OrionXR

K232189 · Hoth Intelligence, Inc. · LLZ · Sep 14, 2023 · Radiology

Device Facts

Record IDK232189
Device NameOrionXR
ApplicantHoth Intelligence, Inc.
Product CodeLLZ · Radiology
Decision DateSep 14, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

OrionXR is a software device for display, manipulation, and evaluation of externally-generated 3D models of patient anatomy through an Augmented Reality Head Mounted Display (HMD) to assist in visualization, planning and communication of treatment options. OrionXR is indicated for use by qualified healthcare professionals including but not restricted to radiologists, non-radiology specialists, physicians, and technologists. Digital models viewed through the HMD are for informational purposes only and not intended for diagnostic use. OrionXR is not intended to guide surgical instrumentation and it is not to be used for stereotactic procedures or surgical navigation. OrionXR software is designed for use with performance-tested hardware specified in the User Manual.

Device Story

OrionXR is a software system for viewing 3D anatomical models via an Augmented Reality Head Mounted Display (HMD). Input consists of externally-generated 3D models (STL/OBJ formats) uploaded to a web server. The system enables users to access these models on a Microsoft Hololens 2 headset, allowing 3D manipulation (translation/rotation) for visualization and treatment planning. Used in clinical settings (hospitals, clinics, ORs, ICU, ER) by healthcare professionals. Output is a 3D visual representation of patient anatomy. The device facilitates communication and planning; it does not provide diagnostic information or surgical guidance. Benefits include improved perception of anatomical shape and scale for preoperative planning.

Clinical Evidence

Bench testing only. Verification and validation included dimensional accuracy of 3D models, optical performance of the HMD (contrast, resolution, field of view, luminance, distortion, frame rate), and qualitative assessment of 3D models. Human factors and usability testing were conducted in simulated environments to identify use errors and confirm risk mitigations.

Technological Characteristics

System includes a web server for model upload and Microsoft Hololens 2 HMD for visualization. Supports STL and OBJ file formats. Connectivity via encrypted WiFi. Display frame rate is 60 fps. Power source is lithium batteries with 18W charger. No sterilization required. Software-based visualization platform.

