The TDS (TECHWIN DENTAL SYSTEM) are prefabricated prosthetic components for direct connection to endosseous dental implants and are intended to be used as an aid in prosthetic rehabilitation. It is intended for use in edentulous and partially edentulous patients with established bone and completed growth period to support single unit prosthetic restorations in the mandible or maxilla. It is intended for one-stage and two-stage surgical procedures. This system is intended for delayed loading.
Device Story
TDS (TECHWIN DENTAL SYSTEM) consists of prefabricated prosthetic components (Dual Abutment, Healing Abutment, Abutment Screw) for connection to endosseous dental implants. Used in dental clinics by clinicians for prosthetic rehabilitation in edentulous/partially edentulous patients. Components support single-unit restorations via delayed loading. Abutments are made of Ti-6Al-4V ELI alloy (ASTM F136), some with TiN coating via Arc Ion Plating. Supplied non-sterile; requires end-user steam sterilization. Device interfaces with specific third-party implants (Chaorum Implant System). Clinical benefit is the restoration of dental function and aesthetics for patients with missing teeth.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-1, -5, -6, -10, -23), surface treatment analysis (SEM, roughness, coating thickness), sterilization validation (ISO 17665), and MRI safety assessment.
Technological Characteristics
Materials: Ti-6Al-4V ELI alloy (ASTM F136). Surface: Uncoated or TiN coated (Arc Ion Plating). Connection: Hex/Non-hex internal conical. Sterilization: Steam (autoclave) by end-user. Dimensions: Various diameters, post heights, and gingival heights. Connectivity: None (mechanical component).
Indications for Use
Indicated for edentulous and partially edentulous patients with established bone and completed growth period requiring single-unit prosthetic restorations in the mandible or maxilla.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
TECHWIN Co., Ltd. % Kyung-Hwan Kim Representative Consultant SMB Korea #606, #607, 7, Boramae-ro 5ga-gil, Donjak-gu Seoul, 07071 REPUBLIC OF KOREA
August 20, 2024
Re: K231932
Trade/Device Name: TDS (TECHWIN DENTAL SYSTEM) Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 16, 2024 Received: July 23, 2024
Dear Kyung-Hwan Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231932
#### Device Name
TDS (TECHWIN DENTAL SYSTEM)
#### Indications for Use (Describe)
The TDS (TECHWIN DENTAL SYSTEM) are prefabricated prosthetic components for direct connection to endosseous dental implants and are intended to be used as an aid in prosthetic rehabilitation. It is intended for use in edentulous and partially edentulous patients with established bone and completed to support single unit prosthetic restorations in the mandible or maxilla. It is intended for one-stage surgical procedures. This system is intended for delayed loading.
The TDS (TECHWIN DENTAL SYSTEM) is compatible with the following devices:
| No. | Subject<br>Device | Implantable<br>Diameters (mm) | Compatible<br>Implants (510k<br>No.)/Manufacture | Type of Implant<br>Abutment<br>Connection/ | Implant<br>Platform<br>Type | Implant Platform<br>Platform<br>Diameters(mm) | Implant Body<br>Diameter<br>(mm) |
|-----|---------------------|-----------------------------------------------------------------------------|-------------------------------------------------------------------|--------------------------------------------|-----------------------------|-----------------------------------------------|--------------------------------------|
| 1. | Dual<br>Abutment | Ø4.0, 4.6 | Chaorum<br>Implant System<br>(K160536)/<br>MEDIMECCA<br>CO., LTD. | Hex or non-hex/<br>2.1 Hex | Narrow<br>(mini) | Ø3.25, 3.89 | Ø3.0, 3.5 |
| | | Ø4.5, 4.6, 5.0, 5.5,<br>6.0, 6.5, 7.0 | | Hex or non-hex/<br>2.5 Hex | Wide<br>(Regular) | Ø3.89, 4.28,<br>4.78, 5.28, 6.28 | Ø3.9, 4.0, 4.4,<br>4.5, 4.9, 5.5, 5. |
| 2. | Healing<br>Abutment | Ø4.3, 4.8 | | Hex or non-hex/<br>2.1 Hex | Narrow<br>(mini) | Ø3.25, 3.89 | Ø3.0, 3.5 |
| | | Ø4.0, 4.3, 4.5, 4.6,<br>4.8, 5.3, 5.5, 5.6, 6.3.<br>6.5, 7.3, 7.5, 8.3, 8.5 | | Hex or non-hex/<br>2.5 Hex | Wide<br>(Regular) | Ø3.89, 4.28,<br>4.78, 5.28, 6.28 | Ø3.9, 4.0, 4.4,<br>4.5, 4.9, 5.5, 5. |
| 3. | Abutment<br>Screw | Ø2.22 | | - | Narrow<br>(mini) | - | - |
| | | Ø2.1, 2.3, 2.32, 2.5 | | | Wide<br>(Regular) | - | - |
Type of Use (Select one or both, as applicable)
∑ Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
TECHWIN Co., Ltd Traditional 510(k) Premarket Submission TDS (TECHWIN DENTAL SYSTEM) K231932
Image /page/5/Picture/1 description: The image shows the logo for SMB Korea, a consulting expert group. The logo features the text "SMB Korea" in a gray, sans-serif font. Below the text is the tagline "A Member of the Consulting Expert Group" in a smaller, lighter gray font. A curved line underlines the "SMB" portion of the logo, adding a visual element to the design.
