2.5 million scans from 5 countries across the world from more than 25 manufacturers.
—
Standalone study: 40 sites across the US, 55.48% females, 90.92% between 22 and 85 years of age, 31.43% from portable X-ray machines, sourced from more than 10 manufacturers.
>1 (radiologists, pulmonologists and ER physicians)
Indications for Use
The qXR-LN (qXR_Lung_nodule) is computer-aided detection software to identify and mark regions in relation to suspected pulmonary nodules from 6 to 30 mm in size. The device is intended to be used in the incidental adult population. It is designed to aid the physician to review the frontal (AP/PA) chest radiographs of adults acquired on digital radiographic systems as a second reader and be used with any DICOM viewer or PACS . qXR-LN provides adjunctive information only and is not a substitute for the original chest radiographic image.
Device Story
qXR-LN is a CADe software device for frontal (AP/PA) chest X-rays; identifies/marks suspected pulmonary nodules (6-30 mm). Input: DICOM chest X-rays from PACS or X-ray systems. Processing: Deep learning algorithm analyzes images to detect/localize nodules. Output: DICOM images with ROI markings indicating nodule presence/location. Used in clinical settings (radiology, ER, pulmonology) by physicians as a second reader. Provides adjunctive information; does not replace clinical assessment or rule out disease. Benefits: Assists clinicians in identifying incidental nodules, potentially improving detection accuracy.
Clinical Evidence
Standalone study (n=40 sites, 8 US states) compared qXR-LN against ground truth (5 ABR-certified radiologists using CT/X-ray). Nodule-level sensitivity: 84.1%; FPPI: 0.18; Scan-level AUC: 94.51%. Multi-reader multi-case (MRMC) study (radiologists, pulmonologists, ER physicians) showed statistically significant improvement in reader performance (aided vs unaided): AFROC improvement 0.0762 (p < 1x10^-5), AUC improvement 0.0697, nodule-level sensitivity improvement 11.96%.
Technological Characteristics
Software-only CADe device. Deep learning algorithm. Inputs: DICOM chest X-rays. Outputs: DICOM with ROI markings. Connectivity: Integrates with PACS/DICOM viewers. No hardware components. Moderate level of concern software.
Indications for Use
Indicated for incidental adult population undergoing frontal (AP/PA) chest radiography. Aids physicians (radiologists, ER physicians, pulmonologists) in detecting and localizing suspected pulmonary nodules (6-30 mm). Not for standalone clinical decision-making or ruling out conditions.
Regulatory Classification
Identification
Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
December 22, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
Qure.ai Technologies % Sri Anusha Matta Senior Regulatory Affairs Manager Level 7, Commerz II, International Business Park Oberoi Garden City, Goregaon(E) Mumbai, Maharashtra 400063 INDIA
Re: K231805
Trade/Device Name: qXR-LN Regulation Number: 21 CFR 892.2070 Regulation Name: Medical Image Analyzer Regulatory Class: Class II Product Code: MYN Dated: June 19, 2023 Received: June 20, 2023
# Dear Sri Anusha Matta:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231805
Device Name qXR-LN
### Indications for Use (Describe)
The qXR-LN (qXR Lung nodule) is computer-aided detection software to identify and mark regions in relation to suspected pulmonary nodules from 6 to 30 mm in size. The device is intended to be used in the incidental adult population. It is designed to aid the physician to review the frontal (AP/PA) chest radiographs of adults acquired on digital radiographic systems as a second reader and be used with any DICOM viewer or PACS . qXR-LN provides adjunctive information only and is not a substitute for the original chest radiographic image.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
Qure.ai's qXR-LN
#### 1 SUBMITTER
Qure.ai Technologies Level 7, Commerz II, International Business Park Oberoi Garden City, Goregaon (E), Mumbai 400 063 Phone: +91-9768123013 Primary Contact Person: Sri Anusha Matta Secondary contact person: Bunty Kundnani
Date Prepared: June 16, 2023
#### 2 DEVICE
| Name of Device: | qXR-LN |
|-----------------------|-------------------------|
| Common or Usual Name: | Analyzer, Medical Image |
| Classification Name: | Medical image analyzer |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 892.2070 |
| Product Code: | MYN |
#### 3 PREDICATE DEVICE
| Name of Device: | Auto Lung Nodule Detection |
|-----------------|------------------------------|
| Manufacturer: | Samsung Electronics Co., Ltd |
| 510(k) Number: | K201560 |
#### INTENDED USE / INDICATIONS FOR USE: 4
The qXR-LN (qXR_Lung_nodule) is computer-aided detection software to identify and mark regions in relation to suspected pulmonary nodules from 6 to 30 mm in size. The device is intended to be used in the incidental adult population. It is designed to aid the physician to review the frontal (AP/PA) chest
{4}------------------------------------------------
radiographs of adults acquired on digital radiographic systems as a second reader and be used with any DICOM viewer or PACS . qXR-LN provides adjunctive information only and is not a substitute for the original chest radiographic image.
