GAMMEX PI Hybrid Micro (340002055); GAMMEX PI Hybrid Micro (340002060); GAMMEX PI Hybrid Micro (340002065); GAMMEX PI Hybrid Micro (340002070); GAMMEX PI Hybrid Micro (340002075); GAMMEX PI Hybrid Micro (340002080); GAMMEX PI Hybrid Micro (340002085); GAMMEX PI Hybrid Micro (340002090)
K231902 · Ansell Healthcare · KGO · Oct 20, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231902
Device Name
GAMMEX PI Hybrid Micro (340002055); GAMMEX PI Hybrid Micro (340002060); GAMMEX PI Hybrid Micro (340002065); GAMMEX PI Hybrid Micro (340002070); GAMMEX PI Hybrid Micro (340002075); GAMMEX PI Hybrid Micro (340002080); GAMMEX PI Hybrid Micro (340002085); GAMMEX PI Hybrid Micro (340002090)
Applicant
Ansell Healthcare
Product Code
KGO · General, Plastic Surgery
Decision Date
Oct 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
Gammex PI Hybrid Micro Surgical Gloves are sterile, non-pyrogenic, are intended to be worn by operating room personnel to protect a surgical wound from contamination and have a shelf life of 3 years. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Sterile, disposable, hand-specific surgical gloves; synthetic polyisoprene and neoprene blend; brown color; beaded cuff. Used by operating room personnel to protect surgical wounds from contamination. Tested for permeation resistance against various chemotherapy drugs per ASTM D6978. Provides barrier protection; does not involve software or algorithms.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D3577 (dimensions, physical properties, freedom from holes, powder residuals), ASTM D6978 (chemotherapy permeation), and ISO 10993 (biocompatibility: irritation, sensitization, systemic toxicity, cytotoxicity). USP <151> testing confirmed non-pyrogenic status.
Technological Characteristics
Synthetic polyisoprene and neoprene blend; polymer coating; sterile (gamma irradiation, 10^-6 SAL); powder-free; hand-specific; beaded cuff. Meets ASTM D3577-19, ASTM D6978-05(2019), and ISO 10993 standards. Brown color.
Indications for Use
Indicated for operating room personnel to protect surgical wounds from contamination. Tested for use with specific chemotherapy drugs per ASTM D6978. Warning: Do not use with Carmustine or ThioTEPA.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Gammex PI Hybrid Surgical Gloves Tested for Use with Chemotherapy Drugs (K182948)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 20, 2023
Ansell Healthcare Carson Delaloye Senior Administrator Quality 2301 Robb Drive Reno, Nevada 89523
## Re: K231902
Trade/Device Name: GAMMEX PI Hybrid Micro (340002055); GAMMEX PI Hybrid Micro (340002060); GAMMEX PI Hybrid Micro (340002065); GAMMEX PI Hybrid Micro (340002070); GAMMEX PI Hybrid Micro (340002075); GAMMEX PI Hybrid Micro (340002080); GAMMEX PI Hybrid Micro (340002085); GAMMEX PI Hybrid Micro (340002090) Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO, OPJ Dated: September 21, 2023 Received: September 21, 2023
Dear Carson Delaloye:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Image /page/2/Picture/3 description: The image shows the text "Bifeng Qian -S" in a large, sans-serif font. The text is black and is set against a white background. The words are arranged horizontally, with "Bifeng" and "Qian" being the most prominent. The "-S" is smaller and slightly offset to the right.
BiFeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231902
#### Device Name
GAMMEX PI Hybrid Micro (340002055): GAMMEX PI Hybrid Micro (340002060): GAMMEX PI Hybrid Micro (340002065): GAMMEX PI Hybrid Micro (340002070): GAMMEX PI Hybrid Micro (340002075): GAMMEX PI Hybrid Micro (340002080): GAMMEX PI Hybrid Micro (340002085); GAMMEX PI Hybrid Micro (340002090)
#### Indications for Use (Describe)
Gammex PI Hybrid Micro Surgical Gloves are sterile, non-pyrogenic, are intended to be worn by operating room personnel to protect a surgical wound from contamination and have a shelf life of 3 years. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Tested chemotherapy drugs are as follows: Test chemotherapy drug and concentration Minimum breakthrough detection time in minutes Bleomycin (15.0 mg/ml) >240 Busulfan (6.0 mg/ml) >240 Carboplatin (10.0 mg/ml) >240 Carmustine (3.3 mg/ml) 24.1 Cisplatin (1.0 mg/ml) >240 Cyclophosphamide (20.0 mg/ml) >240 Cytarabine HCl (100.0 mg/ml) >240 Dacarbazine (10.0 mg/ml) >240 Daunorubicin HCl (5.0 mg/ml) >240 Docetaxel (10.0 mg/ml) >240 Doxorubicin HCl (2.0 mg/ml) >240 Epirubicin (2.0 mg/ml) >240 Etoposide (20.0 mg/ml) >240 Fludarabine (25.0 mg/ml) >240 Fluorouracil (50.0 mg/ml) >240 Gemcitabine (38.0 mg/ml) >240 Idarubicin (1.0 mg/ml) >240 Ifosfamide (50.0 mg/ml) >240 Irinotecan (20.0 mg/ml) >240 Mechlorethamine HCI (1.0 mg/ml) >240 Melphalan (5.0mg/ml) >240 Methotrexate (25.0 mg/ml) >240 Mitomycin C (0.5 mg/ml) >240 Mitoxantrone (2.0mg/ml) >240 Oxaliplatin (2.0 mg/ml) >240 Paclitaxel (6.0 mg/ml) >240 Rituximab (10.0 mg/ml) >240 ThioTEPA (10.0 mg/ml) 26.0 Vincristine Sulfate (1.0 mg/ml) >240 Warning: Do not use with Carmustine or ThioTEPA Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
### 510KSummary
### 510(k) Number:
K231902
### Submitter:
Ansell Healthcare 2301 Robb Drive Reno, NV 89523
### Contact Person(s):
Carson Delaloye Senior Coordinator, Regulatory Affairs Phone: 530-401-8977 Email: carson.delaloye@ansell.com
#### Date Prepared
October 16, 2023
#### Name of Device
| Trade Names: | GAMMEX PI Hybrid Micro (340002055); |
|----------------------------|-------------------------------------|
| | GAMMEX PI Hybrid Micro (340002060); |
| | GAMMEX PI Hybrid Micro (340002065); |
| | GAMMEX PI Hybrid Micro (340002070); |
| | GAMMEX PI Hybrid Micro (340002075); |
| | GAMMEX PI Hybrid Micro (340002080); |
| | GAMMEX PI Hybrid Micro (340002085); |
| | GAMMEX PI Hybrid Micro (340002090) |
| Common Name: | Surgeon's Gloves |
| Classification Name: | Non-Powdered Surgeon's Glove |
| Classification Regulation: | 21 CFR 878.4460 |
| Device Class: | I |
| Product Code: | KGO, OPJ |
| Classification Panel: | General and Plastic Surgery |
#### Legally Marketed Predicate Device
| K182948 | |
|----------------------|-------------------------------------------------------------------------|
| Trade Name: | Gammex PI Hybrid Surgical Gloves Tested for Use with Chemotherapy Drugs |
| Common Name: | Surgeon's Gloves |
| Classification Name: | Non-Powdered Surgeon's Glove |
{6}------------------------------------------------
| Classification Regulation: | 21 CFR 878.4460 |
|----------------------------|-----------------------------|
| Device Class: | I |
| Product Code: | KGO |
| Classification Panel: | General and Plastic Surgery |
### Device Description
Gammex PI Hybrid Micro Surgical Gloves are sterile and disposable devices. The gloves are made of a synthetic rubber polyisoprene and neoprene blend. Ansell has performed independent laboratory testing to support use with Chemotherapy Drugs and Non-Pyrogenic claims. The glove has a shelf life of 3 years, is brown in color and comes in the following sizes: 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0. Each size is assigned with a model number.