# 510(k) Summary
# For
# TDS (TECHWIN DENTAL SYSTEM)
[Complying with 21 CFR 807.92]
#### SUBMISSION SPONSOR l.
TECHWIN Co., Ltd. #401, Eungyenam-gil 40, Siheung-si, Gyeonggi-do, Republic of Korea Office Phone: +82-31-365-5601 Email: tw rnd@techwinkorea.com Contact Person: Mr. Min-jae Jo, QMR
# II. SUBMISSION CORRESPONDENT
SMB Korea #606, #607, 7, Boramae-ro 5ga-gil, Donjak-gu, Seoul, 07071, Republic of Korea Cell Phone: +82-10-2247-5579 Office Phone: +82-2-6241-9001 Kyung-hwan Kim, Representative Consultant RA/QA Contact: Email: info@smbkorea.com
## III. DATE PREPARED
August 19, 2024
## IV. DEVICE
| Trade or Proprietary Name: | TDS (TECHWIN DENTAL SYSTEM) |
|----------------------------|--------------------------------------------------|
| Common or Usual Name: | Endosseous Dental Implant Abutment |
| Classification Name: | Abutment, Implant, Dental, Endosseous (872.3630) |
| Regulatory Class: | II |
| Product Code: | NHA |
| Classification Panel: | Dental |
## V. PREDICATE DEVICE
Primary Predicate Device: K160536, Chaorum Implant System / Medimecca Co., Ltd.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for SMB Korea. The logo is in gray and consists of the text "SMB Korea" with a curved line underneath. Below the line, it says "A Member of the Consulting Expert Group" in a smaller font. The logo is simple and professional.
## VI. DEVICE DESCRIPTION
The TDS (TECHWIN DENTAL SYSTEM), made of Ti-6A1-4V ELI alloy (ASTM F136), is intended for use in conjunction with the appliance in partially or fully edentulous mandibles and maxillae to support single-unit loading (i.e., crown). It consists of the Dual Abutment and Healing Abutment. The surface of the abutments is uncoated or partially or half coated with TiN using AIP (Arc Ion Plating). The implant-to-abutment interface remains uncoated. All abutments are supplied non-sterile and must be autoclaved by the end-user.
