The SimpleSense Platform is intended for use at home, a healthcare facility, or medical research organization under the direction of a licensed medical professional to record, display, and store the following physiological data: a) 2 leads of Electrocardiogram; b) Respiration rate measured through thoracic impedance; c) Heart Sounds; d) Activity including posture; e) Systolic and Diastolic Blood Pressure and f) other validated data sources. The SimpleSense Platform is intended for use when the licensed medical professional decides to evaluate the physiologic signals of adult patients as an aid to diagnosis and treatment. The SimpleSense Platform is intended to be used by patients at rest with a stationary torso. ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances. The SimpleSense Platform does not produce alarms and is not intended for active patient monitoring. The SimpleSense Platform is not intended for use as life supporting equipment on high-risk patients such as critical care patients. The SimpleSense Platform is not intended for use in the presence of a pacemaker. The SimpleSense-BP software application is intended to estimate, display and store blood pressure data on adult patients who are twenty two (22) years and older. The SimpleSense-BP can be used after a clinician determines the user's hypertension classification via an auscultatory blood pressure cuff measurement. The Blood Pressure algorithm uses patient specific information (age, gender, height and weight) and the blood pressure measurement as inputs. SimpleSense-BP is used to provide blood pressure estimations derived from physiological sensors to qualified medical personnel as a complimentary physiological feature for the purposes of assessing a patient's cardiac health and variance.
Device Story
SimpleSense-BP is a software module for the existing SimpleSense wearable garment/electronics system. It processes ECG, heart sounds, and thoracic impedance data captured by textile nanosensors and MEMS microphones. Inputs include patient demographics (age, gender, height, weight) and a clinician-provided baseline blood pressure measurement (required every 21 days). The system operates at home or in clinical settings; data is transmitted via Bluetooth to a mobile app and then to a web server for physician review. The device provides retrospective blood pressure estimations to assist in assessing cardiac health. It does not provide real-time monitoring or alarms. The system rejects data if motion is detected, ensuring measurements occur only at rest.
Clinical Evidence
Clinical validation study (n=149) evaluated performance against auscultatory reference measurements across four hypertension categories (Normal, Prehypertensive, Stage 1, Stage 2). Results met ISO 81060-2 criteria (MD ≤ ±5 mmHg, SD ≤ 8 mmHg). Accuracy maintained over a 21-day calibration period (n=91) with no evidence of drift. Subgroup analysis confirmed performance across diverse demographics and blood pressure ranges.
Technological Characteristics
Wearable chest-worn garment with textile-based nanosensors for ECG and thoracic impedance; MEMS microphone for heart sounds. Materials: Polyester, ABS, Neoprene, Spandex, Silver, Polyurethane. Connectivity: Bluetooth to mobile app, Wi-Fi/cellular to cloud. Software: Fixed computational algorithm for BP estimation. Non-sterile. Standards: AAMI ES 60601-1, IEC 60601-1-6, IEC 60601-1-2, IEC 60601-2-47, ISO 10993-1, ISO 14971.
Indications for Use
Indicated for adult patients (22+ years) requiring evaluation of physiologic signals (ECG, respiration, heart sounds, activity, blood pressure) as an aid to diagnosis and treatment. Contraindicated for high-risk/critical care patients and patients with pacemakers. Requires stationary torso at rest.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 8, 2023
Nanowear Inc. Venkatesh Varadan CEO 53 Boerum Place, Suite 3F Brooklyn, New York 11201
Re: K232053
Trade/Device Name: SimpleSense-BP, SimpleSense-BP Software Application Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN, DXH, DSB, BZQ, DPS, DQD Dated: July 10, 2023 Received: July 11, 2023
Dear Venkatesh Varadan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Robert T. Kazmierski -S
for
LCDR Stephen Browning Assistant Director Division of Cardiac, Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232053
Device Name SimpleSense-BP
#### Indications for Use (Describe)
The SimpleSense Platform is intended for use at home, a healthcare facility, or medical research organization under the direction of a licensed medical professional to record, display, and store the following physiological data: a) 2 leads of Electrocardiogram; b) Respiration rate measured through thoracic impedance; c) Heart Sounds; d) Activity including posture; e) Systolic and Diastolic Blood Pressure and f) other validated data sources. The SimpleSense Platform is intended for use when the licensed medical professional decides to evaluate the physiologic signals of adult patients as an aid to diagnosis and treatment. The SimpleSense Platform is intended to be used by patients at rest with a stationary torso. ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances.
