The LumenEye® X1 device is intended to be used to examine the rectum, and using additional accessories, perform tissue sampling of the rectum, if required. The LumenEye® X1 can enable suitably trained healthcare professionals in clinical environments, when used in conjunction with other tests, to examine the rectum as part of a clinical assessment in adults. The LumenEye® X1 is not intended to determine a final clinical diagnosis without further confirmatory tests. Following examination, the healthcare practitioner may make recommendations for further investigations as appropriate.
Device Story
Reusable handheld rigid endoscope; enables rectal examination and tissue sampling. Distally mounted white LEDs (5300K) illuminate area; HD CMOS sensor captures video. Device connects via USB 3.0 to PC; displays video on screen. Operated by trained healthcare professionals in clinical environments. Provides high-definition visual feedback to assist clinical assessment; not for final diagnosis. Benefits include improved visualization compared to traditional analog viewing windows. Device includes endoscope handle (PC+ABS), sheath/obturator (polypropylene), and stainless steel camera stalk. Powered via USB (5V).
Clinical Evidence
Bench testing only. Performance verified per IEC 60601-1, IEC 60601-1-2, IEC 62471, and ISO 10993. Optical properties (SNR, MTF, distortion, light/spectral distribution) confirmed equivalent to reference device. Reprocessing validated per FDA 2015 guidance and ISO 17664:2017 using V-PRO maX 2 sterilization.
Technological Characteristics
Rigid endoscope; PC+ABS handle, polypropylene sheath/obturator, SS 316L camera stalk. Distal white LED illumination; HD CMOS sensor (1080p, 90° FOV, 60 fps). USB 3.0 powered (5V, 500mA). IP67 rated camera. Sterilization via V-PRO maX 2. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62471, ISO 10993, ISO 14971.
Indications for Use
Indicated for examination of the rectum and tissue sampling in adults by trained healthcare professionals in clinical settings.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
PENTAX Medical VIVIDEO ENT Videoscope Solution, VNL9-CF (K171011)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 30, 2023
SurgEase Innovations Limited Roger Parker Head of Hardware Engineering Pendle Innovation Centre, Brook Street Nelson Lancashire, BB9 9PU United Kingdom
Re: K232262
Trade/Device Name: LumenEye® X1 endoscope (LX1-SCP-203); LumenEye® X1 consumables (LX1-CSB-201); LumenEye® X1 Consumable Carton (25 Sets) (LX1-CSP-201 ) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FAN Dated: October 10, 2023 Received: October 16, 2023
Dear Roger Parker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K232262
Device Name
LumenEye® X1 endoscope (LX1-SCP-203);
LumenEye® X1 consumables (LX1-CSB-201);
LumenEye® X1 Consumable Carton (25 Sets) (LX1-CSP-201 )
Indications for Use (Describe)
The LumenEye® X1 device is intended to be used to examine the rectum, and using additional accessories, perform tissue sampling of the rectum, if required.
The LumenEye® X1 can enable suitably trained healthcare professionals in clinical environments, when used in conjunction with other tests, to examine the rectum as part of a clinical assessment in adults.
The LumenEye® X1 is not intended to determine a final clinical diagnosis without further confirmatory tests. Following examination, the healthcare practitioner may make recommendations for further investigations as appropriate.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for SurgEase Innovations. The logo consists of a circular graphic on the left and the company name on the right. The graphic is a blue circle with a stylized eye in the center. The company name is also in blue, with the word "SurgEase" in a larger font than the word "Innovations" which is located below and to the right.
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. All data included in this document is accurate and complete to the best of SurgEase Innovations' knowledge.
