The Ziehm Solo FD is intended for use in providing medical imaging for adults and pediatric populations, using pulsed and continuous fluoroscopic imaging. The device provides contactless fluoroscopic image capture, temporarily storing, and display of digital subtraction, and acquisition of cine loops during diagnostic, interventional and surgical procedures. Examples of clinical application may include pediatric, cholangiography, endoscopic, urologic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room fluoroscopy procedures. The visualization of such anatomical structures assists the clinician in the clinical outcome. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.
Device Story
Mobile fluoroscopic C-arm system; captures, stores, and displays X-ray images for surgical/interventional guidance. Input: X-rays generated by mono-block generator; detected by 8-inch CMOS or IGZO flat panel detector. Operation: Physician-directed; mobile stand with C-arm; CAN BUS control system; optional viewing station. Output: Real-time fluoroscopic images, cine loops, digital subtraction. Clinical utility: Assists in localizing pathological areas, visualizing anatomy/implants during procedures. Benefits: Real-time visualization, low-dose functionality, improved image quality via IGZO detector. Used in OR, clinic, emergency settings.
Clinical Evidence
Bench testing only. Non-clinical imaging and dose testing performed using anatomical phantoms representing adult and pediatric populations. Radiologist assessment of image sets confirmed image quality is comparable to reference device (K193230). Compliance with 21 CFR 1020.30-32 and IEC 60601 series standards verified.
Technological Characteristics
Mobile C-arm; mono-block X-ray generator; 8-inch CMOS or IGZO flat panel detectors (CsI scintillator); CAN BUS control; laser positioning; wireless/wired footswitch; DICOM 3.0 connectivity; Linux-based OS. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, IEC 60601-2-54, ISO 14971.
Indications for Use
Indicated for medical imaging in adults and pediatric populations using pulsed and continuous fluoroscopic imaging for diagnostic, interventional, and surgical procedures, including pediatric, cholangiography, endoscopic, urologic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room applications. Not for mammography or use near MRI systems.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
December 12, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
Ziehm Imaging GmbH % Tsvetelina Milanova Specialist Regulatory Affairs Lina-Ammon-Strasse 10 Nuremberg, Bavaria 90471 GERMANY
Re: K231669
Trade/Device Name: Ziehm Solo FD Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAA, OXO Dated: November 13, 2023 Received: November 14, 2023
Dear Tsvetelina Milanova:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K231669
Device Name
Ziehm Solo FD
## Indications for Use (Describe)
The Ziehm Solo FD is intended for use in providing medical imaging for adults and pediatric populations, using pulsed and continuous fluoroscopic imaging.
The device provides contactless fluoroscopic image capture, temporarily storing, and display of digital subtraction, and acquisition of cine loops during diagnostic, interventional and surgical procedures. Examples of clinical application may include pediatric, cholangiography, endoscopic, urologic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room fluoroscopy procedures.
The visualization of such anatomical structures assists the clinician in the clinical outcome. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular graphic on the left and the text "ziehmimaging" on the right. The graphic is a stylized representation of a globe with a human figure inside. The text is in a teal color and a serif font.
U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center –W066-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
December, 7 2023
In accordance with 21 CFR §807.92 the following 510(k) K231669 summary information is provided:
| Submitter Address: | Ziehm Imaging GmbH<br>Lina-Ammon-Strasse 10<br>90471 Nuremberg<br>Germany<br>Phone: +49 911 66067 0<br>Fax: +49 911 66067 390 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person | Tsvetelina Milanova<br>Specialist Regulatory Affairs<br>Ziehm Imaging GmbH<br>Phone: +49 911 66067 219<br>Fax: +49 172 8651712 |
| Secondary Contact<br>Person: | Stefan Fiedler<br>Director QM/RA<br>Ziehm Imaging GmbH<br>Phone: +49 911 66067 219<br>Fax: +49 172 8651712 |
| US Agent | Kevin Bridgman<br>VP Quality Assurance & Requlatory Affairs<br>Ziehm-Orthoscan, Inc.<br>14555 N 82nd St<br>Scottsdale AZ, 85260 |
| Device (Trade Name): | Ziehm Solo FD |
| Common /Usual<br>Names: | Mobile Fluoroscopic C-Arm |
| Requlation: | 21CFR 892.1650 |
| Regulation Description: | Image-intensified fluoroscopic x-ray system |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Ziehm Imaging. The logo consists of a stylized globe-like icon on the left, followed by the company name "ziehmimaging" in a teal-colored sans-serif font. The globe icon features a human figure within intersecting circles, giving it a modern and technological appearance.
