Defendo Fujifilm 700 Single Use Biopsy Valve; Defendo Fujifilm 700 Single Use Air/Water Valve; Defendo Fujifilm 700 Single Use Suction Valve; ENDOGATOR Fujifilm 700 Single Use Connector; Defendo Fujifilm 700 Single Use Cleaning Adapter
K232329 · Medivators (A Subsidiary of Steris Corporation) · ODC · Dec 21, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K232329
Device Name
Defendo Fujifilm 700 Single Use Biopsy Valve; Defendo Fujifilm 700 Single Use Air/Water Valve; Defendo Fujifilm 700 Single Use Suction Valve; ENDOGATOR Fujifilm 700 Single Use Connector; Defendo Fujifilm 700 Single Use Cleaning Adapter
Applicant
Medivators (A Subsidiary of Steris Corporation)
Product Code
ODC · Gastroenterology, Urology
Decision Date
Dec 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The Single Use Biopsy Valve is intended for covering the endoscope biopsy port during an endoscopy procedure. The Single Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure. The Single Use Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure. The Single Use Suction Valve is intended to be used to control the suction function on an endoscope during a GI endoscopic procedure. The Single Use ENDOGATOR™ Connector is intended to be used in conjunction with ENDOGATOR™ Irrigation Tubing to provide irrigation via sterile water during GI endoscopic procedures when used with an irrigation pump (or cautery unit). The Single Use Cleaning Adapter is intended to be used only to pre-clean the endoscope's air/water channel post-procedure, and not to be used during a patient procedure.
Device Story
Set of sterile, single-use accessories for Fujifilm 700 series GI endoscopes; includes biopsy, air/water, and suction valves, irrigation connector, and cleaning adapter. Valves manually actuate to control insufflation, suction, and instrument passage; connector facilitates irrigation via pump; cleaning adapter enables post-procedure channel flushing. Used in clinical settings by physicians/endoscopy staff. Single-use design eliminates reprocessing requirements for these components, reducing cross-contamination risk. Output is mechanical control of endoscope functions; assists clinicians in visualization, fluid removal, and therapeutic access during GI procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including gas/water flow rates, backflow prevention, mechanical depression/application/removal forces, instrument access, slit dimensional stability, and button cycling durability. Biocompatibility testing conducted per ISO 10993-1. Packaging integrity validated per ISO 11607/ASTM 4169.
Technological Characteristics
Sterile, single-use GI endoscope accessories. Materials: biocompatible polymers. Sterilization: EtO (SAL 10^-6). Manual actuation. Dimensions specific to Fujifilm 700 series ports. Packaging: PETG tray with Tyvek lid. No energy source or software.
Indications for Use
Indicated for patients undergoing GI endoscopic procedures requiring endoscope accessory valve functions (air/water control, suction control, biopsy port access, irrigation, or post-procedure channel cleaning).
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Defendo Disposable Air/Water Valve Model 100304 (K102409)
Defendo Disposable Suction Valve Model 100305 (K102581)
FUJIFILM Endoscope Models EG-760R, EG-760Z, EC-760R-V/L and EC-760ZP-V/L, and FUJIFILM Water Tank Model WT-603 (K172916)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 21, 2023
Medivators (A Subsidiary of STERIS Corporation) Nick Wang Senior Manager, Regulatory Affairs 3150 Pollok Drive Conroe, Texas 77303
## Re: K232329
Trade/Device Name: Defendo Fujifilm 700 Single Use Biopsy Valve; Defendo Fujifilm 700 Single Use Air/Water Valve; Defendo Fujifilm 700 Single Use Suction Valve; ENDOGATOR Fujifilm 700 Single Use Connector; Defendo Fujifilm 700 Single Use Cleaning Adapter Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: ODC, FDF, OCX, FEQ Dated: November 21, 2023 Received: November 21, 2023
## Dear Nick Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Ouality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
## 510(k) Number (if known) K232329
## Device Name
Defendo Fujifilm 700 Single Use Biopsy Valve; Defendo Fujifilm 700 Single Use Air/Water Valve; Defendo Fujifilm 700 Single Use Suction Valve; ENDOGATOR Fujifilm 700 Single Use Connector; Defendo Fujifilm 700 Single Use Cleaning Adapter
## Indications for Use (Describe)
The Single Use Biopsy Valve is intended for covering the endoscope biopsy port during an endoscopy procedure. The Single Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure.
