Published clinical literature (observational studies, meta-analyses, and RCTs)
The sponsor provided a summary of published clinical literature (33 studies) to demonstrate the class effect of CES for treating anxiety and to provide context for the device's clinical performance.
Cranial Electrotherapy Stimulation; CES; Anxiety; Literature Review; Observational Studies
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
General CES Summary (Literature Review); Systematic review of 33 studies (14 RCTs, 13 observational studies, 2 meta-analyses, 4 reviews); Study Period: January 01, 1970, to November 2022
Patients with anxiety and/or insomnia; Sample Size: Not specified (aggregate of 33 studies)
Not applicable for this study
Reduction in anxiety symptoms
Indications for Use
Modius Stress is a non-invasive, home-use neurostimulation device that is indicated to treat the symptoms of Generalized Anxiety Disorder in adults aged 22 and older, when used for approximately 4 weeks. Durability of effect after 4 weeks has not been established.
Device Story
Modius Stress is a non-invasive, battery-powered cranial electrotherapy stimulator (CES) for home use. It delivers low-level electrical energy (up to 1.5mA) via two self-adhesive electrode pads placed on the skin behind the ears (mastoid area). The device is operated by the patient; intensity is adjustable via buttons on the unit. It uses a symmetrical biphasic rectangular waveform at 100Hz. The device includes a microcontroller, USB connector, and EEPROM memory. Users apply the device for 30-minute sessions. The output is intended to modulate neural activity to reduce anxiety symptoms. Healthcare providers use the device to provide a drug-free treatment option for GAD. Patients benefit from potential symptom reduction without the systemic side effects of pharmacological interventions.
Clinical Evidence
Two pivotal randomized, double-blind, sham-controlled trials (UK and India) evaluated GAD-7 score reduction. UK study (n=84): Modius Stress group showed mean reduction of -5.41 vs -3.96 for sham (p=0.139). India study (n=60): Modius Stress group showed mean reduction of -7.44 vs -2.23 for sham (p<0.001). Post-hoc subgroup analysis of participants meeting GAD-7 baseline criteria (≥10) showed consistent benefit. Adverse events were minor and self-resolving.
Technological Characteristics
Cranial electrotherapy stimulator (CES). Materials: Plastic enclosure, self-adhesive electrodes. Energy: 3.75V Lithium Polymer battery. Waveform: Symmetrical biphasic rectangular. Frequency: 100Hz. Output: 0-1500µA. Connectivity: Micro-USB for charging. Software: Embedded firmware for stimulation control and power management. Biocompatibility: ISO 10993 compliant. Electrical safety: IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for adults aged 22 and older with Generalized Anxiety Disorder to treat symptoms. No specific contraindications listed.
Regulatory Classification
Identification
A cranial electrotherapy stimulator is a prescription device that applies electrical current that is not intended to induce a seizure to a patient's head to treat psychiatric conditions.
Special Controls
In combination with the general controls of the FD&C Act, the virtual reality behavioral therapy device for pain relief is subject to the following special controls:
*Classification.* (1) Class II (special controls) when intended to treat insomnia and/or anxiety. The special controls for this device are:(i) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device to treat insomnia and/or anxiety.
(ii) Components of the device that come into human contact must be demonstrated to be biocompatible.
(iii) The device must be designed and tested for electrical safety and electromagnetic compatibility (EMC) in its intended use environment.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The technical parameters of the device, including waveform, output mode, pulse duration, frequency, train delivery, maximum charge, and energy, must be fully characterized and verified.
(vi) The labeling for the device must include the following:
(A) The intended use population and the intended use environment;
(B) A warning that patients should be monitored by their physician for signs of worsening;
(C) A warning that instructs patients on how to mitigate the risk of headaches, and what to do should a headache occur;
(D) A warning that instructs patients on how to mitigate the risk of dizziness, and what to do should dizziness occur;
(E) A detailed summary of the clinical testing, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device;
(F) Instructions for use that address where to place the electrodes, what stimulation parameters to use, and duration and frequency of treatment sessions. This information must be based on the results of clinical studies for the device;
(G) A detailed summary of the device technical parameters, including waveform, output mode, pulse duration, frequency, train delivery, and maximum charge and energy; and
(H) Information on validated methods for reprocessing any reusable components between uses.
(vii) Cranial electrotherapy stimulator devices marketed prior to the effective date of this reclassification must have an amendment submitted to the previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.
(2) Class III (premarket approval) when intended to treat depression.
(c) Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 19, 2020, for any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 19, 2020, been found to be substantially equivalent to any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 27, 2024
Neurovalens Ltd. Jason Mckeown, MD CEO 8 Carmagrim Road Ballymena, County Antrim BT44 8BP United Kingdom
Re: K232253
Trade/Device Name: Modius Stress Regulation Number: 21 CFR 882.5800 Regulation Name: Cranial electrotherapy stimulator Regulatory Class: Class II Product Code: OJO Dated: February 26, 2024 Received: February 26, 2024
#### Dear Dr. McKeown:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Pamela D. Scott -S
Pamela Scott, MS Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K232253
Device Name
Modius Stress
Indications for Use (Describe)
Modius Stress is a non-invasive, home-use neurostimulation device that is indicated to treat the symptoms of Generalized Anxiety Disorder in adults aged 22 and older, when used for approximately 4 weeks. Durability of effect after 4 weeks has not been established.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in all caps. Below the company name is the tagline "Your Health, Our Technology" in a smaller font size.