| Product Code: | OWB; JAA, OXO |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification: | II |
| Predicate Device: | K161976 Ziehm Solo FD |
| Reference Device | K193230 Ziehm Vision FD |
| Decision Date: | 10/06/2016 |
| Regulation: | 21CFR 892.1650 |
| Regulation Description Name: | Image-intensified fluoroscopic x-ray system |
| Product Code: | OWB; JAA, OXO |
| Summary of Technological Characteristics | The Ziehm Solo FD uses X-ray imaging technology to visualize the human anatomy. The X-ray tube in the generator produces X-rays that penetrate the patient and then hit a special detector that converts them into digital images. This is done under the control of the user and at the direction of a physician who determines the specific clinical procedure. This visualization assists the physician in localizing pathological areas or during surgical procedures. The device enables real-time image acquisition as well as visualization of in vivo surgical procedures and post-operative results.<br>The Ziehm Solo FD consist of one mobile unit, the Mobile Stand. Optionally the device can be ordered with a Viewing Station (Monitor Cart). The Mobile Stand incorporates a small compact design making the positioning of the C-arm in relation to the patient easier for the operator. The generator with X-ray tube, advanced heat management system, X-ray control and collimators are assembled in one housing in a mono-block generator. The system control is handled via CAN BUS control system.<br>The mechanical C-Profile supports the generator, the flat panel detector and an integrated laser positioning device.<br>The optional available Viewing Station (Monitor Cart) provides a remote touch Solo Center that duplicates the touch Solo Center mounted on the Mobile Stand.<br>The proposed modified device Ziehm Solo FD employs the same fundamental control, and substantially equivalent scientific technology as that of our predicate device Ziehm Solo FD (K161976). Software architecture design is substantially equivalent to that of the predicate Ziehm Solo FD<br>The primary modification of the C-Arm includes an 8 inch IGZO (Indium gallium zinc oxide) flat panel detector (FPD The new 8 inch IGZO FPD is an addition to already introduced CMOS FPD. The flat panel detectors have the same outer product design of the housing, both devices use safety shielding |
510(k) Summary
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Ziehm Imaging. The logo consists of a circular graphic on the left and the company name "ziehmimaging" on the right. The graphic features a stylized human figure inside a circle with orbital lines around it. The text is in a teal color and appears to be a sans-serif font.
for radiation suppression and use solid state x-ray image receptors (SSXI / FPD) 8 inch CMOS and the only difference to the predicate Ziehm Solo FD is the additional 8 inch IGZO panel.
The comparison of the predicate device and the modified devices shows that the scientific and technical characteristics of the Ziehm Solo FD are substantially equivalent as those of the Ziehm Solo FD predicate device (K161976).
- Intended Use The Ziehm Solo FD is a mobile C-arm providing image data by means of a noninvasive x-ray technique during medical procedures and stores them temporarily. The Ziehm Solo FD is intended for use in all medical indications requiring fluoroscopy. The system is intended for use with human beings of any age. It is the physician's responsibility to decide whether to use the system with infants, children, and adipose patients. The svstem is intended for use with human bodies covering such structures but not limited to the following, e.q., organs, tissue, bones, implants depending on the medical indication.
Indications for Use: The Ziehm Solo FD is intended for use in providing medical imaging for adults and pediatric populations, using pulsed and continuous fluoroscopic imaging. The device provides contactless fluoroscopic image capture, temporarily storing, and display of digital subtraction, and acquisition of cine loops during diagnostic, interventional and surgical procedures. Examples of clinical application may include pediatric, cholanqiography, endoscopic, urologic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room fluoroscopy procedures. The visualization of such anatomical structures assists the clinician in the clinical outcome. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.
## Device Comparison Table
The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:
| Model | Modified<br>Ziehm Solo FD | Predicate<br>Ziehm Solo FD (K161976) | Comparable<br>Properties<br>Substantial<br>Equivalence<br>Discussion |
|------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| 510(k)<br>Number | K231669 | K161976 | --- |
| Product<br>Codes | OWB (interventional fluoroscopic x-ray<br>system)<br>Subsequent:<br>JAA (system, x-ray, fluoroscopic) | OWB (interventional fluoroscopic x-ray system)<br>Subsequent:<br>JAA (system, x-ray, fluoroscopic) | Identical |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Ziehm Imaging. The logo consists of a circular graphic to the left of the company name. The graphic contains a stylized human figure inside of a circle with curved lines. The company name is written in a teal sans-serif font.