#### The TDS (TECHWIN DENTAL SYSTEM) is compatible with the following devices:
| No. | 510k No. | Trade Name | | Platform Type | Platform Diameters<br>(mm) | Manufacture |
|-----|----------|------------|---------|---------------|----------------------------|---------------------|
| | K160536 | Chaorum | Implant | Narrow | Ø3.25~3.89 | MEDIMECCA CO., LTD. |
| | | System | | Regular | Ø3.89~6.28 | |
| Abutment | Diameter (ø) | Post<br>Height<br>(mm) | Total Length (mm) | Cuff Height<br>(mm) | Angulation<br>(°) | Surface |
|---------------------|-------------------------------------------------------------------------------|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|-------------------|--------------------|
| Dual<br>Abutment | 4.0, 4.5, 4.6, 5.0,<br>5.5, 6.0, 6.5, 7.0 | 4.0, 5.5,<br>7.0 | 7.5, 8.5, 9.0, 9.1, 9.2, 9.5, 9.7, 10.0, 10.1,<br>10.5, 10.6, 10.7, 10.85, 11.0, 11.1, 11.2,<br>11.5, 11.6, 11.7, 11.85, 12.0, 12.1, 12.2,<br>12.5, 12.6, 12.7, 12.85, 13.0, 13.1, 13.2,<br>13.5, 13.6, 13.7, 13.85, 14.0, 14.2, 14.5,<br>14.6, 14.85, 15.2 | 1.0, 1.3, 1.8,<br>2.3, 2.8, 3.3,<br>3.8, 4.3, 4.8,<br>5.3 | - | TiN<br>coating |
| Healing<br>Abutment | 4.0, 4.3, 4.5, 4.6,<br>4.8, 5.3, 5.5, 5.6,<br>6.3, 6.5, 7.3, 7.5,<br>8.3, 8.5 | 2.0, 3.0,<br>3.5, 4.0,<br>5.0, 5.5,<br>7.0 | 7.5, 8.5, 8.7, 9.5, 10.5, 11.0, 11.5, 12.0,<br>12.5, 14.0, 14.5 | 1.0, 2.0, 3.0,<br>4.0, 5.0 | - | Non/TiN<br>coating |
| Abutment<br>Screw | 2.1, 2.22, 2.3,<br>2.32, 2.5 | - | 8, 8.35, 8.5, 8.8, 9.8, 10.1, 10.2 | - | - | Machined |
#### The abutment's dimension range is as follows:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for SMB Korea. The logo is in gray and features the text "SMB Korea" in a sans-serif font. A curved line extends from under the "S" in SMB to under the "K" in Korea. Below the text is the phrase "A Member of the Consulting Expert Group" in a smaller font.
#### VII. INDICATION FOR USE
The TDS (TECHWIN DENTAL SYSTEM) are prefabricated prosthetic components for direct connection to endosseous dental implants and are intended to be used as an aid in prosthetic rehabilitation. It is intended for use in edentulous and partially edentulous patients with established bone and completed growth period to support single unit prosthetic restorations in the mandible or maxilla. It is intended for one-stage and twostage surgical procedures. This system is intended for delayed loading.
| No. | Subject<br>Device | Implantable<br>Diameters (mm) | Compatible<br>Implants (510k<br>No.)/<br>Manufacture | Type of Implant -<br>Abutment<br>Connection/<br>Diameter (mm) | Implant<br>Platform<br>Type | Implant<br>Platform<br>Diameters<br>(mm) | Implant Body<br>Diameters<br>(mm) |
|-----|---------------------|-----------------------------------------------------------------------------|------------------------------------------------------|---------------------------------------------------------------|-----------------------------|------------------------------------------|------------------------------------------|
| 1. | Dual<br>Abutment | Ø4.0, 4.6 | Chaorum<br>Implant System<br>(K160536)/ | Hex or non-hex/<br>2.1 Hex | Narrow<br>(mini) | Ø3.25, 3.89 | Ø3.0, 3.5 |
| | | Ø4.5, 4.6, 5.0, 5.5,<br>6.0, 6.5, 7.0 | MEDIMECCA<br>CO., LTD. | Hex or non-hex/<br>2.5 Hex | Wide<br>(Regular) | Ø3.89, 4.28,<br>4.78, 5.28,<br>6.28 | Ø3.9, 4.0, 4.4,<br>4.5, 4.9, 5.5,<br>5.9 |
| 2. | Healing<br>Abutment | Ø4.3, 4.8 | | Hex or non-hex/<br>2.1 Hex | Narrow<br>(mini) | Ø3.25, 3.89 | Ø3.0, 3.5 |
| | | Ø4.0, 4.3, 4.5, 4.6,<br>4.8, 5.3, 5.5, 5.6, 6.3,<br>6.5, 7.3, 7.5, 8.3, 8.5 | | Hex or non-hex/<br>2.5 Hex | Wide<br>(Regular) | Ø3.89, 4.28,<br>4.78, 5.28,<br>6.28 | Ø3.9, 4.0, 4.4,<br>4.5, 4.9, 5.5,<br>5.9 |
| 3. | Abutment<br>Screw | Ø2.22 | | | Narrow<br>(mini) | - | - |
| | | Ø2.1, 2.3, 2.32, 2.5 | | | Wide<br>(Regular) | - | - |