#### 5 DEVICE DESCRIPTION
qXR-LN is a Computer-Aided Detection (CADe) device that is designed to perform CAD processing in frontal (PA or AP view) Chest X-ray images for indication of locations for high nodule probability, which has an effective detection size from 6 mm to 30 mm. The device is intended to be a secondary aid to the qualified intended user to identify incidental pulmonary lung nodules chest x-ray images.
The device utilizes a deep learning algorithm. The qXR-LN was trained on a large and diverse dataset of 2.5million scans from 5 countries across the world. The training dataset was from more than 25 manufacturers.
Chest X-rays are sent to qXR-LN by the means of transmission functions within the user's image storage system (e.g., Picture Archiving and Communication System (PACS)) or other radiological imaging equipment (e.g., X-ray systems) and processed by the qXR-LN to detect and localise lung nodules. Following receipt of chest radiographs, the software device automatically analyses each image to detect and localise lung nodules.
qXR-LN receives chest x-ray images in digital imaging and communications in medicine (DICOM) as input. The qXR-LN device produces DICOM format outputs that enable users to view the presence and location of lung nodules.
This device is intended to aid the intended user in review of chest x-rays and detect and localise lung nodules as a secondary reader. The results are not intended to be used on a standalone basis for clinical decision-making nor is it intended to rule out the target conditions or otherwise preclude clinical assessment of X-ray cases.
#### б COMPARISON OF THE PREDICATE DEVICE
qXR-LN is technologically similar to the predicate device, Samsung Electronics Auto Lung Nodule Detection in regard to the intended use and technological characteristics. Both are medical image analysers intended to read chest X-rays to detect and localise lung nodules. The algorithms function similarly and with the same purpose of detection and localization of lung nodules. There are minor technological differences between the subject and predicate devices. Samsung Auto Lung Nodule Detects and marks nodules sized 10-30mm, whereas, qXR-LN can detect and mark nodules sized 6-30mm. The device is intended to be used in the incidental adult population. Performance testing demonstrates that even with the difference in nodule size detection the device performance is substantially equivalently to the predicate device. The intended users of Samsung Auto Lung Nodule Detection are physicians, whereas, the intended users of the qXR-LN are radiologists, emergency room physicians or pulmonologists. The difference in intended users does not raise different questions of safety or effectiveness.
In terms of establishing substantial equivalence, the subject and predicate device have intended use, as a medical image analyser that detects and localises lung nodules and produces case-level output.
{5}------------------------------------------------
| | | Predicate Device<br>Auto Lung Nodule Detection | Subject Device<br>qXR-LN | |
|--|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Device Name | Auto Lung Nodule Detection | qXR-LN | |
| | 510(k) Number | K201560 | K231805 | |
| | Regulation | 21 CFR 892.2070 | 21 CFR 892.2070 | |
| | Regulation Description | Medical Image Analyser | Medical Image Analyser | |
| | Product Code | MYN | MYN | |
| | Device type | Medical Image Analyser | Medical Image Analyser | |
| | Manufacturer | Samsung Electronics Co., Ltd | Qure.ai Technologies | |
| | Intended use / Indications<br>for Use | The Auto Lung Nodule Detection<br>is computer-aided detection<br>software to identify and mark<br>regions in relation to suspected<br>pulmonary nodules from 10 to 30<br>mm in size. It is designed to aid<br>the physician to review the PA<br>chest radiographs of adults as a<br>second reader and be used as part<br>of S-Station, which is operation<br>software installed on Samsung<br>Digital X-ray Imaging systems.<br>Auto Lung Nodule Detection<br>cannot be used on the patients<br>who have lung lesions other than<br>abnormal nodules. | The qXR-LN (qXR_Lung_nodule) is<br>computer-aided detection software<br>to identify and mark regions in<br>relation to suspected pulmonary<br>nodules from 6 to 30 mm in size. The<br>device is intended to be used in the<br>incidental adult population. It is<br>designed to aid the physician to<br>review the frontal (AP/PA) chest<br>radiographs of adults acquired on<br>digital radiographic systems as a<br>second reader and be used with any<br>DICOM viewer or PACS. qXR-LN<br>provides adjunctive information only<br>and is not a substitute for the original<br>chest radiographic image. | |
| | Intended User | Physicians. | Radiologists, emergency room<br>physicians or pulmonologists who<br>regularly review chest X-rays as part<br>of their daily practice. | |
| | Modality | Chest X-ray | Chest X-ray | |
| | Target clinical conditions | Lung Nodules on PAview Chest X-<br>rays | Lung Nodules on PA/AP view Chest X-<br>rays | |
| | Technology | Machine learning | Deep/Machine Learning | |
| | Input format | DICOM | DICOM | |
| | Output | ROI marked on the duplicated<br>input image. | ROI marked on the duplicated input<br>image | |
# Table 1 Comparison between qXR-LN and the Predicate Device
{6}------------------------------------------------
| | Predicate Device | Subject Device |
|--------------------------|------------------------------------------------------------|--------------------------------------------------|
| | Auto Lung Nodule Detection | qXR-LN |
| Performance Metrics | | |
| Performance metrics used | Sensitivity, FPPI and JAFROC were calculated | Sensitivity, FPPI and AFROC were calculated |
| | Nodule level sensitivity 80.69% | Nodule level sensitivity 84.10 (77.97-97.24) |
| | JAFROC (aided – unaided) 7.8 (p=0.0003) | AFROC (aided - unaided) 7.62 (p < 1 x 10-5) |
| | FPPI (aided - unaided) +0.019 | FPPI (aided – unaided) -0.0078 |
| | Nodule level Sensitivity (aided- unaided) 10.8 | Nodule level Sensitivity (aided – unaided) 11.96 |
#### 7 TESTING
## Software
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device has a Moderate level of concern.