#### Indication for Use Statement
Gammex PI Hybrid Micro Surgical Gloves are sterile, non-pyrogenic, are intended to be worn by operating room personnel to protect a surgical wound from contamination and have a shelf life of 3 years. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Test chemotherapy drug and concentration | Minimum breakthrough<br>detection time in minutes |
|------------------------------------------|---------------------------------------------------|
| Bleomycin (15.0 mg/ml) | >240 |
| Busulfan (6.0 mg/ml) | >240 |
| Carboplatin (10.0 mg/ml) | >240 |
| Carmustine (3.3 mg/ml) | 24.1 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (20.0 mg/ml) | >240 |
| Cytarabine HCl (100.0 mg/ml) | >240 |
| Dacarbazine (10.0 mg/ml) | >240 |
| Daunorubicin HCl (5.0 mg/ml) | >240 |
| Docetaxel (10.0 mg/ml) | >240 |
| Doxorubicin HCl (2.0 mg/ml) | >240 |
| Epirubicin (2.0 mg/ml) | >240 |
| Etoposide (20.0 mg/ml) | >240 |
| Fludarabine (25.0 mg/ml) | >240 |
| Fluorouracil (50.0 mg/ml) | >240 |
| Gemcitabine (38.0 mg/ml) | >240 |
| Idarubicin (1.0 mg/ml) | >240 |
| Ifosfamide (50.0 mg/ml) | >240 |
| Irinotecan (20.0 mg/ml) | >240 |
| Mechlorethamine HCI (1.0 mg/ml) | >240 |
| Melphalan (5.0mg/ml) | >240 |
| Methotrexate (25.0 mg/ml) | >240 |
Tested chemotherapy drugs are as follows:
{7}------------------------------------------------
| Mitomycin C (0.5 mg/ml) | >240 |
|---------------------------------|------|
| Mitoxantrone (2.0mg/ml) | >240 |
| Oxaliplatin (2.0 mg/ml) | >240 |
| Paclitaxel (6.0 mg/ml) | >240 |
| Rituximab (10.0 mg/ml) | >240 |
| ThioTEPA (10.0 mg/ml) | 26.0 |
| Vincristine Sulfate (1.0 mg/ml) | >240 |
Warning: Do not use with Carmustine or ThioTEPA
### Technological Characteristics:
## Technological Characteristics Comparison Table for Predicate Device
| | Predicate Device | Proposed Subject Device | Comparison |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | Gammex PI Hybrid Surgical Gloves Tested for<br>Use with Chemotherapy Drugs | GAMMEX PI Hybrid Micro (340002055);<br>GAMMEX PI Hybrid Micro (340002060);<br>GAMMEX PI Hybrid Micro (340002065);<br>GAMMEX PI Hybrid Micro (340002070);<br>GAMMEX PI Hybrid Micro (340002075);<br>GAMMEX PI Hybrid Micro (340002080);<br>GAMMEX PI Hybrid Micro (340002085);<br>GAMMEX PI Hybrid Micro (340002090) | Different -<br>Trade Name<br>o The names are<br>different to indicate<br>which of the devices is<br>the smaller version. This<br>does not affect safety or<br>efficacy as the difference<br>is to inform the end user. |
| 510k Number | K182948 | K231902 | Different -New 510(k) |
| Product<br>Owner | Ansell Healthcare | Ansell Healthcare | Same |
| Product Code | KGO | KGO, OPJ | Different -OPJ product code<br>added for chemo |
| Regulation<br>Number | 21 CFR 878.4460 | 21 CFR 878.4460 | Same |
| Regulatory<br>Class | I | I | Same |
| Regulation<br>Name | Non-powdered Surgeon's Gloves | Non-powdered Surgeon's glove | Same |