### 1. Device Details
| Applicant: | SurgEase Innovations Ltd.<br>Pendle Innovation Centre<br>Brook Street, Nelson<br>Lancashire, UK<br>BB9 9PU |
|------------|------------------------------------------------------------------------------------------------------------|
|------------|------------------------------------------------------------------------------------------------------------|
- Roger Parker Contact: Head of Hardware Email: roger.parker@surgease.co.uk Mobile: +44 (0) 7793054510
- Date Prepared: 2023-Jul-31
- Common Name: Rectoscope (Rigid)
- Trade names: LumenEye® X1 Endoscope LumenEye® X1 Consumable Pack LumenEye® X1 Consumable Carton (25 Sets)
- Model Number: LumenEye® X1 Endoscope: LX1-SCP-203 LumenEye® X1 Consumable Pack: LX1-CSB-201 LumenEye® X1 Consumable Carton (25 Sets): LX1-CSP-201
| Regulation Number: | 21 CFR Part 876.1500 |
|--------------------|----------------------------|
| Regulation Name: | Endoscopes and Accessories |
| Regulatory Class: | Class II |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for SurgEase Innovations. The logo consists of a circular graphic on the left and the company name on the right. The graphic is a blue circle with a stylized eye in the center. The company name is also in blue, with the word "SurgEase" in a larger font than the word "Innovations" which is located below and to the right of "SurgEase".
LumenEye® X1 FDA 510(k) – Administrative Documentation – 510(k) Summary
AD001 Ver. 1
| Product Code: | FAN - Sigmoidoscope, electric, rigid. |
|---------------------|---------------------------------------------------------|
| Medical Speciality: | Gastroenterology/Urology |
| Predicate Device: | SIGMOIDOSCOPE, DISPOSABLE, MODEL #53125 |
| Manufacturer: | Welch Allyn, Inc., United States |
| 510(K) Number: | K770291 |
| Reference device: | PENTAX Medical VIVIDEO ENT Videoscope Solution, VNL9-CF |
| Manufacturer: | Pentax Medical |
| 510(K) Number: | K171011 |
### 2. Device Description
The LumenEye" X1 is a re-usable handheld rigid endoscope that permits insufflation of the rectum and endoluminal high-definition video capture to enable an examination of the rectum (Figure 1). The device has no lumens or channels within the reusable elements of the handle.
Image /page/5/Figure/7 description: The image shows a LumenEye device, which is a medical instrument used for eye examinations. The device has a white and blue color scheme, with a cylindrical probe extending from the front. The handle is ergonomically designed for a comfortable grip, and a cable is attached to the rear, suggesting it is powered by an external source.
Figure 1 | Image of device assembled with sheath and obturator attached.
The area of interest is illuminated by a distally mounted ring of LEDs which emit white light (5300K). Light reflected from the subject is captured by the HD CMOS sensor. The LEDs and camera sensor are housed in the distal end of a stainless-steel camera tube. The image is transferred via USB protocol to a PC and displayed on a screen. There is limited firmware capability within the sensor which simply enables the LEDs to turn on when a video feed from the camera is requested by the PC. The device functions with any Off-the-shelf (OTS) software with a camera display capability. The full HD CMOS camera provides 1080p, 90° field of view and a frame rate of 60 frames per second. The camera enclosure is IP67 rated.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for SurgEase Innovations. The logo features a circular design on the left, resembling an eye. To the right of the eye is the company name, "SurgEase", with the word "INNOVATIONS" in smaller font underneath.
The device features a 1.8m long PVC cable and is powered via USB 3.0 Type-A connection (maximum input voltage 5V, rated current 500mA).
## 3. Indications for Use
The LumenEye® X1 device is intended to be used to examine the rectum, and using additional accessories, perform tissue sampling of the rectum, if required.
The LumenEye® X1 can enable suitably trained healthcare professionals in clinical environments, when used in conjunction with other tests, to examine the rectum as part of a clinical assessment in adults.
The LumenEye® X1 is not intended to determine a final clinical diagnosis without further confirmatory tests. Following examination, the healthcare practitioner may make recommendations for further investigations as appropriate.
#### 4. Summary of Technological Characteristics
The LumenEye® X1 is substantially equivalent to the predicate device. The subject and predicate share the same intended use and the indicated patient population and anatomical site for the subject device are also identical. However, there are some differences in technological characteristics, as summarised in Table 1.
