K231660 · Jiangsu Dynamic Medical Technology Co., Ltd. · MUH · Dec 20, 2023 · Dental
Device Facts
Record ID
K231660
Device Name
Digital image scanner of dental
Applicant
Jiangsu Dynamic Medical Technology Co., Ltd.
Product Code
MUH · Dental
Decision Date
Dec 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
Digital image scanner of dental is intended to be used for scanning and processing digital images exposed on IP Imaging Plate in dental applications.
Device Story
Digital image scanner of dental is an intraoral plate scanner used in dental clinics by clinicians/technicians. It processes IP imaging plates (sizes S0-S3) previously exposed to X-rays in a patient's mouth. Operation involves inserting the exposed plate into the device; a 630-645 nm laser, moved by an oscillating mirror, scans the plate line-by-line. The plate's X-ray sensitive coating emits light proportional to the exposure dose; a photomultiplier tube (PMT) captures this light, converting it into an electrical signal. This signal is digitized to create a dental X-ray image. An internal LED-PCB erases the plate for reuse. The device transmits images to a screen for viewing, storage, and manipulation. It provides a preview on an internal touch screen, while diagnostic viewing occurs on an external monitor. The device facilitates dental diagnostics by enabling digital capture and review of intraoral X-rays.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), and dental imaging performance (IEC 61223-3-4). Performance testing verified imaging metrics including spatial resolution (MTF), uniformity, contrast, and bit depth.
Technological Characteristics
Extraoral source X-ray system (Class II, MUH). Uses 630-645 nm laser scanning with oscillating mirror and PMT sensor. Supports IP plate sizes S0, S1, S2, S3. 16-bit image depth. Connectivity via computer interface for display/storage. Power: 100-240VAC. Non-sterile. Software-controlled image processing and plate erasure.
Indications for Use
Indicated for scanning and processing digital images exposed on IP Imaging Plates in dental applications for use by licensed clinicians and authorized technicians.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
December 20, 2023
Jiangsu Dynamic Medical Technology Co., Ltd. % Doris Dong General Manager Shanghai CV Technology Co., Ltd. Room 602, No. 19 Dongbao Road, Songjiang Area SHANGHAI, SHANGHAI 201613 CHINA
Re: K231660
Trade/Device Name: Digital image scanner of dental Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: Class II Product Code: MUH Dated: November 20, 2023 Received: November 24, 2023
Dear Doris Dong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231660
Device Name Digital image scanner of dental
Indications for Use (Describe)
Digital image scanner of dental is intended to be used for scanning and processing digital images exposed on IP Imaging Plate in dental applications.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary [As required by 21 CFR 807.92]
#### 1. Submission Information
| 510(k) Number: | K231660 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | December 08, 2023 |
| Type of 510(k) Submission: | Traditional 510(k) |
| Basis for 510(k) Submission: | New device |
| Owner: | JIANGSU DYNAMIC MEDICAL TECHNOLOGY CO., LTD.<br>No.108 Xingpu Road, Lujia Town, Kunshan, Suzhou, Jiangsu, China<br>215331<br>Tel: 86-512-82627666<br>Email: technic2@dynamicgroup.cn |
| Contact: | Doris Dong<br>[Consultant, from Shanghai CV Technology Co., Ltd.]<br>Add: Room 602, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613<br>China<br>E-mail: doris.d@ceve.org.cn<br>Tel: 86 21-31261348 / Fax: 86 21-57712250 |
#### 2. Device Description
| Proprietary Name: | Digital image scanner of dental |
|----------------------|------------------------------------------------------------------|
| Model: | DDT-100, DDT-100C, DDT-200, DDT-200C |
| Common Name: | Extraoral source x-ray system |
| Classification Name: | Extraoral source x-ray system |
| Regulation Number: | 21 CFR 872.1800 |
| Product Code: | MUH |
| Device Class: | 2 |
| Review Panel: | Radiology |
| Device Description: | Digital image scanner of dental is a dental device that scans IP |
maging plates that have been exposed in place of dental X-ray film and allows the resulting images to be displayed on the screen and stored for later recovery. It will be used by licensed clinicians and authorized technicians for this purpose. The device is an intraoral Plate Scanner, which is designed to read out intraoral Plates of the size S0, S1, S2 and S3. The intraoral plates are put into the mouth of the patient, exposed to X-rays and then are read out with the device. The read-out-process is carried out with 630-645 nm laser. The laser beam is moved across the surface of the plate by an oscillating mirror. The laser beam stimulates the top coating of the plates, which consists of X-rays sensitive material. Depending on the exposed dose, the coating emits different levels of light. These light particles are then requisitioned by an optical sensor (Photo Multiplier tube/ PMT) and transferred into an electrical output signal. This signal is digitalized and is the data for the digital X-ray image. Before the plate is discharged, the remaining data is erased by a LED-PCB. The user chooses which size of plate he must use and prepares the device by inserting the appropriate plate
{4}------------------------------------------------
| | insert into the device. He then exposes the plate and then puts the plate |