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
#### l. Submitter
- Neurovalens Limited 8 Carmagrim Road Portglenone County Antrim BT44 8BP United Kingdom
#### Contact person/ official correspondence:
Dr Jason McKeown CEO Neurovalens Ltd. Tel: +44 (0) 7412508707 Email: Jason.mckeown@neurovalens.com
#### Device ll.
Name of Device: Modius Stress Common or usual name: Modius Stress Classification name: Cranial Electrotherapy Stimulator (CES) Regulatory Class: II Product Code: QJQ Regulation number: 882.5800 Classification panel: Neurology
#### III. Predicate devices
### Primary predicate device
2
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
01 Version Page 2 of 34
### Modius Stress K232253 Summary
CES Ultra, K062284 Secondary predicate device Alpha-Stim® CS, K903014
No reference devices were used in this submission.
#### Device description IV.
Modius Stress is a non-invasive transdermal neurostimulation device to treat the symptoms of Generalized Anxiety Disorder. The proposed mechanism of the device is through a technology known as Cranial Electrotherapy Stimulation (CES).
It consists of a battery-powered device designed to transcutaneously deliver low-level electrical energy (up to 1.5mA) to the skin behind the ears, over the mastoid processes. The delivery of this neurostimulation is through two self-adhesive electrode pads. These pads are placed on the skin behind each ear (mastoid area). The intensity of the electric pulse can be adjusted up or down by the user. When turned on, the device delivers a small electrical impulse, and adjustments to the stimulation level may be made using the up and down buttons on the device, which are located just above the power button. The device can also be paused by pressing the power button twice. When finished the user removes the device and disposes of the electrode pads after each use. When the device is not being used it can be charged through a micro-USB cable. For safety reasons, it is not possible to recharge the battery while the device is in use in stimulation mode.
The key components of the Modius Stress include the Modius Stress device (plastic enclosure and printed circuit board assembly (PCBA), stimulation pads (K210448 and K132588), skin cleansing wipes (K121655) and charging accessories. The PCBA consists of a microcontroller, USB connector, transformer driver, IO expander and EEPROM memory. The embedded software within the device manages overall functionality of the device from Stimulation Control, Power Management, and user interaction (Indication LED and Audio tones).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in a bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
01 Version 3 of 34 Page
### Modius Stress K232253 Summary
Biocompatibility testing in accordance with the endpoints of ISO 10993-1 to be addressed for a surface medical device, with prolonged: > 24 hours less than 30 days contact duration, demonstrated good biocompatibility for all components of the Modius Stress device.
#### Device Usage
Using the Modius Stress device has demonstrated an improvement in the symptoms of Generalized Anxiety Disorder (GAD).
After using Modius Stress there are usually no physical limitations imposed so most users can resume normal activities immediately. Some users may have a response that affect their ability to perform potentially hazardous tasks, such as operating a motor vehicle or heavy machinery for up to several hours after treatment. However, no significant lasting side effects have been reported.
#### V. Indications for use
Modius Stress is a non-invasive, home-use neurostimulation device that is indicated to treat the symptoms of Generalized Anxiety Disorder in adults aged 22 and older, when used for approximately 4 weeks. Durability of effect after 4 weeks has not been established.
### VI. Comparison of the technological characteristics with the predicate device
Modius Stress is substantially equivalent with respect to indications for use, stimulation parameters (i.e., current levels, frequencies, pulse width and amplitude) and electrode placement to the predicate devices CES Ultra (cleared by K062284), and Alpha-Stim® CS (cleared by K903014). 2
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
#### Table 1. Summary of substantial equivalence
N
| Property | Proposed Device | Primary Predicate | Secondary Predicate | Comment |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | Modius Stress | CES Ultra | Alpha-Stim® CS | N/A |
| Indications for Use | Modius Stress is a non-invasive, home-use<br>neurostimulation device<br>that is indicated to treat<br>the symptoms of<br>Generalized Anxiety<br>Disorder in adults aged<br>22 and older, when used<br>for approximately 4<br>weeks. Durability of<br>effect after 4 weeks has<br>not been established. | The CES Ultra is<br>indicated for the<br>treatment of insomnia,<br>depression or anxiety | The Alpha-Stim® 100 is a<br>precision medical<br>instrument used for the<br>management of pain,<br>anxiety, depression, and/or<br>insomnia. | Predicate devices are indicated for the<br>treatment of anxiety and in the case of<br>Modius Stress for symptoms of Generalized<br>Anxiety Disorder. |
| Target Population | Adults 22 and older | Adults | Adults | N/A |
| Environment | Home | Home | Home | N/A |
| Waveform | Symmetrical Biphasic<br>Rectangular Wave | Symmetrical Biphasic<br>Rectangular Wave | Symmetrical Biphasic<br>Rectangular Wave | Identical |
| Current Intensity Range | 0 $ µ $ A – 1500 $ µ $ A | 0 $ µ $ A – 1500 $ µ $ A<br>continually adjustable | 0 $ µ $ A – 500 $ µ $ A | Modius Stress and the CES Ultra have near<br>identical current intensity ranges |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, sans-serif font. Below the wordmark is the tagline "Your Health, Our Technology" in a smaller, lighter font.