The Single Use Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure.
The Single Use Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure.
The Single Use ENDOGATOR™ Connector is intended to be used in conjunction with ENDOGATOR™ Irrigation Tubing to provide irrigation via sterile water during GI endoscopic procedures when used with an irrigation pump (or cautery unit).
The Single Use Cleaning Adapter is intended to be used only to pre-clean the endoscope's air/water channel postprocedure, and not to be used during a patient procedure.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <span style="text-decoration: line-through;">\</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration:overline;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K232329 Page 1 of 11
# 510(k) Summary
## Defendo Fujifilm 700 Single-Use Valves Set
| Manufacturer: | Medivators<br>3150 Pollok Drive<br>Conroe, TX 77303 |
|---------------|-----------------------------------------------------------------------------------|
| Submitter: | Nick Wang, Ph.D. RAC<br>Senior Manager, Regulatory Affairs<br>Phone: 440-392-7482 |
06 Dec 2023 Summary Date:
{5}------------------------------------------------
### 1. Device Name
| Trade Name: | Defendo Fujifilm 700 Single-Use Valves Set |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification: | Class II |
| Common/usual Name: | Defendo Fujifilm 700 Single Use Biopsy Valve;<br>Defendo Fujifilm 700 Single Use Air/Water<br>Valve; Defendo Fujifilm 700 Single Use<br>Suction Valve; ENDOGATOR Fujifilm 700<br>Single Use Connector; Defendo Fujifilm 700<br>Single Use Cleaning Adapter |
| Classification Name: | Endoscope and accessories |
| Classification Number: | 21 CFR 876.1500 |
| Product Code: | ODC, FDF, OCX, and FEQ |
### 2. Predicate Device(s)
Primary predicate is listed first:
K102409 - Defendo Disposable Air/Water Valve Model 100304
K102581 - Defendo Disposable Suction Valve Model 100305
K210342 - BioShield- Biopsy Valve
K220395 - Endogator Endoscopy Irrigation Tubing
K172916 - FUJIFILM Endoscope Models EG-760R, EG-760Z, EC-760R-V/L
and EC-760ZP-V/L, and FUJIFILM Water Tank Model WT-603 (Air/Water Cleaning Adapter)
### 3. Description of Device
The Defendo Fujifilm 700 Single-Use Valves Set consists of a family of ready-to-use, sterile, single-use valves designed to be used on the Fujifilm 700 series GI endoscopes as accessories. The specific valves that are the subject of this 510(k) include the Defendo Fujifilm 700 Single-Use Air/Water Valve, the Defendo Fujifilm 700 Single-Use Suction Valve, the Defendo Fujifilm 700 Single Use Biopsy Valve, the Endogator Fujifilm 700 Connector, and the Defendo Fujifilm 700 Cleaning Adapter.
- . The Air/Water valve allows the end-user to control and maintain air or CO2 insufflation down the endoscope's air/water feed channels and control the water used to wash the lens of the endoscope during an endoscopic procedure.
- The Suction valve allows the user to control suction through the endoscope's accessory channel and serves as a fluid-removal conduit to aspirate fluids from the patient. This allows the clinician to remove fluids/soil from the GI tract during an endoscopic procedure.
{6}------------------------------------------------
- The Biopsy valve allows instruments/devices to be passed through it while maintaining insufflation, allowing suction, and guiding instruments through the biopsy channel during endoscopic procedures.
- The Connector allows attachment between the irrigation tube set via its luer connection and the compatible endoscope at its dedicated auxiliary water connection. This is sometimes also referred to as the auxiliary water jet valve.
- The Cleaning Adapter allows the user to pre-clean the air and water channels at the bedside prior to endoscope reprocessing.
These valves are supplied in various sets. Each set may include combinations of two or more of the valves listed above.
### 4. Intended Use
The Single Use Biopsy Valve is intended for covering the endoscope biopsy port during an endoscopy procedure. The Single Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure.
The Single Use Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure.