The SimpleSense Platform does not produce alarms and is not intended for active patient monitoring. The SimpleSense Platform is not intended for use as life supporting equipment on high-risk patients such as critical care patients. The SimpleSense Platform is not intended for use in the presence of a pacemaker.
The SimpleSense-BP software application is intended to estimate, display and store blood pressure data on adult patients who are twenty two (22) years and older. The SimpleSense-BP can be used after a clinician determines the user's hypertension classification via an auscultatory blood pressure cuff measurement. The Blood Pressure algorithm uses patient specific information (age, gender, height and weight) and the blood pressure measurement as inputs. SimpleSense-BP is used to provide blood pressure estimations derived from physiological sensors to qualified medical personnel as a complimentary physiological feature for the purposes of assessing a patient's cardiac health and variance.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY PER 21 CFR §807.92
## 1. Submitter's information
Nanowear Inc.
Contact: Venkatesh Varadan
53 Boerum Pl, Suite 3F Brooklyn, NY 11201 United States
Phone: 718-637-4815
Date: October 11, 2022
# 2. New Device
- Trade Name: SimpleSense-BP
- Common Name: Blood pressure monitor
- Classification Name: ●
- System, Measurement, Blood-Pressure, Non-Invasive (21 CFR §870.1130 Product O Code DXN)
# 3. PRIMARY PREDICATE DEVICE - K212160
Provided all sensors and the majority of functionality has not changed since the prior clearance (no physical changes and 5 of 6 product codes remain the same), Nanowear's prior clearance was selected as the predicate device.
- Trade Name: Nanowear Inc. – SimpleSense Platform
- Common Name: Vital signs monitoring system
- Classification Names:
- O Transmitters and Receivers, Electrocardiograph, Telephone (21 CFR §870.2920 -Product Code: DXH)
- Plethysmograph, Impedance (21 CFR §870.2770 Product Code: DSB) O
- Monitor, Breathing Frequency (21 CFR §868.2375 Product Code: BZQ) O
- Electrocardiograph (21 CFR §870.2340 Product Code: DPS) O
- Stethoscope, Electronic (21 CFR §870.1875 Product Code: DQD) O
# 4. SECONDARY PREDICATE DEVICE - K190792
Since the predicate device does not include product code DXN, the following reference device is provided to establish equivalence to a previously cleared medical device for the additional product code. This device is used to establish similar use of software-based analysis to obtain blood pressure in a multi-parameter system.
- Trade Name: Biobeat Technologies Ltd. – BB-613 WP
- Common Name: Vital signs monitoring system ●
- Classification Names:
{4}------------------------------------------------
- Oximeter (21 CFR §870.2920 Product Code: DQA) о
- System, Measurement, Blood-Pressure, Non-Invasive (21 CFR §870.1130 Product o Code DXN)
- Transmitters And Receivers, Physiological Signal, Radiofrequency (21 CFR §868.2910 o - Product Code: DRG)
# 5. Indications for Use
The SimpleSense Platform is intended for use at home, a healthcare facility, or medical research organization under the direction of a licensed medical professional to record, display, and store the following physiological data: a) 2 leads of Electrocardiogram; b) Respiration rate measured through thoracic impedance; c) Heart Sounds; d) Activity including posture; e) Systolic and Diastolic Blood Pressure and f) other validated data sources. The SimpleSense Platform is intended for use when the licensed medical professional decides to evaluate the physiologic signals of adult patients as an aid to diagnosis and treatment. The SimpleSense Platform is intended to be used by patients at rest with a stationary torso. ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances.