ry
### Modius Stress K232253 Summary
| Property | Proposed Device | Primary Predicate | Secondary Predicate | Comment | |
|----------------------|-----------------------|-----------------------|-------------------------|----------------------------------------------------|--|
| | | | | | |
| Pulse Width Range | 2.5ms | 2ms | 250ms - 1s | Pulse width varies depending on frequency<br>range | |
| Number of electrodes | Two | Two | Two | | |
| | | | | Identical | |
| | | | | | |
| Electrode placement | Mastoid | | Head – earlobes | The Modius Stress electrodes are placed on | |
| | | Mastoid or earlobes | | mastoids. CES Ultra also supports mastoid only | |
| | | | | electrode placement | |
| Power Source | 3.75 Lithium Polymer | 9V Alkaline Battery | 2 x AAA NiMH | | |
| | Battery | | | | |
| Frequency | 100Hz | 100Hz | 0.5Hz, 1.5Hz, 100Hz | | |
| Treatment Range | 30 min | 30min, 60min or | 10min, 20min, 60 min or | Modius Stress maximum treatment time is within | |
| | | continuous | continuous | the range of the predicates | |
| Unit Controls | Built into the device | Built into the device | Built into the device | Identical | |
| Dimensions | 16.5cm x 15.1cm x | 13.5cm x 6.4cm x | 9.8cm x 6.3cm x 2cm | Modius Stress is larger than the predicate | |
| | 6.6cm | 3.3cm | | devices | |
| Enclosure | Plastic | Plastic | Plastic | Identical | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in a bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
#### Table 2. Safety Information Comparison
N
| | Property | Proposed Device<br>Modius Stress | Primary Predicate<br>CES Ultra | Secondary Predicate<br>Alpha-Stim® CS | Comment |
|-------------------|------------------------------|----------------------------------------------|-------------------------------------------|-------------------------------------------|------------------------------------------------------------|
| Electrical Safety | | Complies with IEC<br>60601-1 | Complies with IEC<br>60601 | Complies with<br>IEC 60601 | The proposed device and predicate devices are<br>identical |
| EMC | | Complies with IEC<br>60601-1-2 | Complies with IEC<br>60601-1-2 | Complies with IEC<br>60601-1-2 | The proposed device and predicate devices are<br>identical |
| Software | Level<br>of<br>Concern | Moderate | Moderate | Moderate | The proposed device and predicate devices are<br>identical |
| | Verification &<br>Validation | Complies with<br>FDA Guidance<br>Requirement | Complies with FDA<br>Guidance Requirement | Complies with FDA<br>Guidance Requirement | |
| Biocompatibility | | Complies with ISO<br>10993 | Complies with ISO<br>10993 | Complies with<br>ISO 10993 | The proposed device and predicate devices are<br>identical |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
N
### Modius Stress K232253 Summary
#### Table 3. Technical Information for the Modius Stress Output parameters
| | Modius Stress | CES Ultra | Alpha-Stim CS | |
|--------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|----------------------------------------------------------------|-------------------------------|
| 510(k) Number | K232253 | K024377 | K903014 | |
| Device Name | Modius Stress | CES ultra | Alpha-Stim CS | |
| Manufacturer | Neurovalens Ltd | Neuro-Fitness LLC | Electromedical Products<br>Inc. | |
| Power Sourcet | 3.75 Lithium Polymer<br>Battery | 9V Alkaline Battery | 2 x AAA NiMH | |
| -Method of Line Current Isolation | DC:DC transformer | N/A Non Rechargeable | N/A Non-Rechargeable | |
| - Patient Leakage Current (as per ANSI/AAMI 60601-1) | 0μA | Unknown | Unknown | |
| -Normal Condition (μΑ) | 0-1500μΑ | 0-1500μΑ | 0-500μΑ | |
| -Single Fault Condition (μA) | 2.3μA | Unknown | 1.3mA | |
| Average DC current through the electrodes when the device is on but no<br>pulses are being delivered (µA) | 0μA | Unknown | Unknown | |
| Number of output channels | 1 | 1 | 1 | |
| -If more than one channel, is the stimulus delivered to<br>each channel synchronous or alternating between each channel? | N/A | N/A | N/A | |
| -If more than one channel, describe method of channel isolation | N/A | N/A | N/A | |
| Software/Firmware/Microprocessor Control | Yes | Yes | Yes | |
| Automatic Overload Trip? | No | No | No | |
| | | Modius Stress | CES Ultra | Alpha-Stim CS |
| Automatic Shut Off? | | Yes | Yes | Yes |
| User Override Control? | | Yes | Yes | Yes |
| Mode or Program Name | | Stimulation mode | Stimulation mode | Stimulation/Treatment<br>Mode |
| Waveform (e.g., pulsed monophasic, biphasic) | | Biphasic | Biphasic | Biphasic |
| Shape (e.g., rectangular, spike, rectified sinusoidal) | | Rectangular | Rectangular | Rectangular |
| Maximum Output Voltage (volts) (+/-1%) | | 750mV @500 Ω | Unknown | @500 Ω |
