K232363 · Viz. Ai, Inc. · QIH · Feb 5, 2024 · Radiology
Device Facts
Record ID
K232363
Device Name
Viz HDS, Viz Volume Plus, Viz ICH+
Applicant
Viz. Ai, Inc.
Product Code
QIH · Radiology
Decision Date
Feb 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Intracranial Hyperdensity Volume
Deep-learning algorithm
MAE < 7.5 mL; DICE > 70%
MAE < 7.5 mL; DICE > 70% (lower CI bound)
—
—
Clinical performance study comparing Viz HDS output to ground truth established by trained radiologists.
>1 (trained radiologists)
Lateral Ventricle Volume
Deep-learning algorithm
MAE < 3 mL; DICE > 70%
MAE < 3 mL; DICE > 70% (lower CI bound)
—
—
Clinical performance study comparing Viz HDS output to ground truth established by trained radiologists.
>1 (trained radiologists)
Midline Shift Distance
Deep-learning algorithm
MAE < 2 mm
MAE < 2 mm
—
—
Clinical performance study comparing Viz HDS output to ground truth established by trained radiologists.
>1 (trained radiologists)
Indications for Use
The Viz HDS device is intended for automatic labeling, visualization, and quantification of segmentable brain structures from a set of Non-Contrast CT (NCCT) head scans. The software is intended to automate the current manual process of identifying, labeling, and quantifying the volume of segmentable brain structures identified on NCCT images. Viz HDS provides volumes from NCCT scans acquired at a single time point. The Viz HDS software is indicated for use in the analysis of the following structures: Intracranial Hyperdensities, Lateral Ventricles and Midline Shift. The device output should be reviewed along with patient's original images by a physician.
Device Story
Viz HDS is a cloud-based, software-only device that processes non-contrast head CT (NCCT) scans in DICOM format. It utilizes a locked AI/ML algorithm to automatically segment and quantify intracranial hyperdensities, lateral ventricles, and midline shift. The device generates a summary series (table of measurements) and a segmentation series (DICOM images with RGB overlays of identified structures). These outputs are transmitted to a hospital's PACS for physician review. By automating the manual identification and measurement process, the device assists clinicians in assessing brain structures, potentially improving workflow efficiency and diagnostic support. It is intended for use in clinical settings where NCCT scans are performed.
Clinical Evidence
Performance validated by comparing algorithm output to radiologist-established ground truth. Metrics included Mean Absolute Error (MAE) and DICE scores. MAE upper 95% CI bounds were <7.5 mL for hyperdensity volume, <3 mL for lateral ventricle volume, and <2 mm for midline shift. DICE score lower CI bound for hyperdensities and lateral ventricles was >70%. Performance was stratified by clinical site, patient demographics (age, gender), scanner parameters (manufacturer, model, slice thickness), and lesion size.
Technological Characteristics
Software-only device; cloud-hosted. Uses deep-learning AI/ML algorithm for image segmentation and quantification. Inputs: DICOM NCCT head scans. Outputs: DICOM-formatted summary series and segmentation series with RGB overlays. Connectivity: Networked integration with hospital PACS. No physical materials or energy sources.
Indications for Use
Indicated for analysis of intracranial hyperdensities, lateral ventricles, and midline shift in patients undergoing non-contrast head CT scans. Intended for use by physicians to aid in identifying, labeling, and quantifying these structures.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Viz.ai, Inc. Pooja Shah Regulatory Affairs Specialist 5000 Center Green Way Cary, NC 27513
February 5, 2024
Re: K232363 Trade/Device Name: Viz HDS. Viz Volume Plus, Viz ICH+ Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: January 10, 2024 Received: January 10, 2024
Dear Pooja Shah:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232363
Device Name
Viz HDS, Viz Volume Plus, Viz ICH+
#### Indications for Use (Describe)
The Viz HDS device is intended for automatic labeling, visualization of segmentable brain structures from a set of Non-Contrast CT (NCCT) head scans. The software is intended to automate the current manual process of identifying, labeling, and quantifying the volume of segmentable brain structures identified on NCCT images. Viz HDS provides volumes from NCCT scans acquired at a single time point. The Viz HDS software is indicated for use in the analysis of the following structures: Intracranial Hyperdensities, Lateral Ventricles and Midline Shift. The device output should be reviewed along with patient's original images by a physician.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K232363
Image /page/3/Picture/1 description: The image shows a blue logo. The logo is composed of several curved lines that are arranged in a symmetrical pattern. The lines are thicker at the top and bottom and thinner in the middle. The overall shape of the logo is reminiscent of a stylized letter V or a bird in flight.
