" in large, green and blue letters. Below the word "Renewal" is the phrase "Full Circle Reprocessing" in smaller, blue letters.
# 510(k) 232511 Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR § 807.92.
| Submitter/<br>Owner | Surgical Instrument Service and Savings (dba Medline ReNewal)<br>1500 NE Hemlock Ave., Redmond, OR 97756 | | |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|--|
| Contact/<br>Prepared by | Stephanie Boyle Mays<br>Senior Regulatory Affairs Specialist, Quality Assurance/Regulatory Affairs<br>P: 541-516-4205 • F: 541-923-3375 • smays@medline.com | | |
| Date Prepared | August 17, 2023 | | |
| Device Name<br>and<br>Classification | Proprietary/Trade<br>Name: | Medline ReNewal Remanufactured LigaSure Exact<br>Dissector, without Nano-coating | |
| | Common or Usual<br>Name | Bipolar sealing device | |
| | Regulatory<br>Name/Reference | Electrosurgical, cutting & coagulation & accessories<br>21 CFR § 878.4400 | |
| | Regulatory Class | 2 | |
| | Product Code | GEI | |
| | Panel | General & Plastic Surgery | |
| | 510(k) Number | K202917 | |
| Predicate<br>Device | Proprietary or<br>Trade Name | LigaSure Exact Dissector, Nano-coated | |
| | Common or Usual<br>Name | Bipolar sealing device | |
| | Regulatory<br>Name/Reference | Electrosurgical, cutting & coagulation & accessories<br>21 CFR § 878.4400 | |
| | Regulatory Class | 2 | |
| | Product Code | GEI | |
| | Panel | General & Plastic Surgery | |
| | Manufacturer | Covidien LLC, 5920 Longbow Dr., Boulder, CO 80301 | |
| Device<br>Description | The Medline ReNewal Remanufactured LigaSure Exact Dissector, Nano-<br>coated (LF2019) device is a sterile, single-use, handle-held bipolar vessel<br>sealing device designed for use with compatible Covidien electrosurgical<br>generators that include vessel sealing capabilities to ligate (seal) and divide<br>(cut) vessels, tissue bundles, and lymphatics clamped between the jaws,<br>grasp tissue, and perform blunt dissection during general surgical<br>procedures (as indicated) using radio frequency (RF) energy.<br>A hemostat style body allows the user to open or close the instrument jaws<br>during vessel sealing and cutting, with a single activation button<br>incorporated into the body of the device. | | |
| Technological<br>Characteristics | The technological characteristics, materials, and the fundamental scientific<br>technology of the subject device is equivalent to the predicate device. The<br>proposed device is a reprocessed version of the predicate device. K202917<br>LigaSure Exact Dissector, Nano-coated was used as the primary predicate<br>to support intended use, technological characteristics, and functional<br>performance specifications. | | |
| | The original equipment manufacturer of the LigaSure Exact Dissector,<br>Nano-coated applied a nano-coating to the device jaws to reduce tissue<br>sticking. Medline ReNewal does not replace the coating on the device jaws<br>in its process of decontamination, inspections, packaging and sterilization. | | |
| Non-clinical<br>Testing<br>Summary | The functional characteristics of the subject device have been evaluated<br>and found to be substantially equivalent to the predicate device based on<br>the following tests:<br>Functional performance: simulated use and artificial soiling; visual inspection; device functionality; in vivo acute and chronic animal studies Electrical Safety Cleaning Validation: Biocompatibility: Sterilization Validation Packaging and shelf life validation; bioburden testing; and ethylene oxide and ethylene chlorohydrin residuals testing Product stability | | |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Medline Renewal Full Circle Reprocessing. The Medline logo is on the left and consists of a blue square with a white cross inside and the word "MEDLINE" below the cross. To the right of the Medline logo is the word "Renewal" in large, bold, green and blue letters. Below "Renewal" is the phrase "Full Circle Reprocessing" in smaller, black letters.
