K232055 · Shanghai Lingfu Technology Co., Ltd. · LHI · Oct 6, 2023 · General Hospital
Device Facts
Record ID
K232055
Device Name
Vial Adapter
Applicant
Shanghai Lingfu Technology Co., Ltd.
Product Code
LHI · General Hospital
Decision Date
Oct 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Vial Adapter is indicated for the transfer and mixing of drugs contained in vials.
Device Story
Vial Adapter is a single-use, sterile fluid transfer device; consists of polycarbonate housing, piercing spike, and luer connector. Used in healthcare facilities or home settings by patients/caregivers. Device pierces elastomeric septum of drug vial; seats securely around vial ferrule via snap-fit. Opposite side features female/male luer fitting for connection to standard needle-free syringe. Facilitates reconstitution and removal of drug contents. Available in 13mm, 20mm, and 28mm configurations for various vial sizes. Enables safe drug transfer and mixing.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included appearance, particulate, tensile strength, leakage, piercing spike functionality, puncture force, chip generation, housing integrity, luer connector (ISO 80369-7), and chemical properties. Biocompatibility evaluated per ISO 10993-1. Sterilization validated per ISO 11137-1/2. Shelf life validated per ISO 11607-1/2 and ASTM F1980-16. Transportation integrity tested per ASTM D4169-DC13.
Technological Characteristics
Materials: Polycarbonate. Principle: Single-lumen piercing spike with snap-fit housing. Dimensions: 13mm, 20mm, 28mm configurations. Connectivity: Standard Luer (female/male). Sterilization: Electron beam irradiation (SAL 10^-6). Biocompatibility: ISO 10993-1 compliant. Mechanical standards: ISO 80369-7 (Luer).
Indications for Use
Indicated for the transfer and mixing of drugs contained in vials. Intended for use in healthcare facilities or in the home environment by the patient or care-giver to aid and support prescribed treatment and therapy.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 6, 2023
Shanghai Ling Fu Technology Co., Ltd. Esther Zhang Regulatory Affairs 4F, No.585-2 Wanyuan Road, Minhang District Shanghai, Shanghai 201102 China
Re: K232055
Trade/Device Name: Vial Adapter Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LHI Dated: July 7, 2023 Received: July 11, 2023
# Dear Esther Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
Daivd Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232055
Device Name Vial Adapter
Indications for Use (Describe)
The Vial Adapter is indicated for the transfer and mixing of drugs contained in vials.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary- K232055
#### l Submitter
| Device submitter: | Shanghai Ling Fu Technology Co., Ltd.<br>4F, No.585-2 Wanyuan Road, Minhang District, Shanghai<br>P.R.China |
|-------------------|-------------------------------------------------------------------------------------------------------------|
| Contact person: | Esther ZHANG<br>Regulatory affairs<br>Phone: 0086-13771505757<br>Email: Esther.zhang@llins-tech.com |
Prepare Date: October 6, 2023
#### II Device
Trade Name of Device: Vial Adapter Common Name: Fluid Transfer IV Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product code: LHI Review Panel: General Hospital
#### III Predicate Devices
| Trade name: | Vial Adapter 15mm |
|-----------------------------------|-------------------------------------|
| Common name: | Set, I.V. Fluid Transfer |
| Classification/Regulation Number: | Class II, 21 CFR 880.5440 |
| Regulation Name: | Intravascular Administration<br>Set |
| Product Code: | LHI |
| Premarket Notification: | K171796 |
| Manufacturer: | Medimop Medical Projects Ltd |
#### IV Device description
The Vial Adapter consists of luer connector, housing and piercing spike, all of which are made of polycarbonate (PC). The sterile device pierces the elastomeric septum of a drug vial with its integrated piercing spike. The device is then pushed fully onto the drug vial and seats securely around the ferrule of the drug vial utilizing the housing of Vial Adapter. The connector on opposite side of the Vial Adapter is for the connection
{5}------------------------------------------------
of a standard Luer needle free syringe for the reconstitution and removal of the contents of the druq vial.
The proposed Vial Adapter is available in 13mm, 20mm and 28mm diameter configurations to accommodate respective size of drug vials. The device is intended for use in healthcare facilities or in the home environment by the patient or care-giver to aid and support prescribed treatment and therapy.
### V Indications for use
The Vial Adapter is indicated for the transfer and mixing of drugs contained in vials.
