Therm-X (Therm-X Home and Therm-X AT) combines cold, heat, contrast, and compression therapy. Therm-X is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated. Therm-X Home systems also provide DVT therapy. Therm-X Home systems with DVT therapy are intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression. Therm-X (Therm-X Home and Therm-X AT) is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.
Device Story
Therm-X is an AC-powered, software-controlled multimodality device for thermal and compression therapy. It uses a centralized coolant reservoir, pumps, fans, and thermoelectric modules to circulate fluid through flexible garments (back, elbow, hand/wrist, shoulder, neck, ankle, hip, knee, calf, foot). The device provides cold, heat, contrast, and pneumatic compression. Therm-X Home includes DVT prophylaxis. Operated via touch screen interface; allows adjustment of treatment time, temperature, and compression settings. Features password protection for cycle compliance. Used in clinical or home settings by healthcare professionals or patients under prescription. Benefits include edema/pain reduction and DVT risk mitigation. The device supports two-patient treatment via split umbilical hose.
Clinical Evidence
Bench testing verified performance, electrical safety (IEC 60601-1, 1-2, 1-11), and biocompatibility (ISO 10993). Clinical testing consisted of an IRB-approved study on healthy volunteers to measure minimum skin temperature (40°F) to ensure no thermal damage. Human factors/usability assessments in simulated environments confirmed safe and effective operation by intended users.
Technological Characteristics
Thermoelectric heating/cooling; pneumatic compression. Materials: 30/200 denier nylon coated in urethane. Dimensions: 15"x10.5"x9". Power: 100-240 VAC, 50/60 Hz. Connectivity: Standalone with touch screen interface. Software: Moderate level of concern (per 2005 guidance) / Enhanced Document Level (per 2023 guidance). Biocompatibility per ISO 10993-5/10. Non-sterile.
Indications for Use
Indicated for patients requiring cold, heat, contrast, or compression therapy for post-surgical/acute injury edema, swelling, and pain reduction. Therm-X Home indicated for DVT prophylaxis via lower extremity compression. Intended for use by or on order of licensed healthcare professionals in clinics, rehab, athletic training, or home settings.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 27, 2023
Zenith Technical Innovations % Rita King Chief Executive Officer MethodSense. Inc. 1 Copley Parkway, Suite 130 Morrisville, North Carolina 27560
Re: K231912
Trade/Device Name: Therm-X Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP. ILO. JOW Dated: June 28, 2023 Received: June 29, 2023
Dear Rita King:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231912
Device Name Therm-X
#### Indications for Use (Describe)
Therm-X (Therm-X Home and Therm-X AT) combines cold, heat, contrast, and compression therapy. Therm-X is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated.
Therm-X Home systems also provide DVT therapy. Therm-X Home systems with DVT therapy are intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.
Therm-X (Therm-X Home and Therm-X AT) is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-family: Arial;"> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
### Zenith Technical Innovations K231912
This 510(k) Summary is in conformance with 21 CFR 807.92
| Submitter: | Zenith Technical Innovations, LLC. (Zenith)<br>1396 St. Paul Ave.<br>Gurnee, IL 60031<br>Phone: (847) 672-7481<br>Fax: (847) 672-8721 |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Rita King, CEO<br>MethodSense, Inc.<br>Email: ritaking@methodsense.com<br>Phone: 919-313-3961 |
| Company Contact: | Greg Binversie<br>Chief Technical Officer |
- Date Prepared: June 28, 2023
## Device Name and Classification
| Trade Name: | Therm-X |
|-----------------------|--------------------------------------------------|
| Common Name: | Massage, Powered Inflatable Tube |
| Classification: | Class II |
| Regulation Number: | 21 CFR 890.5650, Powered inflatable tube massage |
| Classification Panel: | Physical Medicine |
| Product Code: | IRP, ILO, JOW |
### Predicate Device
| Trade Name: | Therm-X |
|----------------------------|---------------------------------------------|
| Common Name: | Heat and/or Cold and Compression Therapy |
| 510(k) Submitter / Holder: | Zenith Technical Innovations, LLC. (Zenith) |
| 510(k) Number: | K193550 |
| Classification: | Class II |
| Requlation Number: | 890.5650, Powered Inflatable tube massager |
| Classification Panel: | Physical Medicine |
| Product Code: | IRP, ILO, JOW |
The predicate device has not been subject to a design-related recall.