The SimpleSense Platform does not produce alarms and is not intended for active patient monitoring. The SimpleSense Platform is not intended for use as life supporting equipment on high-risk patients such as critical care patients. The SimpleSense Platform is not intended for use in the presence of a pacemaker.
The SimpleSense-BP software application is intended to estimate, display and store blood pressure data on adult patients who are twenty two (22) years and older. The SimpleSense-BP can be used after a clinician determines the user's hypertension classification via an auscultatory blood pressure cuff measurement. The Blood Pressure algorithm uses patient specific information (age, gender, height and weight) and the blood pressure measurement as inputs. SimpleSense-BP is used to provide blood pressure estimations derived from physiological sensors to qualified medical personnel as a complimentary physiological feature for the purposes of assessing a patient's cardiac health and variance.
# 6. Device Description
The SimpleSense-BP Software Application accesses the physiological parameters like ECG, heart sounds, and thoracic impedance captured by the SimpleSense Device for processing into the vital sign outputs of the product which includes estimation of Systolic and Diastolic blood pressure. The software uses recorded data from the SimpleSense electronics module as inputs into a validated computational model for estimating blood pressure over the period of wear. The system samples blood pressure while the user is at rest. In addition, SimpleSense-BP Software utilizes inputs such as demographic information (age, weight, height, and gender) and a blood pressure measurement for clinical stratification to the algorithm. The blood pressure outputs are returned to the SimpleSense Mobile Application and/or SimpleSense webserver for display, review and interpretation by a physician.
The Nanowear SimpleSense system is a non-invasive, wearable, and portable medical device for the evaluation and monitoring of patients. It utilizes physiologic and biometric sensors embedded in a garment and an electronics module to gather the heart health data.
{5}------------------------------------------------
The specific physiological parameters recorded by the device include: two vectors of Electrocardiogram (ECG), respiratory rate though thoracic impedance, heart sounds, and activity including posture. The signals are recorded by the electronics module on a removable data storage card and are periodically transferred to a smartphone mobile application that connects to the electronics module over a wireless Bluetooth connection. The mobile application provides the functionality of transferring the data collected by the electronics module then relaying the data to the Nanowear web server for display of the data by a physician.
# 7. Intended Use
The Nanowear SimpleSense-BP blood pressure software module is intended to estimate systolic and diastolic blood pressure using the SimpleSense device data and demographic information of the patient inclusive of age, weight, and gender. In addition, an initial blood pressure value for clinical stratification must be determined and entered by a clinician. The blood pressure value is also input into the device as an input to the blood pressure algorithm. A new blood pressure measurement needs to be taken and updated into the system every 21 days.
# 8. CLINICAL USE MODEL
The SimpleSense-BP software application allows medical professionals to remotely obtain, and review blood pressure measurements computed from the SimpleSense device data. Under the care of a physician, a medical professional prescribes the SimpleSense device which captures data via the garment, stores and transmits data from a Signal Acquisition Unit (SAU) to a smartphone device via Bluetooth and wirelessly transmits via Wi-Fi or cellular data for review by a medical professional. The data that is transmitted to the Nanowear servers is used as inputs to the SimpleSense-BP software algorithm. The output of the algorithm is estimated Systolic and Diastolic blood pressure.
### 9. Substantial Equivalence Comparison
See §4 (Indications for Use) above to review bold font that demarks the difference between the predicate device and the SimpleSense-BP software.
The table below delineates the differences between the cleared SimpleSense Platform, and the proposed SimpleSense-BP software algorithm. The table includes comparisons of the system features of the devices. Any differences between the two designs have been marked with bold font which are further discussed below the table.