The Single Use Suction Valve is intended to be used to control the suction function on an endoscope during a GI endoscopic procedure.
The Single Use ENDOGATOR™ Connector is intended to be used in conjunction with ENDOGATOR™ Irrigation Tubing to provide irrigation via sterile water during GI endoscopic procedures when used with an irrigation pump (or cautery unit).
The Single Use Cleaning Adapter is intended to be used only to pre-clean the endoscope's air/water channel post-procedure, and not to be used during a patient procedure.
{7}------------------------------------------------
### 5. Technological Characteristic Comparison Table
Table 1. Proposed/Predicate Device Technological Characteristics Comparison Table – Air / Water Valve
| Feature | Defendo Fujifilm 700<br>Single-Use Air/Water<br>Valve<br>(Subject Device) | Defendo Disposable<br>Air/Water Valve<br>(Predicate K102409) | Comparison |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Intended use | The Single Use<br>Air/Water Valve is<br>intended to be used to<br>control the air/water<br>function on an<br>endoscope during a GI<br>endoscopic procedure. | The Defendo Disposable<br>Air/Water Valve is<br>intended to be used to<br>control the air/water<br>function on an<br>endoscope during a GI<br>endoscopic procedure. | Same |
| Sterile/Non-<br>sterile | Sterile | Sterile | Same |
| Sterilization<br>Method | EtO | EtO | Same |
| Sterilization<br>Assurance Level | 10-6 | 10-6 | Same |
| Usage | Single-Use | Single-Use | Same |
| Main Device<br>Dimensions<br>(lengths/widths) | Air/Water Valve:<br>Length: ~43.5 mm<br>Diameter: ~19.2 mm | Air/Water Valve:<br>Length: ~46 mm<br>Diameter: ~17 mm | Similar, the<br>slight<br>differences are<br>due to the<br>differences in<br>the endoscope<br>port. |
| Target Population | Patients undergoing an<br>endoscopic procedure | Patients undergoing an<br>endoscopic procedure | Same |
| Energy<br>Used/Delivered | None | None | Same |
| Method of<br>Application | Manual actuation | Manual actuation | Same |
| Compatible<br>Endoscopes | Fujifilm 700 series GI<br>endoscope | Olympus endoscope | Reason for this<br>510(k) |
| Packaging | Supplied as a kit in<br>PETG Tray with Tyvek<br>Lid | Supplied individually in<br>Tyvek Pouch | Package<br>systems all<br>maintain sterile<br>barrier |
| Table 2. Proposed/Predicate Device Technological Characteristics |
|------------------------------------------------------------------|
| Comparison Table – Suction Valve |
| Feature | Defendo Fujifilm 700<br>Single-Use Suction<br>Valve<br>(Subject Device) | Defendo Disposable<br>Suction Valve<br>(Predicate K102581) | Comparison |
|--------------|-------------------------------------------------------------------------|------------------------------------------------------------|------------|
| Intended Use | The Single Use<br>Suction Valve is | The Defendo<br>Disposable Suction | Same |
{8}------------------------------------------------
| Feature | Defendo Fujifilm 700<br>Single-Use Suction<br>Valve<br>(Subject Device) | Defendo Disposable<br>Suction Valve<br>(Predicate K102581) | Comparison |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| | intended to be used to<br>control the suction<br>function on an<br>endoscope during a GI<br>endoscopic procedure. | Valve is intended to be<br>used to control the<br>suction function on an<br>endoscope during a GI<br>endoscopic procedure. | |
| Sterile/Non-sterile | Sterile | Sterile | Same |
| Sterilization Method | EtO | EtO | Same |
| Sterilization<br>Assurance Level | 10-6 | 10-6 | Same |
| Usage | Single-Use | Single-Use | Same |
| Main Device<br>Dimensions<br>(lengths/widths) | Suction Valve:<br>Length: ~28.0 mm<br>Diameter: ~17.2 mm | Suction Valve:<br>Length: ~27 mm<br>Diameter: ~18 mm | Similar, the<br>slight<br>differences<br>are due to the<br>differences in<br>the<br>endoscope<br>port. |