| Indications for<br>Use | Gammex PI Hybrid Surgical Gloves are<br>intended to be worn by operating room<br>personnel to protect a surgical wound from<br>contamination. These gloves were tested for<br>use with Chemotherapy Drugs as per ASTM<br>D6978-05 Standard Practice for Assessment of<br>Medical Gloves to Permeation by<br>Chemotherapy Drugs.<br>Tested chemotherapy drugs are as follows:<br>Test Chemotherapy drug & Concentration<br>Minimum Breakthrough Detection Time in<br>Minutes<br>Bleomycin sulfate (15.0 mg/ml) >240 | Gammex PI Hybrid Micro Surgical Gloves are sterile,<br>non-pyrogenic, are intended to be worn by<br>operating room personnel to protect a surgical<br>wound from contamination and have a shelf life of 3<br>years. These gloves were tested for use with<br>Chemotherapy Drugs as per ASTM D6978 Standard<br>Practice for Assessment of Medical Gloves to<br>Permeation by Chemotherapy Drugs.<br>Tested chemotherapy drugs are as follows:<br>Test Chemotherapy drug & concentration Minimum<br>Breakthrough detection time in minutes<br>Bleomycin sulfate (15.0 mg/ml) >240<br>Busulfan (6.0 mg/ml) >240 | Different<br>Itemized List of<br>Differences Between<br>Predicate and Proposed:<br>• Indications for Use<br>o The indications for use<br>differ as the proposed<br>device carries non-<br>pyrogenic claims. The<br>permeation times also<br>differ for Carmustine and<br>ThioTEPA. |
{8}------------------------------------------------
| | Busulfan (6.0 mg/ml) >240 | Carboplatin (10.0 mg/ml) >240 | Carboplatin (10.0 mg/ml) >240 | Carmustine (3.3 mg/ml) 24.1 | Same | |
|------------------------------------------|-------------------------------------------------------------------------------------------|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|------------------------------|--|
| | Carboplatin (10.0 mg/ml) >240 | Carmustine (3.3 mg/ml) 23.6 | Carmustine (3.3 mg/ml) 24.1 | Cisplatin (1.0 mg/ml) >240 | | |
| | Carmustine (3.3 mg/ml) 23.6 | Cisplatin (1.0 mg/ml) >240 | Cisplatin (1.0 mg/ml) >240 | Cyclophosphamide (20.0 mg/ml) >240 | | |
| | Cisplatin (1.0 mg/ml) >240 | Cyclophosphamide (20.0 mg/ml) >240 | Cyclophosphamide (20.0 mg/ml) >240 | Cytarbine (100.0 mg/ml) >240 | | |
| | Cyclophosphamide (20.0 mg/ml) >240 | Cytarabine (100.0 mg/ml) >240 | Cytarbine (100.0 mg/ml) >240 | Dacarbazine (10 mg/ml) >240 | | |
| | Cytarabine (100.0 mg/ml) >240 | Dacarbazine (10 mg/ml) >240 | Dacarbazine (10 mg/ml) >240 | Daunorubicin (5.0 mg/ml) >240 | | |
| | Dacarbazine (10 mg/ml) >240 | Daunorubicin (5.0 mg/ml) >240 | Daunorubicin (5.0 mg/ml) >240 | Docetaxel (10.0 mg/ml) >240 | | |
| | Daunorubicin (5.0 mg/ml) >240 | Docetaxel (10.0 mg/ml) >240 | Docetaxel (10.0 mg/ml) >240 | Doxorubicin Hydrochloride (2.0 mg/ml) >240 | | |
| | Docetaxel (10.0 mg/ml) >240 | Doxorubicin Hydrochloride (2.0 mg/ml) >240 | Doxorubicin Hydrochloride (2.0 mg/ml) >240 | Epirubicin (2.0 mg/ml) >240 | | |
| | Doxorubicin Hydrochloride (2.0 mg/ml) >240 | Epirubicin (2.0 mg/ml) >240 | Epirubicin (2.0 mg/ml) >240 | Etoposide (20.0 mg/ml) >240 | | |
| | Epirubicin (2.0 mg/ml) >240 | Etoposide (20.0 mg/ml) >240 | Etoposide (20.0 mg/ml) >240 | Fludarabine (25.0 mg/ml) >240 | | |
| | Etoposide (20.0 mg/ml) >240 | Fludarabine (25.0 mg/ml) >240 | Fludarabine (25.0 mg/ml) >240 | Fluorouracil (50.0 mg/ml) >240 | | |
| | Fludarabine (25.0 mg/ml) >240 | Fluorouracil(50.0 mg/ml) >240 | Fluorouracil (50.0 mg/ml) >240 | Gemcitabine (Gemzar) (38.0 mg/ml) >240 | | |
