K231875 · Advanced Tactile Imaging, Inc. · HIR · Feb 22, 2024 · Obstetrics/Gynecology
Device Facts
Record ID
K231875
Device Name
Vaginal Tactile Ultrasound Imager
Applicant
Advanced Tactile Imaging, Inc.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Feb 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1425
Device Class
Class 2
Indications for Use
The Vaginal Tactile Ultrasound Imager obtains high-resolution mapping of pressures, ultrasound images and assesses the strength of pelvic floor muscles within the vagina. It is used in a medical setting to acquire the pressures, ultrasound images and store the corresponding data. It also provides visualization, analysis tools and information. The real time data as well as the analysis information can then be viewed with an intention of assisting in the diagnosis and evaluation. The device is intended for use by physicians, surgeons and medically trained personnel.
Device Story
The Vaginal Tactile Ultrasound Imager (TIUSv) is a cart-based diagnostic system used by physicians and surgeons to assess pelvic floor muscle function. The device utilizes a handheld probe equipped with 96 capacitive pressure sensors, 192 ultrasound transducers, an orientation accelerometer, and temperature sensors with micro-heaters. During a transvaginal procedure, the probe acquires dynamic ultrasound images and pressure responses from vaginal walls under load. The system integrates pressure and positioning data to generate real-time pressure maps (tactile images) and B-mode ultrasound visualizations. Software provides measurement, imaging, and reporting tools, including playback of stored sessions and calculation of pressure gradients. The device assists clinicians in the quantitative biomechanical assessment of the vagina and pelvic floor conditions, aiding in diagnosis and evaluation. The probe surfaces are preheated to body temperature for patient comfort.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation included biocompatibility (ISO 10993-1), cleaning/disinfection validation, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and ultrasound diagnostic equipment safety (IEC 60601-2-37). System-level performance testing confirmed pressure accuracy (±0.7 kPa), temperature accuracy (±0.5 °C), and ultrasound imaging resolution (2.0 mm lateral, 1.0 mm axial).
Technological Characteristics
Cart-based system with handheld probe. Probe materials: plastic, silicone. Sensors: 96 capacitive pressure sensors, 192 ultrasound transducers (4-7 MHz), accelerometer, temperature sensors. Power: 110-240V AC. Connectivity: ArtUs ultrasound electronics interface. Software: Moderate level of concern, IEC 62304 compliant. Sterilization: Non-sterile, high-level disinfection required for probe. Dimensions: 51cm x 51cm x 140cm. Weight: 37 kg.
Indications for Use
Indicated for physicians, surgeons, and medically trained personnel to obtain high-resolution pressure mapping, ultrasound imaging, and assessment of pelvic floor muscle strength within the vagina in a medical setting.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
July 9, 2024
Advanced Tactile Imaging Inc. Brendan Francy Biomedical Engineer 1457 Lower Ferry Rd Ewing, New Jersey 08618
Re: K231875
Trade/Device Name: Vaginal Tactile Ultrasound Imager Regulation Number: 21 CFR 884.1425 Regulation Name: Perineometer Regulatory Class: Class II Product Code: HIR
Dear Brendan Francy:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change for your device cleared on February 22, 2024. Specifically, FDA is updating this SE Letter because FDA inadvertently indicated that the SE determination also included review and clearance of a predetermined change control plan (PCCP). However, your 510(k) submission did not include a PCCP, so FDA is providing this administrative correction. Please see the attached revised clearance letter.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jason Roberts, OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices, 240-402-6400, jason.roberts(@fda.hhs.gov.
Sincerely,
**Jason Roberts -S**
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
July 9, 2024
Advanced Tactile Imaging Inc. Vladimir Egorov CEO 1457 Lower Ferry Rd Ewing, New Jersey 08618
Re: K231875 Trade/Device Name: Vaginal Tactile Ultrasound Imager Regulation Number: 21 CFR 884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: HIR Received: June 26, 2023
Dear Vladimir Egorov:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change for your device cleared on February 22, 2024.