{4}------------------------------------------------
### Device Description
Therm-X is an AC powered, software-controlled multimodality device, designed to be used in a clinical or home-use setting, and under the direction, prescription, or supervision of a licensed healthcare professional. The device is available in two configurations: Therm-X Home and Therm-X AT.
Therm-X (Therm-X Home and Therm-X AT) features iceless cold therapy, heat therapy, and contrast (alternating heat and cold) therapy. The Therm-X Home system also provides DVT prophylaxis therapy and continuous therapy for users who wish to receive treatment over an extended period of time.
Therm-X consists of various reusable inflatable wraps for thermal treatment of the back, elbow, hand/wrist, shoulder, ankle, hip, lower leg, or knee and DVT prophylactic treatment applied to the foot or calf. Multi-patient use garments are available for all anatomical areas that can be cleaned and disinfected in between uses to be reused for different patients. Single-patient use garments are available for thermal treatment of the back, hand/wrist, shoulder, neck, ankle, hip, and knee and can be disposed of after patient treatment is concluded. The thermal garments are flexible coolant circulating garments that apply to the body to deliver cold, heat, or contrast therapy in combination with or without pneumatic compression. The foot and calf DVT prophylactic garments apply pneumatic compression alone and are intended for use by the Therm-X Home system only.
Therm-X is controlled by an intuitive touch screen computer interface, allowing the user to manage the therapy modalities as well as easily adjust and monitor treatment times, temperature, and compression settings. Therm-X AT and Therm-X Home models provide an optional password protection feature that allows for a user to use only stored cycles, giving health care providers an ability to ensure compliance to a chosen cycle. The device also provides functionality to allow the health care provider to assign a date at which the user will be able to access a second stored cycle instead of the first. The Therm-X device also allows users to change treatment pressure and temperatures to less extreme therapies (e.g. changing a warm cycle to a lower warm temperature or changing a cold cycle to a higher cold temperature).
The Therm-X is approximately 15 lbs. when filled with coolant and has a handle placed on the top of the device. It has a centralized coolant reservoir accessible through a cap located at the back of the device that supplies its coolant and radiator systems. The reservoir, pumps, fans, circuit board, and other components of the Therm-X are located inside a covered enclosure made out of plastic and metal components, accessible only using a specialized tool.
### Indications for Use
Therm-X (Therm-X Home and Therm-X AT) combines cold, heat, contrast, and compression therapy. Therm-X is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surqical conditions for which localized thermal therapy (hot or cold) are indicated.
{5}------------------------------------------------
Therm-X Home systems also provide DVT therapy. Therm-X Home systems with DVT therapy are intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.
Therm-X (Therm-X Home and Therm-X AT) is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.
### Risk Analysis Method
The Therm-X was assessed to determine the risks to health associated with the device modifications and evaluate risks related to safety, effectiveness, and usability. A risk analysis was conducted in accordance with ISO 14971:2019, Medical devices -- Application of risk manaqement to medical devices. No additional risks were associated with the release of the four new thermal garments for this submission.
### Substantial Equivalence
Therm-X is substantially equivalent to Therm-X (K193550) by Zenith Technical Innovations, LLC. (Zenith).
Therm-X has the same intended use and indications for use as the predicate device and uses equivalent overall design and operating principals as the predicate device.