| | | 3V @ 2 kΩ | Unknown | @ 2 kΩ |
| | | 15V @10 kΩ | Unknown | @10 KΩ |
| Maximum Output Current (specify units) (+/-2%) | | 1500μA @500 Ω | Unknown | 500μA @ 500 Ω |
| | | 1500μA @ 2 kΩ | 1.5mA @ 2kΩ | @ 2 kΩ |
| | | 1500μA @10 kΩ | Unknown | @10 KΩ |
| Duration of primary (depolarizing) phase† (msec) | | 10ms | 10ms | 2000ms, 1000ms, 10ms |
| Pulse Duration†† (msec) | | 5ms | 5ms | 1000ms, 500ms, 5ms |
| Frequency†† (Hz) [or Rate†† (pps)] | | 100Hz | 100Hz | 0.5Hz, 1Hz, 100Hz |
| | For interferential modes only: Beat Frequency† (Hz) | | | |
| | For multiphasic waveforms Symmetrical phases? | | | |
| only: | Phase Duration† (include units),<br>(state range, if applicable), (both phases, if<br>asymmetrical) | Not Applicable for these devices | | |
| | | | | |
| | | Modius Stress | CES Ultra | Alpha-Stim CS |
| | Net Charge (microcoulombs (mC) per pulse) (If zero, state method of achieving zero net charge.) | 3.75μC @2k Ω /<br>3.75 μC @ 500 Ω | 0.75μC @500 Ω | 300μC, 150 μC, 1.5 μC |
| | Maximum Phase Charge, (μC) | 3.75μC @2k Ω /<br>3.75 μC @ 500 Ω | 0.75μC@ 2 kΩ | 600μC @ 500Ω |
| Maximum Current Density,††† (mA/cm², r.m.s.) | | 0.75mA /cm² @500 Ω | 1.91mA/cm² @500 Ω | ~1.2mA/cm² @ 500Ω |
| Maximum Average Current (average absolute value), mA | | 1.5mA @500 Ω<br>1.5mA @ 2kΩ | 3.07mA/cm² @ 2 kΩ | Unknown |
| | Maximum Average Power Density, ††† (W/cm²), (using smallest electrode conductive surface area) | 0.28mW/cm² @500Ω<br>2.24mW/cm² @2kΩ | 9.19mW/cm² @2 kΩ | Unknown |
| Burst Mode††† (i.e., pulse trains): | (a) Pulses per burst | 2 | 2 | 2, 2, 2 |
| | (b) Bursts per second | 100 | 100 | 0.5, 1, 100 |
| | (c) Burst duration (seconds) | 0.01 | 0.01 (10ms) | 2, 1, 0.01s |
| | (d) Duty Cycle: Line (b) x Line (c) | 0.5 (50%) | 0.2(20%) | 0.5 (50%) |
| ON Time (seconds) | | 5ms | 10ms | 1sec, 0.5sec, 10ms |
| OFF Time (seconds) | | 5ms | 40ms | 1sec, 0.5sec, 10ms |
| Output value<br>Load conditions | Modius stress | CES Ultra | Alpha Stim CS | |
| Output value at 500Ω, 2kΩ and 10kΩ | 1.5mA at all loads | 1.5mA at all loads | 0.6mA at all loads | |
| Electrode surface<br>area in cm2 | Conductive Area (16mm diameter)<br>2.0cm2 | Ear clips<br>Diameter 7.9mm<br>0.5cm2 Area | 1.5cm Diameter<br>1.77cm2 | |
| Current density | 500Ω<br>0.75mA/cm2,<br><br>2kΩ<br>0.75mA/cm2<br><br>10kΩ<br>0.75mA/cm2 | 500Ω<br>1.91mA/cm2<br><br>2kΩ<br>3.07mA/cm2<br><br>10kΩ<br>Unknown | 500Ω<br>1.2mA/cm2<br><br>2kΩ<br>Unknown<br><br>10kΩ<br>Unknown | |
| Charge density | 500Ω<br>1.87µC – Per Pulse<br><br>2kΩ<br>1.87µC – Per Pulse<br><br>10kΩ<br>1.87µC – Per Pulse | 500Ω<br>Unknown<br><br>2kΩ<br>Unknown<br><br>10kΩ<br>Unknown | 500Ω<br>Unknown<br><br>2kΩ<br>Unknown<br><br>10kΩ<br>Unknown | |
| Power density | 500Ω,<br>0.56mW/cm2<br><br>2kΩ<br>2.24mW/cm2<br><br>10kΩ<br>11.19mW/cm2 | 500Ω,<br>Unknown<br><br>2kΩ<br>9.19ma/cm2<br><br>10kΩ<br>Unknown | 500Ω,<br>Unknown<br><br>2kΩ<br>Unknown<br><br>10kΩ<br>Unknown | |
| Max phase charge<br>(pulse width x peak current) | 500Ω,<br>3.75 µC – Per Pulse<br><br>2kΩ<br>3.75 µC – Per Pulse<br><br>10kΩ<br>3.75 µC – Per Pulse | 500Ω,<br>Unknown<br><br>2kΩ<br>0.75 µC<br><br>10kΩ<br>Unknown | 500Ω,<br>600 µC<br><br>2kΩ<br>Unknown<br><br>10kΩ<br>Unknown | |
| Max phase charge density (pulse width x peak current) / electrode surface area | 500Ω<br>1.87µC – Per Pulse<br><br>2kΩ<br>1.87µC – Per Pulse | 500Ω<br>Unknown<br><br>2kΩ<br>Unknown | 500Ω<br>Unknown<br><br>2kΩ<br>Unknown | |
| | 10kΩ | 10kΩ | 10kΩ | |
| | 1.87µC – Per Pulse | Unknown | Unknown | |
| Max average power<br>density (Duty cycle x<br>peak current)^2 x<br>(load Ω) ÷ electrode<br>surface area) | 500Ω,<br>0.28mW/cm²<br><br>2kΩ<br>1.12mW/cm²<br><br>10kΩ<br>5.6mW/cm² | 500Ω,<br>Unknown<br><br>2kΩ<br>9.19mW/cm²<br><br>10kΩ<br>Unknown | 500Ω,<br>Unknown<br><br>2kΩ<br>Unknown<br><br>10kΩ<br>Unknown | |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
Page 8 of 34
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in a bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
ry
### Modius Stress K232253 Summary
Image /page/12/Picture/4 description: The image shows the text "Page 9 of 34". The text is written in a simple, sans-serif font and is positioned in the center of the image. The numbers indicate that this is page 9 of a document that has a total of 34 pages.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in a bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
Version 01 Page 10 of 34
## Modius Stress K232253 Summary
### Table 4. Comparison of Device Outputs
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in bold, uppercase letters. Below the company name is the tagline "Your Health, Our Technology" in a smaller font.
