← Product Code [FMF](/productcode/FMF) · K231792

# VeriSafe Safety sterile syringes (K231792)

_Promisemed Hangzhou Meditech Co., Ltd. · FMF · Sep 28, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K231792

## Device Facts

- **Applicant:** Promisemed Hangzhou Meditech Co., Ltd.
- **Product Code:** [FMF](/productcode/FMF.md)
- **Decision Date:** Sep 28, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

It is intended to inject fluids into the body or withdraw blood for medical purposes. After injection, the needle tube will be retracted into the barrel when the safety features are manually activated to minimize the risk of accidental sharps injury.

## Device Story

VeriSafe Safety Sterile Syringes are single-use, piston-style syringes consisting of a barrel, plunger, piston, spring, and needle. Available in Type F (fixed needle) and Type D (detachable needle). Operated by healthcare professionals in clinical or transport settings. During use, the plunger fills and discharges fluid. Upon completion, the user presses the push-button to fully depress the plunger, triggering a spring-loaded mechanism that retracts the needle tube, hub, and cylinder into the barrel. This action shields the needle to prevent accidental sharps injuries. The device is provided sterile via ethylene oxide. Clinical benefit includes reduced risk of needlestick injuries for healthcare workers.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified against ISO 7886-1 (syringes), ISO 9626 (needle tubing), ISO 7864 (needles), and ISO 23908 (sharps injury protection). Biocompatibility testing per ISO 10993 series (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis, pyrogenicity) confirmed safety. Sterility and shelf-life (3 years) validated per ISO 11135 and ASTM F1980-16.

## Technological Characteristics

Materials: Polypropylene (barrel, push-button, hub, cap), stainless steel (needle, spring), silicone rubber (sealing ring), nitrile butadiene rubber/TPE (piston). Energy: Manual operation with spring-loaded retraction. Connectivity: None. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 7886-1, ISO 9626, ISO 7864, ISO 23908, ISO 10993, ASTM F1980-16, ASTM D4169-16.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Verifine Safety Syringe with Fixed Needle ([K213560](/device/K213560.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

September 28, 2023

Promisemed Hangzhou Meditech Co., Ltd. % Wei Hsu Regulatory Manager Vee Care (Asia) Limited 17th Chung Pont Commercial Building, 300 Hennessy Road Hong Kong, Hong Kong China

Re: K231792

Trade/Device Name: VeriSafe Safety sterile syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, MEG Dated: August 29, 2023 Received: August 30, 2023

Dear Wei Hsu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Juliane C. Lessard -S

Juliane C. Lessard, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K231792

Device Name VeriSafe Safety Sterile Syringes

Indications for Use (Describe)

It is intended to inject fluids into the body or withdraw blood for medical purposes. After injection, the needle tube will be retracted into the barrel when the safety features are manually activated to minimize the risk of accidental sharps injury.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K231792-510(K)Summary

#### Date Prepared 1

September 18, 2023

## 2 Submitter's Information

Name of Sponsor:

Promisemed Hangzhou Meditech Co., Ltd.

Address:

No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou

City, 311121 Zhejiang, China

Contact Name:

Zearou Yang

Telephone No.:

+86 571 88772985

Fax No.:

+86 571 88772985

Email Address:

zearou.yang@promisemed.ca

#### ന Trade Name, Common Name, Classification

Trade/Product Name: VeriSafe Safety sterile syringes Common Name: Piston Syringe Classification name: Syringe, Piston; Syringe, Antistick Regulation Number: 21 CFR 880.5860 Device Class: Class II Product Code: FMF, MEG

- ব ldentification of Predicate Device K213560 Verifine Safety Syringe with Fixed Needle

{4}------------------------------------------------

#### Description of the Device റ

Safety Sterile Syringes is a sterile device consisting of a calibrated barrel, plunger, piston, push button, retraction cylinder, sealing ring, spring and a needle. It is made of plastic, stainless steel and silicone materials. Safety Sterile Syringes have two types: Type F and type D. Type F is a syringe with a fixed needle. Type D is a syringe with a detachable needle. The detachable needle is secured by rotating it through the front needle hub into the back needle hub located on the syringe.