| Target Population | Patients undergoing an<br>endoscopic procedure | Patients undergoing an<br>endoscopic procedure | Same |
| Energy<br>Used/Delivered | None | None | Same |
| Method of<br>Application | Manual actuation | Manual actuation | Same |
| Compatible<br>Endoscopes | Fujifilm 700 series GI<br>endoscope | Olympus endoscope | Reason for<br>this 510(k) |
| Packaging | Supplied as a kit in<br>PETG Tray with Tyvek<br>Lid | Supplied individually in<br>Tyvek Pouch | Package<br>systems all<br>maintain<br>sterile barrier |
Table 5-3. Proposed/Predicate Device Technological Characteristics Comparison Table – Biopsy Valve
| Feature | Defendo Single-Use<br>Fujifilm 700 Biopsy<br>Valves<br>(Subject Device) | BioShield Biopsy<br>Valve (for Fujifilm<br>Endoscopes)<br>(Predicate K210342) | Comparison |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Single Use Biopsy<br>Valve is intended for<br>covering the endoscope<br>biopsy port during an<br>endoscopy procedure.<br>The single use Biopsy<br>Valve provides access<br>for endoscopic device | The single-use<br>BioShield biopsy<br>valve is used to cover<br>the opening to the<br>biopsy/suction<br>channel of<br>gastrointestinal<br>endoscopes. It | Similar<br>The intended use<br>for the subject<br>device biopsy valve<br>is identical to that<br>of other Defendo<br>Biopsy Valves |
| Feature | Defendo Single-Use<br>Fujifilm 700 Biopsy<br>Valves<br>(Subject Device) | BioShield Biopsy<br>Valve (for Fujifilm<br>Endoscopes)<br>(Predicate K210342) | Comparison |
| | passage and exchange,<br>helps maintain<br>sufflation, and<br>minimizes leakage of<br>biomaterial from the<br>biopsy port throughout<br>the endoscopic<br>procedure. | provides access for<br>endoscopic device<br>passage and<br>exchange, helps<br>maintain insufflation,<br>minimizes leakage of<br>biomaterial from the<br>biopsy port<br>throughout the<br>endoscopic procedure<br>and provides access<br>for irrigation. | currently marketed<br>by Medivators. The<br>minor differences in<br>language between<br>the BioShield and<br>Defendo valves are<br>mainly editorial in<br>nature and do not<br>represent a<br>different intended<br>use. |
| Sterile/Non-<br>sterile | Sterile | Sterile and Non-<br>Sterile | See note below |
| Sterilization<br>Method | EtO | EtO | Same |
| Sterilization<br>Assurance<br>Level | 10-6 | 10-6 | Same |
| Usage | Single-Use | Single-Use | Same |
| Main Device<br>Dimensions<br>(lengths/widths<br>) | Length: ~16.8 mm<br>Width (handle to tip):<br>~32.4 mm | Length: ~16.8 mm<br>Width (handle to tip):<br>~32.4 mm | Same |
| Target<br>Population | Patients undergoing an<br>endoscopic procedure | Patients undergoing<br>an endoscopic<br>procedure | Same |
| Energy Used/<br>Delivered | None | None | Same |
| Compatible<br>Endoscopes | Fujifilm 700 series GI<br>endoscope | Fuji Gen 5 or later<br>endoscopes | Same |
| Packaging | Continue to be supplied<br>individually in Tyvek<br>Pouch (predicate) and<br>supplied as a set in<br>PETG Tray with Tyvek<br>Lid (new). | Supplied individually in<br>Tyvek Pouch | Package systems<br>all maintain sterile<br>barrier |
{9}------------------------------------------------
Note: Defendo Fujifilm 700 Single Use Biopsy Valve, the subject valve, is identical to the black non-sterile BioShield Biopsy Valve from K210342, with the only difference is that the subject device is EO sterilized in the valves set. Device testing included in K210342 on both the sterile and non-sterile blue BioShield Biopsy Valves provides evidence that EO sterilization does not negatively affect device performance and both the sterile and non-sterile versions of the device met the performance requirements and performed as expected as the biopsy valve. In this submission, STERIS performed additional confirmatory testing to ensure the biopsy valve included in the subject device (black sterile) continued to perform as intended.