|----------------------|------------------------------------------------------------------------------|
| | directly into the insert by pushing it out of the light protection envelope. |
| | The user closes the light protection over and starts the read-out process. |
| | After the read out process the picture is transmitted to the screen, the |
| | picture can be viewed, and the IP is erased and ready to use for the next |
| | acquisition. |
| Indications for use: | Digital image scanner of dental is intended to be used for scanning and |
| | processing digital images exposed on IP Imaging Plate in dental |
| | applications. |
## 3. Predicate Device Information
| Predicate 510(k) Number: | K212080 |
|---------------------------|------------------------------------------------|
| Marketing clearance date: | 27/09/2021 |
| Product name: | i-Scan Imaging Plate Scanner |
| Manufacturer: | Guilin Woodpecker Medical Instrument Co., Ltd. |
| Parameters | New Device | Predicate Device | Comparison |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| 510(k) number | K231660 | K212080 | -- |
| Submitter | JIANGSU DYNAMIC<br>MEDICAL TECHNOLOGY<br>CO.,LTD. | Guilin Woodpecker Medical<br>Instrument Co., Ltd. | -- |
| Device name | Digital image scanner of dental | i-Scan Imaging Plate Scanner | -- |
| Model | DDT-100,DDT-100C,DDT-200,<br>DDT-200C | i-Scan | -- |
| Regulation number | 21 CFR 872.1800 | 21 CFR 872.1800 | Same |
| Device class | II | II | Same |
| Product code | MUH | MUH | Same |
| Classification name | Extraoral source x-ray system | Extraoral source x-ray system | Same |
| Indications for use | Digital image scanner of dental<br>is intended to be used for<br>scanning and processing digital<br>images exposed on IP Imaging<br>Plate in dental applications. | i-Scan Imaging Plate Scanner is<br>intended to be used for scanning<br>and processing digital images<br>exposed on IP Imaging Plate in<br>dental applications. | Same |
| Prescription Use | Yes | Yes | Same |
| Site of Use | Hospitals and clinics | Hospitals and clinics | Same |
| Intended user | Operated and used by trained<br>professionals with physician's<br>guidance | Operated and used by trained<br>professionals with physician's<br>guidance | Same |
| Sterile | Non-Sterile | Non-Sterile | Same |
| Product Dimension<br>(H*W*D) | DDT-100,DDT-100C:365*262*<br>287mm<br>DDT-200,DDT-200C:320*170*<br>384mm | 138*245.5*290.2mm | Similar<br>Note 1 |
| Product Weight | DDT-100,DDT-100C:9.3kg<br>DDT-200.DDT-200C:9.3kg | 4.5kg | |
| Mechanical design | The exposed and unwrapped plates are scanned using a laser with a certain wavelength of 630-645 nm. The scan pattern is line by line | The exposed and unwrapped plates are scanned using a laser with a certain wavelength of 651-665 nm. The scan pattern is line by line. | Similar Note 2 |
| Electrical design | Light with a wavelength of approximately 380 nm is from the plate in proportion to the number of captured x-ray photons. This light is collected and formed into an image that may be viewed on a video display and stored for later recovery in a computer memory. | Light with a wavelength of approximately 380 nm is from the plate in proportion to the number of captured x-ray photons. This light is collected and formed into an image that may be viewed on a video display and stored for later recovery in a computer memory. | Same |
| Viewing the image | The touch screen only shows a preview which serves to provide an initial impression of the final x-ray image. For the purposes of diagnosis, the x-ray image must be viewed on a diagnostic monitor. The scanned images are displayed on an internal LCD or an external monitor using a computer and user software including image storage, retrieval and manipulation. | The touch screen only shows a preview which serves to provide an initial impression of the final x-ray image. For the purposes of diagnosis, the x-ray image must be viewed on a diagnostic monitor. The scanned images are displayed on an internal LCD or an external monitor using a computer and user software including image storage, retrieval and manipulation. | Same |
| Image scanning | Laser/Photomultiplier Tube | Laser/Photomultiplier Tube | Same |
| Erasing the residual image following scanning for plate reuse | The residual image is erased in the scanner by an inline erasing function. | The residual image is erased in the scanner by an inline erasing function. | Same |
| Transport/feed mechanism of imaging plate | The Plate realizes transmission movement through a conveyor belt, and the movement of the laser and the IP image board provides two orthogonal scanning directions. | The Plate realizes transmission movement through a conveyor belt, and the movement of the laser and the IP image board provides two orthogonal scanning directions. | Same |
| Imaging plate size | S0:22×35 mm<br>S1:24×40 mm<br>S2:31×41 mm<br>S3:27×54 mm | Size 0: 22×35 mm<br>Size 1: 24×40 mm<br>Size 2: 31×41 mm<br>Size 3: 27×54 mm | Same |
| Image data bit depth | 16 bits | 16 bits | Same |
| Contact body site | The device does not directly contact with human body or organ. | The device does not directly contact with human body or organ. | Same |
| Patient Contamination | Single patient use barrier | Single patient use barrier | Same |
| prevention | envelope encloses the imaging plate while in the patient's mouth. | envelope encloses the imaging plate while in the patient's mouth. | |
