The Electric Breast Pump Model: YM-8166 is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.
Device Story
Electric breast pump; diaphragm-type vacuum pump; software-controlled suction and cycle speed. User selects modes (Stimulation, Expression, Pseudo-true Expression, Array) and vacuum levels via buttons; device provides visual feedback on display. Powered by USB-rechargeable battery. Used by lactating women in home settings to express and collect milk. Output is expressed breast milk; device prevents backflow into motor. Benefits include efficient milk expression and collection for the user.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 19, 2024
Guangdong Youmeng Electrical Technology Co., Ltd. % Candice Oiu Registration Engineer Feiying Drug & Medical Consulting Technical Service Group Rm. 2401 Zhenye International Business Center, No. 3101-90. Qianhai Road Shenzhen, Guangdong 518052 China
Re: K232466
> Trade/Device Name: Electric Breast Pump Model: YM-8166 Regulation Number: 21 CFR\$ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: December 5, 2023 Received: December 21, 2023
# Dear Candice Qiu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
Page
3
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for mhre information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological HealtH
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232466
Device Name Electric Breast Pump Model: YM-8166
Indications for Use (Describe)
The Electric Breast Pump Model: YM-8166 is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K232499 510(k) Summary
"510(k) Summary" as required by 21 CFR Part 807.92.
Date Prepared: January 17, 2023
#### I. Submitter
Guangdong Youmeng Electrical Technology Co., Ltd. One of Area A, level 3, Building 2, No. 17, Keyuan 3rd Road, Shunde Hi Tech Industrial Park, Xiaohuangpu Community, Ronggui Street, Shunde District, Foshan, Guangdong, China Post code: 528303 Tel.: +86 13567769890 Liu Zongjun General Manager Tel: +86 13567769890 Email: 223316378@qq.com
# II. Device
Name of Device: Electric Breast Pump Model: YM-8166 Common or Usual Name: Powered breast pump Regulation Name: Powered breast pump Regulation Number: 21 CFR 884.5160 Regulatory Class: II Product Code: HGX (Powered, Breast, Pump)
# III. Predicate Device
Predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Clearance Date |
|----------------------------|----------------------------------------------------------------------------------------------------------------------|---------------|----------------|
| Zhejiang Carebao Co., Ltd. | Wearable Breast Pump (YD-1193, YD-1193S, YD1195, YD-1195S, YD-1196, YD-1198, YD-1199, YD-1196S, YD- 1198S, YD-1199S) | K222782 | Jan 13, 2023 |
The predicate device has not been subject to a design-related recall.
{5}------------------------------------------------
# IV. Device Description
The Electric Breast Pump Model: YM-8166 is an over-the-counter, non-sterile, single-user, powered breast pump intended to be used by lactating women to express and collect milk from their breasts. The device is intended for daily use in a home environment. The device uses a diaphragm-type vacuum pump driven by software embedded in the device to generate suction required to stimulate and express breast milk. The software provides control over vacuum pressure and cycle speed.
The following patient-contacting materials are used in the subject device:
| Component<br>name | Material of Component | Body Contact Category | Contact Duration |
|---------------------------------------------------------------------|-----------------------|-------------------------------------------|--------------------|
| Electric Breast<br>Pump Model:<br>YM-8166<br>(Major<br>ingredients) | PP & Silicone | Surface-contacting<br>device: Intact skin | Less than 24 hours |
Detailed device specifications can be found in the comparison chart below.
# V. Indications for Use
The Electric Breast Pump Model: YM-8166 is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.
# VI. Comparison of Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristic of the subject and predicate device.
| Comparison<br>Elements | Subject Device | Predicate Device | Remark |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) Number | K232499 | K222782 | / |
| Trade name | Electric Breast Pump<br>Model: YM-8166 | Wearable Breast Pump<br>(YD-1193, YD-1193S,<br>YD1195, YD-1195S, YD-<br>1196 YD-1198 YD-1199) | / |
| Comparison<br>Elements | Subject Device | Predicate Device | Remark |
| | | YD-1196S, YD- 1198S, YD-<br>1199S) | |
| Manufacturer | Guangdong Youmeng<br>Electrical Technology<br>Co., Ltd. | Zhejiang Carebao Co., Ltd. | / |
| Regulation number | 21 CFR 884.5160 | 21 CFR 884.5160 | Same |
| Product code | HGX | HGX | Same |
| Device<br>classification | Class II | Class II | Same |
| Indication for use/<br>Intended use | The Electric Breast<br>Pump Model: YM-8166<br>is intended to express<br>milk from lactating<br>women in order to<br>collect milk from their<br>breasts. The device is<br>intended for a single<br>user. | The Wearable Breast Pump is<br>intended to express milk from<br>lactating women in order to<br>collect milk from their<br>breasts. The device is<br>intended for a single user. | Same |
| Prescription or<br>OTC | OTC | OTC | Same |
| Pump Options | Single, Double | Single | Different |
| Provide Non-sterile | Yes | Yes | Same |