Safety Sterile Syringes is using plunger to fill syringe as well as discharge the fluid. After use, by continuing to press the push-button down to depress the plunger all of the way within the barrel, the safety mechanism will be triggered, then the needle tube, needle hub and retraction cylinder will be retracted into the barrel completely under the force of the spring to minimize the risk of accidental sharps injury.

Product is delivered sterilization process is validated according to EN ISO 11135. This is a single-use device. Safety Sterile Syringes is intended to be used in professional healthcare facility/hospital and transport environment.

| Coding rule: |              |                                                               |                               |                   |
|--------------|--------------|---------------------------------------------------------------|-------------------------------|-------------------|
| MWS-032525   |              |                                                               |                               |                   |
| Type         | Capacity     | Gauge                                                         | Nominal length of needle tube | Wall              |
| Type D       | 1mL,2mL,     | 18G, 20G, 22G,<br>23G, 25G,<br>27G,28G, 29G,<br>30G, 31G, 32G | 6mm,8mm,12mm,<br>25mm,38mm    | Regular wall (RW) |
| Type F       | 3mL,5mL,10mL |                                                               |                               | Thin wall (TW)    |

#### 6 Indication

lt is intended to inject fluids into the body or withdraw blood for medical purposes. After injection, the needle tube will be retracted into the barrel when the safety features are manually activated to minimize the risk of accidental sharps injury.

{5}------------------------------------------------

### 7 Similarities and Differences of the Proposed Devices to the Predicate Devices

The VeriSafe Safety sterile syringe is substantially equivalent to the predicate device, Verifine Safety Syringe with Fixed Needle (K213560) in that these devices have same intended use and technological characteristics. Using plunger to fill syringe as well as discharge the fluid is the technological principle of both the subject device and predicate device. The needle of both devices is made of stainless steel (X5CrNi18-10). The performance of both devices fulfills the requirement of ISO 7886-1, ISO 9626, ISO 7864 and ISO 23908. Both the subject and predicate devices are disposable, EO sterilized, single patient use devices. Both the subject and predicate device are used in Professional healthcare facility and transport environment.