Indications for Use

Indicated for qualified healthcare professionals (radiologists, specialists, physicians, technologists) to visualize, plan, and communicate treatment options using 3D anatomical models. Not for diagnostic use, surgical guidance, stereotactic procedures, or surgical navigation.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue. Hoth Intelligence Inc. % Kelliann Payne Partner Hogan Lovells US LLP 1735 Market Street, Suite 2300 PHILADELPHIA, PENNSYLVANIA 19103 September 14, 2023 Re: K232189 Trade/Device Name: OrionXR Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: July 24, 2023 Received: July 24, 2023 Dear Kelliann Payne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica Lamb. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232189 Device Name #### OrionXR Indications for Use (Describe) OrionXR is a software device for display, manipulation, and evaluation of externally-generated 3D models of patient anatomy through an Augmented Reality Head Mounted Display (HMD) to assist in visualization, planning and communication of treatment options. OrionXR is indicated for use by qualified healthcare professionals including but not restricted to radiologists, nonradiology specialists, physicians, and technologists. Digital models viewed through the HMD are for informational purposes only and not intended for diagnostic use. OrionXR is not intended to guide surgical instrumentation and it is not to be used for stereotactic procedures or surgical navigation. OrionXR software is designed for use with performance-tested hardware specified in the User Manual. Type of Use (Select one or both, as applicable) 区 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) SUMMARY Hoth Intelligence, Inc.'s OrionXR K232189 ### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared: Hoth Intelligence Inc. 1700 Market Street Unit 1005, Philadelphia, PA 19103 Contact Person: Kelliann Pavne, Hogan Lovells US LLP Phone: (267) 675-4600 Date Prepared: September 14, 2023 Name of Device: OrionXR Common or Usual Name: Medical image management and processing system ## Classification and Product Code: 21 CFR 892.2050; LLZ Predicate Device: Xironetic, LLC's IntraOpVSP Software Device (K213128) ### Intended Use / Indications for Use OrionXR is a software device for display, manipulation of externally-generated 3D models of patient anatomy through an Augmented Reality Head Mounted Display (HMD) to assist in visualization, planning and communication of treatment options. OrionXR is indicated for use by qualified healthcare professionals including but not restricted to radiologists, non-radiology specialists, physicians, and technologists. Digital models viewed through the HMD are for informational purposes only and not intended for diagnostic use. OrionXR is not intended to guide surgical instrumentation and it is not to be used for stereotactic procedures or surgical navigation. OrionXR software is designed for use with performance-tested hardware specified in the User Manual. ## Technological Characteristics OrionXR includes a server for uploading pre-acquired 3D annotations of patient anatomy and the Microsoft Hololens 2 head mounted display for visualizing the models via a Mixed reality platform. The components of the device include: - 1. Web Server Users can load 3D annotations of anatomy to OrionXR web server which can then be accessed on the Head mounted display. - 2. Head Mounted Display OrionXR is compatible with the Microsoft Hololens 2. A user is able to access 3D digital models on the headset. User can manipulate the model in three dimensions of translational and rotational space. #### Performance Data Software verification and validation were successfully conducted and is summarized below. {4}------------------------------------------------ Design verification and validation was performed to ensure that output specifications meet design input requirements: - Dimensional Accuracy of 3D Models ● - Optical Performance of Headset Display: Contrast ratio, Resolution, Field of View, ● Luminance Uniformity, Eyebox, Distortion, Frame Rate - Qualitative Assessment of 3D Anatomic Models. Human Factors and Usability enqineering testing was also performed to identify use errors that could result in serious harm to the patient or user, as well as to develop relevant risk mitigation measures. This testing included simulated use replicative of both the intended use and the intended environment of use. No additional use-related risks to the safety or effectiveness of the device were identified. ### Substantial Equivalence The OrionXR is as safe and effective as the IntraOpVSP (K213128). The OrionXR has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the OrionXR and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the OrionXR is as safe and effective as IntraOpVSP (K213128). Thus, the OrionXR is substantially equivalent. | | Subject Device | Predicate Device | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | OrionXR | IntraOpVSP (K213128) | | Indication of Use | OrionXR is a software device for<br>display, manipulation, and<br>evaluation of externally-generated<br>3D models of patient anatomy<br>through an Augmented Reality<br>Head Mounted Display (HMD) to<br>assist in visualization, planning<br>and communication of treatment<br>options. | IntraOpVSP is a software device<br>that is indicated for use with an<br>augmented reality head-mounted<br>display which allows for<br>visualization and orientation of<br>3D digital models of selected<br>structures of a patient's anatomy. | | | OrionXR is indicated for use by<br>qualified healthcare professionals<br>including but not restricted to<br>radiologists, non-radiology<br>specialists, physicians, and<br>technologists. | IntraOpVSP is intended to<br>supplement conventional Virtual<br>Surgical Planning (VSP) by<br>facilitating perception of the<br>shape and scale of a patient's<br>anatomical targets for use in<br>preoperative planning and heads-<br>up 3D visualization during<br>surgery. | | | Digital models viewed through the<br>HMD are for informational<br>purposes only and not intended<br>for diagnostic use. OrionXR is not<br>intended to guide surgical<br>instrumentation and it is not to be<br>used for stereotactic procedures<br>or surgical navigation. | IntraOpVSP is not intended to<br>provide diagnosis or to guide<br>surgical instrumentation. It is not<br>to be used for stereotactic<br>procedures or surgical<br>navigation. | | | | IntraOpVSP is intended for use<br>by surgeons who have been | | | OrionXR software is designed for<br>use with performance-tested<br>hardware specified in the User<br>Manual. | trained to operate IntraOpVSP.<br>IntraOpVSP software is designed<br>for use with performance-tested<br>hardware specified in the User<br>Manual. | | | | | | Intended Use<br>Environment | Not for intraoperative use. The<br>software is intended to be used in<br>the following environments:<br>• Operating rooms<br>• Office environment within<br>hospitals or any other<br>clinical setting<br>• Intensive Care unit,<br>Emergency room, or any<br>other location where<br>medical care is provided | Not for intraoperative use. For<br>use in healthcare settings, such<br>as hospitals, clinics and<br>operating rooms | | Intended users | Qualified healthcare professionals<br>including but not restricted to<br>radiologists, non-radiology<br>specialists, physicians, and<br>technologists. | IntraOpVSP is intended for use<br>by surgeons who have been<br>trained to operate IntraOpVSP. | | System<br>Components | • Microsoft Hololens 2 Headset<br>with near eye see-through display<br>• OrionXR Software application | • Microsoft Hololens 2 Headset<br>with near eye see-through<br>display<br>• Software application | | Display Frame<br>Rate | 60 fps | 60 fps | | Medical Device<br>Interfaces | Pre-operative planning<br>workstation OrionXR server | Pre-operative planning<br>workstation Virtual Surgery<br>Planning server | | Communication<br>between<br>Headset and<br>Computer | Wireless—WiFi communication<br>with OrionXR server, encrypted | Wireless, encrypted | | HMD Power<br>Source | Lithium Batteries + 18W charger | Lithium Batteries + 18W charger | | Biocompatibility | N/A | N/A | | Sterilization | Sterilization not required | Sterilization not required | | Data Type<br>Supported | • STL<br>• OBJ | • STL<br>• OBJ | {5}------------------------------------------------ {6}------------------------------------------------ # Conclusions OrionXR is substantially equivalent to the predicate device, IntraOpVSP. OrionXR has the same intended use, similar indications for use, technological characteristics and principles of operation as its predicate device. Testing demonstrates that OrionXR is capable of accurately uploading and visualizing 3D anatomic models on a HMD. Thus, OrionXR does not raise different questions of safety and effectiveness and is substantially equivalent to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%