| Re-usable | Yes | Yes | Same |
| Direct user contact | Yes | Yes | Same |
| Backflow<br>Protection | Yes | Yes | Same |
| Suction Modes | Stimulation Mode,<br>Expression Mode,<br>Pseudo-true Expression<br>Mode, Array Mode,<br>and Single/double | Stimulation Mode and<br>Expression Mode<br>Mixed Mode | Different |
| Comparison<br>Elements | Subject Device | Predicate Device | Remark |
| | Pumping Mode | | |
| Suction levels | 16 | 9 | Different |
| Adjustable suction<br>levels | Yes | Yes | Same |
| Vacuum range:<br>Stimulation mode | 60-165±20 (mmHg) | YD-1193,YD -1193S, YD-<br>1195,YD-1195S: 36 mmHg<br>to 241 mmHg<br>YD-1196,YD-1198, YD-<br>1199, YD-1196S, YD-<br>1198S, YD-1199S: 37 mmHg<br>to 242 mmHg | Similar |
| Vacuum range:<br>Expression mode | 85-280±20 (mmHg) | YD-1193,YD -1193S, YD-<br>1195,YD-1195S: 69 mmHg<br>to 282 mmHg<br>YD-1196,YD-1198, YD-<br>1199, YD-1196S, YD-<br>1198S, YD-1199S: 71 mmHg<br>to 282 mmHg | Similar |
| Cycle Speed:<br>Stimulation | Single Suction: 66 to<br>138±5 cycles/minute<br>Double suction: 64 to<br>180±5 cycles/minute | 37.5 to 150 cycles/minute | Different |
| Cycle Speed:<br>Expression | Single Suction: 32 to<br>112±5 cycles/minute<br>Double suction:28 to<br>132±5 cycles/minute | YD-1193,YD -1193S, YD-<br>1195,YD-1195S: 23 to 72<br>cycles/minute<br>YD-1196, YD-1198, YD-<br>1199, YD-1196S, YD-<br>1198S, YD-1199S: 24 to 69<br>cycles/minute | Different |
| Controls | Power button;<br>Mode<br>adjustment<br>button;<br>Vacuum level increase/ | On/Off button; Mode<br>selection Increase/decrease<br>vacuum button; | Similar |
| Comparison<br>Elements | Subject Device | Predicate Device | Remark |
| | decrease button;<br>Night light button<br>(optional) ;<br>Single/ double pumping<br>change button | | |
| Indicator | Yes. General display | Yes. LED | Same |
| Power Supply | USB connector (Only<br>for charging) by DC 5<br>V, 2 A | Li-Ion Battery (internally<br>powered by 3.7Vdc lithium<br>battery or externally powered<br>by 5Vdc USB.) | Same |
| Materials | Milk collection bottle:<br>PP and PPSU<br>(Optional)<br>Breast shield: Silicone<br>rubber<br>Pump unit: ABS and PC | Milk Container:<br>Polypropylene<br>Flange: Silicone Pump Outer<br>Housing: Acrylonitrile<br>Butadiene Styrene (ABS)<br>plastic | Same |
| Software/<br>Firmware/<br>Microprocessor<br>Control? | Yes | Yes | Same |
| Electrical safety | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 62133-2 | AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 62133-2 | Same |
| Biocompatibility | ISO 10993-5<br>ISO 10993-10<br>ISO 10993-23 | ISO 10993-5<br>ISO 10993-10 | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
The Electric Breast Pump Model: YM-8166 has the same intended use, mode of action and similar operational characteristics as the predicate device. The subject device differs from the predicate device in suction modes, strength, and levels, cycle speed, and a double pumping option. The different technological characteristics between the subject device and the predicate device do not raise different questions of safety and effectiveness.
{9}------------------------------------------------
# VII.Summary of Non-Clinical Performance Testing
### 1) Biocompatibility Testing
The biocompatibility evaluation for the patient-contacting components of the Electric Breast Pump Model: YM-8166 was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as follows:
- > ISO 10993-5:2009. Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-10:2021, Biological evaluation of medical devices -Part 10: Tests for skin sensitization
- > ISO 10993-23:2021, Biological evaluation of medical devices -Part 23: Tests for skin irritation
The patient-contacting components of the subject device were shown to be non-cytotoxic, nonirritating, and non-sensitizing.
### 2) Electrical Safety and EMC
Electrical safety and EMC testing was performed per the following standards:
- > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility
- > ANSI AAMI ES60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- > IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid electrolytes -Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
### 3) Software Verification and Validation
Software verification and validation consistent with a basic level of concern per the FDA guidance document "Content of Premarket Submissions for Device Software Functions" dated June 14, 2023. System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met, and all software hazards have been mitigated to acceptable risk levels.
{10}------------------------------------------------
# 4) Additional Non-Clinical Testing
Suction strength and cycle speed of subject devices were tested. All the test results complied with the design specifications of the subject device throughout the use life.
Backflow testing was conducted to ensure that no liquid will backflow into the tubing, and therefore no liquid can backflow into the pump motor. The test results showed that there was no backflow during the test.
Battery performance testing was conducted to demonstrate that the battery remains functional during its stated use-life.
Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.
#### VIII. Conclusions
The results of the performance testing described above demonstrate that the Electric Breast Pump Model: YM-8166 is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.