← Product Code [OHT](/productcode/OHT) · K231717

# IPL Home Use Hair Removal Device, Model(s): JD-TM002, JD-TM003, JD-TM012, JD-TM016, JD-TM022 (K231717)

_Foshan Jindi Electric Appliance Co., Ltd. · OHT · Sep 6, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K231717

## Device Facts

- **Applicant:** Foshan Jindi Electric Appliance Co., Ltd.
- **Product Code:** [OHT](/productcode/OHT.md)
- **Decision Date:** Sep 6, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

IPL Home Use Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

## Device Story

IPL Home Use Hair Removal Device is an over-the-counter, single-person-use device for hair reduction. It uses Intense Pulsed Light (IPL) generated by a Xenon arc flashlamp to target hair follicles below the skin surface without cutting or pulling. The device is powered by an external adapter and operated via a finger switch. It includes a skin sensor to ensure full contact before light pulse emission. Four models (JD-TM003, JD-TM012, JD-TM016, JD-TM022) feature a cooling function to reduce burning sensations, and four models (JD-TM003, JD-TM012, JD-TM016, JD-TM022) include an auxiliary head for small areas like lips or underarms. Users operate the device at home to treat various body areas. The output is intended to reduce hair regrowth over time, providing a benefit of long-term hair reduction with minimal pain.

## Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes biocompatibility testing (ISO 10993-5, ISO 10993-10), electrical safety and EMC testing (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57), photobiological eye safety (IEC 62471), software verification and validation, and usability engineering per FDA guidance.

## Technological Characteristics

Intense Pulsed Light (IPL) device using a Xenon arc flashlamp. Enclosure and flash window made of ABS and PC. Powered by external 100-240V AC adapter. Features include skin contact sensor, cooling function (select models), and auxiliary heads (select models). Wavelength range 550-1100nm or 600-1100nm depending on model. Energy density 2-6 J/cm². Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-57, and IEC 62471.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- iPulse Smoothskin Gold Hair Removal System ([K160968](/device/K160968.md))
- IPL Cooling Hair Removal Device (NBB01, NBB02) ([K230360](/device/K230360.md))

## Reference Devices

- IPL Hair Removal Device, Model(s): AP10, AP20, AP30, AP32 ([K230021](/device/K230021.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 6, 2023

Foshan Jindi Electric Appliance Co.,Ltd % Yvonne Liu Registration engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90, Qianhai Road Shenzhen, Guangdong 518052 China

Re: K231717

Trade/Device Name: IPL Home Use Hair Removal Device, Model(s): JD-TM002, JD-TM003, JD-TM012, JD-TM016, JD-TM022 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: June 12, 2023 Received: June 13, 2023

Dear Yvonne Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) /K231717

Device Name IPL Home Use Hair Removal Device Model(s): JD-TM002, JD-TM003, JD-TM012, JD-TM016, JD-TM022

#### Indications for Use (Describe)

IPL Home Use Hair Removal Device is indicated for the removal of unwanted harr. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |
|                                                 |  |

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K231717

# "510(k) Summary" as required by 21 CFR Part 807.92.

#### I. Submitter

Foshan Jindi Electric Appliance Co.,Ltd

No.13 Baiye Avenue, Xiqiao Science and Technology Industrial Park, Xiqiao Town, Nanhai District, Foshan City, Guangdong Province, China Post code: 528211 Tel.: +86 18316578883 Zhaozhi Li Legal Person Tel: +86 18316578883 Email: renyigea@126.com

#### II. Device

Name of Device: IPL Home Use Hair Removal Device Model(s): JD-TM002, JD-TM003, JD-TM012, JD-TM016, JD-TM022 Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810

#### III. Predicate Device

#### Primary predicate device:

| Manufacturer   | Predicate Device                              | 510(k) Number | Approval Date  |
|----------------|-----------------------------------------------|---------------|----------------|
| CyDen Limited. | iPulse Smoothskin Gold Hair<br>Removal System | K160968       | April 14, 2016 |

Secondary predicate device:

| Manufacturer                             | Reference Device                                  | 510(k) Number | Approval Date  |
|------------------------------------------|---------------------------------------------------|---------------|----------------|
| Shenzhen Jizhimei<br>Technology Co., Ltd | IPL Cooling Hair Removal<br>Device (NBB01, NBB02) | K230360       | April 10, 2023 |

Reference device:

| Manufacturer                           | Reference Device                                          | 510(k) Number | Approval Date  |
|----------------------------------------|-----------------------------------------------------------|---------------|----------------|
| Shenzhen Yangyi<br>Technology Co., Ltd | IPL Hair Removal Device,<br>Model(s): AP10,AP20,AP30,AP32 | K230021       | March 31, 2023 |

#### IV. Device Description

{4}------------------------------------------------

IPL Home Use Hair Removal Device (Models:JD-TM003, JD-TM003, JD-TM012, JD-TM016, JD-TM022),is an over-the-counter, home-use and single-person-use device for hair reduction by using Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.