|                                            | Subject Device                                                                                                                                                                                                                                                                     | Predicate Device (K213560)                                                                                                                                                                                                   |                                 |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Trade Name                                 | VeriSafe Safety Sterile Syringes                                                                                                                                                                                                                                                   | Verifine Safety Syringe<br>with Fixed Needle                                                                                                                                                                                 |                                 |
| Manufacturer                               | Promisemed Hangzhou<br>Meditech Co., Ltd                                                                                                                                                                                                                                           | Promisemed Hangzhou<br>Meditech Co., Ltd                                                                                                                                                                                     |                                 |
| Device Class                               | Class II                                                                                                                                                                                                                                                                           | Class II                                                                                                                                                                                                                     | Same                            |
| Product Code                               | FMF, MEG                                                                                                                                                                                                                                                                           | FMF, MEG                                                                                                                                                                                                                     | Same                            |
| Regulation number                          | 880.5860                                                                                                                                                                                                                                                                           | 880.5860                                                                                                                                                                                                                     | Same                            |
| Regulation Name                            | Piston syringe                                                                                                                                                                                                                                                                     | Piston syringe                                                                                                                                                                                                               | Same                            |
| Intended Use/<br>Indications for Use       | It is intended to inject fluids<br>into the body or withdraw<br>blood for medical purposes.<br>After injection, the needle<br>tube will be retracted into the<br>barrel when the safety<br>features are manually<br>activated to minimize the risk<br>of accidental sharps injury. | It is intended to inject<br>fluids into or withdraw<br>fluids from the body for<br>medical purposes.                                                                                                                         | Same                            |
| Operating<br>Principle                     | Plunger is used to fill syringe<br>as well as discharge the fluid.<br>After use, the needle tube will<br>be retracted into the barrel by<br>pressing the push-button to<br>depress the plunger all of the<br>way within the barrel.                                                | Plunger is used to fill<br>syringe as well as discharge<br>the fluid.<br>The protective shield will<br>permanently be locked in<br>place by pulling forward<br>till click, providing<br>protection against needle<br>sticks. | Different-<br>See Comment #1    |
| Safety mechanism                           | The safety mechanism is<br>activated by pressing the push-<br>button down to depress the<br>plunger all of the way within                                                                                                                                                          | The safety mechanism is<br>activated by pushing the<br>sliding sleeves. When                                                                                                                                                 | Different<br>See Comment#2      |
|                                            | the barrel. When activated,<br>needle is retracted into the<br>barrel.                                                                                                                                                                                                             | activate, needle is covered<br>by sliding sleeves.                                                                                                                                                                           |                                 |
| Volume                                     | 1mL,2mL,3mL,5mL,10mL                                                                                                                                                                                                                                                               | 0.5ml,1.0ml                                                                                                                                                                                                                  | Different                       |
|                                            |                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                              | See Comment #3                  |
| Tip type                                   | Fixed Needle and detachable<br>needle                                                                                                                                                                                                                                              | Fixed Needle                                                                                                                                                                                                                 | Different                       |
|                                            |                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                              | See Comment #4                  |
| Needle and<br>syringe tip<br>configuration | Type F<br>Image: Type F syringe<br>Type D<br>Image: Type D syringe                                                                                                                                                                                                                 | Image: Syringe                                                                                                                                                                                                               | Different                       |
|                                            |                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                              | See Comment #5                  |
| Gauge                                      | 18G, 20G, 22G, 23G, 25G,<br>27G,28G, 29G, 30G, 31G, 32G                                                                                                                                                                                                                            | 21G, 22G, 23G, 24G, 25G,<br>26G, 27G, 28G                                                                                                                                                                                    | Different                       |
|                                            |                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                              | See Comment #6                  |
| Needle Length                              | 6mm,8mm,12mm,<br>25mm,38mm                                                                                                                                                                                                                                                         | 12mm, 16mm, 25 mm,<br>32mm, 38mm                                                                                                                                                                                             | Different                       |
|                                            |                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                              | See Comment #7                  |
| Gradations<br>legibility                   | Legible                                                                                                                                                                                                                                                                            | Legible                                                                                                                                                                                                                      | Same                            |
| Lubricant                                  | Silicone oil                                                                                                                                                                                                                                                                       | Silicone oil                                                                                                                                                                                                                 | Same                            |
| Barrel<br>transparency                     | Transparent                                                                                                                                                                                                                                                                        | Transparent                                                                                                                                                                                                                  | Same                            |
| Reuse durability                           | Single Use                                                                                                                                                                                                                                                                         | Single Use                                                                                                                                                                                                                   | Same                            |