#### The TDS (TECHWIN DENTAL SYSTEM) is compatible with the following devices:
#### VIII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
<Substantial Equivalence to Predicate device Table – TDS (TECHWIN DENTAL SYSTEM)>
| | SUBJECT Device | Primary PREDICATE Device<br>(K160536) | Significant<br>Difference |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Manufacturer | TECHWIN Co., Ltd. | Medimecca Co., Ltd. | - |
| Trade Name | TDS (TECHWIN DENTAL SYSTEM) | Chaorum Implant System | - |
| Regulation<br>Description | Endosseous Dental Implant System | Endosseous Dental Implant System | - |
| Regulation<br>Number | 21 CFR 872.3630 | 21 CFR 872.3630 | No difference. |
| Product Code | NHA | NHA | No difference. |
| Class | II | II | No difference. |
| Indications | The TDS (TECHWIN DENTAL<br>SYSTEM) are prefabricated<br>prosthetic components for direct<br>connection to endosseous dental<br>implants and are intended to be<br>used as an aid in prosthetic<br>rehabilitation. It is intended for use<br>in edentulous and partially<br>edentulous patients with | Chaorum Implant System is<br>intended for use in partially or fully<br>edentulous mandibles and maxilla,<br>in support of single of multiple-unit<br>restorations including; cemented<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or intermediate abutment<br>support for fixed bridgework. | Similarities |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for SMB Korea, a consulting expert group. The logo features the text "SMB Korea" in a stylized font, with a curved line extending from the bottom of the "B" in SMB. Below the main text, in a smaller font, is the phrase "A Member of the Consulting Expert Group."
| A Member of the Consulting Expert Group | | | | | |
|-----------------------------------------|--|--|--|--|--|
|-----------------------------------------|--|--|--|--|--|
| | SUBJECT Device | Primary PREDICATE Device<br>(K160536) | Significant<br>Difference |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| | established bone and completed<br>growth period to support single<br>unit prosthetic restorations in the<br>mandible or maxilla. It is intended<br>for one-stage and two-stage<br>surgical procedures. This system is<br>intended for delayed loading. | Chaorum Implant System is for<br>single stage and two stage surgical<br>procedures. This system is<br>intended for delayed loading. | |
| | Dual Abutment | | |
| Design | | | Similarities |
| | Image: [Hex/Non-hex O-type] | Image: [Hex/Non-hex] | |
| | Image: [Hex/Non-hex D-type] | | |
| | Image: [Hex I-type] | | |
| Diameter (ø) | [Hex/Non-hex O-type]<br>: 4.0, 4.6, 5.0, 6.0, 7.0 mm<br>[Hex/Non-hex D-type]<br>: 4.5, 5.5, 6.5 mm<br>[Hex I-type]<br>: 4.5, 5.5 mm | 3.5, 4.0, 4.5, 5.5, 5.5, 6.0, 6.5 mm | Similarities |
| Post Height | 4.0, 5.5, 7.0 mm | 4.5, 5.5, 7.0 mm | Similarities |
| Gingival Height | [Hex/Non-hex O-type]<br>: 1.0, 1.8, 2.8, 3.8, 4.8 mm<br>[Hex/Non-hex D-type]<br>: 1.0, 1.3, 2.3, 3.3, 4.3, 5.3 mm<br>[Hex I-type]<br>: 1.0, 1.8, 2.8, 3.8, 4.8 mm | 0.85, 1.35, 2.35, 3.35, 4.35, 5.35 mm | Similarities |
| Raw material | Ti-6Al-4V ELI Titanium Alloy | Titanium Grade 4 | Similarities |
| Connection<br>Interface | Hex, Non-hex | Hex, Non-hex | No difference |
| Surface<br>Treatment | TiN Coating | TiN Coating | No difference. |
| Sterilization | Non-sterile; Steam sterilization<br>prior to use | Non-sterile; Steam sterilization<br>prior to use | No difference. |
| Single<br>Use Only | Single use Only | Single use Only | No difference. |
| | SUBJECT Device | Primary PREDICATE Device<br>(K160536) | Significant<br>Difference |
| Biocompatibility | Biocompatible according to ISO<br>10993-1 | Biocompatible according to ISO<br>10993-1 | No difference. |