{6}------------------------------------------------
# Table 1: Predicate Comparison: System Design
| | Nanowear Inc.<br>SimpleSense Platform | Nanowear Inc.<br>SimpleSense-BP (on SimpleSense Platform) | BioBeat Technologies LTD<br>BB-613 | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| 510(k)<br>Number | K212160 | New Device - K214069 | K190792 | |
| Indications<br>for Use<br>Statement | The SimpleSense Platform is<br>intended for use at home, or at a<br>healthcare facility, under the<br>direction of a licensed medical<br>professional, to record, display and<br>store the following physiological<br>data: a) 2 leads of<br>Electrocardiogram; b) Respiration<br>rate measured through thoracic<br>impedance; c) Heart Sounds; d)<br>Activity including posture; and e)<br>other validated data sources. The<br>SimpleSense Platform is intended<br>for use when the licensed medical<br>professional decides to evaluate<br>the physiologic signals of adult<br>patients as an aid to diagnosis and<br>treatment. The SimpleSense<br>Platform is intended to be used by<br>patients at rest and not performing<br>any activities or movements. ECG<br>recordings are indicated for the<br>manual assessment of cardiac<br>rhythm disturbances. The<br>SimpleSense Platform does not<br>produce alarms and is not intended<br>for active patient monitoring (real-<br>time).<br>The SimpleSense Platform is not<br>intended for use as life supporting<br>equipment on high-risk patients<br>such as critical care patients. The<br>SimpleSense Platform is not<br>intended for use in the presence of<br>a pacemaker. | The SimpleSense Platform is intended for use at home,<br>a healthcare facility, or medical research organization<br>under the direction of a licensed medical professional<br>to record, display, and store the following physiological<br>data: a) 2 leads of Electrocardiogram; b) Respiration<br>rate measured through thoracic impedance; c) Heart<br>Sounds; d) Activity including posture; e) Systolic and<br>Diastolic Blood Pressure and f) other validated data<br>sources. The SimpleSense Platform is intended for use<br>when the licensed medical professional decides to<br>evaluate the physiologic signals of adult patients as an<br>aid to diagnosis and treatment. The SimpleSense<br>Platform is intended to be used by patients at rest with<br>a stationary torso. ECG recordings are indicated for the<br>manual assessment of cardiac rhythm disturbances.<br>The SimpleSense Platform does not produce alarms<br>and is not intended for active patient monitoring.<br>The SimpleSense Platform is not intended for use as life<br>supporting equipment on high-risk patients such as<br>critical care patients. The SimpleSense Platform is not<br>intended for use in the presence of a pacemaker.<br>The SimpleSense-BP software application is intended to<br>estimate, display and store blood pressure data on<br>adult patients who are twenty two (22) years and<br>older. The SimpleSense-BP can be used after a clinician<br>determines the user's hypertension classification via a<br>FDA cleared blood pressure cuff and enters patient<br>specific information (age, gender, height and weight).<br>SimpleSense-BP is used to provide blood pressure<br>calculations derived from physiological sensors to<br>qualified medical personnel as a complimentary<br>physiological feature for the purposes of assessing a<br>patient's cardiac health and variance. | The BB-613 WP is a<br>wrist-worn or skin<br>attached device<br>indicated for use in<br>measuring and<br>displaying functional<br>oxygen saturation of<br>arterial hemoglobin<br>(%SpO2) and pulse rate.<br>The BB-613WP can also<br>track changes in blood<br>pressure based on Pulse<br>Wave Transit Time<br>(PWTT) which is<br>obtained utilizing pulse<br>measurements from the<br>integrated SpO2 sensor,<br>following a calibration<br>process using<br>oscillometric blood<br>pressure monitor.<br>The BB-613WP is<br>intended for spot-<br>checking of adult<br>patients in hospitals,<br>clinics, long-term care,<br>and home use. | |