Product stability
#### Summary Table: Predicate and Medline ReNewal Reprocessed LigaSure Exact Dissector, Nano-coated comparison chart.
| | Predicate | Proposed | Comparison |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| View | Covidien LigaSure Exact<br>Dissector, Nano-coated | Medline ReNewal<br>Remanufactured LigaSure<br>Exact Dissector, without<br>Nano-coating | As Stated |
| 510(k) | K202917 | K232511 | N/A |
| Model Number | LF2019 | LF2019 | As stated |
| Common Name | Bipolar sealing device | Bipolar sealing device, | Same |
| Regulation No. | 21 CFR § 878.4400 | 21 CFR § 878.4400 | Same |
| Regulatory<br>Class | 2 | 2 | Same |
| Product Code | GEI | GEI | Same |
| Indications for<br>Use | LigaSure Sealer/Divider is a<br>bipolar electrosurgical<br>instrument intended for use in<br>open surgical procedures<br>where ligation and division of<br>vessels, tissue bundles, and<br>lymphatics is desired. The<br>LigaSure Sealer/ Divider can<br>be used on vessels (arteries<br>and veins) up to and including<br>7 mm. It is indicated for use in<br>general surgery and in such<br>surgical specialties as<br>urologic, thoracic, plastic, and<br>reconstructive. Procedures<br>may include, but are not<br>limited to, bowel resections,<br>gall bladder procedures,<br>Nissen fundoplication, breast<br>procedures, and adhesiolysis.<br>The instrument is also<br>indicated for open ENT<br>procedures in adults<br>(thyroidectomy, radical neck<br>dissection, parotidectomy, and<br>tonsillectomy) for ligation and<br>division of vessels, lymphatics<br>and tissue bundles 2-3 mm<br>away from unintended<br>thermally-sensitive structures<br>such as nerves and<br>parathyroid glands.<br>The LigaSure system has not<br>been shown to be effective for<br>tubal sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the<br>LigaSure system for these<br>procedures. | The Medline ReNewal<br>Remanufactured LigaSure<br>Sealer/Divider is a bipolar<br>electrosurgical instrument<br>intended for use in open<br>surgical procedures where<br>ligation and division of<br>vessels, tissue bundles, and<br>lymphatics is desired. The<br>LigaSure Sealer/ Divider can<br>be used on vessels (arteries<br>and veins) up to and including<br>7 mm. It is indicated for use in<br>general surgery and in such<br>surgical specialties as<br>urologic, thoracic, plastic, and<br>reconstructive. Procedures<br>may include, but are not<br>limited to, bowel resections,<br>gall bladder procedures,<br>Nissen fundoplication, and<br>adhesiolysis.<br>The instrument is also<br>indicated for open ENT<br>procedures in adults<br>(thyroidectomy, radical neck<br>dissection, parotidectomy, and<br>tonsillectomy) for ligation and<br>division of vessels, lymphatics<br>and tissue bundles 2-3 mm<br>away from unintended<br>thermally-sensitive structures<br>such as nerves and<br>parathyroid glands.<br>The LigaSure system has not<br>been shown to be effective for<br>tubal sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the<br>LigaSure system for these<br>procedures. | Same |
| Technological<br>Characteristics | To operate, the device is<br>connected to a compatible<br>generator and is operated by<br>either a hand or a food switch. | To operate, the device is<br>connected to a compatible<br>generator and is operated by<br>either a hand or a food switch. | Same |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains two logos. The first logo is a blue square with a white star-like shape in the center and the word "MEDLINE" written in white below the shape. The second logo is the word "ReNewal" in green and blue, with the words "Full Circle Reprocessing" written in smaller font below it.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "ReNewal Full Circle Reprocessing". The word "ReNewal" is in a large font, with "Re" in green and "Newal" in blue. Below that is the phrase "Full Circle Reprocessing" in a smaller font. To the left of the text is the Medline logo, which consists of the word "MEDLINE" in blue, next to a blue symbol that looks like a stylized star or compass.
| a Neither K202917 nor the current submission included the generators as part of their<br>respective submissions. Only device LF2019 is included in the project scope. | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | The predicate and proposed devices in this application have the same<br>indications for use and technological characteristics. Based on this and the<br>non-clinical testing data presented in this 510(k) submission, the Medline<br>ReNewal Remanufactured LigaSure Exact Dissector, without Nano-coating<br>device model LF2019 is substantially equivalent to the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.