| SE | The subject device has a similar intended use, material, and design as the predicate device,<br>with only minor differences in dimensions. While the subject device is indicated for single-<br>unit loading only, as opposed to single and multiple-unit loading for the predicate device,<br>this represents a more conservative approach that doesn't introduce additional risks. The<br>abutment gingival height range of the subject device falls within the range of the predicate<br>device, and the minor differences in diameter and post height are intended to meet<br>individual patient needs without compromising performance or safety. The 4.0 mm<br>minimum abutment post height of the subject device, compared to the 4.5 mm of the<br>predicate device, allows for greater flexibility to accommodate patient-specific anatomical<br>variations without affecting safety or effectiveness. Although the subject device uses Ti-6Al-<br>4V ELI titanium alloy instead of titanium grade 4 used in the predicate, both materials are<br>well-established in dental implant applications, with Ti-6AI-4V ELI offering improved<br>mechanical properties and biocompatibility. The connection interface remains the same<br>between the devices. These differences do not raise any new safety or effectiveness issues,<br>maintaining substantial equivalence between the subject and predicate devices. | | |
| | | | |
| Design | Image: [O-type]<br><br>Image: [D-type]<br><br>Image: [I-type] | Image: [O-type] | Similarities |
| Diameter (ø) | [O-type]<br>: 4.3, 4.8, 5.3, 6.3, 7.3, 8.3 mm<br>[D-type]<br>: 4.0, 4.5, 5.5, 6.5, 7.5, 8.5 mm<br>[I-type]<br>: 4.6, 5.6 mm | 3.5 ~ 8.5 mm | Similarities |
| Post Height | [O-type]<br>: 3.0, 4.0, 5.0, 7.0 mm<br>[D-type]<br>: 2.0, 3.5, 5.0, 7.0 mm<br>[I-type] | 2.0 ~ 7.0 mm | Similarities |
| | SUBJECT Device | Primary PREDICATE Device<br>(K160536) | Significant<br>Difference |
| | : 4.0, 5.5, 7.0 mm | | |
| Gingival Height | [O-type]<br>: 1.0, 2.0, 3.0, 5.0 mm<br>[D, I-type]<br>: 2.0, 3.0, 4.0 mm | 0.0 ~ 5.5 mm | Similarities |
| Raw material | Ti-6Al-4V ELI Titanium Alloy | Titanium Grade 4 | Similarities |
| Surface<br>Treatment | Non/TiN Coating | TiN Coating | Similarities |
| Sterilization | Non-sterile; Steam sterilization<br>prior to use | Non-sterile; Steam sterilization<br>prior to use | No difference. |
| Single<br>Use/Reuse | Single use Only | Single use Only | No difference. |
| Biocompatibility | Biocompatible according to ISO<br>10993-1 | Biocompatible according to ISO<br>10993-1 | No difference. |
| SE | The subject device has a similar intended use, material, and design as the predicate device, with only minor differences in dimensions. The abutment post height and gingival height range of the subject device fall within the range of the predicate device. While there are minor differences in diameter, these variations are intended to meet individual patient needs and do not compromise performance or safety due to their minimal nature. The subject device uses Ti-6Al-4V ELI titanium alloy, whereas the predicate uses titanium grade 4. Both materials are well-established in dental implant applications, with Ti-6Al-4V ELI offering improved mechanical properties and biocompatibility. This material difference maintains substantial equivalence without raising new safety or efficacy concerns. Additionally, the subject device offers both non-coated and TiN-coated options for the Healing Abutment, while the predicate device uses only TiN coating. This flexibility in surface treatment allows for customization based on specific clinical needs without compromising the device's overall performance or biocompatibility. Furthermore, the connection interface remains the same between the devices. These minor variations do not introduce any new issues in performance or safety, thus maintaining substantial equivalence between the subject and predicate devices. | | |
| | Abutment Screw | | |
| Design | <Abutment Screw><br>Image: [Abutment Screw O-type] | <Abutment Screw><br>Image: [Abut…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.