CoreLink Navigation Instruments

K232116 · Corelink, LLC · OLO · Aug 16, 2023 · Neurology

Device Facts

Record IDK232116
Device NameCoreLink Navigation Instruments
ApplicantCorelink, LLC
Product CodeOLO · Neurology
Decision DateAug 16, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2

Intended Use

CoreLink Navigation Instruments is indicated for use during the preparation and placement of Siber Ti Sacroiliac Joint Fusion System implants and Entasis Dual-Lead Sacroiliac Implant System implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation S8 System (V1.2.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Device Story

CoreLink Navigation Instruments are non-sterile, reusable manual surgical instruments; constructed from stainless steel. Designed to interface with Medtronic StealthStation S8 System; used during spinal surgery to assist surgeons in precisely locating anatomical structures for placement of CoreLink sacroiliac fusion implants. Instruments function as passive markers or tracked tools within the navigation environment; providing spatial reference to the Medtronic system. Surgeon uses navigation output to guide instrument placement; improving accuracy during open or minimally invasive procedures. Benefits include enhanced precision in implant positioning relative to patient anatomy.

Clinical Evidence

Bench testing only. Positional accuracy evaluated per ASTM F2554. Engineering analysis demonstrated that previous testing for K212825 is representative of the subject instruments and that no new worst-case scenarios for accuracy were introduced.

Technological Characteristics

Non-sterile, reusable manual surgical instruments; stainless steel construction. Designed for use with Medtronic StealthStation S8 System (V1.2.0). Positional accuracy testing performed per ASTM F2554. Connectivity is via physical interface with the navigation system tracking hardware.

Indications for Use

Indicated for use during spinal surgery to assist surgeons in locating anatomical structures for the preparation and placement of Siber Ti Sacroiliac Joint Fusion System and Entasis Dual-Lead Sacroiliac Implant System implants in open or minimally invasive procedures.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue. August 16, 2023 CoreLink, LLC % Nathan Wright Engineer & Regulatory Specialist Empirical Technologies 4628 Northpark Drive Colorado Springs, Colorado 80918 Re: K232116 Trade/Device Name: CoreLink Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: July 14, 2023 Received: July 17, 2023 #### Dear Nathan Wright: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Teien D. Soni -S For Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K232116 Device Name CoreLink Navigation Instruments #### Indications for Use (Describe) CoreLink Navigation Instruments is indicated for use during the preparation and placement of Siber Ti Sacroiliac Joint Fusion System implants and Entasis Dual-Lead Sacroiliac Implant System implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation S8 System (V1.2.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. Type of Use (Select one or both, as applicable) � Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection including suggestions for reducing this burden to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K232116 510(K) SUMMARY | Submitter's Name: | CoreLink, LLC | |--------------------------------|------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Address: | 2072 Fenton Logistics Park<br>St. Louis, Missouri 63026 | | Submitter's Telephone: | 888-349-7808 | | Contact Person: | Nathan Wright MS, RAC<br>Empirical Technologies<br>719-351-0248<br>nwright@empiricaltech.com<br>Image: Empirical Technologies Logo | | Date Summary was Prepared: | August 16, 2023 | | Trade or Proprietary Name: | CoreLink Navigation Instruments | | Common Name: | Orthopedic Stereotaxic Instruments | | Device Classification Name: | Stereotaxic Instrument | | Classification & Regulation #: | Class II per 21 CFR §882.4560 | | Product Code: | OLO | | Classification Panel: | Orthopedic - Stereotaxic, Trauma, and Restorative Devices (DHT6C) | ## Description of the Device Subject to Premarket Notification: CoreLink Navigation Instruments are non-sterile, re-usable manual surgical instruments made from stainless steel for use with CoreLink pedicle screw and sacroiliac screw systems. These instruments are designed to interface with the Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures. The purpose of this submission is to offer compatibility of the previously cleared CoreLink Navigation Instruments for sacroiliac fusion implants with additional CoreLink Sacroiliac Screws and to update the sacroiliac navigation instrument set. ## INDICATIONS FOR USE CoreLink Navigation Instruments is indicated for use during the preparation and placement of CoreLink implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medronic StealthStation S8 System (V1.2.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. ### TECHNOLOGICAL CHARACTERISTICS The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are similar between the subject and predicates: - Indications for use - Materials of manufacture - Principles of operation ● - Sizes ● {4}------------------------------------------------ ## K232116 510(K) SUMMARY #### Predicate Devices | 510k Number | Trade or Proprietary or Model Name | Manufacturer | Predicate Type | |-------------|-------------------------------------------------|-------------------------|----------------| | K212825 | CoreLink Navigation Instruments | CoreLink, LLC | Primary | | K161210 | Medtronic Navigated Manual Reusable Instruments | Medtronic Sofamor Danek | Additional | ### Performance Data Performance testing per ASTM F2554 has been conducted to evaluate positional accuracy of the CoreLink Navigation Instruments as submitted under previous submissions of the CoreLink Navigation Instruments (K212825). An engineering analysis has been performed to show that the subject instruments present no new worst case for positional accuracy and that the previous testing is representative of the subject instruments. ### CONCLUSION The overall technology characteristics and mechanical performance data lead to the conclusion that the CoreLink Navigation Instruments are substantially equivalent to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%