| Biocompatibility                           | Complied with ISO10993 series<br>standards, and the following<br>tests are performed                                                                                                                                                                                               | Complied with ISO10993<br>series standards, and the<br>following tests are<br>performed                                                                                                                                      | Same                            |
|                                            | - Cytotoxicity: No cytotoxicity                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                              |                                 |
|                                            | - Skin Irritation: No evidence<br>of skin irritation                                                                                                                                                                                                                               | - Cytotoxicity: No<br>cytotoxicity                                                                                                                                                                                           |                                 |
|                                            |                                                                                                                                                                                                                                                                                    | - Skin Irritation: No<br>evidence of skin irritation                                                                                                                                                                         |                                 |
|                                            | - Skin Sensitization: No<br>evidence of sensitization                                                                                                                                                                                                                              | - Skin Sensitization: No<br>evidence of sensitization                                                                                                                                                                        |                                 |
|                                            | -Acute Systemic Toxicity: No<br>systemic toxicity                                                                                                                                                                                                                                  | -Acute Systemic Toxicity:<br>No systemic toxicity                                                                                                                                                                            |                                 |
|                                            | -Hemolysis: No evidence of<br>hemolysis                                                                                                                                                                                                                                            | -Hemolysis: No evidence of<br>hemolysis                                                                                                                                                                                      |                                 |
|                                            | -Pyrogen: Non pyrogenic                                                                                                                                                                                                                                                            | -Pyrogen: Non pyrogenic                                                                                                                                                                                                      |                                 |
|                                            | -USP <788> Particulate matter                                                                                                                                                                                                                                                      | -USP <788> Particulate<br>matter                                                                                                                                                                                             |                                 |
| Configuration and<br>Materials             | Needle, spring: Stainless Steel<br>(X5CrNi18-10)                                                                                                                                                                                                                                   | Needle: Stainless Steel<br>(X5CrNi18-10)                                                                                                                                                                                     | Similar                         |
|                                            | Barrel/push-button:<br>Polypropylene                                                                                                                                                                                                                                               | Barrel/push-button:<br>Polypropylene                                                                                                                                                                                         | See Comment# 8                  |
|                                            | Plunger: Polypropylene                                                                                                                                                                                                                                                             | Plunger: Polypropylene                                                                                                                                                                                                       |                                 |
|                                            | Needle cap: Polypropylene                                                                                                                                                                                                                                                          | Needle cap: Polyethylene                                                                                                                                                                                                     |                                 |
|                                            | Piston: Nitrile butadiene<br>rubber/Thermoplastic<br>elastomer                                                                                                                                                                                                                     | Piston: Polyisoprene rubber                                                                                                                                                                                                  |                                 |
|                                            |                                                                                                                                                                                                                                                                                    | Sliding Sleeve, Cover:<br>Polypropylene                                                                                                                                                                                      |                                 |
|                                            | Retraction cylinder sealing<br>ring: Silicone rubber                                                                                                                                                                                                                               | Plunger stopper, Tight ring:<br>Acrylonitrile Butadiene<br>Styrene                                                                                                                                                           |                                 |
|                                            | Needle hub: Polypropylene                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                              |                                 |
|                                            | Needle hub sealing ring:<br>Thermoplastic elastomer                                                                                                                                                                                                                                |                                                                                                                                                                                                                              |                                 |
|                                            | Retraction cylinder:<br>Acrylonitrile Butadiene Styrene                                                                                                                                                                                                                            |                                                                                                                                                                                                                              |                                 |
|                                            |                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                              |                                 |
|                                            |                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                              |                                 |
| Performance                                | Complied with ISO 7886-1, ISO<br>9626, ISO 7864, ISO 23908                                                                                                                                                                                                                         | Complied with ISO 7886-1,<br>ISO 9626, ISO 7864, ISO<br>23908                                                                                                                                                                | Same                            |
| Sterilization<br>method and SAL            | EO Sterilization<br>SAL = 10 $-6$                                                                                                                                                                                                                                                  | EO Sterilization<br>SAL = 10 $-6$                                                                                                                                                                                            | Same                            |
| Environment of<br>use                      | Professional healthcare facility<br>and transport environment.                                                                                                                                                                                                                     | Professional healthcare<br>facility and transport<br>environment.                                                                                                                                                            | Same                            |
| shelf life                                 | 3 years                                                                                                                                                                                                                                                                            | 5 years                                                                                                                                                                                                                      | Different $9$<br>See Comment# 9 |

