K231677 · Edgecare, Inc. · IYO · Mar 6, 2024 · Radiology
Device Facts
Record ID
K231677
Device Name
EdgeFlow UH10
Applicant
Edgecare, Inc.
Product Code
IYO · Radiology
Decision Date
Mar 6, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
AI/ML, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Bladder presence classification
MobileNetV2 feature extraction and fully connected neural network
F1 score ≥ 0.90; PR AUC ≥ 0.95
F1 score 0.979 (95% CI 0.974-0.984)
Training Data: 9,422 images; Test Data: 3,711 images
>1 (evaluators)
—
—
Bladder segmentation
DeepLabv3+ semantic segmentation network
Dice Score ≥ 0.89
Dice score 0.896 (95% CI 0.890-0.901)
Training Data: 7,115 images; Test Data: 1,528 images
>1 (evaluators)
—
—
Indications for Use
The EdgeFlow UH10 is an ultrasound device intended to be used for measuring the urine volume in the bladder noninvasively. It is intended for use in professional healthcare facilities, such as hospitals, clinics, by qualified and trained healthcare professionals. The EdgeFlow UH10 supports B-mode and harmonic imaging modes.
Device Story
Handheld ultrasound system with integrated probe and 3.2-inch touchscreen; used by healthcare professionals in clinical settings. Inputs: live B-mode ultrasound images. Processing: deep learning model (MobileNetV2 feature extraction, DeepLabv3+ semantic segmentation, binary classification) identifies bladder presence and delineates bladder area. Output: bladder contours (green lines) and calculated volume displayed on screen. Data management: results saved on-device; exportable via USB or Wi-Fi. Benefits: noninvasive, automated bladder volume quantification to support clinical decision-making.
Clinical Evidence
Bench testing only. Algorithm performance validated using 90-subject clinical trial dataset (IRB No: 1-2022-0076). Classification network (3,711 test images) achieved F1 score 0.979 (95% CI 0.974-0.984) and PR AUC ≥0.95. Segmentation network (1,528 test images) achieved Dice score 0.896 (95% CI 0.890-0.901). Biocompatibility per ISO 10993-1, -5, -10, -23. Electrical safety per IEC 60601-1 and IEC 60601-1-2.
Indicated for noninvasive measurement of urine volume in the bladder for male, female, and pediatric patients in professional healthcare settings. Contraindicated for fetal use, pregnant patients, patients with ascites, or patients with open skin/wounds in the suprapubic region. Not for use by those allergic to coupling agents or with skin disease.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
March 6, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue.
Edgecare Inc. % Milly Regulatory Affairs Consultant KMC, Inc. Room no. 1709, 123, Digital-ro 26-gil, Guro-gu Seoul. 08390 SOUTH KOREA
#### Re: K231677
Trade/Device Name: EdgeFlow UH10 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO, ITX, QIH Dated: February 6, 2024 Received: February 7, 2024
#### Dear Milly:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K231677
Device Name EdgeFlow UH10
#### Indications for Use (Describe)
The EdgeFlow UH10 is an ultrasound device intended to be used for measuring the urine volume in the bladder noninvasively. It is intended for use in professional healthcare facilities, such as hospitals, clinics, by qualified and trained healthcare professionals. The EdgeFlow UH10 supports B-mode and harmonic imaging modes.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY (K231677)
# K231677
This summary of 510(k) -safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Feb 29, 2024
#### 1. INFORMATION
1.1 Submitter Information
- I Submitter Name: Edgecare Inc.
- . Address : 403, Teihard Hall, 35, Baekbeom-ro, Mapo-gu, Seoul, 04107, Republic of Korea
- 트 Telephone Number: +82-70-4290-9046
- . Fax: +82-504-845-0217
#### 1.2 Official Correspondent
- . Name: Milly (Consultant / KMC, Inc.)