Version 01 Page 11 of 34
# Modius Stress K232253
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in a bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
#### VII. Performance Data
### Non-clinical tests
The Modius Stress was evaluated for its safety and effectiveness based on the following testing:
#### Table 5: Non-clinical tests
| Test Name | Test Description | Results |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Device Ship/Transport<br>Testing | Ensure device, enclosed in the selected shipping container,<br>meets ASTM D4169 specifications. | Passed |
| Biocompatibility Testing | Testing and analysis of the Modius Stress device has<br>demonstrated compliance to ISO 10993-1: Biological<br>evaluation of medical devices - Guidance | Passed |
| Electrical Safety | The Modius Stress device was tested to confirm that it met<br>the applicable standards for electrical safety (IEC 60601-1) | Passed |
| EMC | The Modius Stress device was tested to confirm that it met<br>the applicable standards for electromagnetic compatibility<br>(EMC) (IEC 60601-1-2) | Passed |
| Usability Testing | Modius Stress was assessed with regards to usability for<br>compliance with IEC 62366 - Medical devices - Application<br>of usability engineering to medical devices | Passed |
| Maximum Current &<br>Phase Charge Testing | The Maximum Current & Phase Charge Testing was to detail<br>the Max Current / Power output of the system under different<br>load conditions and determine the Max Phase Charge, Max<br>Current Density and Max Power Density of the Modius<br>Stress device. | Passed |
| System Verification and<br>Validation Testing | The system verification and validation testing was performed<br>to verify the hardware and firmware of the Modius Stress<br>Device | Passed |
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, sans-serif font. Below the wordmark is the tagline "Your Health, Our Technology" in a smaller, lighter font.
01 Version 13 of 34 Page
### Modius Stress K232253 Summary
### Labeling
The labeling for the device includes the intended use population and the intended use environment; warning that patients should be monitored by their physician for signs of worsening; A warning that instructs patients on how to mitigate the risk of headaches, and what to do should a headache occur; a warning that instructs patients on how to mitigate the risk of dizziness, and what to do should dizziness occur; a detailed summary of the clinical testing, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device; Instructions for use that address where to place the electrodes, what stimulation parameters to use, and duration and frequency of treatment sessions based on the results of clinical studies for the device; A detailed summary of the device technical parameters, including waveform, output mode, pulse duration, frequency, train delivery, and maximum charge and energy; and Information on validated methods for reprocessing any reusable components between uses.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
The software for this device was considered as a "moderate" level of concern, since a failure or latent design flaw could directly result in minor injury to the patient or a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
#### Animal Studies
No animal studies were conducted as part of submission to prove substantial equivalence.
#### Clinical Studies
CES devices collectively demonstrate a class effect of CES for treating anxiety and/or insomnia. However, it cannot be concluded, based on available information alone, that specific CES devices or stimulation parameters are effective for treating anxiety and/or insomnia. As such, individuals using this device should work with the prescribing medical provider to
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in a bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
determine the best treatment settings to use. As such, please see General CES Summary below for a general summary of pertinent clinical literature that has been published using various combinations of CES devices, stimulation settings, and electrode positions.
### General CES Summary (January 01, 1970, to November 2022)
As of November 2022, 33 studies investigated the impact of Cranial Electrotherapy Stimulation (CES) on anxiety (14 randomized controlled trials (RCTs), 13 observational studies, 2 metaanalysis, and 4 reviews). Of the RCTs that were evaluated, some trials reported superiority of CES treatment versus placebo (Rosenthal, S H et al. 1972) (Philip, P et al. 1991) (Lee, S H, et al. 2013) (Sousa, A D et al. 1975) (Gibson, T H et al. 1987) (Ryan, J J. et al. 1976) (Kim, J, et al. 2021) or control (Kanq, H W, et al. 2020) (Park, B S, et al. 2022) in reducing anxiety symptoms, while other studies demonstrated no impact on anxiety (Levitt, E A et al. 1975) (Passini, F G et al. 1976) (Scallet, A, et al. 1976) (13). One study noted transient improvement in symptoms (Hearst, E D, et al. 1974). The majority of observational studies reported a positive association between CES treatment and reduction in anxiety symptoms (Flemenbaum, A et al. 1974,) (Overcash, S J. et al 1999) (Rosenthal, S et al. 1970) (Rosenthal, S H et al. 1970) (Smith, R B et al. 1999) (Ryan, J J et al. 1977) (Matteson, M T et al. 1986) (Morriss, R, et al. 2019) (Morriss, R, et al. 2020) (Bystritsky, A, et al. 2008). One observational study reported that CES was as effective as usual care (Royal, S, et al., 2022). Only 2 observational studies reported that CES did not have a significant impact on anxiety based on clinical assessment and standard inventories (Moore, J A, et al. 1975) (Von Richthofen et al 1980). A meta-analysis of 10 studies evaluating the effectiveness of cranial stimulation in treating depression noted an effect on anxiety as a secondary study outcome compared to sham (Cheng, Y, et al. 2022). Another meta-analysis of 14 RCTs evaluating the effectiveness of cranial stimulation indicated that CES versus sham treatment was associated with significantly improved anxiety (Klawansky, S, et al. 1995).
