← Product Code [OWN](/productcode/OWN) · K231854

# 1788 4K Camera System with Advanced Imaging Modality; L12 LED Light Source with AIM (K231854)

_Stryker · OWN · Sep 20, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K231854

## Device Facts

- **Applicant:** Stryker
- **Product Code:** [OWN](/productcode/OWN.md)
- **Decision Date:** Sep 20, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Pediatric

## Indications for Use

The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ arthroscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repar, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, orthopedic surgeons, ENT/heurosurgeons and urologists. Upon intravenous administration of SPY AGENT GREEN (indocyanine green for injection, USP), the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L12 Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month and older, and visualization of at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct) in adults and pediatric patients 12 to 17 years of age, using near-infrared imaging. Fluorescence imaging of biliary ducts with the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. Additionally, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging. Upon interstitial administration of SPY AGENT GREEN, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. Upon administration and use of pafolacianine consistent with its approved label, the L12 LED Light Source with Advanced Imaging Modality and SafeLight™ Cable are used to perform intraoperative fluorescence imaging of tissues that have taken up the drug. The L12 LED Light Source with Advanced Imaging Modality is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

## Device Story

Endoscopic 4K imaging system; utilizes white light, near-infrared (NIR) fluorescence, and cyan spectral imaging. Inputs: light projected via endoscope; CMOS sensor acquires image data. Processing: digital video stream generation; NIR fluorescence imaging via SPY AGENT GREEN or pafolacianine; NIR transillumination for ureters; Cyan Spectral Imaging for mucosal surface/capillary visualization. Output: real-time video displayed for surgeon. Used in OR by surgeons (general, pediatric, gynecological, cardiac, thoracic, plastic, orthopedic, ENT, neuro, urology). Assists in intraoperative visualization of vessels, blood flow, tissue perfusion, lymphatics, and biliary ducts. Benefits: enhanced anatomical visualization during minimally invasive procedures.

## Clinical Evidence

Bench testing only. No clinical data required. Performance testing included electromagnetic compatibility (IEC 60601-1-2), electrical safety (ANSI AAMI ES 60601-1), laser safety (IEC 60825-1), sterilization (ISO 17664, ISO 14937), and software validation (IEC 62304). Bench performance metrics included spatial uniformity, minimum detectable fluorescence, fluorescence detection depth, signal-to-noise ratio, and photobiological safety.

## Technological Characteristics

System includes 4K CMOS camera and L12 LED light source with infrared laser (780nm/830nm). Connectivity: digital video output. Sterilization: per ISO 17664/14937. Safety standards: IEC 60601-1, IEC 60601-2-18, IEC 60825-1, IEC 62471. Imaging modes: White light, NIR fluorescence, NIR transillumination, Cyan Spectral Imaging. Laser class 1M.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Advanced Imaging Modality (AIM) System ([K230605](/device/K230605.md), [K230754](/device/K230754.md))
- Olympus VISERA ELITE Video System Center and Xenon Light Source ([K111425](/device/K111425.md))
- PINPOINT Endoscopic Fluorescence Imaging System ([K182606](/device/K182606.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right, the FDA logo is displayed in blue, with the acronym "FDA" stacked above the words "U.S. FOOD & DRUG ADMINISTRATION."

September 20, 2023

Stryker April Malmborg Senior Director, Regulatory Affairs 5900 Optical Court San Jose, California 95138

Re: K231854

Trade/Device Name: 1788 4K Camera System with Advanced Imaging Modality; L12 LED Light Source with AIM Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, GWG, NWB, OWN, FCS, FCW Dated: June 23, 2023 Received: June 23, 2023

Dear April Malmborg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tanisha L. Tanisha L. Hithe -S 2023.09.20 Hithe -S 09:19:05 -04'00'

Tanisha Hithe, MS, MHS Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

### K231854

#### Device Name

1788 4K Camera System with Advanced Imaging Modality

#### Indications for Use (Describe)

The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ arthroscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients.

A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repar, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.

The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, orthopedic surgeons, ENT/heurosurgeons and urologists.

