SmartBolus Calculator

K231824 · Insulet Corporation · QRX · Oct 18, 2023 · Clinical Chemistry

Device Facts

Record IDK231824
Device NameSmartBolus Calculator
ApplicantInsulet Corporation
Product CodeQRX · Clinical Chemistry
Decision DateOct 18, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1358
Device ClassClass 2
AttributesSoftware as a Medical Device, Therapeutic, Pediatric

Indications for Use

The SmartBolus Calculator is software intended for the management of diabetes in persons aged 2 and older requiring rapid-acting U-100 insulin. The SmartBolus Calculator calculates a suggested bolus dose based on user-entered carbohydrates, most recent sensor glucose value (or blood glucose reading if using fingerstick), rate of change of the sensor glucose (if applicable), insulin on board (IOB), and programmable correction factor, insulin to carbohydrate ratio, and target glucose value. The SmartBolus Calculator is intended for single patient, home use and requires a prescription.

Device Story

SmartBolus Calculator is a software component of the Omnipod 5 Automated Insulin Delivery System; operates on mobile app platforms (Android/iOS). Inputs include user-entered carbohydrates, CGM sensor glucose values (or manual blood glucose readings), sensor glucose rate of change, insulin on board (IOB), and programmable settings (correction factor, insulin-to-carbohydrate ratio, target glucose). Device processes these inputs via an algorithm to calculate suggested meal or correction insulin bolus doses. Output is displayed to the user, who may accept, modify, or cancel the suggested dose. Used in home settings by patients; facilitates insulin therapy adjustment. Benefits include automated, personalized bolus calculations to assist in glycemic management.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by software verification and human factors validation testing conducted on the Omnipod 5 App (iOS) as part of the broader system evaluation.

Technological Characteristics

Software-only device residing within the Omnipod 5 App. Operates on iOS and Android platforms. Integrates with iCGM systems or manual blood glucose meter inputs. Complies with IEC 62304:2015-05 (software life cycle) and ANSI AAMI ISO 14971:2019 (risk management).

Indications for Use

Indicated for diabetes management in persons aged 2 and older requiring rapid-acting U-100 insulin.

Regulatory Classification

Identification

An insulin therapy adjustment device is a device intended to incorporate biological inputs, including glucose measurement data from a continuous glucose monitor, to recommend insulin therapy adjustments as an aid in optimizing insulin therapy regimens for patients with diabetes mellitus.

Special Controls

(b) Classification: Class II (special controls). The special controls for this device are - 1. Design verification and validation must include the following: - (i) A complete description of the required data inputs, including timeframe over which data inputs must be collected and number of data points required for accurate recommendations. - (ii) A complete description of the types of device outputs and insulin therapy adjustment recommendations, including how the recommendations are generated. - (iii)Robust data demonstrating the clinical validity of the device outputs and insulin therapy recommendations. - (iv)A robust assessment of all input data specifications, including accuracy requirements for continuous glucose monitors and other devices generating data inputs, to ensure accurate and reliable therapy adjustment recommendations. This assessment must include adequate clinical justification for each specification. - (v) A detailed strategy to ensure secure and reliable means of data transmission to and from the device, including data integrity checks, accuracy checks, reliability checks, and security measures. - (vi)Robust data demonstrating that users can understand and appropriately interpret recommendations generated by the device. - (vii) An appropriate mitigation strategy to minimize the occurrence of dosing recommendation errors, and to mitigate the risk to patients of any residual dosing recommendation errors to a clinically acceptable level. - 2. The device must not be intended for use in implementing automated insulin dosing. - 3. Your 809.10(b) labeling must include: - (i) The identification of specific insulin formulations that have been demonstrated to be compatible with use of the device. - (ii) A detailed description of the specifications of compatible devices that provide acceptable input data (e.g., continuous glucose monitors, insulin pumps) used to provide accurate and reliable therapy adjustment recommendations. - (iii) A detailed description of all types of required data (inputs) and dosing recommendations (outputs) that are provided by the device. - (iv) A description of device limitations, and instructions to prevent possible disruption of accurate therapy adjustment recommendations (e.g., time zone changes due to travel).