510(k) SUMMARY
Viz.ai, Inc.'s Viz HDS
- Applicant Name: Viz.ai, Inc. 5000 Center Green Way, Cary, NC 27513
- Contact Person: Pooja Shah
- Regulatory Affairs Specialist
- 5000 Center Green Way,
- Cary, NC 27513
- Tel. (669)288-4344
- pooja.shah@viz.ai
- Date Prepared: January 30, 2024
## Device Name and Classification:
- Name of Device: Viz HDS, Viz Volume Plus, Viz ICH+
- Common or Usual Name: Automated Radiological Image Processing Software
- Classification Panel: Radiology
- Regulation No: 21 C.F.R. § 892.2050
- Regulatory Class: Class II
- Product Code: QIH
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a blue logo that is composed of several curved and angular shapes. The logo has a stylized, abstract appearance, resembling a letter 'V' or a bird in flight. The shapes are arranged in a symmetrical pattern, creating a sense of balance and visual harmony. The blue color is consistent throughout the logo.
### Predicate Devices
| Manufacturer | Device Name | Application No. |
|----------------------|-------------|-----------------|
| Qure.ai technologies | qER-Quant | K211222 |
## Device Description
Viz HDS is a software-only device that uses a locked artificial intelligence machine learning (AI/ML) algorithm to processes non-contrast head CT scans to outline intracranial hyperdensity areas, lateral ventricles (right and left), midline shift, and then quantify the volume of intracranial hyperdensity(ies), volume of lateral ventricle asymmetry ratio and distance of midline shift.
Viz HDS analyzes the head NCCT series in DICOM format and produces a summary series and a segmentation series in DICOM format. The summary series is a two-slice output: a single slice from the NCCT series with segmented areas overlaid on it, and a summary table providing the calculated measurements. The segmentation series shows an RGB overlay, on each slice of the input series, of the lateral ventricles and hyperdensity(ies) segmentation masks and a midline shift. For slices including hyperdensities or ventricle/ventricles, its volume would be mentioned in a color legend that is also overlaid on the slice. The colors are only for visual differentiation between the segmented regions, the colors don't have a meaning on their own. The device output is exported in DICOM format, which is sent to a pre-configured PACS destination together with the original NCCT series for review by a physician to aid in the assessment of measuring intracranial hyperdensity(ies), lateral ventricles, and midline shift.
#### Workflow:
Viz HDS is hosted in Viz.ai's cloud server. For customers that have Viz HDS enabled, when they send NCCT head scans to the server, the scans are directed to the HDS processor for analysis. At the end of the processing the results series are created and sent back to the customer's site to a DICOM node (e.g. PACS) defined for that customer.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a blue logo. The logo is composed of several curved lines that are stacked on top of each other. The top line is the widest, and the lines get progressively narrower as they go down. The overall shape of the logo is similar to an inverted triangle or a stylized letter V.
## Workflow Diagram:
Image /page/5/Figure/2 description: This image shows a diagram of the Viz.ai cloud server system. The diagram shows three healthcare facilities uploading head NCCT scans to the Viz.ai cloud server. The scans are processed on the Viz server using the Viz HDS algorithm. The processed results are sent to the healthcare facilities' PACS.
Figure 1. Workflow diagram of Viz HDS
## Segmentation and Summary Series:
Following are two examples for the output series generated by the Viz HDS device. Figure 1 shows a sample image from the Segmentation series and next to it the original slice from the head NCCT series (without the overlay and legend). Figure 2 shows an example of the summary slice and next to it the summary table image that is part of the summary series.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a blue logo with a stylized letter V. The V is composed of several horizontal lines that curve slightly inward, creating a sense of depth and dimension. The logo has a modern and clean design, with the blue color giving it a professional and trustworthy appearance.
Image /page/6/Picture/1 description: The image shows two axial CT scans of a brain. The left scan has colored overlays indicating different regions and their volumes: HD1 is 53.58ml, HD2 is 41.92ml, RLV is 11.8ml, and LLV is 7.7ml. The right scan shows the same axial view without the colored overlays, providing a clear view of the brain's anatomical structures.