- 트 Address: Room no. 1709, 123, Digital-ro 26-gil, Guro-gu, Seoul, 08390, Republic of Korea
- I Telephone Number: +82-70-8965-5554
- . Fax: +82-2-2672-0579
- . E-mail: milly@kmcerti.com
#### 2. DEVICE INFORMATION
- 2.1 Trade Name / Proprietary Name
- : EdgeFlow UH10
- 2.2 Common Name: Ultrasonic Pulsed Echo Imaging System
- 2.3 Classification Name: Ultrasonic pulsed echo imaging system
- 2.4 Product Code: IYO, ITX, QIH
- 2.5 Classification Regulation: 21CFR 892.1560, 21CFR892.1570, 21CFR892.2050
- 2.6 Device Class: Class II
- 2.7 Classification Panel: Radiology
{5}------------------------------------------------
#### 3. PREDICATE DEVICE
- l Primary Predicate Device: BladderScan® PRIME PLUS system (K172356)
- . Reference Device: Butterfly Auto 3D Bladder Volume Tool (K200980)
#### 4. SUBJECT DEVICE DESCRIPTION
This device is an ultrasound system that measures urinary bladder volume noninvasively using ultrasound. The system consists of a handheld unit (console) with a 3.2inch touchscreen display and a permanently attached probe. The system calculates the bladder volume based on ultrasound images using the probe. The images and the bladder volume are displayed on the screen. The results of the exam are automatically saved and saved exams are managed on the screen. Data can be exported to a standard PC using a wired USB export or Wi-Fi communications.
The system includes a rechargeable lithium-ion battery can be charged in the system directly connecting a USB cable with a power adapter. Also, a cradle is available for charging the battery or printing the results.
The deep learning model employed in EdgeFlow UH10 comprises three components: feature extraction, binary classification, and semantic segmentation networks. B-mode ultrasound images undergo classification in the network to determine the presence of the bladder, while the segmentation network is responsible for delineating the bladder area. Live B-mode ultrasound images are acquired once the scan button is pressed to start scanning. The output of the deep learning model manifests as the bladder contours displayed as green lines in the ultrasound images, and the bladder volume is subsequently calculated based on these lines when the scan button is pressed.
Image /page/4/Figure/8 description: The image shows a diagram of a bladder segmentation process. The process starts with an input B-mode image, which undergoes preprocessing. The preprocessed image is then fed into a feature extraction module using MobileNetV2, followed by a semantic segmentation network using DeepLabv3+. The outputs of the semantic segmentation network and a binary classification network (fully connected neural network) are combined in a postprocessing step to produce the final output: bladder segmentation.
{6}------------------------------------------------
All data used for training and test dataset were collected under a clinical trial approved by Institutional Review Board with training data being independent from the test data.
- Sample Size -
| Item | Training Data | Test Data |
|------------------------|---------------|-----------|
| Classification Network | 9,422 | 3,711 |
| Segmentation Network | 7,115 | 1,528 |
Bladder volume, which is the measurement output of the subject device, is reflected by the size and shape of the bladder. The size of the bladder is a result of urine accumulation in the bladder. However, bladder shape can be influenced by various factors, including pelvic structures, capacity, and bladder contents, which implies it varies from person to person and changes time to time within the same person. There is no specific relationship between bladder shape and race proven by sufficient clinical studies. Thus, demographics are only collected as auxiliary references in data collection.
Since the subject device, EdgeFlow UH10, is intended to measure bladder volume or residual urine volume, the data acquired by the target device (EdgeFlow UH10) are expected to show an imbalance between the presence of bladder and the case of bladder absence. Thus, the PR AUC (Precision-Recall Area Under Curve) is considered as a secondary endpoint for the performance test of the classification network by drawing a precision-recall curve.
For data truthing, the test dataset is independently reviewed by two evaluators who have in clinical experiences. The review results are transformed into ground truths to assess the performance of two networks.
#### 5. INDICATION FOR USE
The EdgeFlow UH10 is an ultrasound device intended to be used for measuring the urine volume in the bladder noninvasively. It is intended for use in professional healthcare facilities, such as hospitals, clinics, by qualified and trained healthcare professionals. The EdgeFlow UH10 supports B-mode and harmonic imaging modes.
{7}------------------------------------------------
## 6. SUBSTANTIAL EQUIVALENCE
Comparison of the technical characteristics of the subject device and predicate devices is shown in the Table of Substantial Equivalence Below.