The major difference between the subject device and predicate device is that the LumenEye® X1 is a device with a HD camera allowing HD digital visualisation of the rectum, whereas the predicate provides a method for direct viewing of the rectum via a proximally placed viewing window. Owing to the differences between the subject and predicate device (i.e. analogue versus digital visualisation), a reference device with a similar technological profile was selected (PENTAX Medical VIVIDEO ENT Videoscope Solution, Model VNL9-CP). The technical (optical) specifications of the chosen reference device closely match those of the subject device.
While there are some differences in the dimensional specifications in the subject device, these differences have been evaluated through performance testing and raise no issue of
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for SurgEase Innovations. The logo consists of a circular graphic on the left, resembling an eye or a stylized swirl. To the right of the graphic, the text "SurgEase" is displayed in a larger, bolder font. Below "Ease", the word "INNOVATIONS" is written in a smaller, sans-serif font.
AD001 Ver. 1
safety and effectiveness of the subject device as these differences have no effect on the performance, function or general intended use of the device.
| | Subject<br>Device | Predicate<br>Device | Reference<br>Device |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Indications For<br>Use | Used to examine<br>the rectum, and<br>using additional<br>accessories,<br>perform tissue<br>sampling of the<br>rectum, if required. | Used to examine the<br>rectum and lower bowel,<br>and using additional<br>accessories, perform<br>various diagnostic and/or<br>therapeutic procedures. | N/A |
| Reusable | Yes | Yes | N/A |
| Reprocessing | Manual cleaning<br>Sterilisation (V-PRO<br>maX 2)¹ | Manual cleaning<br>Intermediate Level<br>Disinfection | N/A |
| Power Source | 5V (USB-A<br>connection to<br>PC/Laptop) | 3.5V (Battery) | N/A |
| Material | Endoscope Handle:<br>PC+ABS<br><br>Sheath and<br>Obturator:<br>Polypropylene<br><br>Endoscope Camera<br>Stalk: SS 316L | Sigmoidoscope Handle:<br>Stainless Steel<br><br>Obturator: Plastic | N/A |
| | Subject<br>Device | Predicate<br>Device | Reference<br>Device |
| Insufflation | Insufflation Bulb<br>(integrated)<br>Single use filter | Insufflation Bulb<br>(separate)<br>Single use filter | N/A |
| Diameter<br>(Sheath) | 19mm | 19mm | N/A |
| Length<br>(Sheath) | 20cm | 25cm | N/A |
| Integrated LED | Yes (tip) | No (Halogen light) | Yes (tip) |
| Imaging<br>Technology | White Light Imaging | N/A | White Light<br>Imaging |
| Depth of Field | 40mm | N/A | 5-50mm |
| Field of View | 90° | N/A | 90° |
| MTF50 | 0.12 (@40mm)<br>[cycle/mm] | N/A | 3.58 (@30mm)<br>[cycle/mm] |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for SurgEase Innovations. The logo consists of a stylized eye symbol on the left, followed by the text "SurgEase" in a larger, bolder font. Below "SurgEase" is the word "INNOVATIONS" in a smaller, thinner font. The logo is in a dark blue color and is set against a white background.
AD001 Ver. 1
Table 1 | Comparison table of technological characteristics between subject, predicate and reference devices.
Note 1: Recommended by the FDA
### 5. Non-Clinical Performance Data
The LumenEye® X1 has been successfully tested for its functions, performance and safety as per FDA recognized consensus standards. Testing has been conducted as per IEC60601-1, IEC 60601-1-2, IEC 62471, and ISO 10993. Risk analysis was carried out in accordance with ISO 14971:2019. Design verifications tests and acceptance criteria were defined and performed as a result of this risk analysis assessment. Verification bench testing of the LumenEye® X1 was performed including operational and functional testing, to ensure that the subject device meets the design specifications and performs as intended. These following performance data provided in support of the substantial equivalence determination.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for SurgEase Innovations. The logo consists of a circular graphic on the left and the company name on the right. The graphic is a blue circle with a stylized eye in the center. The company name is also in blue, with the word "SurgEase" in a larger font than the word "Innovations" which is located below and to the right of "SurgEase".