Similar findings were reported in a review that examined 34 controlled trials involving a total of 767 patients receiving CES and an additional 867 patients serving as controls (De Felice, E A. 1997). Twenty-six (Moore, J A, et al. 1975,) of 34 studies (77%) reported decreased anxiety after treatment with CES and the remaining 8 of 34 studies (24%) reported no such benefit. Two other reviews of trials using CES in patients with anxiety and depression concluded that CES may provide a modest benefit in the treatment of anxiety (31) (Brunyé, T
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in a bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
01 Version Page 15 of 34
### Modius Stress K232253 Summary
T, et al. 2021). A fourth review found several small studies reporting decreased anxiety with CES use in patients with various conditions, including 1 small study in individuals with anxiety disorder (Freire, R C, et al. 2020). In studies that reported improvement in anxiety with use of CES, the reported stimulation parameters, electrode placement, and treatment schedule varied widely as listed below and were only evaluated in a small number of combinations. Please also note that not all CES devices are capable of providing the same stimulation parameters or combination of stimulation parameters.
Electrode placement:
- . Over orbits and mastoids
- . Over evelid and mastoids
- . Over orbits and occiput
- Earlobes
- . Anterior and posterior regions of the head; in ear
Waveform:
- . Square
- . Rectangular
- Monophasic ●
- Bipolar, asymmetric, rectangular waves ●
- . Sine wave
- . Biphasic modified square wave
- . Modified sine wave
Frequency:
- . 100 Hertz (Hz)
- 350 Hz
- 0.5 Hz ●
- . 10 Hz
Pulse Width:
- 0.7 millisecond (ms)
- 1 ms .
- . 2 ms
Maximum current:
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in a bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
- . 0.01-4 milliamperes (mA)
Treatment schedule:
- . 30 minutes, 5 times a week for 1-2 weeks
- . 30-minute sessions twice a day
- . single 20-minute session
- 5 times a week for 3 weeks ●
- . 20 minutes a day for 2 days
- . 60 minutes a day for 6-12 weeks
- . 5 minutes
- 45 minutes daily for 3 weeks .
2
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in all caps. Below the wordmark is the tagline "Your Health, Our Technology".
#### Modius Stress – Clinical Studies
A summary of the clinical performance testing is given below with detailed information provided in Section 20.
Clinical testing of the Modius Stress device (MA1500) included two pivotal randomized controlled trials to evaluate the safety and efficacy of the Modius Stress device, for the treatment of anxiety. Originally a multi-site study was planned; however, logistical challenges prevented the research site in India from adhering to the precise study design outlined in the UK protocol. As a result, the studies are documented separately.
The stimulation sessions received at the India site were of the same duration (30 minutes) as the sessions participants completed at home in the UK. UK participants were instructed to use their device daily or no less than 5 times per week; however, at the India site, participants were asked to complete 20 stimulation sessions at a rate of ~3 -5 sessions per week. The difference in the treatment duration was to allow flexibility to facilitate the additional travel and time spent going to the India research site.
#### Study Design
Both studies were randomised, double blinded, sham-controlled trials, one of which was conducted in the UK and the other in India. The UK study was undertaken from July 2022 to April 2023 and has been registered at https://clinicaltrials.gov (ClinicalTrials.gov Identifier: NCT05907967). The UK study protocol (version 6.0, 25/11/2022) was approved at the UK site by East of England - Cambridge Central Research Ethics Committee (Ref: 22/EE/0009). The study in India was conducted from October 2022 to June 2023 and has also been registered at ClinicalTrials.gov (Identifier: NCT05845658). The Indian study protocol (version 1.0, 25/07/2022) was approved by the Institutional Ethics Committee – R.D.Gardi Medical College, Ujjain, Madhya Pradesh, India (Ref: 14/2022). Informed consent was obtained from all participants.
In both studies, adults with a General Anxiety Disorder 7th edition (GAD-7) score of 10 or higher (Spitzer et al., 2006) were recruited using social media advertising and were randomly
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in all caps. Below the company name is the tagline "Your Health, Our Technology" in a smaller font size. The logo is simple and modern, with a focus on the company name and its mission.
01 Version 18 of 34 Page
### Modius Stress K232253 Summary
allocated (1:1 allocation ratio) to receive the Modius Stress device (intervention group) or the sham device (active control group). In the UK study, both groups were advised to use their allocated devices for 30 minutes per day for a four-week duration at home. In the India study, both groups were asked to complete 20 stimulation sessions (30 minutes duration) at a rate of ~3-5 sessions per week at the Department of Physiology at R.D. Gardi Medical College.
The Modius stress and sham devices are identical in appearance and therefore neither the participant nor the clinical study team knew whether the Modius Stress device or sham device was allocated.
### Primary Effectiveness Endpoint
The efficacy of the Modius Stress device was quantified by change in GAD-7 score at baseline and week 4 in the UK study, and when each participant finished their 20 stimulation sessions in the India study, between the Modius Stress group and sham control group.
### Primary Safety Endpoint
In both studies, an evaluation of the safety of the Modius Stress device was quantified by the occurrence in adverse events between the Modius Stress group and sham control group.