{8}------------------------------------------------
| | Subject Device | Primary Predicate Device | Reference Device |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Edgecare Inc. | Verathon Inc. | Butterfly Network, Inc. |
| Trade Name | EdgeFlow UH10 | BladderScan® PRIME PLUS system | Auto 3D Bladder Volume Tool |
| 510(k) Number | | K172356 | K200980 |
| Product Code | IYO, ITX, QIH | IYO, ITX | IYO, ITX |
| Indications for Use | The EdgeFlow UH10 is an ultrasound<br>device intended to be used for<br>measuring the urine volume in the<br>bladder noninvasively. It is intended<br>for use in professional healthcare<br>facilities, such as hospitals, clinics, by<br>qualified and trained healthcare<br>professionals. The EdgeFlow UH10<br>supports B-mode and harmonic<br>imaging modes. | The BladderScan® Prime PLUS<br>System is an ultrasound device<br>intended to be used for measuring the<br>urine volume in the bladder<br>noninvasively. | The Butterfly Auto 3D Bladder<br>Volume Tool is a software application<br>package. It is designed to view,<br>quantify and report results acquired on<br>Butterfly Network ultrasound systems<br>for noninvasive volume measurements<br>of the bladder, to support physician<br>diagnosis. Indicated for use in adult<br>populations |
| Contraindications | It is contraindicated for fetal use and<br>for use on pregnant patients. And it<br>should not be used by those who are<br>allergic to coupling agent and who<br>have abdomen wound and skin disease. | The BladderScan® Prime PLUS<br>System is not intended for fetal use or<br>for use on pregnant patients, patients<br>with ascites, or patients with open skin<br>or wounds in the suprapubic region. | The Auto 3D Bladder Volume Tool is<br>not intended for fetal or pediatric use<br>or for use on pregnant patients,<br>patients with ascites, or patients with<br>open skin or wounds in the suprapubic<br>region. |
| User | Physicians/Medical Professionals | Physicians/Medical Professionals | Physicians/Medical Professionals |
| Target Population | Male<br>Female<br>Pediatric patients | Male<br>Female<br>Pediatric patients | Male<br>Female |
| Anatomical Site | Bladder | Bladder | Bladder |
| Technology | Neural network technology | Neural network technology | Neural network technology |
| Sterility | Non-sterile | Non-sterile | Non-sterile |
| Power Source | Battery Powered<br>(Lithium-ion battery) | Battery Powered<br>(Lithium-ion battery) | Battery Powered<br>(Lithium-ion battery) |
| Energy Delivered | Ultrasound | Ultrasound | Ultrasound |
| Measurement Accuracy | $0-100mL = \pm7.5mL$<br>$100-999 mL = \pm7.5%$ | $0-100mL = \pm7.5mL$<br>$100-999 mL = \pm7.5%$ | $0-100mL = \pm7.5mL$<br>$100-999 mL = \pm7.5%$ |
| Measurement Range | 0 to 999 mL | 0 to 999 mL | 0 to 740 mL |
| Automatically Calculating Function | Yes | Yes | Yes |
| 2D/3D Image | 2D | 2D | 3D |
| Mode of operation | B-mode | B-mode | B-mode |
| Transducer Type | Electronic Sector Scanning<br>(Phased Array) | Mechanical Sector Probe | Electronic Sector Scanning<br>(Phased Array) |
| Sector Angle | 120 degrees | 120 degrees | 100 degrees |
| Number of Scan Planes | 2 | 12 | 25 |
| Portable | Yes | Yes | Yes |
| Display | LCD | LCD | LCD |
| Live Scan Image | Yes | Yes | Yes |
| Touch Screen | Yes | Yes | Yes |
| Calibration | No Calibration recommended | No Calibration recommended | No Calibration recommended |
| Data Connections | USB, Wireless (to PC) | USB, SD card | Wireless ( to Mobile App, Cloud) |
| Accessories | Cradle<br>Power Adapter<br>USB Cable<br>Thermal Paper | Printer, battery<br>battery charger<br>power cord<br>mobile cart | Printer, battery<br>battery charger<br>power cord<br>mobile cart |
| FDA Ultrasound Track | Track 3 | Track 1 | Track 3 |
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Edgecare" in a sans-serif font. The word is black, except for a blue line above the "E". The word is horizontally oriented and centered in the image. The background is white.
{10}------------------------------------------------
The subject devices are substantially equivalent to the predicate and reference devices with respect to indications for use, technology and construction. The differences between the predicate devices and the subject devices are minor and any risks have been mitigated through testing. The difference does not raise new safety or effectiveness concerns for the provided intended use.
#### 7. NON-CLINICAL DATA
As part of demonstrating substantial equivalence of the EdgeFlow UH10 to the predicate device and reference device, Edgecare Inc. conducted performance testing on the subject devices. Although there are slightly different points such as technical parts (data connection type, accessories, ultrasound track), it does not impact the ability to determine substantial equivalence of the subject devices because the substantial equivalence of performance for EdgeFlow UH10 is demonstrated by the following verification and validation data to demonstrate the safety and performance effectiveness.
