1788 4K Camera System with Advanced Imaging Modality; L12 LED Light Source with AIM
Applicant
Stryker
Product Code
OWN · Gastroenterology, Urology
Decision Date
Sep 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ arthroscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repar, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, orthopedic surgeons, ENT/heurosurgeons and urologists. Upon intravenous administration of SPY AGENT GREEN (indocyanine green for injection, USP), the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L12 Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month and older, and visualization of at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct) in adults and pediatric patients 12 to 17 years of age, using near-infrared imaging. Fluorescence imaging of biliary ducts with the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. Additionally, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging. Upon interstitial administration of SPY AGENT GREEN, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. Upon administration and use of pafolacianine consistent with its approved label, the L12 LED Light Source with Advanced Imaging Modality and SafeLight™ Cable are used to perform intraoperative fluorescence imaging of tissues that have taken up the drug. The L12 LED Light Source with Advanced Imaging Modality is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.
Device Story
Endoscopic 4K imaging system; utilizes white light, near-infrared (NIR) fluorescence, and cyan spectral imaging. Inputs: light projected via endoscope; CMOS sensor acquires image data. Processing: digital video stream generation; NIR fluorescence imaging via SPY AGENT GREEN or pafolacianine; NIR transillumination for ureters; Cyan Spectral Imaging for mucosal surface/capillary visualization. Output: real-time video displayed for surgeon. Used in OR by surgeons (general, pediatric, gynecological, cardiac, thoracic, plastic, orthopedic, ENT, neuro, urology). Assists in intraoperative visualization of vessels, blood flow, tissue perfusion, lymphatics, and biliary ducts. Benefits: enhanced anatomical visualization during minimally invasive procedures.
Clinical Evidence
Bench testing only. No clinical data required. Performance testing included electromagnetic compatibility (IEC 60601-1-2), electrical safety (ANSI AAMI ES 60601-1), laser safety (IEC 60825-1), sterilization (ISO 17664, ISO 14937), and software validation (IEC 62304). Bench performance metrics included spatial uniformity, minimum detectable fluorescence, fluorescence detection depth, signal-to-noise ratio, and photobiological safety.
Technological Characteristics
System includes 4K CMOS camera and L12 LED light source with infrared laser (780nm/830nm). Connectivity: digital video output. Sterilization: per ISO 17664/14937. Safety standards: IEC 60601-1, IEC 60601-2-18, IEC 60825-1, IEC 62471. Imaging modes: White light, NIR fluorescence, NIR transillumination, Cyan Spectral Imaging. Laser class 1M.
Indications for Use
Indicated for general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery, and plastic surgery in adults and pediatric patients. Used for visualization of vessels, blood flow, tissue perfusion, biliary ducts (12-17 years), lymphatic system, and ureter transillumination. Used with SPY AGENT GREEN or pafolacianine for fluorescence imaging. Contraindicated for standalone biliary duct visualization.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Advanced Imaging Modality (AIM) System (K230605, K230754)
Olympus VISERA ELITE Video System Center and Xenon Light Source (K111425)
PINPOINT Endoscopic Fluorescence Imaging System (K182606)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right, the FDA logo is displayed in blue, with the acronym "FDA" stacked above the words "U.S. FOOD & DRUG ADMINISTRATION."
September 20, 2023
Stryker April Malmborg Senior Director, Regulatory Affairs 5900 Optical Court San Jose, California 95138
Re: K231854
Trade/Device Name: 1788 4K Camera System with Advanced Imaging Modality; L12 LED Light Source with AIM Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, GWG, NWB, OWN, FCS, FCW Dated: June 23, 2023 Received: June 23, 2023
Dear April Malmborg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tanisha L. Tanisha L. Hithe -S 2023.09.20 Hithe -S 09:19:05 -04'00'
Tanisha Hithe, MS, MHS Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
### K231854
#### Device Name
1788 4K Camera System with Advanced Imaging Modality
#### Indications for Use (Describe)
The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ arthroscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients.
A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repar, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.
The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, orthopedic surgeons, ENT/heurosurgeons and urologists.
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|------------------------------------------------------------|
| <span></span> | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known)
### K231854
#### Device Name
L12 LED Light Source with Advanced Imaging Modality
### Indications for Use (Describe)
Upon intravenous administration of SPY AGENT GREEN (indocyanine green for injection, USP), the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L12 Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month and older, and visualization of at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct) in adults and pediatric patients 12 to 17 years of age, using near-infrared imaging.