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231875
#### Device Name
Vaginal Tactile Ultrasound Imager (TIUSv)
#### Indications for Use (Describe)
The Vaginal Tactile Ultrasound Imager obtains high-resolution mapping of pressures, ultrasound images and assesses the strength of pelvic floor muscles within the vagina. It is used in a medical setting to acquire the pressures, ultrasound images and store the corresponding data. It also provides visualization, analysis tools and information. The real time data as well as the analysis information can then be viewed with an intention of assisting in the diagnosis and evaluation. The device is intended for use by physicians, surgeons and medically trained personnel.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 1. IDENTIFICATION
| Submission Date: | February 22, 2024 |
|--------------------------------|--------------------------------------------------------------------------------|
| Applicant's/Owner's Name: | Advanced Tactile Imaging, Inc.<br>1457 Lower Ferry Rd<br>Trenton, NJ 08618 USA |
| Contact Person: | Vladimir Egorov |
| Phone number: | 609 883-0100 |
| Fax number: | n/a |
| Email: | egorov@tactile-imaging.com |
| Device Common Name: | Pressure Mapping Perineometer |
| Device Trade/Proprietary Name: | Vaginal Tactile Ultrasound Imager (TIUSv) |
| Device Class: | Class II |
| Classification Name: | Perineometer/Ultrasonic pulsed echo imaging system |
| Product Code: | HIR/IYO |
| Regulation Number: | 884.1425/892.1560 |
## 2. LEGALLY MARKETED DEVICES FOR SUBSTANTIAL EQUIVALENCE
1. K142355. Vaginal Tactile Imager (Advanced Tactile Imaging, Inc.),
# 3. DESCRIPTION
The Vaginal Tactile Ultrasound Imager (TIUSv) acquires ultrasound images and tactile images (pressure maps) from within the vagina and assesses the strength of pelvic floor muscles within the vagina. The device provides data on the pressures applied to vaginal walls along with the probe location to visualize pelvic floor support structures, and to record pelvic floor muscle contraction. The TIUSv software provides measurement, imaging, and reporting tools. The acquired data and the analysis results provided by TIUSv software can be used by a physician for quantitative biomechanical assessment of the vagina and pelvic floor conditions.
The TIUSv probe is equipped with 96 pressure (tactile) sensors, 192 ultrasound transducers, an orientation sensor (accelerometer), temperature sensors, and micro-heaters. During the clinical procedure, the probe is used to acquire dynamic ultrasound images and pressure responses from the vaginal walls under applied loads. The TIUSv examination procedure includes data collection from all segments of the vagina. Real-time data are sampled from the probe sensors via the interface electronics. The TIUSv software displays the acquired data in real time. The resulting pressure maps (tactile images) of the vagina integrate all the acquired pressure and positioning data for each of the pressure sensing elements. The TIUSv records the dynamic contraction for pelvic floor muscle(s). The probe surfaces that contact the vaginal walls are preheated to human body temperature.
{6}------------------------------------------------
The TIUSv supports physician data analysis by means of a playback function, which replays a stored session using previously recorded data instead of the real time data. The TIUSv also provides data and graphs such as pressure applied to the probe and location, resting pressures, muscle contraction pressures and calculated pressure gradients.
### 4. INTENDED USE
The Vaginal Tactile Ultrasound Imager obtains high-resolution mapping of pressures, ultrasound images and assesses the strength of pelvic floor muscles within the vagina. It is used in a medical setting to acquire the pressures, ultrasound images and store the corresponding data. It also provides visualization, analysis tools and information. The real time data as well as the analysis information can then be viewed with an intention of assisting in the diagnosis and evaluation. The device is intended for use by physicians, surgeons and medically trained personnel.
| Device & Predicate Device(s): | K231875 | K142355 | Comment |
|----------------------------------|----------------------------------------------------|-----------------------------------------------|-----------|
| Device Name | Vaginal Tactile<br>Ultrasound Imager | Vaginal Tactile<br>Imager | - |
| Primary Product Code | HIR | HIR | Same |
| Duration of Procedure | Up to 5 minutes | 1 - 2 minutes | Similar |
| Method of Data Collection | Contact tissue surface<br>(vaginal mucosa) | Contact tissue<br>surface (vaginal<br>mucosa) | Same |
| Handheld Probe | Yes | Yes | Same |
| Probe Size | 17 - 31mm | 17 - 31mm | Same |
| Probe Length | 120mm | 120mm | Same |
| Probe Material | Plastic, Silicone | Plastic, Silicone | Same |
| Pressure Sensor Technology | Capacitive pressure<br>sensor | Capacitive pressure<br>sensor | Same |
| Number of Pressure Sensors | 96 | 96 | Same |
| Pressure Range | 30 kPa | 33 kPa | Similar |
| Pressure Accuracy | 0.5 kPa | 0.5 kPa | Same |
| Accelerometer | Yes | Yes | Same |
| Ultrasound Technology | Yes | No | Different |
| Number of Ultrasound Transducers | 192 | N/A | Different |
| Probe Sterilization | Non-sterile | Non-Sterile | Same |
| Method of Data Presentation | Pressure color<br>mapped and<br>ultrasound images. | Tactile Images | Similar |
# 5. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
{7}------------------------------------------------
| | 110 – 240V AC | 110 – 240V AC | |
|--------------|---------------------------|---------------------------|------|
| Power Source | 47 – 63 Hz<br>1.25 – 0.5A | 47 – 63 Hz<br>1.25 – 0.5A | Same |
The Vaginal Tactile Imager has very similar technological characteristics to the subject device, including, the size and material of the probe, the number of pressure sensors, use of an accelerometer, ability to preheat, and the medical diagnostic cart platform.
The Vaginal Tactile Imager (Predicate Device) contains very similar intended use as the subject device except it does not include the use of Ultrasound Imaging capabilities or technology. Both devices develop a pressure map of the vagina to measure the strength of female pelvic floor. Both devices have the same intended user and environment.