- . Biocompatibility
The biocompatibility tests were performed to protect patients from undue risks arise from biological hazards associated with materials of manufacture and final device. The tests were established in accordance with ISO 10993-1 and FDA Guidance - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
Biological Safety Assessment was prepared by using test report and others in accordance with ISO 10993-5, 10, 23
- Electrical Safety and EMC
The electrical safety tests were performed to protect patients from undue risks arise from any hazards associated with final device. The tests were performed in accordance with the following standards.
| No. | Test Items | Standards |
|-----|-------------------------------------------------------------------|------------------------------|
| 1 | General requirement for basic safety and<br>essential performance | IEC 60601-1:2005+A1:2012 |
| 2 | General requirement for safety -<br>Electromagnetic disturbances | IEC 60601-1-2:2014/AMD1:2020 |
{11}------------------------------------------------
#### Wireless .
The following tests were performed to assess effectiveness of software of the device. The test was performed in accordance with following standards.
| No. | Test Items | Standards |
|-----|------------------------|-----------------------|
| 1 | Wireless Coexistence | ANSI IEEE C63.27-2017 |
| 2 | Wi-Fi Performance Test | In house hold |
#### Software .
The following tests were performed to assess effectiveness of software of the device. The test was performed in accordance with following standards.
| No. | Test Items | Standards |
|-----|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | General requirement for safety –<br>Programmable electrical medical systems<br>(PEMS) | IEC 62304:2006/A1:2015 FDA Guidance (“Guidance for the<br>Content of Premarket Submissions for<br>Software Contained in Medical<br>Devices”) FDA Guidance (“Off-the-Shelf Software<br>Use in Medical Devices”) |
| 2 | Cybersecurity Test | AAMI/UL 29001-:2017 IEC 81001-5-1:2021 FDA Guidance (“Cybersecurity in<br>Medical Devices: Quality System<br>Considerations and Content of Premarket<br>Submissions”) |
#### Performance Test
The following tests were performed to assess effectiveness of the product performance including characteristic properties.
| No. | Test Items | Standards |
|-----|-------------------------------------------|-------------------------------|
| 1 | Acoustic Output | IEC 60601-2-37:2007/AMD1:2015 |
| 2 | Measurement Accuracy of Bladder<br>Volume | Manufacturing SOP |
| 3 | Deep Neural Network Performance | Manufacturing SOP |
{12}------------------------------------------------
The performance of the deep neural network was validated with a patient's dataset. Demographics are only collected as auxiliary references in data collection. All data used for training and test datasets were collected under a clinical trial approved by Yonsei University Institutional Review Board in Severance Hospital (IRB No: 1-2022-0076). Subject's demographics are described below:
| Gender | Male | 34 (37.78%) |
|--------|--------------|-------------|
| | Female | 56 (62.22%) |
| Ages | Less than 19 | 1 (1.11%) |
| | 19 or more | 89 (98.89%) |
| BMI | Maximum | 36.36 |
| | Minimum | 17.41 |
| | Average | 24.34 |
| | Variance | 10.78 |
Demographics of the clinical trial (Total number of subjects: 90)
A total of 1,528 images were used for Segmentation, and 3,711 images were used for Classification in the ultrasound images of patients.
{13}------------------------------------------------
As information on the protocol used for testing the algorithm:
- Classification Network: -
- . Input a test dataset into the deep learning model to generate the outputs of the classification network.
- . Compare the outputs (bladder presence, bladder absence) from the model's classification network with the ground truths.
- . Evaluate the performance of the classification network with 95% confidence intervals using the test criteria for the primary endpoint (F1 score: ≥0.90) and secondary endpoint (PR AUC: ≥0.95).
- Segmentation Network: -
- . Input a test dataset into the deep learning model to generate the outputs of the segmentation network.
- Compare the outputs generated from the model's segmentation network with the ground truths.
- . Evaluate the performance of the segmentation network with 95% confidence intervals using the test criteria for the primary endpoint (Dice Score: ≥0.89).
The accuracy of the classification and segmentation networks were measured respectively with the test dataset. The test criteria of classification accuracy and segmentation accuracy were both satisfied with an F1 score of 0.979 (95% CI 0.974-0.984), and a Dice score of 0.896 (95% CI 0.890-0.901). In conclusion, the deep learning model of EdgeFlow UH10 satisfies performance criteria established by Edgecare Inc.
The non-clinical overall data results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device.
{14}------------------------------------------------
#### 8. CONCLUSION
The comparison between the subject devices and the predicate devices shows that the general information, some technical and material information are the same. Although there are some differences, the performance test reports are supported to the substantial equivalence of the subject device, the performance test reports are provided to demonstrate substantial equivalence of the subject devices. Therefore, we conclude that the different characteristics do not raise different questions of safety and effectiveness, and thus the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.