The device is only powered by the external power adapter and its IPL emission activation is by finger switch. This product adopts irreplaceable flash window and is suitable for multiple hair removal areas (such as: lips, underarms, bikini lines, arms, legs, etc.). It contains a skin sensor to detect appropriate skin contact, if the device is not in full contact with the skin, the device cannot emit the treatment light pulses.

The subject device includes five models. The three models of JD-TM002,JD-TM003,JD-TM012 are white , JD-TM016 has two colors of white and purple while three colors of rose gold, green and champagne for JD-TM022. Except JD-TM022, the other four models have cooling function to relieve pain of burning sensation during operation. Except JD-TM002, you can use auxiliary head to limit the spot area and get better effect with the other four models. The auxiliary head is suitable for small parts such as underarms and lips.

IPL Home Use Hair Removal Device, models: JD-TM002, JD-TM003, JD-TM012, JD-TM016, JD-TM022 have the same indication for use, performance, structure design and operation. The main difference mainly contains product appearance, size, weight, energy density , number of levels and auxiliary head. The little difference won't affect the safety and effectiveness of the device.

#### V. Indications for Use

IPL Home Use Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

## VI. Materials

| Component<br>name                                                         | Material of Component | Body Contact Category                     | Contact Duration   |
|---------------------------------------------------------------------------|-----------------------|-------------------------------------------|--------------------|
| IPL Home Use<br>Hair Removal<br>Device<br>(Enclosure and<br>flash window) | ABS, PC               | Surface-contacting<br>device: Intact skin | Less than 24 hours |

We have tested the device for biocompatibility by a reliable third-party lab. For details, please refer to Section 16 "Biocompatibility Discussion".

## VII. Comparison of Technological Characteristics With the Predicate Device

The IPL Home Use Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices and reference device do no raise any issues of safety or efficacy. Performance 

{5}------------------------------------------------

data supports that the device is safe and as effective as the predicate devices and reference device for its intended use. Therefore, the IPL Home Use Hair Removal Device may be found substantially equivalent to its predicate devices and reference device.

IPL Home Use Hair Removal Device is compared with the following Predicate Device and Reference Devices in terms of intended use, design, material,specifications, and performance:

| Comparison<br>Elements                                | Subject Device                                                                                                                                                                                                                                                                                                                                                   | Primary Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                  | Secondary Predicate<br>Device                                                                                                | Reference Device                                                                                                 | Remark              |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------|
| 510(k) Number                                         | Pending                                                                                                                                                                                                                                                                                                                                                          | K160968                                                                                                                                                                                                                                                                                                                                                                      | K230360                                                                                                                      | K230021                                                                                                          | /                   |
| Trade name                                            | IPL Home Use Hair<br>Removal Device (JD-<br>TM002,JD-TM003,JD-<br>TM012,JD-TM016,JD-<br>TM022)                                                                                                                                                                                                                                                                   | iPulse Smoothskin Gold Hair Removal System                                                                                                                                                                                                                                                                                                                                   | IPL Cooling Hair<br>Removal Device<br>( NBB01, NBB02 )                                                                       | IPL Hair Removal<br>Device,<br>Model(s):AP10,AP20<br>,AP30,AP32                                                  | /                   |
| Manufacturer                                          | Foshan Jindi Electric<br>Appliance Co.,Ltd                                                                                                                                                                                                                                                                                                                       | CyDen Limited.                                                                                                                                                                                                                                                                                                                                                               | Shenzhen<br>Jizhimei<br>Technology Co.,Ltd                                                                                   | Shenzhen<br>Yangyi<br>Technology Co.,Ltd                                                                         |                     |
| Regulation<br>number                                  | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                  | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                              | 21 CFR 878.4810                                                                                                              | 21 CFR 878.4810                                                                                                  | Same                |
| Product code                                          | OHT                                                                                                                                                                                                                                                                                                                                                              | OHT, GEX                                                                                                                                                                                                                                                                                                                                                                     | OHT                                                                                                                          | OHT                                                                                                              | Same                |
| Device<br>classification                              | Class II                                                                                                                                                                                                                                                                                                                                                         | Class II                                                                                                                                                                                                                                                                                                                                                                     | Class II                                                                                                                     | Class II                                                                                                         | Same                |
| Indication<br>for<br>use/<br>Intended<br>use          | IPL Home Use Hair<br>Removal Device is<br>indicated for the<br>removal of unwanted<br>hair. The device is also<br>indicated for the<br>permanent reduction in<br>hair regrowth, defined<br>as the long-term, stable<br>reduction in the number<br>of hairs regrowing<br>when measured at 6, 9<br>and 12 months after the<br>completion of a<br>treatment regime. | The<br>iPulse<br>Smoothskin<br>Gold Hair<br>Removal<br>System is indicated<br>for the removal of<br>unwanted hair. The<br>iPulse Smoothskin<br>Gold is also<br>indicated for the<br>permanent reduction<br>in hair regrowth,<br>defined as the long-<br>term, stable<br>reduction in the<br>number of hairs<br>regrowing when<br>measured at 6, 9 and<br>12 months after the | The IPL Cooling Hair Removal Device is an over-the-counter device intended for removal of unwanted body and /or facial hair. | IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and /or facial hair. | Same                |
|                                                       |                                                                                                                                                                                                                                                                                                                                                                  | completion<br>of a<br>treatment regime.                                                                                                                                                                                                                                                                                                                                      |                                                                                                                              |                                                                                                                  |                     |
| Prescription or<br>OTC                                | OTC                                                                                                                                                                                                                                                                                                                                                              | OTC                                                                                                                                                                                                                                                                                                                                                                          | OTC                                                                                                                          | OTC                                                                                                              | Same                |
| Device design                                         |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                              |                                                                                                                  |                     |
| Source energy                                         | Supplied by external<br>adapter                                                                                                                                                                                                                                                                                                                                  | Supplied by external<br>adapter                                                                                                                                                                                                                                                                                                                                              | Supplied by external<br>adapter                                                                                              | Supplied by external<br>adapter                                                                                  | Same                |
| Power Supply                                          | 100-240V~, 50/60Hz                                                                                                                                                                                                                                                                                                                                               | 110V or 230V,<br>50/60Hz                                                                                                                                                                                                                                                                                                                                                     | 100-240V, 50/60Hz                                                                                                            | 100-240V AC                                                                                                      | Same                |
| Dimension                                             | JD-TM002:<br>138*78*47mm<br>JD-TM003:<br>145*75*50mm<br>JD-TM012:<br>195*73*47mm<br>JD-TM016:<br>180*100*47mm<br>JD-TM022:<br>176*147*91mm                                                                                                                                                                                                                       | Unknown                                                                                                                                                                                                                                                                                                                                                                      | NBB01:<br>47.6*54.3*240mm<br>NBB02-MAX:<br>235*76*43mm                                                                       | AP10:<br>201*184*80mm<br>AP20:<br>209*161*78mm<br>AP30:<br>166*62*40mm<br>AP32: 167*63*41mm                      | Different<br>Note 1 |
| Sterilization                                         | Not required                                                                                                                                                                                                                                                                                                                                                     | Not required                                                                                                                                                                                                                                                                                                                                                                 | No                                                                                                                           | Not required                                                                                                     | Same                |
| Output specification                                  |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                              |                                                                                                                  |                     |
| Light Source                                          | Intense Pulsed Light                                                                                                                                                                                                                                                                                                                                             | Intense Pulsed Light                                                                                                                                                                                                                                                                                                                                                         | Intense Pulsed Light                                                                                                         | Intense Pulsed Light                                                                                             | Same                |
| Energy<br>medium                                      | Xenon Arc Flashlamp                                                                                                                                                                                                                                                                                                                                              | Xenon Arc<br>Flashlamp                                                                                                                                                                                                                                                                                                                                                       | Xenon Arc Flashlamp                                                                                                          | Xenon Arc Flashlamp                                                                                              | Same                |
| Wavelength<br>range                                   | JD-TM002,JD-TM003:<br>550-1100nm<br>JD-TM012,JD-TM016:<br>600-1100nm<br>JD-TM022:<br>640-1100nm                                                                                                                                                                                                                                                                  | 510-1100nm                                                                                                                                                                                                                                                                                                                                                                   | 550-1200mm                                                                                                                   | 530-1100nm                                                                                                       | Same                |
| Spot Size<br>( cm2 )                                  | JD-TM002,JD-<br>TM003,JD-TM012:<br>3 cm2<br>JD-TM016:<br>3.3 cm2<br>JD-TM022: 4 cm2                                                                                                                                                                                                                                                                              | 3cm2                                                                                                                                                                                                                                                                                                                                                                         | 3.6cm2                                                                                                                       | 3.3cm2<br>3.96cm2 ,3.63cm2                                                                                       | Similar<br>Note 2   |
| Energy<br>Density                                     | JD-TM002,JD-<br>TM003,JD-TM012,<br>JD-TM016: $2-5 J/cm^2$<br>JD-TM022:<br>$2-6 J/cm^2$                                                                                                                                                                                                                                                                           | $3-6 J/cm^2$                                                                                                                                                                                                                                                                                                                                                                 | $2-5 J/cm^2$ (applicable<br>for model NBB01)<br>$2-4 J/cm^2$ (applicable<br>for model NBB02)                                 | Max $4.3 J/cm^2$                                                                                                 | Similar<br>Note 3   |
| Pulse duration                                        | 8-12ms                                                                                                                                                                                                                                                                                                                                                           | 2-10 ms                                                                                                                                                                                                                                                                                                                                                                      | 6.4-7.2ms                                                                                                                    | 8.8-13.2 ms                                                                                                      | Similar<br>Note 4   |
| Pulsing control                                       | Finger switch                                                                                                                                                                                                                                                                                                                                                    | Finger switch                                                                                                                                                                                                                                                                                                                                                                | Finger switch                                                                                                                | Finger switch                                                                                                    | Same                |
| Output<br>intensity level                             | JD-TM002,JD-<br>TM003,JD-TM012:<br>5 Levels<br>JD-TM016: 9 Levels<br>JD-TM022: 10 Levels                                                                                                                                                                                                                                                                         | Unknown                                                                                                                                                                                                                                                                                                                                                                      | NB001: 5 Levels<br>NBB02: 6 Levels                                                                                           | 5 (applicable for<br>model AP10,<br>AP30,AP32)<br>9 (applicable for<br>model AP20)                               | Similar<br>Note 3   |
| Delivery<br>Device                                    | Direct Illumination to<br>Tissue                                                                                                                                                                                                                                                                                                                                 | Direct Illumination<br>to Tissue                                                                                                                                                                                                                                                                                                                                             | Direct Illumination to<br>Tissue                                                                                             | Direct Illumination to<br>Tissue                                                                                 | Same                |
| Software/<br>Firmware/<br>Microprocesso<br>r Control? | Yes                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                          | Yes                                                                                                              | Same                |
| Additional features                                   |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                              |                                                                                                                  |                     |
| Electrical<br>safety                                  | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57                                                                                                                                                                                                                                                                                                 | IEC 60601-1<br>IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-83                                                             | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-83                                                 | Similar             |
| Eye safety                                            | IEC 62471                                                                                                                                                                                                                                                                                                                                                        | IEC 62471                                                                                                                                                                                                                                                                                                                                                                    | IEC 62471                                                                                                                    | IEC 62471                                                                                                        | Same                |
| Biocompatibili<br>ty                                  | ISO 10993-5<br>ISO 10993-10                                                                                                                                                                                                                                                                                                                                      | ISO 10993-5<br>ISO 10993-10<br>ISO 10993-12                                                                                                                                                                                                                                                                                                                                  | ISO 10993-5<br>ISO 10993-10                                                                                                  | ISO 10993-5<br>ISO 10993-10                                                                                      | Same                |