| | Fluorouracil(50.0 mg/ml) >240 | Gemcitabine (Gemzar) (38.0 mg/ml) >240 | Gemcitabine (Gemzar) (38.0 mg/ml) >240 | Idarubicin (1.0 mg/ml) >240 | | |
| | Gemcitabine (Gemzar) (38.0 mg/ml) >240 | Idarubicin (1.0 mg/ml) >240 | Idarubicin (1.0 mg/ml) >240 | Ifosfamide (50.0 mg/ml) >240 | | |
| | Idarubicin (1.0 mg/ml) >240 | Ifosfamide (50.0 mg/ml) >240 | Ifosfamide (50.0 mg/ml) >240 | Irinotecan (20.0 mg/ml) >240 | | |
| | Ifosfamide (50.0 mg/ml) >240 | Irinotecan (20.0 mg/ml) >240 | Irinotecan (20.0 mg/ml) >240 | Mechlorethamine HCI (1.0 mg/ml) >240 | | |
| | Irinotecan (20.0 mg/ml) >240 | Mechlorethamine HCl(1.0 mg/ml) >240 | Mechlorethamine HCI (1.0 mg/ml) >240 | Melphalan (5.0 mg/ml) >240 | | |
| | Mechlorethamine HCl(1.0 mg/ml) >240 | Melphalan (5.0 mg/ml) >240 | Melphalan (5.0 mg/ml) >240 | Methotrexate (25.0 mg/ml >240 | | |
| | Melphalan (5.0 mg/ml) >240 | Methotrexate (25.0 mg/ml) >240 | Methotrexate (25.0 mg/ml) >240 | Mitomycin C (0.5 mg/ml) >240 | | |
| | Methotrexate (25.0 mg/ml) >240 | Mitomycin C (0.5 mg/ml) >240 | Mitomycin C (0.5 mg/ml) >240 | Mitoxantrone (2.0 mg/ml) >240 | | |
| | Mitomycin C (0.5 mg/ml) >240 | Mitoxantrone (2.0 mg/ml) >240 | Mitoxantrone (2.0 mg/ml) >240 | Oxaliplatin (2.0 mg/ml) >240 | | |
| | Mitoxantrone (2.0 mg/ml) >240 | Oxaliplatin (2.0 mg/ml) >240 | Oxaliplatin (2.0 mg/ml) >240 | Paclitaxel (Taxol) (6.0 mg/ml) >240 | | |
| | Oxaliplatin (2.0 mg/ml) >240 | Paclitaxel (Taxol) (6.0 mg/ml) >240 | Paclitaxel (Taxol) (6.0 mg/ml) >240 | Rituximab (10.0 mg/ml) >240 | | |
| | Paclitaxel (Taxol) (6.0 mg/ml) >240 | Rituximab (10.0 mg/ml) >240 | Rituximab (10.0 mg/ml) >240 | ThioTEPA (10.0 mg/ml) 26.0 | | |
| | Rituximab (10.0 mg/ml) >240 | ThioTEPA (10.0 mg/ml) 36.1 | ThioTEPA (10.0 mg/ml) 26.0 | Vincristine Sulfate (1.0 mg/ml) >240 | | |
| | ThioTEPA (10.0 mg/ml) 36.1 | Vincrinstine Sulfate(1.0 mg/ml) >240 | Vincristine Sulfate (1.0 mg/ml) >240 | Warning: Do not use with Carmustine or ThioTEPA | | |
| | Vincrinstine Sulfate(1.0 mg/ml) >240 | | Warning: Do not use with Carmustine or ThioTEPA | | | |
| Material<br>Composition | Synthetic polyisoprene rubber | | Synthetic polyisoprene rubber | | Same | |
| Coating | Polymer Coating | | Polymer Coating | | Same | |
| Design | Single use | | Single use | | Same | |
| | Powder-free | | Powder-free | | Same | |
| | Hand Specific | | Hand Specific | | Same | |
| | Beaded cuff | | Beaded cuff | | Same | |
| | Natural | | Brown | | Different - Changed<br>Color | |
| Color | | | | | | |
| Labeling | Surgeon's Gloves | | Surgeon's Gloves | | Same | |
| Shelf Life | 3 Years | | 3 Years | | Same | |
| Performance<br>a. Dimensions | Meets ASTM D3577-19 (2015) requirements | | Meets ASTM D3577-19 requirements | | Same | |
| b. Physical<br>Properties | Meets ASTM D3577-19 (2015) requirements | | Meets ASTM D3577-19 requirements | | Same | |
| | | | Meets ASTM D3577-19 (2015) | | | |
| c. Freedom from | Meets ASTM D3577-19 (2015) | | | | Same | |
| holes | AQL Meets CFR 800.20 Requirements | | AQL Meets CFR 800.20 Requirements | | | |
| d. Powder | Meet applicable definition for powder free per | | Meet applicable definition for powder free per | | Same | |