| Model | Modified<br>Ziehm Solo FD | Predicate<br>Ziehm Solo FD (K161976) | Comparable<br>Properties<br>Substantial<br>Equivalence<br>Discussion |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Image/<br>General<br>Overview | image-intensified),<br>OXO (image-intensified fluoroscopic x-<br>ray system, mobile)<br>Image: Modified Ziehm Solo FD | image-intensified),<br>OXO (image-intensified fluoroscopic<br>x-ray system, mobile)<br>Image: Predicate Ziehm Solo FD (K161976) | The flat panel<br>detectors are<br>identical in the<br>design of the<br>housing, both<br>devices use<br>safety shielding<br>for radiation<br>suppression and<br>use solid state x-<br>ray image<br>receptor (SSXI /<br>FPD) 8 inch CMOS<br>and the only<br>difference to the<br>predicate Ziehm<br>Solo FD is the<br>new 8 inch IGZO<br>FPD. |
| Model | Modified<br>Ziehm Solo FD | Predicate<br>Ziehm Solo FD (K161976) | Comparable<br>Properties<br>Substantial<br>Equivalence<br>Discussion |
| X-ray Generator | | | |
| Pulsed<br>Fluoroscopy:<br>Operating<br>values | kV range:<br>variant B0: 40 - 110 kV<br>variant B1+B2: 40 - 120 kV mA range:<br>variant B0: 0.2 - 16 mA<br>variant B1: 0.2 - 16 mA<br>variant B2: 0.2 - <b>20 mA</b> | kV range:<br>variant B0: 40 - 110 kV<br>variant B1+B2: 40 - 120 kV mA range:<br>variant B0: 0.2 - 16 mA<br>variant B1: 0.2 - 16 mA<br>variant B2: 0.2 - 16 mA | Substantial<br>equivalent.<br>The characteristic<br>line was changed<br>to 20mA, which<br>has no effect on<br>the safety as it is<br>below the<br>maximum dose. |
| X-ray Tube | | | |
| Tube Type | stationary anode | stationary anode | Identical. |
| X-Ray<br>Tubes | for variants B0, B1, and B2 Avg. Power: 0.6 kW Peak Power: 2,4kW Max. Voltage: <b>125kV</b> for variants B0+B1 Avg. Power:0.6 kW Peak Power: 2.02 kW Max. Voltage: 120 kV | for variants B0, B1, and B2 Avg. Power: 0.6 kW Peak Power: 2,4kW Max. Voltage: 120kV for variants B0+B1 Avg. Power:0.6 kW Peak Power: 2.02 kW Max. Voltage: 120 kV | Substantial<br>equivalent<br>The maximal<br>voltage of the X-<br>Ray tube for<br>variants B0, B1,<br>and B2 have the<br>max. Voltage up<br>to 125kV, but as<br>the generator is<br>up to 120kV the<br>value was given<br>as 120kV. |
| Monitors | | | |
| Display<br>Monitor | The device can be equipped with<br>monitors of different sizes 19" Duo flat-screen monitor<br>(optionally with Quick Release) 27" flat screen monitor | The device can be equipped with<br>monitors of different sizes 19" DUO flat-screen monitor 24" flat screen monitor | Substantial<br>equivalent.<br>The 24" flat<br>screen monitor<br>has been<br>discontinued by<br>the manufacturer<br>and have been<br>replaced by 27"<br>monitor. |
| | Radiation Switches | | |
| X-Ray<br>foot<br>switch | Cable bound footswitch (with or<br>without protective bracket) optional: Wireless footswitch with<br>an emergency cord the if battery is | Cable bound footswitch optional: Wireless footswitch<br>with an emergency cord the if<br>battery is low | Substantial<br>equivalent<br>All footswitches<br>are available with |
| Model | Modified<br>Ziehm Solo FD | Predicate<br>Ziehm Solo FD (K161976) | Comparable<br>Properties<br>Substantial<br>Equivalence<br>Discussion |
| | low (with or without protective<br>bracket) | | protective<br>bracket. |
| | Digital Imaging Processing | | |
| Real-Time<br>processing<br>functions | • Recursive filter: 4 levels<br>• Last Image Hold<br>• Edge enhancement filter: 5 levels<br>• Windowing and step windowing<br>• Digital image rotation and reversal<br>without radiation<br>• Grayscale inversion<br>• Digital collimators<br>• Ziehm Adaptive Image Processing<br>(ZAIP) | • Recursive filter: 4 levels<br>• Stack filter 'Last Image Hold': 5<br>levels<br>• Last Image Hold<br>• Edge enhancement filter:<br>5 levels<br>• Windowing and step windowing<br>• Digital image rotation and<br>reversal without radiation<br>• Grayscale inversion<br>• Digital collimators<br>• Ziehm Adaptive Image<br>Processing (ZAIP) | Substantial<br>equivalent.<br>Stack filter 'Last<br>Image Hold' was<br>discontinued. |
| Appli-<br>cation-<br>Oriented<br>Anato-<br>mical<br>Programs<br>(AOAP) | • Bone: Extremities, Trunk<br>• Heart, Abdomen, Soft, Uro (option)<br>• Vascular: Extremities, Trunk, Bolus | • Bone: Extremities, Trunk<br>• Abdomen, Soft, Uro (option),<br>Long Procedure Spine<br>• Vascular: Extremities, Trunk,<br>Bolus | Substantial<br>equivalent.<br>Long procedure<br>spine was<br>discontinued. |
| Model | Modified<br>Ziehm Solo FD | Predicate<br>Ziehm Solo FD (K161976) | Comparable<br>Properties<br>Substantial<br>Equivalence<br>Discussion |
| Data<br>Organizati<br>on | • Patient-based data management<br>with 16-image mosaic display<br>• Pre-registration via DICOM Worklist<br>(option)<br>• Manual input or emergency<br>registration<br>• Calculated Dose Area Product (DAP)<br>• DAP value tagged to stored image<br>• Air Kerma dose display<br>• Air kerma value tagged to the<br>stored image<br>• HIPAA security package (option)<br>• Radiation Dose Structured Report<br>(RDSR)<br>• Video Output<br>- Full HD SDI (split display)<br>- 1920 x 1080p<br>- 60Hz<br>• Measured dose area product (DAP)<br>with digital display (option)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.