A detailed comparison to the predicate is provided in Table 1.

{6}------------------------------------------------

{7}------------------------------------------------

## Discussions of differences in technological characteristics

Comment #1 and #2 : The subject device has different safety mechanisms to protect against sharps injury. The sharps protection feature of both devices are addressed through testing per ISO 23908. The difference in sharps protection feature does not affect the effectiveness and safety of the device.

{8}------------------------------------------------

Comment #3 : The subject device has a different capacity range. The specification of capacity of both devices fulfils the requirement of ISO 7886-1. The difference in capacity does not affect the effectiveness and safety of the device.

Comment # 4 : The subject device has an additional type of detachable needle tip. The performance of detachable needle type fulfils the requirement of ISO 9626:2016, ISO 7864:2016 and ISO 7886-1. The difference in tip type does not affect the effectiveness and safety of the device.

Comment # 5: Needle and syringe tip configuration of the subject device is different from that of predicate device. The performance of the subject device fulfill the requirements of ISO 9626:2016, ISO 7886-1:2016, ISO 7864:2016, ISO 23908:2011. The difference in needle and syringe tip configuration does not affect the effectiveness and safety of the device.

Comment #6 and #7 : The gauge and needle length of the subject device is different from the predicate device. The specification of gauge and needle length of both devices fulfil the requirement of ISO 9626:2016 and ISO 7864:2016. The difference in gauge and needle length does not affect the effectiveness and safety of the device.

Comment #8: The syringe bodies of subject device and predicate device have material differences. The biocompatibility tests conducted demonstrated the subject device. The difference in materials does not affect the effectiveness and safety of the device.

Comment # 9 : The subject device has shorter shelf life. The performance of aged subject device has been validated in accordance with ASTM F1980-16, ISO 9626:2016, ISO 7886-1:2016, ISO 7864:2016, ISO 23908:2011. The difference in shelf life does not affect the effectiveness and safety of the device.

The differences above between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question 

{9}------------------------------------------------

is raised regarding effectiveness and safety. The subject device is substantially equivalent to the identified predicate device.

#### 8 Performance Testing Summary

The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:

- ISO 7886-1:2017, Sterile hypodermic syringes for single use Part 1: Syringes for manual use
	- Extraneous matter, Limits for acidity or alkalinity, Limits for extractable metals, Lubrication, Tolerance on graduated capacity, Graduated Scale, Numbering of Scales, Overall length of scale to Nominal capacity line, Position of scale, Barrel, Plunger stopper/ plunger assembly, Nozzle, Dead space, Freedom from air and liquid leakage past plunger stopper, Force to operate the piston, Fit of the plunger stopper/plunger in barrel.
- · ISO 9626:2016, Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods
- Stiffness, Resistance to breakage, Resistance to corrosion
- · ISO 7864:2016, Sterile hypodermic needles for single use Requirements and test methods
- . Cleanliness, Colour coding, Needle tube-Outside diameter, Tolerance on length,Freedom from defects, Needle points, Fragmentation test, Penetration force and drag force, Bond between hub and needle tube, Patency od lumen
- ISO 23908:2011, Sharps injury protection Requirements and test methods Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
	- Safety mechanism activation. Safety overriding and unlocking force after activation, Simulated clinical Use
- USP<85> Bacterial Endotoxin Test

{10}------------------------------------------------

## Biocompatibility

In accordance with ISO 10993-1 the device is classified as Externalcommunicating device, Blood path, indirect, with limited contactduration (≤ 24h). The biosafety tests that required consideration were validated:

- ISO 10993-5:2009 Biological Evaluation of Medical Devices -- Part 5:Tests for in Vitro Cytotoxicity
- ISO 10993-10:2021 Biological Evaluation of Medical Devices -- Part 10: Tests for Skin Sensitization
- ISO 10993-23:2021 Biological Evaluation of Medical Devices -Part 23: Test for irritation
- ISO 10993-11:2017- Biological evaluation of medical devices Part 11: Tests for systemic toxicity-Acute systemic toxicity
- ASTM F756-17- Standard practice for assesment of hemolytic properties of materials
- ISO 10993-7:2008- Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals
- USP42-NF37<151> Pyrogen Test
- USP<788> Particulate matter in injections

## Sterility, Shipping, and Shelf-Life

- ISO 11135:2014, Sterilization of health-care products Ethylene oxide Requirements for the development, validation and routine control of a sterilization process for medical devices
- Package integrity testing, after environmental conditioning and simulated transportation in accordance with ASTM D4169-16, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance.
- · Sterile Barrier Packaging Testing performed on the proposed device:
	- Seal strength ASTM F88/F88-15
	- Dye penetration ASTM F1929-15
	- Sterility test USP38-NF33_C71
- Shelf life of 3 years is validated using the FDA recognized standard ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- の Conclusion

Based on the information provided within this 510(k) submission, the differences between the predicate and the subject device do not raise any new or different questions of safety

{11}------------------------------------------------

or effectiveness. The proposed subject device is substantially equivalent to the predicate device and is as safe and as effective as the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K231792](https://fda.innolitics.com/device/K231792)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
