K231829 · Ht Medical D.B.A. Xenix Medical · OUR · Aug 15, 2023 · Orthopedic
Device Facts
Record ID
K231829
Device Name
Xenix Medical Sacroiliac Fixation System
Applicant
Ht Medical D.B.A. Xenix Medical
Product Code
OUR · Orthopedic
Decision Date
Aug 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Xenix Medical Sacroiliac Fixation System is indicated for sacroiliac joint fusion for: - Sacroiliac joint dysfunction including degenerative sacroiliac joint disruptions - Augmenting immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion The Xenix Medical Sacroiliac Fixation System is also indicated for fracture fixation of acute, and nontraumatic fractures involving the sacroiliac joint. The Xenix Medical Sacroiliac Fixation System Navigation Instruments are intended to be used with the Xenix Medical Sacroiliac Fixation System during surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation System S8, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.
Device Story
System comprises 3D-printed titanium alloy implants and surgical instrumentation for sacroiliac joint fusion and fracture fixation. Navigation instruments facilitate precise anatomical localization during open or minimally invasive surgery by integrating with Medtronic StealthStation System S8. Surgeon uses navigation to register rigid anatomical structures (pelvis/vertebra) against CT/MRI models, fluoroscopy, or digitized landmarks. System provides mechanical stabilization of sacroiliac joint; navigation aids surgeon in accurate implant placement. Benefits include improved surgical precision and joint stabilization.
Clinical Evidence
Bench testing only. Mechanical testing conducted per ASTM F2193 and ASTM F543 (static/dynamic cantilever bending, axial pull-out). Sterilization validation per ISO 11137-1/2. Bacterial endotoxin testing per ANSI/AAMI ST-72. Navigation instrument accuracy confirmed via registration/function testing with Medtronic StealthStation S8 and dimensional analysis against predicate instruments.
Technological Characteristics
Implants: 3D-printed Ti-6Al-4V ELI titanium alloy (ASTM F-3001). Navigation instruments: stereotaxic, compatible with Medtronic StealthStation S8. Implants provided sterile; reusable instruments provided non-sterile for steam sterilization. Mechanical testing per ASTM F2193, ASTM F543.
Indications for Use
Indicated for skeletally mature patients requiring sacroiliac joint fusion due to dysfunction (including degenerative disruptions) or stabilization during sacropelvic fixation (lumbar/thoracolumbar fusion). Also indicated for fracture fixation of acute, non-acute, and non-traumatic fractures of the sacroiliac joint.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
August 15, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
HT Medical d.b.a. Xenix Medical % Justin Eggleton Vice President, Regulatory Affairs Mcra LLC 803 7th Street NW, 3rd Floor Washington, District of Columbia 20001
Re: K231829
Trade/Device Name: Xenix Medical Sacroiliac Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR, OLO, HWC Dated: June 21, 2023 Received: June 21, 2023
## Dear Justin Eggleton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image contains the name "Colin O'neill -S" in black font on the left side of the image. On the right side of the image is the letters "FDA" in a light blue color. The letters "FDA" are much larger than the name on the left side of the image. The background of the image is white.
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231829
Device Name
Xenix Medical Sacroiliac Fixation System
#### Indications for Use (Describe)
The Xenix Medical Sacroiliac Fixation System is indicated for sacroiliac joint fusion for:
- · Sacroiliac joint dysfunction including degenerative sacroiliac joint disruptions
· Augmenting immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion
The Xenix Medical Sacroiliac Fixation System is also indicated for fracture fixation of acute, and nontraumatic fractures involving the sacroiliac joint.