Figure 2. Left image: A sample image from the Segmentation series showing overlays of the identified hyperdensities and lateral ventricles segmentation provided by the Viz HDS device, legend contains the HDS volume measurements, sorted from high to low volume, and the lateral ventricles volume measurements. Right image: the corresponding image from the original head NCCT series.
Image /page/6/Figure/3 description: This image shows a brain scan with a "Viz Hyperdensity Summary" on the right side. The summary lists HD1 as 53.58ml, HD2 as 41.92ml, and a total of 95.5ml. It also lists RLV Volume as 11.8ml, LLV Volume as 7.7ml, Midline Shift as 2.2mm, and LV Asymmetry Ratio as 1.5.
Figure 3. Left image: An example of a single slice summary image showing as overlays the identified hyperdensity, lateral ventricles and midline shift segmentation provided by the Viz HDS device. The legend contains the hyperdensities volume measurements (mL), sorted from high to low volume, the right and left lateral ventricles volume measurements (mL) and the midline shift value (mm). Right image: An example of the summary table image in the Summary series generated by the Viz HDS device. The table shows all of the
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a blue logo that is composed of several curved and angular shapes. The logo is symmetrical and has a stylized, abstract design. The shapes are arranged in a way that creates a sense of depth and movement. The logo is simple and modern, and the blue color gives it a sense of calmness and stability.
identified hyperdensities, the right and left lateral ventricles, their asymmetry ratio and midline shift that was measured by the algorithm in the original NCCT series.
## Intended Use / Indications for Use
The Viz HDS device is intended for automatic labeling, visualization, and quantification of segmentable brain structures from a set of Non-Contrast CT (NCCT) head scans. The software is intended to automate the current manual process of identifying, labeling, and quantifying the volume of segmentable brain structures identified on NCCT images. Viz HDS provides volumes from NCCT scans acquired at a single time point. The Viz HDS software is indicated for use in the analysis of the following structures: Intracranial Hyperdensities, Lateral Ventricles and Midline Shift. The device output should be reviewed along with patient's original images by a physician.
## Summary of Technological Characteristics
The subject device, Viz HDS, is substantially equivalent to the predicate device, qER-Quant (K211222). In comparing the technological characteristics, both the subject and predicate devices use an artificial intelligence algorithm to identify, label and quantify clinical findings in NCCT imaging. Where the subject and predicate device differ is that the software algorithm for the subject device provides volumes from NCCT scans acquired at single time point whereas, predicate device provides volumes from NCCT images acquired at a single time point and provides comparative analysis for two or more images that were acquired on the same scanner with the same acquisition protocol for the same individual at multiple time points.
Both the subject device and the predicate device provide findings through back to PACS to hospital's server that allow user to preview images and patients identified, labeled and quantified with segmentable brain structure to automate the manual process of measurement. Both the devices have DICOM annotated series image and summary report as an output. Where the subject and predicate device differ is that the subject device have RGB overlay as the output for abnormal hyperdensities whereas the predicate device represent hyperdensities is a singular color. This difference does not raise any safety or efficacy question as the colors are only for visual differentiation between the segmented regions, the colors don't have a meaning on their own.
Both the subject and predicate device use a deep-learning algorithm that analyzes NCCT images for identifying, labeling and quantifying volume of segmentable brain structures. Both the devices are intended to be used as aid to physicians for identifying, labeling and quantifying segmentable brain structures on NCCT images. Because of the similarities regarding the type of algorithms, the type of imaging data processed by the devices, the intended use of the devices, and the outputs of the image analysis, the difference in the output visual does not raise new or different questions regarding safety or efficacy.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a stylized blue logo. The logo is composed of several curved and angled shapes that resemble a stylized letter 'V' or a bird in flight. The shapes are arranged in a symmetrical pattern, with the top shapes angled outwards and the bottom shapes curving inwards. The overall design is modern and minimalist.