Type of Use (Select one or both, as applicable)

| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|------------------------------------------------------------|
| <span></span> | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### Indications for Use

510(k) Number (if known)

### K231854

#### Device Name

L12 LED Light Source with Advanced Imaging Modality

### Indications for Use (Describe)

Upon intravenous administration of SPY AGENT GREEN (indocyanine green for injection, USP), the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L12 Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month and older, and visualization of at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct) in adults and pediatric patients 12 to 17 years of age, using near-infrared imaging.

Fluorescence imaging of biliary ducts with the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.

Additionally, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.

Upon interstitial administration of SPY AGENT GREEN, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Upon administration and use of pafolacianine consistent with its approved label, the L12 LED Light Source with Advanced Imaging Modality and SafeLight™ Cable are used to perform intraoperative fluorescence imaging of tissues that have taken up the drug.

The L12 LED Light Source with Advanced Imaging Modality is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R Part 807.92(c).

### 510(k) Number :

### Submitter:

| Applicant:      | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138                                                    |
|-----------------|------------------------------------------------------------------------------------------------------------------|
| Contact Person: | April Malmborg<br>Senior Director, Regulatory Affairs<br>Email: april.malmborg@stryker.com<br>Ph: (408) 754-2472 |
| Date Prepared:  | June 23, 2023                                                                                                    |

### Subject Device:

The subject device is the Advanced Imaging Modality (AIM) System and specifically the following system components:

| Name of Device:      | 1788 4K Camera System with Advanced Imaging Modality                                            |
|----------------------|-------------------------------------------------------------------------------------------------|
| Common or Usual Name | 3-chip Video Camera                                                                             |
| Classification Name: | Endoscope and Accessories (21 C.F.R. §876.1500)<br>Neurological Endoscope (21 C.F.R. §882.1480) |
| Regulatory Class:    | II                                                                                              |
| Product Code:        | GCJ<br>GWG<br>NWB                                                                               |
| 510(k) Review Panel: | General & Plastic Surgery<br>Neurology<br>Gastroenterology/ Urology                             |

| Name of Device:      | L12 LED Light Source with AIM                            |
|----------------------|----------------------------------------------------------|
| Common or Usual Name | Light Source, Illuminator                                |
| Classification Name: | Endoscope and Accessories (21 C.F.R. §876.1500)          |
|                      | Fiberoptic light ureteral catheter (21 C.F.R. §876.4020) |
|                      | Neurological Endoscope (21 C.F.R. §882.1480)             |
| Regulatory Class:    | II                                                       |
| Product Code:        | OWN                                                      |
|                      | FCS                                                      |
|                      | FCW                                                      |
|                      | GWG                                                      |
|                      | NWB                                                      |
| 510(k) Review Panel: | General & Plastic Surgery                                |
|                      | Gastroenterology/ Urology                                |
|                      | Neurology                                                |

{5}------------------------------------------------

# Predicate Device(s):

| Advanced Imaging<br>Modality (AIM) System                          | 1788 4K Camera System<br>L12 LED Light Source | K230605 (primary)<br>K230754 (primary) |
|--------------------------------------------------------------------|-----------------------------------------------|----------------------------------------|
| Olympus VISERA ELITE Video System Center and<br>Xenon Light Source |                                               | K111425 (secondary)                    |
| PINPOINT Endoscopic Fluorescence Imaging System                    |                                               | K182606 (secondary)                    |

### Device Description:

Stryker's Advanced Imaging Modality (AIM) System is an endoscopic real-time 4K visible white light, near-infrared illumination and transillumination, and cyan spectral imaging system. Near-infrared illumination is used for fluorescence imaging using SPY AGENT™ GREEN (indocyanine green for injection, USP) or CYTALUX™ (pafalocianine) injection. Near-infrared illumination is also intended for use during transillumination of the ureters using the IRIS Ureteral Kit during minimally invasive and open surgical procedures. Cyan Spectral Imaging is intended as an alternative mode of visualization that uses narrow bands for illumination during endoscopic surgery to provide greater visualization of surface structures and fine capillary patterns of the mucosal membrane.