The table below provides a detailed comparison of Therm-X to the predicate device:
{6}------------------------------------------------
# Detailed Comparison of the Subject and Predicate Device
| Characteristic | Therm-X | Therm-X (K193550) | Comparison |
|----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Subject Device<br>Therm-X (Therm-X Home and Therm-X AT) combines cold, heat, contrast, and compression therapy. Therm-X is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated.<br><br>Therm-X Home systems also provide DVT therapy. Therm-X Home systems with DVT therapy are intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.<br><br>Therm-X (Therm-X Home and Therm-X AT) is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings. | Primary Predicate<br>Therm-X (Therm-X Home and Therm-X AT) combines cold, heat, contrast, and compression therapy. Therm-X is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated.<br><br>Therm-X Home systems also provide DVT therapy. Therm-X Home systems with DVT therapy are intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.<br><br>Therm-X (Therm-X Home and Therm-X AT) is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings. | The indications for use of the Therm-X (subject device) are identical to the Therm-X (K193550). |
| Characteristic | Therm-X | Therm-X (K193550) | Comparison |
| | | | |
| | Subject Device | Primary Predicate | |
| Intended Users | Health Care Professionals and lay<br>users (under prescription) | Health Care Professionals and lay<br>users (under prescription) | Therm-X (subject device) is identical<br>to Therm-X (K193550). |
| Number of Patients<br>that can be treated<br>at one time | Two | One | Therm-X (subject device) is different<br>from Therm-X (K193550). |
| | | | This difference is limited to the ability<br>of the Therm-X (subject device) to<br>treat two patients instead of one via a<br>split umbilical hose accessory.<br>Verification and Validation testing<br>have been conducted to confirm this<br>difference does not affect the intended <br>use or raise different questions of<br>safety and effectiveness. |
| Two Programmable<br>Cycles | Configuration of two programmable<br>cycles are available for all Therm-X<br>Models | Configuration of two programmable<br>cycles are available for all Therm-X<br>Models | Therm-X (subject device) is identical<br>to Therm-X (K193550). |
| Continuous<br>Treatment Cycle | Available on Therm-X Home | Available on Therm-X Home | Therm-X (subject device) is identical<br>to Therm-X (K193550). |
| Heat Therapy | Default: 105°F, 107°F, 110°F | Default: 105°F, 107°F, 110°F | Therm-X (subject device) is identical<br>to Therm-X (K193550). |
| | Custom: 105°F - 110°F | Custom: 105°F — 110°F | |
| | Default, continuous: 105°F, 107°F | Default, continuous: 105°F, 107°F | |
| | Custom, continuous: 105°F - 107°F | Custom, continuous: 105°F - 107°F | |
| Characteristic | Therm-X | Therm-X (K193550) | Comparison |
| | | | |
| | Subject Device | Primary Predicate | |
| Cold Therapy | Default: 34°F, 45°F, 55°F | Default: 34°F, 45°F, 55°F | Therm-X (subject device) is identical<br>to Therm-X (K193550). |
| | Custom: 34°F - 55°F | Custom: 34°F - 55°F | |
| | Default, continuous: 40°F, 45°F, 50°F | Default, continuous: 40°F, 45°F, 50°F | |
| | Custom, continuous: 40°F – 50°F | Custom, continuous: 40°F – 50°F | |
| Edema Pressure<br>Levels | Available in four levels:<br>Lite (5 mm Hg)<br>Low (20 mm Hg)<br>Medium (45 mm Hg)<br>High (70 mm Hg)<br>For continuous treatment, available in<br>three levels:<br>Lite (5 mm Hg)<br>Low (20 mm Hg)<br>Medium (45 mm Hg) | Available in four levels:<br>Lite (5 mm Hg)<br>Low (20 mm Hg)<br>Medium (45 mm Hg)<br>High (70 mm Hg)<br>For continuous treatment, available in<br>three levels:<br>Lite (5 mm Hg)<br>Low (20 mm Hg)<br>Medium (45 mm Hg)<br>High (70 mm Hg) | Therm-X (subject device) is different<br>to Therm-X (K193550).<br>The difference is limited to the Therm-<br>X (subject device) not providing the<br>option for High pressure during<br>Continuous treatment and allowing for<br>Lite treatment during continuous<br>treatment.<br>This difference in options of<br>compression levels for continuous<br>treatments does not affect the<br>intended use or raise different<br>questions of safety and effectiveness. |
| Static or<br>Intermittent<br>Pressure | Both | Both | Therm-X (subject device) is identical<br>to Therm-X (K193550). |
| DVT Only | Available for Therm-X Home Model | Available for Therm-X Home Model | Therm-X (subject device) is identical<br>to Therm-X (K193550). |
| Characteristic | Therm-X | Therm-X (K193550) | Comparison |
| | | | |
| | Subject Device | Primary Predicate…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.