| | Predicate Device | Subject Device |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | qER-Quant | Viz HDS |
| Application No. | K211222 | K232363 |
| Product Code | QIH | QIH |
| Regulation No. | 21 C.F.R. § 892.2050 | 21 C.F.R. § 892.2050 |
| Intended Use /<br>Indications for Use | The qER-Quant device is intended for<br>automatic labeling, visualization and<br>quantification of segmentable brain<br>structures from a set of Non-Contrast head<br>CT (NCCT) images. The software is intended<br>to automate the current manual process of<br>identifying, labeling and quantifying the<br>volume of segmentable brain structures<br>identified on NCCT images.<br>qER-Quant provides volumes from NCCT<br>images acquired at a single time point and<br>provides a table with comparative analysis for<br>two or more images that were acquired on the<br>same scanner with the same image<br>acquisition protocol for the same individual at<br>multiple time points.<br>The qER-Quant software is indicated for use<br>in the analysis of the following structures:<br>Abnormal Intracranial Hyperdensities, Lateral<br>Ventricles and Midline Shift. | The Viz HDS device is intended for automatic<br>labeling, visualization, and quantification of<br>segmentable brain structures from a set of<br>Non-Contrast CT (NCCT) head scans. The<br>software is intended to automate the current<br>manual process of identifying, labeling, and<br>quantifying the volume of segmentable brain<br>structures identified on NCCT images. Viz<br>HDS provides volumes from NCCT scans<br>acquired at a single time point. The Viz HDS<br>software is indicated for use in the analysis of<br>the following structures: Intracranial<br>Hyperdensities, Lateral Ventricles and<br>Midline Shift. The device output should be<br>reviewed along with patient's original images<br>by a physician. |
| Anatomical Region | Head | Head |
| Independent<br>Standard of Care<br>Workflow | Yes | Yes |
| Input images | Non-contrast CT from a single or multiple<br>time points | Non-contrast CT from a single time point |
| Clinical Condition | Intracranial hyperdensities, lateral ventricles<br>and midline shift | Intracranial hyperdensities, lateral ventricles<br>and midline shift |
| Data Acquisition | Acquires medical image data from DICOM<br>compliant imaging devices and modalities. | Acquires medical image data from DICOM<br>compliant imaging devices and modalities. |
| Supported Imaging<br>Modality | Non-contrast CT (NCCT) | Non-contrast CT (NCCT) |
| Alteration of<br>Original Image | No | No |
| Artificial Intelligence<br>Algorithm | Yes | Yes |
| Output | Multiple electronic reports with volumetric<br>information of brain structures and midline<br>shift AND Annotated DICOM Images | Multiple electronic reports with volumetric<br>information of brain structures and midline<br>shift AND Annotated DICOM Images |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a blue logo. The logo is a stylized letter 'V' or a downward-pointing arrow. The logo is made up of several horizontal lines that are stacked on top of each other. The lines are thicker at the top and thinner at the bottom.
## Performance Data
Clinical testing was performed as a study comparing the Viz HDS's output to the ground truth as established by trained radiologists. The study demonstrated that the MAE (Mean absolute error) for hyperdensities total volume, right lateral ventricle volume, left lateral ventricle volume, and midline shift distance upper 95% Cl bounds were less than 7.5 mL, 3 mL, and 2 mm, respectively between the algorithm and the established ground truth, which was aligned with the performance goal.
Additionally, DICE score was calculated to describe the degree of agreement between the measurements by the Viz HDS algorithm in comparison to the measurements that were obtained manually. The study demonstrated that the DICE score for hyperdenity(ies) and both lateral ventricles lower CI bound was greater than 70%, which was aligned with the performance goal.
Stratification of device performance was divided by clinical site, gender, age, slice thickness, scanner manufacturer,model, scanning parameters and size of estimated quantity.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a blue logo with a stylized "V" shape. The logo consists of several curved lines that create the "V" shape. The top of the "V" is formed by two separate curved lines, while the bottom is formed by three curved lines. The logo has a modern and abstract design.
## Conclusions
Viz HDS is as safe and effective as the predicate device. The subject device and the predicate have the same intended use and similar indications, technological characteristics, and principles of operation. The subject device differs in that it provides a report acquired at a single time point whereas, predicate device provides a report acquired at a single time point and provides comparative analysis. Both subject and predicate provide findings through back to PACS destination of hospital. These differences do not present new or different questions of safety or effectiveness with respect to the predicate device. Viz.ai has provided supportive clinical data and software testing which demonstrates that the subject device can perform effective labeling, visualization of intracranial hyperdensities, lateral ventricles and midline shift. Thus, Viz HDS is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.