# Performance Testing - Clinical
A standalone study to assess device performance was conducted. These scans were obtained from 8 states (Ohio, New York, South Carolina, Iowa, Wisconsin, Texas, Oklahoma and Maryland) and 40 sites (each state had multiple sites) across the US. There were 55.48% females. Most scans (90.92%) were between 22 and 85 years of age. 31.43% scans were obtained from portable X-ray machines. The scans were sourced from more than 10 manufacturers. The data also consisted of commonly existing confounding conditions (mimickers, hardware or others). The standalone study was performed to compare qXR-LN's performance against a ground truth determined by 5 ABR certified ground truthers. They read the Chest X-rays with the accompanying CT scans and reports and the ground truth was based on the nodules visible on the Chest Xray. qXR-LN achieved a nodule-level sensitivity of 84.1% which we believe is substantially equivalent to the predicate's performance. The overall FPPI was 0.18 (0.14 - 0.22). As secondary metrics, scan level analyses were performed. qXR-LN achieved a scan level AUC of 94.51 (92.64 - 96.66). Sensitivity of 93.83 (88.94 – 97) and specificity of 81.09 (76.30 – 85) was achieved. Subgroup analyses was performed and presented for relevant subgroups including age, gender, race, nodule characteristics, manufacturer, location of nodule, scan view and confounders.
A fully crossed multi-case, multi-reader, retrospectively study design was utilized. The MRMC study was conducted in a sequential study design as a second read aid. An image viewer with AI
{7}------------------------------------------------
algorithms for lung nodule detection was used for chest X-ray scans. The ground truth was previously established by board-certified radiologists in the standalone study.
The readers consisted of radiologists, pulmonologists and ER physicians who had experience of less 3, 3-7 and more than 7 years.
The multi-reader multi-case demonstrated that aided readers performance for lung nodule detection was improved with statistical significance compared to unaided. We have performed subgroup analyses for several key subgroups to demonstrate generalizability.
| Modality | AFROC Estimate (95% CI) | AFROC (aided-unaided) | P - Value |
|----------|-------------------------|---------------------------|-----------|
| Aided | 0.8095 (0.7695-0.8494) | 0.07621 (0.0497 – 0.1026) | P <1×10-5 |
| Unaided | 0.7333 (0.6892-0.7774) | | |
Table 2 Performance of readers aided vs unaided by qXR-LN
As secondary metrics, AUC and nodule level sensitivity were estimated. The AUC of the readers improved by an estimate of 0.0697 (0.442 – 0.0953) and this result was significant. qXR-LN indicated 11.96% improvement for nodule level sensitivity. Subgroup analyses was performed and presented for relevant subgroups including age, gender, race, nodule characteristics, manufacturer, reader type, reader experience,location of nodule, scan view and confounders.
#### 8 CONCLUSION
The comparison in Table 1 and the software and performance testing presented above demonstrate that the qXR-LN device is substantially equivalent to the predicate device. The qXR-LN is a software only device, similar to the predicate (Samsung Auto Lung Nodule Detection). The qXR-LN has similar indications, technological characteristics, and principles of operation as the predicate device. There are minor technological differences between the subject and predicate devices. Samsung Auto Lung Nodule Detection detects and marks nodules sized 10-30mm, whereas, qXR-LN can detect and mark nodules sized 6 mm to 30 mm. The qXR-LN device can be used to detect and localize smaller nodules than the predicate and the performance of qXR-LN has been demonstrated. The intended users of Samsung Auto Lung Nodule Detection are physicians, whereas, the intended users of the qXR-LN are radiologists, emergency room physicians or pulmonologists. This does not pose any additional risks. Both devices operate as second reads in the standard workflow. The performance testing demonstrates that the qXR-LN device performs as intended and is therefore substantially equivalent to the predicate. Software and Clinical testing supports that the device performs in according with the device requirements.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.