Fluorescence imaging of biliary ducts with the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.
Additionally, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.
Upon interstitial administration of SPY AGENT GREEN, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
Upon administration and use of pafolacianine consistent with its approved label, the L12 LED Light Source with Advanced Imaging Modality and SafeLight™ Cable are used to perform intraoperative fluorescence imaging of tissues that have taken up the drug.
The L12 LED Light Source with Advanced Imaging Modality is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R Part 807.92(c).
### 510(k) Number :
### Submitter:
| Applicant: | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138 |
|-----------------|------------------------------------------------------------------------------------------------------------------|
| Contact Person: | April Malmborg<br>Senior Director, Regulatory Affairs<br>Email: april.malmborg@stryker.com<br>Ph: (408) 754-2472 |
| Date Prepared: | June 23, 2023 |
### Subject Device:
The subject device is the Advanced Imaging Modality (AIM) System and specifically the following system components:
| Name of Device: | 1788 4K Camera System with Advanced Imaging Modality |
|----------------------|-------------------------------------------------------------------------------------------------|
| Common or Usual Name | 3-chip Video Camera |
| Classification Name: | Endoscope and Accessories (21 C.F.R. §876.1500)<br>Neurological Endoscope (21 C.F.R. §882.1480) |
| Regulatory Class: | II |
| Product Code: | GCJ<br>GWG<br>NWB |
| 510(k) Review Panel: | General & Plastic Surgery<br>Neurology<br>Gastroenterology/ Urology |
| Name of Device: | L12 LED Light Source with AIM |
|----------------------|----------------------------------------------------------|
| Common or Usual Name | Light Source, Illuminator |
| Classification Name: | Endoscope and Accessories (21 C.F.R. §876.1500) |
| | Fiberoptic light ureteral catheter (21 C.F.R. §876.4020) |
| | Neurological Endoscope (21 C.F.R. §882.1480) |
| Regulatory Class: | II |
| Product Code: | OWN |
| | FCS |
| | FCW |
| | GWG |
| | NWB |
| 510(k) Review Panel: | General & Plastic Surgery |
| | Gastroenterology/ Urology |
| | Neurology |
{5}------------------------------------------------
# Predicate Device(s):
| Advanced Imaging<br>Modality (AIM) System | 1788 4K Camera System<br>L12 LED Light Source | K230605 (primary)<br>K230754 (primary) |
|--------------------------------------------------------------------|-----------------------------------------------|----------------------------------------|
| Olympus VISERA ELITE Video System Center and<br>Xenon Light Source | | K111425 (secondary) |
| PINPOINT Endoscopic Fluorescence Imaging System | | K182606 (secondary) |
### Device Description:
Stryker's Advanced Imaging Modality (AIM) System is an endoscopic real-time 4K visible white light, near-infrared illumination and transillumination, and cyan spectral imaging system. Near-infrared illumination is used for fluorescence imaging using SPY AGENT™ GREEN (indocyanine green for injection, USP) or CYTALUX™ (pafalocianine) injection. Near-infrared illumination is also intended for use during transillumination of the ureters using the IRIS Ureteral Kit during minimally invasive and open surgical procedures. Cyan Spectral Imaging is intended as an alternative mode of visualization that uses narrow bands for illumination during endoscopic surgery to provide greater visualization of surface structures and fine capillary patterns of the mucosal membrane.
### Indications for Use:
### 1788 4K Camera System with Advanced Imaging Modality:
The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ sinuscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients.
A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection. Laparoscopically assisted hysterectomy. Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.
The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT/neurosurgeons and urologists.
{6}------------------------------------------------
### L12 LED Light Source with Advanced Imaging Modality:
Upon intravenous administration of SPY AGENT GREEN (indocyanine green for injection, USP), the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month and older, and visualization of at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct) in adults and pediatric patients 12 to 17 years of age, using near-infrared imaging.
Fluorescence imaging of biliary ducts with the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.
Additionally, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.
Upon interstitial administration of SPY AGENT GREEN, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
Upon administration and use of pafolacianine consistent with its approved label, the L12 LED Light Source with Advanced Imaging Modality and SafeLight™ Cable are used to perform intraoperative fluorescence imaging of tissues that have taken up the drug.
The L12 LED Light Source with Advanced Imaging Modality is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.