The subject device and the Acuson (Reference device) have similar transvaginal ultrasound transducers with B-mode imaging. Additionally, the subject device and the Acuson have very similar Ultrasound arrays with the same number of elements, imaging depth, and frequency bandwidth.
The ArtUs Ultrasound System is used in the subject device to provide B-mode ultrasound imaging. ArtUs ultrasound electronics allows for the reception of ultrasound signals from the probe's ultrasound array and transmission to the computer on the cart.
The differences between the subject device and the predicate device does not raise any different questions of safety or effectiveness.
## 6. OVERVIEW OF THE PERFORMANCE DATA
The components of the Vaginal Tactile Ultrasound Imager (TIUSv) were tested individually and together at the system level and the necessary test reports were generated. Other verification reports include the biocompatibility test reports, cleaning and disinfection test reports, applied IEC 60601-1-2 reports and bench verification/validation test reports.
Device Protection against Electrical Shock: Class IIa TIUSv probe: Type B Applied Part Device IP Code: IP20 TIUSv probe IP Code: IP47 Mode of Operation: Continuous Operation Sanitary Requirement: High-level disinfection of TIUSv probe Power Consumption: 110 W Input Voltage: 110 - 240 V AC, 47 - 63 Hz, 1.25 - 0.5 A 37 kg Weight:
{8}------------------------------------------------
| Size: | 51 cm (width) x 51 cm (depth) x 140 cm (height) | |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--|
| Operational Conditions: | Temperature Range: from 20 °C to 30 °C,<br>Relative Humidity Range: 0% - 90%, non-condensing | |
| Methods for device securing<br>against unexpected motion: | Four brakes on cart casters | |
| Measurement Values: | Pressure (kPa)<br>Orientation of the probe (degree),<br>Temperature of the micro-heaters (°C). | |
| Measurement Ranges: | Pressure: 0 – 30 kPa<br>Orientation: 0±360 degrees for rotation<br>0 ± 45 degrees for elevation<br>Temperature: from +20 °C to +40 °C | |
Table 1. Measurement Performance (accuracy) for Pressure
| Pressure range | Accuracy<br>(Calculated by standard deviation<br>for all pressure sensors) | Maximum deviation for<br>individual sensor |
|----------------|----------------------------------------------------------------------------|--------------------------------------------|
| 0 — 30 kPa | ± 0.7 kPa | ± 2.5 kPa |
| Measurement Performance:<br>(accuracy) | Temperature: ±0.5 °C<br>Orientation: ± 1.0 degrees |
|-----------------------------------------|------------------------------------------------------------|
| Number Ultrasound Transducers | 192 Elements |
| Ultrasound Frequency | 4-7 MHz |
| Ultrasound lateral resolution | 2.0 mm |
| Ultrasound lateral positioning accuracy | ±1.0 mm |
| Axial resolution | 1.0 mm |
| Ultrasound axial positioning accuracy | ±2.0 mm |
| Calculated Values: | Pressure gradient (kPa/mm)<br>Distance (mm), |
| Calculated Ranges: | Pressure gradient: 0 – 4 kPa/mm<br>Distance:<br>0 – 120 mm |
{9}------------------------------------------------
Intended Users:
Physicians Surgeons Medically trained personnel
## 7. PERFORMANCE TESTING
The performance testing completed on the Vaginal Tactile Ultrasound Imager included the following:
## Cleaning and High-Level Disinfection
- Cleaning and high-level disinfection validation of vaginal probe ●
## Biocompatibility testing
- The biocompatibility evaluation was conducted in accordance with the FDA Guidance for . Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
- Biocompatibility testing (cytotoxicity, sensitization, and irritation) per ISO 10993-1:2009 ●
## Electromagnetic Compatibility and Electrical Safety
- The subject device models were assessed for conformity with the relevant requirements of ● the following standards and found to comply:
- Electromagnetic compatibility testing per IEC 60601-1-2 : 2014/AMD1:2021 ●
- Basic safety and essential performance of medical electrical equipment per IEC 60601-1 ● 2005/AMD1:2012
- Basic safety and essential performance of ultrasonic medical diagnostic and monitoring ● equipment IEC 60601-2-37: 2007/AMD1:2015
## Software Verification and Validation Testing
Software verification and validation testing was conducted and completed, and software ● documentation was provided as recommended by FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005 for a moderate software level of concern.
- . Software verification and validation IEC/EN 62304:2006+A1:2015
The following quality assurance measures are applied to the development of the system:
- Risk analysis
- Component testing (Verification) ●
- System performance testing (Verification & Validation) ●
- Safety testing (Verification) ●
The nonclinical performance data described above demonstrate the Vaginal Tactile Ultrasound Imager is safe and effective.
{10}------------------------------------------------
## 8. CONCLUSION
The Vaginal Tactile Ultrasound Imager has the same intended use as the predicate. Performance testing has demonstrated that the subject device is as safe and effective as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.