{6}------------------------------------------------

{7}------------------------------------------------

Note 1:

Though the dimensions are different from the predicate device, this difference is insignificant and do not raise any safety/ effectiveness problems.

## Note 2:

There is minor difference in spot size between the subject device and the predicate devices and reference device. The spot size is related to light intensity and since the difference in light intensity is not significant as explained in note 3, so this difference will not raise any safety or effectiveness issue.

{8}------------------------------------------------

#### Note 3:

Though the energy density and output intensity level are a little different from the predicate devices and reference device, the energy density of subject device is within the range of the minimum and maximum value of the predicate devices and reference device, and they all comply with IEC 60601-2-57 and IEC 62471 requirements, so this difference will not raise any safety or effectiveness issue.

#### Note 4:

Though the pulse duration of subject device is a little different from the predicate devices and reference device, and the subject device complies with IEC 60601-2-57 and IEC 62471 requirements, so this difference will not raise any safety or effectiveness issue.

#### VIII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

#### 1) Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of the IPL Home Use Hair Removal Device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as recognized by FDA. The following testing was performed to, and passed, including:

- > ISO 10993-5:2009, Biological Evaluation of Medical Devices -Par t 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10:2010, Biological Evaluation of Medical Devices –Par t 10: Tests for Irritation and Skin Sensitization

#### 2) Electrical Safety and EMC

Electrical safety and EMC testing was performed to, and passed, as per the following standards:

- > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility
- > ANSI AAMI ES60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- > IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- > IEC 60601-2-57 Medical electrical equipment Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic,monitoring and cosmetic/aesthetic use.

#### 3) Eve Safetv

- > IEC 62471 Photobiological safety of lamps and lamp systems
#### 4) Software Verification and Validation

{9}------------------------------------------------

Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

# 5) Usability

The product usability has been evaluated and verified according to the following FDA guidance > Applying Human Factors and Usability Engineering to Medical Devices, issued on FEBRUARY 2016

#### IX. Conclusions

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device IPL Home Use Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate device and reference devices.

---

**Source:** [https://fda.innolitics.com/device/K231717](https://fda.innolitics.com/device/K231717)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