#### UK study results
2
A total of 87 eligible participants (female, n=65) were recruited and randomised to the Modius Stress group (n=43) or the sham control group (n=44). Nine participants withdrew from the study before the 4-week visit. Three of these participants withdrew (lost to follow-up, n=2 and self-withdrawal, n=1) prior to the baseline visit and therefore were not included in the Intention to treat (ITT) analysis (no baseline data and did not receive at least one stimulation session). The mean age of the sample was 30 years (age range: 18 to 69 years). One participant was aged over 65 years, and 14 participants were aged under 22 years.
| Stage | Modius Stress Device | Sham control device | Total |
|-----------|----------------------|---------------------|-------|
| Enrolment | 43 | 44 | 87 |
| Treatment | 42 | 42 | 84 |
#### Table 6: Participant accountability (UK study)
{22}------------------------------------------------
Image /page/22/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in all caps. Below the company name is the tagline "Your Health, Our Technology".
| Primary safety<br>endpoint analysis | 42 | 42 | 84 |
|--------------------------------------------|------------------------------------------------|-----------------------------------------------|----------|
| Primary effectiveness<br>endpoint analysis | aIntention to treat= 42<br>bComplete case = 36 | Intention to treat = 42<br>Complete case = 39 | 84<br>75 |
ª Intention to treat analysis was undertaken for all participants who were randomised and had baseline GAD-7 scores. * Complete case analysis was undertaken on all participants with available data at baseline and at the 4-week visit.
#### Effectiveness
The effect of the Modius Stress device on change in GAD-7 scores from baseline to the 4week visit is shown in Table 7 below. The ITT (multiple imputation (MI) and baseline carried forward (BOCF)) and complete case analyses demonstrated that the Modius Stress group had a greater reduction in GAD-7 score compared with the sham control group; however, the mean difference among the groups was not statistically different.
More participants in the Modius Stress group, compared with the sham control group, achieved a 4 point or more reduction in GAD-7 (Toussaint et al., 2020) from baseline to the 4week visit (complete case analysis: 27 out of 36 [75%] vs 20 out of 39 [51%]).
{23}------------------------------------------------
Image /page/23/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, uppercase letters. Below the company name is the tagline "Your Health, Our Technology" in a smaller font.
### Table 7. The effect of Modius Stress on GAD-7 scores according to randomization group (primary outcome) (UK Study)
| Cohort | Modius Stress | | | Sham Control | | | | P |
|------------------|-----------------------|----------------------------|-----------------------------------------|-----------------------|----------------------------|-----------------------------------------|--------------------------------------------------|-------|
| | Baseline<br>Mean (SD) | 4<br>weeks<br>Mean<br>(SD) | Within-group<br>mean change<br>(95% CI) | Baseline<br>Mean (SD) | 4<br>weeks<br>Mean<br>(SD) | Within-group<br>mean change<br>(95% CI) | Between-group<br>difference<br>mean (95%<br>CI)a | |
| ITT (MI) | 11.36<br>(0.66) | 5.95<br>(0.59) | -5.41<br>(-6.83, -3.99) | 10.90<br>(0.54) | 6.94<br>(0.59) | -3.96<br>(-5.25, -6.06) | -1.45<br>(-3.37, 0.47) | 0.139 |
| ITT (BOCF) | 11.36<br>(4.27) | 6.83<br>(4.03) | -4.52<br>(-5.91, -3.14) | 10.90<br>(3.47) | 7.10<br>(3.75) | -3.81<br>(-5.08, -2.54) | -0.71<br>(-2.57, 1.14) | 0.445 |
| Complete<br>Case | 11.17<br>(4.32) | 5.89<br>(3.20) | -5.28<br>(-6.75, -3.80) | 11.15<br>(3.31) | 7.05<br>(3.74) | -4.10<br>(-5.43, -2.78) | -1.18<br>(-3.12, 0.77) | 0.233 |
ªBetween-group difference in GAD-7 scores from baseline to week 4 analysed using Independent Samples t-tests. ITT analysis (MI and BOCF) includes all participants randomized and who had baseline GAD-7 scores (Modius Stress, n=42 and sham control n=42). Three participants did not have baseline GAD-7 scores and were not included in the ITT analysis. Complete case analysis includes participants who had baseline and 4-week GAD-7 scores (Modius Stress n=36 and sham control n=39).
#### Safety
A total of 7 Adverse Events were reported during the 4-week study, 5 reported by the Modius Stress group and 2 reported by the sham control group. The adverse events were all minor, caused a minimal discomfort to participants, and all self-resolved during the period of the study. No serious adverse events were reported.
{24}------------------------------------------------
Image /page/24/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, uppercase letters. Below the company name is the tagline "Your Health, Our Technology" in a smaller font size. The logo is simple, modern, and professional.
Table 8. Number (%) of Adverse Events (Safety Data Analysis) according to Randomization Group (all participants, n=87) (UK Study)
| | | Total<br>n (%) | Modius Stress<br>n (%) | Sham Control<br>n (%) |
|----------------------------|---------------------------------------|----------------|------------------------|-----------------------|
| Nervous<br>System | Migraine | 1 (1) | 1 (1) | 0 (0) |
| Disorders | Sudden sharp<br>movements | 1 (1) | 0 (0) | 1 (1) |
| Ear Disorders | Cartilage swelling | 1 (1) | 1 (1) | 0 (0) |
| | Itchiness | 1 (1) | 1 (1) | 0 (0) |
| | Pressure in ears | 1 (1) | 1 (1) | 0 (0) |
| Mouth/ dental<br>disorders | Metallic fillings pulsing/<br>buzzing | 1 (1) | 0 (0) | 1 (1) |
| Other | Study related anxiety | 1 (1) | 1 (1) | 0 (0) |
#### Conclusion
Participants using the Modius Stress device, 5-7 times per week, had a 5.41 reduction in GAD-7 score, while participants using the sham control device had a 3.96 reduction in GAD-7 score. More participants using the Modius Stress device daily had a 4-point or more reduction in GAD-7 score when assessed at 4 weeks.
{25}------------------------------------------------
Image /page/25/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, sans-serif font. Below the wordmark is the tagline "Your Health, Our Technology" in a smaller, lighter font.