| Residual | ASTM D3577-19; ≤ 2mg per glove | | ASTM D3577-19; ≤ 2mg per glove | | | |
| e. Sterility | Sterile | | Sterile | | Same | |
| Bio-<br>compatibility<br>Skin Irritation | Passes Primary Skin Irritation test per ISO<br>10993-10, Biological Evaluation of medical | | Passes Primary Skin Irritation test per ISO 10993-10,<br>Biological Evaluation of medical devices, Part 10:<br>Test for irritation and skin sensitization | | Same | |
{9}------------------------------------------------
| Bio-<br>compatibility<br>Sensitization | Passes Dermal Sensitization test per ISO 10993-10, Biological Evaluation of medical devices, Part 10: Test for irritation and skin sensitization | Passes Dermal Sensitization test per ISO 10993-10, Biological Evaluation of medical devices, Part 10: Test for irritation and skin sensitization | Same |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Bio-<br>compatibility<br>Acute Systemic<br>Toxicity | Passes Acute Systemic Toxicity Test per ISO 10993-11, Biological Evaluation of medical devices, Part 11: Test for systemic toxicity | Passes Acute Systemic Toxicity Test per ISO 10993-11, Biological Evaluation of medical devices, Part 11: Test for systemic toxicity | Same |
| Bio-<br>compatibility<br>Cytotoxicity | Under the conditions of the study, the device extract was found to be cytotoxic and therefore the device extracts were evaluated by ISO 10993-11 - Test for systemic toxicity. From the Acute Systemic Toxicity device extracts, the device extracts did not elicit a systemic response in the animal model | Under the conditions of the study, the device extract was found to be cytotoxic and therefore the device extracts were evaluated by ISO 10993-11 - Test for systemic toxicity. From the Acute Systemic Toxicity device extracts, the device extracts did not elicit a systemic response in the animal model | Same |
| Non-pyrogenic | Not reported | Passes Material Mediated Pyrogenicity Test per USP <151>. Under the conditions of the study, non-pyrogenic.<br>Less than 20.0 EU/device. Under the conditions of the study, low endotoxin levels. | Different |
| Sterilization<br>Method | Gamma Irradiation | Gamma Irradiation | Same |
| Sterilization<br>Information | Meets ANSI/AAMI/ISO 11137-1:2006 requirement of 10-6 SAL. | Meets ANSI/AAMI/ISO 11137-1:2006 requirement of 10-6 SAL. | Same |
| Prescription or<br>Over The<br>Counter | Over the Counter | Over the Counter | Same |
# ltemized List of Differences Between Predicate and Proposed:
- Trade Name
The names are different to indicate which of the devices is the smaller version. This does not affect safety or efficacy as the difference is to inform the end user.
- . Indications For Use The indications for use differ as the proposed device carries non-pyrogenic claims. The permeation times also differ for Carmustine and ThioTEPA.
- . Color
The color is listed as different for the devices. Brown pigmentation was added to the Proposed Subject device.
- Non-pyrogenic
The proposed device has demonstrated low endotoxin levels and did not elicit a pyrogenic response in testing. Testing was not reported for the predicate device.