{7}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device:
| Item | | Subject Device | Predicate Devices | | |
|-------------------------|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| | | | Primary | Secondary | Secondary |
| | | Advanced Imaging<br>Modality (AIM) System | Primary<br>Advanced Imaging<br>Modality (AIM) System | VISERA ELITE | PINPOINT System |
| Manufacturer | | Stryker | Same as subject device | Olympus Medical<br>System<br>Corporation<br>K111425 | Same as subject<br>device |
| Submission Reference(s) | | Current Submission | K230605; K230754 | K111425 | K182606 |
| Intended Use | | Endoscopic illumination<br>and imaging during<br>endoscopic procedures | Same as subject device | Same as subject<br>device | Same as subject<br>device |
| Indications for Use | | NOTE 1 | Same as subject device | NOTE 2 | NOTE 3 |
| System Components | | Camera System (subject<br>of submission)<br>Light Source (subject of<br>submission) and SafeLight<br>Cable<br>Endoscopes<br>IRIS Ureteral Kit<br>Imaging Agents | Same as subject device | Video Center<br>Xenon Light<br>Source<br>Videoscope | Video Processor/<br>Illuminator<br>Camera Head<br>Light Guide Cable<br>Laparoscopes<br>Imaging Agent |
| Imaging Agents | | SPY AGENT GREEN<br>Pafolacianine | Same as subject device | Not applicable. | SPY AGENT<br>GREEN |
| Principles of Operation | | Via an endoscope, light is<br>projected by a light source<br>onto one or more image<br>sensor(s) which acquire a<br>continuous stream of<br>image data. The image<br>data is processed to<br>provide a video stream<br>that is then sent to a<br>display for viewing. | Same as subject device | Same as subject<br>device | Same as subject<br>device |
| Safety Standards | | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-1-2<br>IEC 60601-4-2<br>IEC 60825-1<br>IEC 62471 | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-1-2<br>IEC 60601-4-2<br>IEC 60825-1 | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-2<br>IEC 62471 | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-2<br>IEC 60825-1 |
| Imaging<br>Modes | White Light | White Light | Same as subject device | Same as subject<br>device | Same as subject<br>device |
| | NIR<br>Fluorescence | ENV<br>Contrast<br>Overlay<br>- without IRIS<br>- with IRIS<br>Color Segmented<br>Fluorescence<br>- without IRIS<br>- with IRIS | ENV<br>Contrast<br>Overlay<br>- without IRIS<br>- with IRIS | Not applicable. | Contrast<br>Overlay<br>Color Segmented<br>Fluorescence |
| | NIR Trans-<br>illumination | IRIS | Same as subject device. | Not applicable. | Not applicable. |
| | Cyan Spectral<br>Imaging<br>(CSI) | Cyan Spectral Imaging<br>(CSI) | Not Applicable. | Narrow Band<br>Imaging | Not applicable. |
| Camera<br>System | Image Sensor | CMOS image sensor | Same as subject device. | CCD image<br>sensor | Same as subject<br>device. |
| Item | | Subject Device | Predicate Devices | | |
| | | Advanced Imaging<br>Modality (AIM) System | Primary<br>Advanced Imaging<br>Modality (AIM) System | Secondary<br>VISERA ELITE | Secondary<br>PINPOINT System |
| | Image Processing/<br>Video Output | Digital | Same as subject device. | Same as subject device. | Same as subject device. |
| | Resolution | 4K (up to 3840 x 2160)<br>Up to 1920 x 1080 | Same as subject device. | Up to 1920 x 1080 | Up to 1920 x 1080 |
| | Frame Rate | 60 frames per second | Same as subject device. | Same as subject device. | Same as subject device. |
| Light<br>Source | Light Source/<br>Laser | RGB LEDs/<br>Infrared Laser | Same as subject device. | Xenon | Same as subject device. |
| | Excitation<br>Wavelength | 780 nm (used for NIR<br>fluorescence)<br>830 nm (used for NIR<br>transillumination) | Same as subject device. | Not applicable. | 806 nm (used for<br>NIR fluorescence) |
| | Laser Safety<br>Classification | Class 1M | Same as subject device. | Not applicable. | Class 3R |
{8}------------------------------------------------
NOTE 1: The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ arthroscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectorny, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectorny, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectorny, Pleurodesis, Internal mammary artery dissection for coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT/neurosurgeons and urologists.