The Xenix Medical Sacroiliac Fixation System Navigation Instruments are intended to be used with the Xenix Medical Sacroiliac Fixation System during surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation System S8, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <div> <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Device Trade Name: | Xenix Medical Sacroiliac Fixation System |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | HT Medical d.b.a. Xenix Medical<br>111 W Jefferson St., Suite 100<br>Orlando, FL 32801 |
| Contact: | Teresa Cherry<br>Vice President of Quality Assurance and Regulatory Affairs<br>Xenix Medical<br>Phone: 888.594.8633 |
| Prepared by: | Mr. Justin Eggleton<br>Vice President, Regulatory Affairs<br>MCRA, LLC<br>803 7th Street NW, 3rd Floor<br>Washington, DC 20001<br>jeggleton@mcra.com<br>Phone: 202-552-5804 |
| Date Prepared: | August 15, 2023 |
| Classifications: | 21 CFR §888.3040; Smooth or threaded metallic bone fixation<br>fastener<br>21 CFR §882.4560; Stereotaxic Instruments |
| Class: | II |
| Product Codes: | OUR, OLO, HWC |
| Device Common<br>Name: | Sacroiliac Joint Screw, Orthopedic Stereotaxic Instrument |
{4}------------------------------------------------
# Indications for Use:
The Xenix Medical Sacroiliac Fixation System is indicated for sacroiliac joint fusion for:
- Sacroiliac joint dysfunction including degenerative sacroiliitis and sacroiliac joint ● disruptions
- Augmenting immobilization and stabilization of the sacroiliac joint in skeletally mature ● patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion
The Xenix Medical Sacroiliac Fixation System is also indicated for fracture fixation of acute, non-acute, and non-traumatic fractures involving the sacroiliac joint.
The Xenix Medical Sacroiliac Fixation System Navigation Instruments are intended to be used with the Xenix Medical Sacroiliac Fixation System during surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation System S8, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.
# Device Description:
The Xenix Medical Sacroiliac Fixation System consists of 3D printed medical grade Titanium Alloy Implants (Ti-6Al-4V ELI per ASTM F-3001) and surgical instrumentation for implantation. Implants are provided sterile in various lengths and diameters. The reusable instrumentation is provided non-sterile in a steam sterilization instrument tray.
## Predicate Devices:
| Device Name(s) | Manufacturer | K-Number |
|---------------------------------------------------------------|--------------------------------------|----------|
| <b>Primary Predicate Device</b> | | |
| iFuse TORQ Implant System | SI-Bone | K222605 |
| <b>Additional Predicate Devices</b> | | |
| Rialto™ SI Fusion System | Medtronic | K161210 |
| Sacrix Sacroiliac Fixation<br>(SacroFuse/SIJFuse) | SpineFrontier (Sacrix) | K150017 |
| Synthes 6.5mm Cannulated Screw | Synthes | K021932 |
| <b>Reference Device</b> | | |
| Quantum Anterior Cervical Plate (for manufacturing reference) | Nvision Biomedical Technologies, Inc | K210424 |
{5}------------------------------------------------
## Performance Testing Summary:
Mechanical testing, including static and dynamic cantilever bending and axial pull-out, were conducted in accordance with ASTM F2193 and ASTM F543 to demonstrate substantial equivalence of the subject Xenix Medical Sacroiliac Fixation System Implants to the predicate devices. Sterilization validation of the subject implants was performed in accordance with ISO 11137-1 and ISO 11137-2. Bacterial endotoxin testing (BET) was conducted in accordance with ANSI/AAMI ST-72.
Accuracy testing was performed on the Xenix Medical Sacroiliac Fixation System Navigation instruments to confirm that the instruments register and function properly with the Medtronic StealthStation S8 Navigation System. Dimensional analysis of the Xenix Medical Sacroiliac Fixation System Navigation instruments was performed against the predicate instruments.
#### Substantial Equivalence:
The subject device was demonstrated to be substantially equivalent to the predicate cited in the passage above with respect to indications, design, materials, function, manufacturing, and performance.
#### Conclusion:
The Xenix Medical Sacroiliac Fixation System is substantially equivalent to the cited predicate devices with respect to intended use, indications for use, design, function, materials, and performance. The differences in the technological characteristics between the subject device and the predicate devices do not raise new or different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.