### Indications for Use:

### 1788 4K Camera System with Advanced Imaging Modality:

The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ sinuscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients.

A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection. Laparoscopically assisted hysterectomy. Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.

The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT/neurosurgeons and urologists.

{6}------------------------------------------------

### L12 LED Light Source with Advanced Imaging Modality:

Upon intravenous administration of SPY AGENT GREEN (indocyanine green for injection, USP), the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month and older, and visualization of at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct) in adults and pediatric patients 12 to 17 years of age, using near-infrared imaging.

Fluorescence imaging of biliary ducts with the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.

Additionally, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.

Upon interstitial administration of SPY AGENT GREEN, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Upon administration and use of pafolacianine consistent with its approved label, the L12 LED Light Source with Advanced Imaging Modality and SafeLight™ Cable are used to perform intraoperative fluorescence imaging of tissues that have taken up the drug.

The L12 LED Light Source with Advanced Imaging Modality is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

{7}------------------------------------------------

# Comparison of Technological Characteristics with the Predicate Device:

| Item                    |                                   | Subject Device                                                                                                                                                                                                                                                        | Predicate Devices                                                              |                                                           |                                                                                                      |
|-------------------------|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------|
|                         |                                   |                                                                                                                                                                                                                                                                       | Primary                                                                        | Secondary                                                 | Secondary                                                                                            |
|                         |                                   | Advanced Imaging<br>Modality (AIM) System                                                                                                                                                                                                                             | Primary<br>Advanced Imaging<br>Modality (AIM) System                           | VISERA ELITE                                              | PINPOINT System                                                                                      |
| Manufacturer            |                                   | Stryker                                                                                                                                                                                                                                                               | Same as subject device                                                         | Olympus Medical<br>System<br>Corporation<br>K111425       | Same as subject<br>device                                                                            |
| Submission Reference(s) |                                   | Current Submission                                                                                                                                                                                                                                                    | K230605; K230754                                                               | K111425                                                   | K182606                                                                                              |
| Intended Use            |                                   | Endoscopic illumination<br>and imaging during<br>endoscopic procedures                                                                                                                                                                                                | Same as subject device                                                         | Same as subject<br>device                                 | Same as subject<br>device                                                                            |
| Indications for Use     |                                   | NOTE 1                                                                                                                                                                                                                                                                | Same as subject device                                                         | NOTE 2                                                    | NOTE 3                                                                                               |
| System Components       |                                   | Camera System (subject<br>of submission)<br>Light Source (subject of<br>submission) and SafeLight<br>Cable<br>Endoscopes<br>IRIS Ureteral Kit<br>Imaging Agents                                                                                                       | Same as subject device                                                         | Video Center<br>Xenon Light<br>Source<br>Videoscope       | Video Processor/<br>Illuminator<br>Camera Head<br>Light Guide Cable<br>Laparoscopes<br>Imaging Agent |
| Imaging Agents          |                                   | SPY AGENT GREEN<br>Pafolacianine                                                                                                                                                                                                                                      | Same as subject device                                                         | Not applicable.                                           | SPY AGENT<br>GREEN                                                                                   |
| Principles of Operation |                                   | Via an endoscope, light is<br>projected by a light source<br>onto one or more image<br>sensor(s) which acquire a<br>continuous stream of<br>image data. The image<br>data is processed to<br>provide a video stream<br>that is then sent to a<br>display for viewing. | Same as subject device                                                         | Same as subject<br>device                                 | Same as subject<br>device                                                                            |