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following: (i) A complete description of the required data inputs, including timeframe over which data inputs must be collected and number of data points required for accurate recommendations; (ii) A complete description of the types of device outputs and insulin therapy adjustment recommendations, including how the recommendations are generated; (iii) Robust data demonstrating the clinical validity of the device outputs and insulin therapy recommendations; (iv) A robust assessment of all input data specifications, including accuracy requirements for continuous glucose monitors and other devices generating data inputs, to ensure accurate and reliable therapy adjustment recommendations. This assessment must include adequate clinical justification for each specification; (v) A detailed strategy to ensure secure and reliable means of data transmission to and from the device, including data integrity checks, accuracy checks, reliability checks, and security measures; (vi) Robust data demonstrating that users can understand and appropriately interpret recommendations generated by the device; and (vii) An appropriate mitigation strategy to minimize the occurrence of dosing recommendation errors, and to mitigate the risk to patients of any residual dosing recommendation errors to a clinically acceptable level. (2) The device must not be intended for use in implementing automated insulin dosing. (3) Your 21 CFR 809.10(b) labeling must include: (i) The identification of specific insulin formulations that have been demonstrated to be compatible with use of the device; (ii) A detailed description of the specifications of compatible devices that provide acceptable input data (e.g., continuous glucose monitors, insulin pumps) used to provide accurate and reliable therapy adjustment recommendations; (iii) A detailed description of all types of required data (inputs) and dosing recommendations (outputs) that are provided by the device; and (iv) A description of device limitations, and instructions to prevent possible disruption of accurate therapy adjustment recommendations (e.g., time zone changes due to travel).

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K231824 B Applicant Insulet Corporation C Proprietary and Established Names SmartBolus Calculator D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QRX | Class II | 21 CFR 862.1358 - Insulin Therapy Adjustment Device | CH - Clinical Chemistry | | NDC | Class II | 21 CFR 868.1890 - Predictive pulmonary-function value calculator | CH - Clinical Chemistry | ## E Purpose for Submission: Modifications to the device to add compatibility with the iOS version of the Omnipod 5 App component of the Omnipod 5 ACE Pump. ## II Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: The SmartBolus Calculator is software intended for the management of diabetes in persons aged 2 and older requiring rapid-acting U-100 insulin. The SmartBolus Calculator calculates a suggested bolus dose based on user-entered carbohydrates, most recent sensor glucose value (or blood glucose reading if using fingerstick), rate of change of the sensor glucose (if applicable), insulin on board (IOB), and programmable correction factor, insulin to carbohydrate ratio, and Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {2} target glucose value. The SmartBolus Calculator is intended for single patient, home use and requires a prescription. ## C Special Conditions for Use Statement(s): Rx – For prescription use only. This device resides in the Omnipod 5 App and requires the App to function. The SmartBolus Calculator works with the following rapid-acting U-100 insulins: NovoLog® (insulin aspart), Humalog® (insulin lispro), and Admelog® (insulin lispro). ## III Device Description The SmartBolus Calculator is a software device that resides in the Omnipod 5 App. It requires input parameters and settings from the SmartAdjust™ Technology, a glycemic controller (iAGC) and glucose values from a compatible integrated continuous glucose monitoring system (iCGM) or blood glucose (BG) values from a blood glucose meter to calculate suggested insulin bolus doses. The SmartBolus Calculator can be used in both open-loop (manual mode) and closed-loop (automated mode). When used with a compatible iCGM, the SmartBolus Calculator can use the sensor glucose values and trend information to calculate a suggested bolus dose. When the SmartBolus Calculator is used with manually entered BG readings it suggests a bolus dose based on the same calculations as the currently cleared Omnipod DASH Insulin Management System bolus calculator (K180045, K192659). ## IV Substantial Equivalence Information: A Predicate Device Name(s): SmartBolus Calculator B Predicate 510(k) Number(s): K222239 C Comparison with Predicate(s): | Device & Predicate Device(s): | K231824 | K222239 | | --- | --- | --- | | Device Trade Name | SmartBolus Calculator | SmartBolus Calculator | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | The SmartBolus Calculator is software intended for the management of diabetes in persons aged 2 and older requiring rapid-acting U-100 | Same | K231824 - Page 2 of 4 {3} | | insulin. The SmartBolus Calculator calculates a suggested bolus dose based on user-entered carbohydrates, most recent sensor glucose value (or blood glucose reading if using fingerstick), rate of change of the sensor glucose (if applicable), insulin on board (IOB), and programmable correction factor, insulin to carbohydrate ratio, and target glucose value. The SmartBolus Calculator is intended for single patient, home use and requires a prescription. | | | --- | --- | --- | | General Device Characteristic Differences | | | | Omnipod 5 App Operating System | iOS and Android Operating System | Android Operating System | V Standards/Guidance Documents Referenced: ANSI AAMI ISO 14971:2019 Medical devices – Application of Risk Management to Medical Devices IEC 62304:2015-05 Medical devices Software – (Software life cycle processes) VI Performance Characteristics: A. Analytical Performance The SmartBolus Calculator is a software-only device, and therefore analytical performance characteristics are not applicable. B. Other Supportive Instrument Performance Characteristics Data 1. Human Factors: The SmartBolus Calculator is a component of the Omnipod 5 System. Human factors validation testing was conducted with the Omnipod 5 App (iOS) installed on a compatible mobile device. Human Factors validation study on the Omnipod 5 System as a whole was performed with 60 representative participants and is described in the Decision Summary for K231826. K231824 - Page 3 of 4 {4} 2. Software: Detailed information on software of the device was reviewed and found to be acceptable. VII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. VIII Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K231824 - Page 4 of 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...