{10}------------------------------------------------
| Feature | Endogator Fujifilm 700<br>Connector<br>(Subject Device) | Endogator<br>Connector<br>(Predicate K220395) | Comparison |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Single Use<br>ENDOGATOR™<br>Connector is intended to<br>be used in conjunction<br>with ENDOGATOR™<br>Irrigation Tubing to<br>provide irrigation via<br>sterile water during Gl<br>endoscopic procedures<br>when used with an<br>irrigation pump (or<br>cautery unit). | The Endogator<br>system is intended to<br>provide irrigation via<br>sterile water supply<br>during GI endoscopic<br>procedure when sued<br>in conjunction with an<br>irrigation pump (or<br>cautery unit). | Similar<br>(Language<br>modified for the<br>subject device<br>to reflect the<br>connector<br>provided does<br>not include the<br>Endogator<br>tubing) |
| Sterile/Non-sterile | Sterile | Sterile | Same |
| Sterilization<br>Method | EtO | EtO | Same |
| Sterilization<br>Assurance Level | 10-6 | 10-6 | Same |
| Usage | Single-Use | Single-Use | Same |
| Main Device<br>Dimensions<br>(lengths/widths) | Length: ~22 mm<br>Diameter: ~12.7 mm | Length: ~22 mm<br>Diameter: ~12.7 mm | Same |
| Target Population | Patients undergoing an<br>endoscopic procedure | Patients undergoing an<br>endoscopic procedure | Same |
| Energy<br>Used/Delivered | None | None | Same |
| Compatible<br>Endoscopes | Fujifilm 700 series GI<br>endoscope | None specified<br>(commonly used with<br>Fujifilm endoscopes) | Same |
| Packaging | Continue to be supplied<br>individually in Tyvek<br>Pouch (predicate) and<br>supplied as a set in<br>PETG Tray with Tyvek<br>Lid (new). | Supplied individually in<br>Tyvek Pouch | Package<br>systems all<br>maintain sterile<br>barrier |
Table 5-4. Proposed/Predicate Device Technological Characteristics Comparison Table - Connector
Note: Endogator Fujifilm 700 Connector (subject device) is the same as the Endogator Backflow Valve in the Endogator system (predicate device). The backflow valve is commonly used as the Auxiliary Water Jet Connector for the Fujifilm 700 Series endoscopes. Testing has been performed to provide evidence that the valve is compatible with and functions as intended when used with the Fujifilm 700 series GI endoscopes
{11}------------------------------------------------
Table 5-5. Proposed/Predicate Device Technological Characteristics Comparison Table – Cleaning Adapter
| Feature | Defendo Fujifilm<br>700 Single-Use<br>Cleaning<br>Adapter<br>(Subject Device) | FUJIFILM Endoscope<br>Models EG-760R, EG-<br>760Z, EC-760R-V/L<br>and EC-760ZP-V/L,<br>and FUJIFILM Water<br>Tank Model WT-603<br>(Predicate K172916) | Comparison |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The Single Use<br>Cleaning Adapter<br>is intended to be<br>used only to pre-<br>clean the<br>endoscope's<br>air/water channel<br>post-procedure,<br>and not to be<br>used during a<br>patient procedure. | The predicate reusable<br>air/water channel<br>cleaning adapter is an<br>accessory to Fujifilm<br>700 endoscopes. While<br>no indications for use<br>for the accessory is<br>directly stated. The<br>adapter's intended<br>purpose is to pre-clean<br>the endoscope's<br>air/water channel post-<br>procedure.<br>The cleared indications<br>for use statement<br>provide in the predicate<br>510(k) is the indications<br>for use statement of the<br>entire endoscope, not<br>just the cleaning<br>adapter accessory.# | Same, the intended<br>purposes of the<br>predicate and<br>subject cleaning<br>adapters are the<br>same. |
| Sterility | Sterile | Non-Sterile | While sterility is not<br>critical for the<br>adapter application,<br>the subject adapter<br>is provided sterile as<br>it is in the same<br>packaging<br>configuration as the<br>other valves in the<br>set. |
| Usage | Single-Use | Reprocessed | Reason for this<br>510k, subject<br>adapter is a single-<br>use adapter,<br>whereas the<br>predicate is reusable<br>and requires<br>reprocessing.<br>Single-use<br>configuration<br>mitigates the risk of |