| Product Code | DXH, DSB, BZQ, DPS, DQD | DXH, DSB, BZQ, DPS, DQD, DXN | DXN, DQA, DRG | |
| Acquired<br>Data | • Electrocardiogram<br>• Respiration Rate from<br>thoracic impedance<br>• Heart Sounds<br>• Activity | • Electrocardiogram<br>• Respiration Rate from thoracic impedance<br>• Heart Sounds<br>• Activity<br>• Systolic and Diastolic Blood Pressure from a<br>combination of above data and patient<br>information | • Oxygen Saturation<br>(SpO2)<br>• Pulse Rate (PR)<br>• Blood Pressure (BP) | |
| Target<br>Population | Adult patients requiring<br>evaluation of physiologic signals | No change - Same | Adults | |
| | Nanowear Inc.<br>SimpleSense Platform | Nanowear Inc.<br>SimpleSense-BP (on SimpleSense Platform) | BioBeat Technologies<br>LTD<br>BB-613 | |
| 510(k)<br>Number | K212160 | New Device - K214069 | K190792 | |
| Monitoring<br>Type | • Intermittent, Spot-checking | No change - Same | Spot-checking | |
| Sensor<br>Technology | • Electrocardiogram: Textile-based<br>nanosensors<br>• Respiration Rate: Textile-based<br>nanosensors measure thoracic<br>impedance and respiration is<br>derived from thoracic impedance<br>• Heart Sound:<br>Microelectromechanical (MEMs)<br>microphone | No change - Same | Pulse reflectance<br>technology, Four LED<br>(red + IR) and photo<br>diode absorbs reflected<br>light.<br>[Photoplethysmograp<br>hy PPG sensor] | |
| Anatomical<br>Sites | Chest worn sash with multiple<br>embedded sensors | No change - Same | Wrist worn device or<br>adhesive patch on<br>chest | |
| Materials | Polyester, ABS Hook and Loop,<br>Neoprene, Spandex, Silver and<br>Polyurethane | No change - Same<br>The materials used in Nanowear hardware have been<br>consistent across all Nanowear products, including this 510(k)<br>for the SimpleSense-BP. The SimpleSense system hardware<br>(K201669) was used for validating the SimpleSense-BP<br>software and is the exclusive hardware to be used with<br>SimpleSense-BP software. | Polycarbonate,<br>Photodiode window,<br>Silicone<br>Adhesive patch | |
| Signal<br>Acquisition<br>Method | • Electrocardiogram (EKG/ECG):<br>Standard Bipolar lead<br>instrumentation amplifier and<br>Sigma-Delta Analog to Digital<br>Converter.<br>• Respiration rate: Thoracic<br>Impedance is measured using<br>four-point probe using low<br>amplitude current applied to the<br>body and impedance measured<br>from voltage drop derived from<br>thoracic impedance.<br>• Heart Sound: solid state<br>Microelectromechanical (MEMs)<br>microphone embedded in<br>garment and located near Apex<br>of heart. | • Electrocardiogram (EKG/ECG): Standard Bipolar lead<br>instrumentation amplifier and Sigma-Delta Analog to<br>Digital Converter.<br>• Respiration rate: Thoracic Impedance is measured<br>using four-point probe using low amplitude current<br>applied to the body and impedance measured from<br>voltage drop derived from thoracic impedance.<br>• Heart Sound: solid state Microelectromechanical<br>(MEMs) microphone embedded in garment and<br>located near Apex of heart.<br>• Blood Pressure: software algorithm model for<br>estimating systolic and diastolic blood pressure<br>based on SimpleSense device data. | • Pulse Wave Transit<br>Time (PWTT)<br>calculation based on<br>measurement of<br>functional oxygen<br>saturation | |
| Display Type | No on-device display<br>Software application for iOS and<br>Android | No change - Same | LCD on device or<br>handheld display unit | |
| | Nanowear Inc.<br>SimpleSense Platform | Nanowear Inc.<br>SimpleSense-BP (on SimpleSense Platform) | BioBeat Technologies<br>LTD<br>BB-613 | |
| 510(k)<br>Number | K212160 | New Device - K214069…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.