### VI Comparison of technological characteristic with the predicate device
The Vial Adapter has the same intended use as the legally marketed predicate device. The differences in technological characteristics between the subject and predicate device do not raise new or different questions of safety and effectiveness.
| Device feature | Subject Device<br>K232055 | Predicate Device<br>K171796 | Comments |
|---------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|------------------------|
| Indications for use | The Vial Adapter is<br>indicated for the<br>transfer and mixing<br>of drugs contained<br>in vials. | The Vial Adapter<br>15mm is indicated<br>for the transfer and<br>mixing of drugs<br>contained in vials | Identical |
| Product code | LHI | LHI | Identical |
| Regulation number | 21 CFR 880.5440 | 21 CFR 880.5440 | Identical |
| Class | CLASS II | CLASS II | Identical |
| Principle of<br>operation | Single use | Single use | Identical |
| Size | 13mm, 20mm,<br>28mm | 15mm | Different<br>Comment 1 |
| Material | Polycarbonate | Polycarbonate | Identical |
| Connector | Female Luer fitting;<br>Male Luer fitting | Luer fitting | Different<br>Comment 2 |
| Piercing Spike | Plastic - Single<br>Lumen | Plastic - Single<br>Lumen | Identical |
| Vial Adapter Fit<br>(Vial Side) | Snap Fit to Vial | Snap Fit to Vial | Identical |
| Sterilization Method | Electron beam<br>Irradiation | Gamma Irradiation | Different<br>Comment 3 |
| Sterility Assurance<br>Level | SAL 10-6 | SAL 10-6 | Identical |
| Biocompatibility | Conforms to ISO<br>10993 | Conforms to ISO<br>10993 | Identical |
| Labeling | Proposed device<br>labeling (IFU)<br>includes transfer | Proposed device<br>labeling (IFU)<br>includes transfer | Identical |
{6}------------------------------------------------
| | and mixing<br>instructions | and mixing<br>instructions | |
|-----------------|----------------------------|----------------------------|------------------------|
| Expiration Date | 3 years | 5 years | Different<br>Comment 4 |
Discussion:
### Comment 1
There are differences in the size of Vial Adapters of the subject device which consist of 13mm, 20mm, 28mm configurations. The different sizes are for different diameter standard vials. The subject device performance testing demonstrates that the difference does not affect the intended use and does not raise new questions of safety and effectiveness.
# Comment 2
The connector of Vial Adapter was divided into a female Luer fitting and a male Luer fitting, while the predicated device has only one luer fitting. This difference was addressed through ISO 80369-7 as well as biocompatibility, sterility and performance testing. The difference does not raise new or different questions of safety and effectiveness.
# Comment 3
The Vial Adapter is provided sterilized by an Electron beam Irradiation method rather than gamma irradiation for the predicate device. However, the validation of the sterilization process in compliance with ISO 11137-1 and ISO 11137-2 ensures the device is adequately sterilized. Therefore, the difference does not raise new or different questions of safety and effectiveness.
# Comment 4
The expiration date of subject device is 3 years which is shorter than the predicated device. The performance of the device over the proposed shelf life was tested to demonstrate that the difference does not affect the intended use and does not raise new questions of safety and effectiveness.
# VII Data
The following data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
Biocompatibility of the Vial Adapter was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External communication device – Blood path
{7}------------------------------------------------
indirect" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended:
| Cytotoxicity | ISO 10993-5: 2009 |
|---------------------------|--------------------|
| Skin sensitization | ISO 10993-10: 2010 |
| Hemolysis | ISO 10993-4: 2017 |
| Intracutaneous reactivity | ISO 10993-10: 2010 |
| Acute systemic toxicity | ISO 10993-11: 2017 |
| Pyrogenicity | ISO 10993-11: 2017 |
# Sterilization and shelf life testing
The sterilization method has been validated to ISO 11137-1 and ISO 11137-2, which has thereby determined the routine control and monitoring parameters. The sterilization process is validated to a minimum SAL 10-6.
The shelf life of the Vial Adapter is determined based on stability study which includes ageing test. The testing is performed according to the following standards:
- A ISO 11607-1: 2019 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- A ISO 11607-2: 2019 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
- A ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
| Performance testing | |
|---------------------|--|
|---------------------|--|
| Performance testing Summary | | | |
|-----------------------------|--------------------------------------|--------------------------------|------|
| Items | Testing standard | Result | |
| Appearance | Internal performance standards | Pass | |
| Particulate | Internal performance standards | Pass | |
| Tensile strength | Internal performance standards | Pass | |
| Leakage | Internal performance standards | Pass | |
| Unobstructed | Internal performance standards | Pass | |
| Piercing Spike | Internal performance standards | Pass | |
| Puncture force | Internal performance standards | Pass | |
| Chips after puncture | Internal performance standards | Pass | |
| Housing | Internal performance standards | Pass | |
| Luer Connector | ISO 80369-7 | Pass | |
| Chemical<br>Properties | Reducing substances<br>(easy oxides) | Internal performance standards | Pass |
| | Metal ions | Internal performance standards | |
| | pH | Internal performance standards | |
| | Evaporation residues | Internal performance standards | |
| | UV absorbance | Internal performance standards | |
| Sterile | Internal performance standards | Pass | |
| Bacterial endotoxin | Internal performance standards | Pass | |
{8}------------------------------------------------
#### Simulated transportation testing
The transportation package has been tested according to ASTM D4169-DC13 test procedure to ensure its integrity during transportation.
#### VIII Conclusion
The Vial Adapter is substantially equivalent to the predicate device, Vial Adapter 15mm cleared under K171796. The non-clinical testing demonstrates that the device is as safe and effective as the predicate device and that the differences in technological characteristics do not raise new or different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.