| Feature | Defendo Fujifilm<br>700 Single-Use<br>Cleaning<br>Adapter<br>(Subject Device) | FUJIFILM Endoscope<br>Models EG-760R, EG-<br>760Z, EC-760R-V/L<br>and EC-760ZP-V/L,<br>and FUJIFILM Water<br>Tank Model WT-603<br>(Predicate K172916) | Comparison |
| Main Device<br>Dimensions<br>(lengths/widths) | Length: ~45.7 mm<br>Diameter: ~19.2<br>mm | Length: ~45.8 mm<br>Diameter: ~18.7 mm | Similar, the minor<br>difference do not<br>affect device<br>performance. |
| Target<br>Population | Patients<br>undergoing an<br>endoscopic<br>procedure. The<br>cleaning adapter<br>does not have any<br>patient contact as<br>it is used post-<br>procedure. | Patients undergoing an<br>endoscopic procedure.<br>The cleaning adapter<br>does not have any<br>patient contact as it is<br>used post-procedure. | Same |
| Energy<br>Used/Delivered | None | None | Same |
| Method of<br>Application | Manual actuation | Manual actuation | Same |
| Compatible<br>Endoscopes | Fujifilm 700 Series<br>GI endoscope | Fujifilm 700 Series Gl<br>endoscope | Same |
| Packaging | Supplied as a set<br>in PETG Tray with<br>Tyvek Lid | Supplied with<br>endoscope or<br>individually | Subject device<br>package system<br>maintains sterile<br>barrier |
{12}------------------------------------------------
# The indications for use listed in the predicate 510k summary is the following: FUJIFILM Endoscope Models EG-760R, EG-760Z, EC-760R-V/L, and EC-760ZP-V/L have following indications for use.
FUJIFILM Endoscope Models EG-760R and EG-760Z are upper gastrointestinal endoscopes intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.
FUJIFILM Endoscope Models EC-760R-V/L and EC-760ZP-V/L are lower gastrointestinal endoscopes intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine. FUJIFILM Water Tank Model WT-603 has following indications for use. The Water Tank Model WT-603 is intended for use in combination with FUJIFILM 700 system scopes to deliver air and water through the endoscope under the management of a physician in medical facilities. Do not use this product for any other purpose.
### 4. Non-Clinical Performance Testing
Non-clinical performance testing was performed on the subject device to support the substantial equivalence determination. The following tests were performed:
Air / Water Valve
- Gas (Air and CO2) flow rate testing ●
- Water flow rate testing ●
- Backflow prevention testing ●
{13}------------------------------------------------
- Mechanical testing ●
- Depression Force O
- Valve Application Force O
- o Valve Removal Force
- o Force to Disassemble via Axial Pull
- Valve Operation / Procedure Duration Test .
Suction Valve
- Suction Bypass ●
- Suction Rate (water and soil) ●
- Mechanical testing .
- o Depression Force
- o Valve Application Force
- o Valve Removal Force
- o Force to Disassemble via Axial Pull
- . Valve Operation / Procedure Duration Test
Biopsy Valve
- Workability / Instrument Access Testing .
- Slit Length / Dimensional Stability .
Auxiliary Waterjet Valve (Connector)
- Flow Rate / Durability Testing ●
- Backflow Prevention Test ●
Cleaning Adapter
- . Gas Flow Rate
- Water Flow Rate ●
- Mechanical Testing ●
- Depression Force o
- Adapter Application Force o
- Adapter Removal Force O
- Button Cycling / Durability Testing ●
In addition to performance testing, ISO 11607/ASTM 4169 testing was completed to ensure the packaging of the subject device kit maintained sterile barrier after shipping/transit.
{14}------------------------------------------------
### 5. Biocompatibility
The biocompatibility of the subject devices was assessed in accordance with the FDA guideline "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Biocompatibility testing conducted on the patient-contacting portions of the subject device in accordance with the ISO 10993 standard. The test result shows that the subject devices are biocompatible.
### 6. Conclusion
Based on the intended use, technological characteristics, non-clinical performance testing, and biocompatibility assessment, the subject device has shown to be substantially equivalent to the predicate and having met the acceptance criteria based on its indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.