#### India study results
A total of 60 eligible participants were recruited and randomised to the Modius Stress group (n=34) or the sham control group (n=26). One participant randomised to the sham control group withdrew from the study due to difficulty attending the research site for treatment and assessments. The mean age of the sample was 35.6 years. The age range was 19-70 years, with one participant aged over 65 years and two participants aged under 22 years.
| Stage | Modius Stress Device | Sham control device | Total |
|--------------------------------------------|------------------------------------------------|-----------------------------------------------|----------|
| Enrolment | 34 | 26 | 60 |
| Treatment | 34 | 26 | 60 |
| Primary safety endpoint<br>analysis | 34 | 26 | 60 |
| Primary effectiveness<br>endpoint analysis | aIntention to treat= 34<br>bComplete case = 34 | Intention to treat = 26<br>Complete case = 25 | 60<br>59 |
#### Table 9: Participant accountability (India study)
case analysis was undertaken on all participants with available data at baseline and at end of study assessment.
#### Effectiveness
The study met the primary endpoint. The Modius Stress group had a mean reduction in GAD-7 score of -7.44 (p<0.001) while the sham control group reduced their GAD-7 mean score by -2.23 (p=0.002). The mean difference between the groups was statistically significant (ITT, mean difference: -5.21 [95% Cls: -6.57, -3.85]; p<0.001).
Furthermore, more participants in the Modius Stress group, compared with the sham control group, achieved a 4 point or more reduction in GAD-7 from baseline to the follow up visit (33 out of 34 [97%] vs 6 out of 25 [24%]).
{26}------------------------------------------------
Image /page/26/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol to the left of the word "NEUROVALENS" in a bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
### Table 10. The effect of Modius Stress on GAD-7 scores according to randomization group (India Study)
| | Modius Stress | | | Sham Control | | | | |
|------------------|-----------------------|--------------------------------|------------------------------------------------|--------------------------|--------------------------------|------------------------------------------------|-----------------------------------------------------|--------|
| Cohort | Baseline<br>Mean (SD) | 20<br>sessions<br>Mean<br>(SD) | Within-<br>group<br>mean<br>change<br>(95% CI) | Baseline<br>Mean<br>(SD) | 20<br>sessions<br>Mean<br>(SD) | Within-<br>group<br>mean<br>change<br>(95% CI) | Between-<br>group<br>difference<br>mean (95%<br>CI) | P |
| ITT (BOCF) | 12.71<br>(1.59) | 5.26<br>(2.06) | -7.44 (-8.12,<br>-6.76) | 12.77<br>(1.48) | 10.54<br>(3.17) | -2.23 (-3.56,<br>-0.91) | -5.21 (-6.57,<br>-3.85) | <0.001 |
| Complete<br>Case | 12.71<br>(1.59) | 5.26<br>(2.06) | -7.44 (-8.12,<br>-6.76) | 12.88<br>(1.39) | 10.56<br>(3.23) | -2.32 (-3.69,<br>-0.95) | -5.12 (-6.63,<br>-3.62) | <0.001 |
9Between-group difference in GAD-7 scores from baseline to end of study (completion of 20 stimulation sessions) analyzed using Independent Samples t-tests. ITT analysis (BOCF) includes all participants randomized and who had baseline GAD-7 scores (Modius Stress, n=34 and sham control n=26). Complete case analysis includes participants who had baseline and end of study (completion of 20 stimulation sessions) GAD-7 scores (Modius Stress n=34 and sham control n=25).
#### Safety
No causally related adverse events or serious adverse events were reported by the Modius Stress group or the sham control group during the study period.
{27}------------------------------------------------
Image /page/27/Picture/0 description: The image shows the logo for Neurovalens. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVALENS" in bold, sans-serif font. Below the company name is the tagline "Your Health, Our Technology" in a smaller, lighter font.
01 Version 24 of 34 Page
### Modius Stress K232253 Summary
#### Conclusion
ITT and complete case analyses of the pivotal study demonstrates that participants using the Modius Stress device for 20 sessions had a significantly greater reduction in GAD-7 score, compared with the sham group. Additionally, more participants using the Modius Stress had a 4-point or more reduction in GAD-7 score compared to the sham group, when assessed after 20 sessions.
#### Study Limitations
As with all clinical trials, confounding factors can make it difficult to interpret how well the intervention has worked. Therefore, to ensure a robust dataset with a low number of variables, several conditions and medications were excluded in this clinical trial.
- Persons currently taking over the counter or prescribed medications for anxiety, . antidepressant medications, beta blockers, or who regularly use antihistamines.
- . Persons with a history of stroke, or severe injury to the brain.
- . Persons with a cognitive impairment.
- Persons with any disease which may cause vestibular neuropathy. ●
- . Persons with myelodysplastic syndromes, or history of malignancy in the past
- 12-months. ●
- . In the UK study, with the same protocol, participants using the Modius Stress device, 5-7 times per week, had a 5.41 reduction in GAD-7 score, while participants using the sham device had a 3.96 reduction in GAD-7 score, when assessed at 4-weeks. While the Modius Stress group had a greater reduction in GAD-7 score compared with the sham control group, the primary endpoint was not met i.e., the mean difference among the groups was not statistically different. In contrast, the mean difference between the groups was statistically significant in the India study. The difference in results between the two studies is a limitation when extrapolating clinical meaning from this study.
Three participants in the UK study who were excluded from the ITT analysis were randomised (active device, n=1 and sham device, n=2); however, they were withdrawn from…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.