# Non-Clinical Testing
Gammex® Pl Plus Glove-in-Glove™ System Tested for Use with Chemotherapy Drugs has the following technological characteristics as compared to ASTM or equivalent standards:
{10}------------------------------------------------
| Results | Title of Test | Purpose of Test | Acceptance Criteria |
|---------|------------------------------------|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PASS | ASTM<br>D3767-03 | Dimensions | Acceptance criteria in accordance with ASTM D3577-19: Standard<br>Specification for Rubber Surgical Gloves |
| PASS | ASTM<br>D3577-19 | Physical Properties | Acceptance criteria for tensile strength, ultimate elongation and stress<br>at 500% elongation before and after accelerated aging for synthetic<br>surgical gloves per ASTM D3577-19: Standard Specification for Rubber<br>Surgical Gloves |
| PASS | ASTM<br>D5151-19 | Freedom from holes | Acceptance criteria in accordance with ASTM D3577-19: Standard<br>Specification for Rubber Surgical Gloves with AQL requirements of 1.5 |
| PASS | ASTM D6124 | Powder-Free | Meets applicable acceptance criteria for powder free ≤ 2mg per glove<br>per ASTM D3577-19: Standard Specification for Rubber Surgical Gloves |
| PASS | ANSI/AAMI/ ISO<br>11137-<br>1:2006 | Sterility | Meets acceptance criteria requirement of 10-6 SAL per ISO 11137-1:<br>Sterilization for Health Care Products - Radiation - Part 1: Requirements<br>for Development, Validation and Routine Control of A Sterilization<br>Process For Medical Devices |
| PASS | ASTM<br>D6978- 05(2019) | Chemotherapy Drug<br>Permeation Test | Acceptance Criteria in accordance with ASTM D6978-05(2019):<br>Standard Practice for Assessment of Resistance of Medical Gloves to<br>Permeation by Chemotherapy Drugs.<br>*Please note that the following drugs have extremely low permeation<br>times: Carmustine (24.1 min) and Thiotepa (26.0 min) |
| FAIL | ISO 10993-5 | ISO in vitro Cytotoxicity<br>Study | Surfaces at undiluted were found to be cytotoxic (grade 4) to L-929<br>cells and non-cytotoxic with grade 0 at 1:2, 1:4, 1:8, 1:16, 1:32 and 1;64<br>dilutions. Acceptance criteria in accordance with ISO 10993-5:<br>Biological Evaluation of Medical Devices, Part 5: Tests for In Vitro<br>Cytotoxicity, 2009. |
| PASS | ISO 10993-<br>10:2010 | ISO Skin<br>Irritation Study | Passes Primary Skin Irritation test per ISO 10993-10, Biological<br>Evaluation Of Medical Devices, Part 10: Test for Irritation and Skin<br>Sensitization. Under the conditions of the study, not an irritant. |
| PASS | ISO 10993-<br>10:2010 | ISO<br>Maximization<br>Sensitization Study | Passes Dermal Sensitization test per ISO 10993-10, Biological<br>Evaluation of medical devices, Part 10: Test for Irritation and Skin<br>Sensitization. Under the conditions of the study, not a sensitizer |
| PASS | ISO 10993-<br>11:2017 | ISO Acute Systemic<br>Toxicity Study - ISO<br>10993-11:2017 | Passes Acute Systemic Toxicity Test per ISO 10993-11, Biological<br>Evaluation of medical devices, Part 11: Test for Systemic Toxicity.<br>Under the conditions of the study, there was no mortality or evidence<br>of acute systemic toxicity |
| PASS | USP <151> | Material Mediated<br>Pyrogenicity Study -<br>USP <151> | Passes Material Mediated Pyrogenicity Test per USP <151>. Under the<br>conditions of the study, non-pyrogenic. |
| PASS | Endotoxin Study | Demonstrate low<br>endotoxin levels | Less than 20.0 EU/device |
### Clinical Studies
A clinical study was not conducted on the subject device.
#### Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device, GAMMEX Pl Hybrid Micro, is as safe, as effective, and performs as well as or better than the legally marketed device K182948.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.