| Safety Standards        |                                   | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-1-2<br>IEC 60601-4-2<br>IEC 60825-1<br>IEC 62471                                                                                                                                                                           | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-1-2<br>IEC 60601-4-2<br>IEC 60825-1 | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-2<br>IEC 62471 | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-2<br>IEC 60825-1                                          |
| Imaging<br>Modes        | White Light                       | White Light                                                                                                                                                                                                                                                           | Same as subject device                                                         | Same as subject<br>device                                 | Same as subject<br>device                                                                            |
|                         | NIR<br>Fluorescence               | ENV<br>Contrast<br>Overlay<br>- without IRIS<br>- with IRIS<br>Color Segmented<br>Fluorescence<br>- without IRIS<br>- with IRIS                                                                                                                                       | ENV<br>Contrast<br>Overlay<br>- without IRIS<br>- with IRIS                    | Not applicable.                                           | Contrast<br>Overlay<br>Color Segmented<br>Fluorescence                                               |
|                         | NIR Trans-<br>illumination        | IRIS                                                                                                                                                                                                                                                                  | Same as subject device.                                                        | Not applicable.                                           | Not applicable.                                                                                      |
|                         | Cyan Spectral<br>Imaging<br>(CSI) | Cyan Spectral Imaging<br>(CSI)                                                                                                                                                                                                                                        | Not Applicable.                                                                | Narrow Band<br>Imaging                                    | Not applicable.                                                                                      |
| Camera<br>System        | Image Sensor                      | CMOS image sensor                                                                                                                                                                                                                                                     | Same as subject device.                                                        | CCD image<br>sensor                                       | Same as subject<br>device.                                                                           |
| Item                    |                                   | Subject Device                                                                                                                                                                                                                                                        | Predicate Devices                                                              |                                                           |                                                                                                      |
|                         |                                   | Advanced Imaging<br>Modality (AIM) System                                                                                                                                                                                                                             | Primary<br>Advanced Imaging<br>Modality (AIM) System                           | Secondary<br>VISERA ELITE                                 | Secondary<br>PINPOINT System                                                                         |
|                         | Image Processing/<br>Video Output | Digital                                                                                                                                                                                                                                                               | Same as subject device.                                                        | Same as subject device.                                   | Same as subject device.                                                                              |
|                         | Resolution                        | 4K (up to 3840 x 2160)<br>Up to 1920 x 1080                                                                                                                                                                                                                           | Same as subject device.                                                        | Up to 1920 x 1080                                         | Up to 1920 x 1080                                                                                    |
|                         | Frame Rate                        | 60 frames per second                                                                                                                                                                                                                                                  | Same as subject device.                                                        | Same as subject device.                                   | Same as subject device.                                                                              |
| Light<br>Source         | Light Source/<br>Laser            | RGB LEDs/<br>Infrared Laser                                                                                                                                                                                                                                           | Same as subject device.                                                        | Xenon                                                     | Same as subject device.                                                                              |
|                         | Excitation<br>Wavelength          | 780 nm (used for NIR<br>fluorescence)<br>830 nm (used for NIR<br>transillumination)                                                                                                                                                                                   | Same as subject device.                                                        | Not applicable.                                           | 806 nm (used for<br>NIR fluorescence)                                                                |
|                         | Laser Safety<br>Classification    | Class 1M                                                                                                                                                                                                                                                              | Same as subject device.                                                        | Not applicable.                                           | Class 3R                                                                                             |