Upon intravenous administration of SPY AGENT™ GREEN (indocyanine green for injection, USP), the L12 LED Light Source with AIM and SafeLight™ Cable are used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L12 LED Light Source with AIM and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month and older, and visualization of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct) in adults and pediatic patients 12 to 17 years of age, using nearinfrared imaging. Fluorescence imaging of biliary ducts with the L12 Light Source with AIM and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. Additionally, the L12 Light Source with AIM and SafeLight Cable enable surgeons to perform minimally invasive cranial neurosurgery in adults and pediative patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using. Upon interstitial administration of SPY AGENT GREEN, the L12 LED Light Source with AIM and SafeLight Cable is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. Upon administration and use of pafolacianine consistent with its approved label, the L12 LED Light Source and SafeLight™ Cable is used to perform intraoperative fluorescence imaging of tissues that have taken up the drug. The L12 Light Source with AIM is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.
NOTE 2: The video system center has been designed to be used with Olympus cameras heads, endoscopes, light source and monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation. The light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
NOTE 3: Upon intravenous administration of SPY AGENT™ GREEN, the PINPOINT Endoscopic Fluorescence Imaging System is used with SPY AGENT GREEN to perform intraoperative fluorescence anglography, and it is also indicated for use in fluorescence imaging of biliary ducts, and when intraoperative cholangiography. The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near infrared imaging of billary ducts with the
{9}------------------------------------------------
PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for bilization. Upon interstitial administration of SPY AGENT GREEN, the PINPOINT System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
### Performance Data:
| Test | Method | Result |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Electromagnetic<br>Compatibility | In accordance with FDA-recognized voluntary consensus standard<br>IEC 60601-1-2:2014 (19-8)<br>In accordance with FDA-recognized voluntary consensus standard<br>IEC 60601-4-2:2016 (19-19) | Pass<br>Pass |
| Electrical Safety | In accordance with FDA-recognized voluntary consensus standard<br>ANSI AAMI ES 60601-1:2005 + A1:2012 + A2: 2021 (19-46)<br>In accordance with FDA-recognized voluntary consensus standard<br>IEC 60601-1-6:2010 + A1:2013 + A2:2020 (5-132)<br>In accordance with FDA-recognized voluntary consensus standard<br>IEC 60601-2-18:2009 (9-114) | Pass<br>Pass<br>Pass |
| Laser Safety | In accordance with FDA-recognized voluntary consensus standard<br>IEC 60825-1:2014 (12-273)<br>In accordance with FDA-recognized voluntary consensus standard<br>IEC 62471 (12-249); and,<br>Comparative testing to legally marketed predicate device | Pass<br>Pass |
| Sterilization | In accordance with FDA-recognized voluntary consensus standard<br>ISO 17664-1:2021 (14-578)<br>In accordance with FDA-recognized voluntary consensus standard<br>ISO 17664-2:2021 (14-579)<br>In accordance with FDA-recognized voluntary consensus standard<br>ISO 14937:2009 (14-337)<br>In accordance with AAMI TIR12:2020<br>In accordance with AAMI TIR30:2011<br>In accordance with FDA-recognized voluntary consensus standard<br>AAMI ST58:2013/ (R)2018 (14-432) | Pass<br>Pass<br>Pass<br>Pass<br>Pass<br>Pass |
| Software Verification<br>and Validation | In accordance with FDA-recognized voluntary consensus standard<br>IEC 62304:2015 (13-79) | Pass |
| Usability | In accordance with FDA-recognized voluntary consensus standard<br>IEC 62366:2015 + A1:2020 (5-129) | Pass |
| Performance Testing -<br>Bench | In accordance with device input specifications<br>Comparative testing to currently legally marketed predicate devices:<br>Spatial Uniformity<br>Minimum Detectable Fluorescence<br>Fluorescence Detection Depth<br>Clinically Meaningful Limits of Detection<br>Signal to Noise<br>Photobiological Safety (IEC 62417)<br>Contrast | Pass<br>Pass |
| Performance Testing | In accordance with user needs and intended uses | Pass |
NOTE: The Advanced Imaging Modality (AIM) System is not patient contacting; therefore, biocompatibility testing was not required to support the determination of substantial equivalence.
{10}------------------------------------------------
NOTE: The Advanced Imaging Modality (AIM) System does not require clinical studies to support the determination of substantial equivalence.
# Conclusions:
The Advanced Imaging Modality (AIM) System is the same or similar in design, intended use, principles of operation, technological characteristics and safety features to the predicate devices. In summary, the Advanced Imaging Modality (AIM) System is the same or similar with respect to safety and effectiveness to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.