AD001 Ver. 1
# Optical Testing
Optical properties including signal to noise ratio, spatial resolution (Modulation Transfer Function, MTF), distortion, liqht distribution, and spectral distribution were measured for the Lumen Eye X1's HD CMOS camera. All results show that there are no differences between the subject and the reference device.
#### Reprocessing Validation
Simulated use testing, pre-cleaning, cleaning, rinsing and sterilization studies of the Lumen Eye X1 were conducted and confirmed the effectiveness of reprocessing procedures in accordance with FDA's 2015 Final Guidance, Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling ("FDA's 2015 Reprocessing Guidance") and ISO 17664:2017. Acceptance criteria were established in accordance with AAMI TIR 30:2011 for amount of residual soil accumulation and extraction efficiency. Acceptance criteria were met after each phase of reprocessing. Note that sterilization is performed using the V-PRO maX 2 method and replaces high level disinfection as the Lumen Eye X1 endoscope cannot be fully immersed in liquid.
### Sterilization and Shelf Life
Shelf-life is not applicable to the subject device and consumables as they are not provided sterile as they are indicated for use in an anatomical region inherently contaminated (high burden of microbes), and therefore the risk of harm is negligible. Furthermore, the predicate device is not provided sterile. Additionally, shelf-life is not applicable to the subject device and consumables because the materials do not degrade significantly over time (i.e. performance is unaffected).
### Biocompatibility
Biocompatibility of direct and indirect contact materials was confirmed by assessing the cytotoxicity, sensitization and intracutaneous reactivity of the surface device with limited
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the title of a document. The title is "LumenEye® X1" and the subtitle is "FDA 510(k) - Administrative Documentation - 510(k) Summary". The text is centered on a white background.
Image /page/10/Picture/1 description: The image shows the logo for SurgEase Innovations. The logo consists of a circular graphic on the left and the company name on the right. The graphic is a blue circle with a stylized eye in the center. The company name is written in a modern, sans-serif font, with the word "SurgEase" in a larger font size than the word "Innovations" which is placed below it.
(less than 24 hours) contact with mucosal membrane in accordance with ISO 10993-1, 5, and 10: Biological Evaluation of Medical Devices.
#### Software
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software level of concern is "Moderate".
Cybersecurity risks have been assessed and mitigated according to the FDA Guidance for Industry and Staff "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."
#### EMC and Electrical Safety
The acceptable level of electromagnetic compatibility (EMC) and electrical safety (ES) for the Lumen Eye X1 were confirmed by the following standards:
- -IEC 60601-1 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AMD1 (2012)
- IEC 60601-1-2:2014 (4th Edition) -
- -IEC 60601-2-18:2009
### 6. Substantial Equivalence Discussion
After analysing the intended use, indications for use, technological characteristics (including fundamental operating principle, energy source, scientific technology, functional characteristics, design features, performance characteristics, and constituent materials), labelling, and sterilization method, we conclude that the subject device is as safe and effective as the predicate device. The differences in indications and intended use between the subject and predicate devices, do not raise new concerns of safety and effectiveness and are therefore, substantially equivalent. The technological differences in terms of design features, performance characteristics and constituent materials are not substantive.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image contains the logo for SurgEase Innovations. The logo consists of a circular graphic on the left, resembling an eye, followed by the text "SurgEase" in a larger, bolder font. Below "SurgEase", the word "INNOVATIONS" is written in a smaller, thinner font.
LumenEye® X1 FDA 510(k) – Administrative Documentation – 510(k) Summary AD001 Ver. 1
# 7. Conclusion
It can be concluded that the LumenEye X1® is substantially equivalent to the predicate Welch Allyn disposable sigmoidoscope with respect to the intended use, principles of operation and technical characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.