{8}------------------------------------------------

NOTE 1: The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ arthroscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectorny, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectorny, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectorny, Pleurodesis, Internal mammary artery dissection for coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT/neurosurgeons and urologists.

Upon intravenous administration of SPY AGENT™ GREEN (indocyanine green for injection, USP), the L12 LED Light Source with AIM and SafeLight™ Cable are used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L12 LED Light Source with AIM and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month and older, and visualization of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct) in adults and pediatic patients 12 to 17 years of age, using nearinfrared imaging. Fluorescence imaging of biliary ducts with the L12 Light Source with AIM and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. Additionally, the L12 Light Source with AIM and SafeLight Cable enable surgeons to perform minimally invasive cranial neurosurgery in adults and pediative patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using. Upon interstitial administration of SPY AGENT GREEN, the L12 LED Light Source with AIM and SafeLight Cable is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. Upon administration and use of pafolacianine consistent with its approved label, the L12 LED Light Source and SafeLight™ Cable is used to perform intraoperative fluorescence imaging of tissues that have taken up the drug. The L12 Light Source with AIM is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

NOTE 2: The video system center has been designed to be used with Olympus cameras heads, endoscopes, light source and monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation. The light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

NOTE 3: Upon intravenous administration of SPY AGENT™ GREEN, the PINPOINT Endoscopic Fluorescence Imaging System is used with SPY AGENT GREEN to perform intraoperative fluorescence anglography, and it is also indicated for use in fluorescence imaging of biliary ducts, and when intraoperative cholangiography. The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near infrared imaging of billary ducts with the

{9}------------------------------------------------

PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for bilization. Upon interstitial administration of SPY AGENT GREEN, the PINPOINT System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

### Performance Data:

| Test                                    | Method                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Result                                       |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Electromagnetic<br>Compatibility        | In accordance with FDA-recognized voluntary consensus standard<br>IEC 60601-1-2:2014 (19-8)<br>In accordance with FDA-recognized voluntary consensus standard<br>IEC 60601-4-2:2016 (19-19)                                                                                                                                                                                                                                                                               | Pass<br>Pass                                 |
| Electrical Safety                       | In accordance with FDA-recognized voluntary consensus standard<br>ANSI AAMI ES 60601-1:2005 + A1:2012 + A2: 2021 (19-46)<br>In accordance with FDA-recognized voluntary consensus standard<br>IEC 60601-1-6:2010 + A1:2013 + A2:2020 (5-132)<br>In accordance with FDA-recognized voluntary consensus standard<br>IEC 60601-2-18:2009 (9-114)                                                                                                                             | Pass<br>Pass<br>Pass                         |
| Laser Safety                            | In accordance with FDA-recognized voluntary consensus standard<br>IEC 60825-1:2014 (12-273)<br>In accordance with FDA-recognized voluntary consensus standard<br>IEC 62471 (12-249); and,<br>Comparative testing to legally marketed predicate device                                                                                                                                                                                                                     | Pass<br>Pass                                 |
| Sterilization                           | In accordance with FDA-recognized voluntary consensus standard<br>ISO 17664-1:2021 (14-578)<br>In accordance with FDA-recognized voluntary consensus standard<br>ISO 17664-2:2021 (14-579)<br>In accordance with FDA-recognized voluntary consensus standard<br>ISO 14937:2009 (14-337)<br>In accordance with AAMI TIR12:2020<br>In accordance with AAMI TIR30:2011<br>In accordance with FDA-recognized voluntary consensus standard<br>AAMI ST58:2013/ (R)2018 (14-432) | Pass<br>Pass<br>Pass<br>Pass<br>Pass<br>Pass |
| Software Verification<br>and Validation | In accordance with FDA-recognized voluntary consensus standard<br>IEC 62304:2015 (13-79)                                                                                                                                                                                                                                                                                                                                                                                  | Pass                                         |
| Usability                               | In accordance with FDA-recognized voluntary consensus standard<br>IEC 62366:2015 + A1:2020 (5-129)                                                                                                                                                                                                                                                                                                                                                                        | Pass                                         |
| Performance Testing -<br>Bench          | In accordance with device input specifications<br>Comparative testing to currently legally marketed predicate devices:<br>Spatial Uniformity<br>Minimum Detectable Fluorescence<br>Fluorescence Detection Depth<br>Clinically Meaningful Limits of Detection<br>Signal to Noise<br>Photobiological Safety (IEC 62417)<br>Contrast                                                                                                                                         | Pass<br>Pass                                 |
| Performance Testing                     | In accordance with user needs and intended uses                                                                                                                                                                                                                                                                                                                                                                                                                           | Pass                                         |

NOTE: The Advanced Imaging Modality (AIM) System is not patient contacting; therefore, biocompatibility testing was not required to support the determination of substantial equivalence.

{10}------------------------------------------------

NOTE: The Advanced Imaging Modality (AIM) System does not require clinical studies to support the determination of substantial equivalence.

# Conclusions:

The Advanced Imaging Modality (AIM) System is the same or similar in design, intended use, principles of operation, technological characteristics and safety features to the predicate devices. In summary, the Advanced Imaging Modality (AIM) System is the same or similar with respect to safety and effectiveness to the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K231854](https://fda.innolitics.com/device/K231854)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