| Electronic User<br>Manual | User manual is provided in printed form to the user. | User manual is provided in printed form to the user. | Same |
| Image Quality | Theoretical resolutions: 17 LP/mm | Theoretical resolutions: 10 or 33 LP/mm | Similar<br>Note 3 |
| RFID | No | No | Same |
| MTF | More than 45% at 3 lp/mm | More than 45% at 3 lp/mm | Same |
| DQE | More than 21.0% at 3 lp/mm | More than 21.0% at 3 lp/mm | Same |
| Software/Firmware/M<br>icroprocessor<br>Control? | Digital Image Scanner Of Dental | Air-dental-woodpecker | Similar<br>Note 4 |
| External power source | 100 to 240VAC, 50/60 Hz | 100 to 240VAC, 50/60 Hz | Same |
| Compliance standards | ANSI AAMI ES60601-1<br>IEC 60601-1-2<br>IEC 61223-3-4<br>IEC 60825-1 | ANSI AAMI ES60601-1<br>IEC 60601-1-2<br>IEC 62220-1<br>IEC 61223-3-4<br>IEC 60825-1<br>IEC 61010-1 | Similar<br>Note 5 |
| Picture | Image: DDT-100, DDT-100C, DDT-200, DDT-200C | Image: Scanner | Similar<br>Note 1 |
| Labeling | Meet FDA's requirement | Meet FDA's requirement | Same |
### 4. Comparison to predicate device
{5}------------------------------------------------
{6}------------------------------------------------
Differences between New device and Predicate Device:
## Note 1:
The weight, dimensions and appearance of the proposed device differ slightly from the predicate device,depending on the design and sales requirements of these differences are insignificant in terms of safety or effectiveness.
## Note 2:
The operating principle of scanning the image plates with a laser of approximately the same wavelength,both the proposal device and the predicate device scans the plates with a line-by-line pattern. The scans collect the same data and process this for visual imaging. They support the same indications for use; the differences in the mechanical design does not raise issues of safety and effectiveness.
## Note 3:
The subject device displays theoretical resolutions of 17-Line Pairs per mm, whereas the predicate shows theoretical resolutions of 10- or 33-Line Pairs per mm. And the image resolution of the proposed device is between the image resolution of the predicate device. The differences do not affect the operating principle and the Image Quality distinctness is such that it does not affect the effectiveness of the device. Note 4:
{7}------------------------------------------------
The software is not the same between the subject device and the predicate, but both the subject and predicate have followed the FDA Guidance for the Content of Premarket for Submissions for Software Contained in Medical Devices.
### Note 5:
According to the latest requirements, IEC 6220-1 is not applicable to digital X-ray imaging equipment for dental radiography. IEC 61010-1 is the test for laboratory equipment, we have done the test for ANSI AAMI ES60601-1, the standard is more stringent than IEC 61010-1.Other tests are the same as the predicate equipment, so it does not cause safety problems.
### 5. Summary of Non-clinical Testing
The safety and effectiveness of the Digital image scanner of dental were established and the substantial equivalence determination was supported by a series of performance testing, including Electrical safety and electromagnetic compatibility (EMC), Software Verification Testing, image reading and checking testing.
### Electrical safety and electromagnetic compatibility (EMC)
General Requirements For Basic Safety And Essential Performance:This product has passed the IEC 60601-1 test and does not raise questions about safety effectiveness.
General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests:This product has passed the IEC 60601-1-2 test and does not raise questions about safety effectiveness.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed device complies with the following standards:
1) ANSI AAMI ES60601-1: 2005/(R)2012 And A1:2012, C1:2009(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD);
2) IEC 60601-1-2:2014, Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests;
### Performance Testing
Acceptance tests - Imaging performance of dental X-ray:Imaging standard requirements were met for X-ray equipment.
Safety of laser products:Laser products have passed safety test.
1) IEC 61223-3-4 First edition 2000-03, Evaluation and routine testing in medical imaging departments -Part 3-4: Acceptance tests - Imaging performance of dental X-ray;
2) IEC 60825-1 Edition 2.0 2007-03, Safety of laser products - Part 1: Equipment classification, and requirements [Including: Technical Corrigendum 1 (2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)].
### Software Verification and Validation Testing
Software testing and V&V:As a result of unit level test, integration review and system level test, we hereby certify that all the functions those support system specification are proper. We confirm that software verification and validation test is properly accomplished.
{8}------------------------------------------------
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
### Image Reading and Checking Testing
Regarding image reading and checking testing, we performed image plates test and image accuracy and fidelity test respectively. Image plate test contains imaging area, degree of attenuation, uniformity, image contrast, bit depth and spatial resolution or MTF test, etc..Image accuracy and fidelity test contains accuracy test, uniformity, image contrast,spatial resolution or MTF test,and image fidelity test. The accuracy of image reading and checking results was verified by above testing.
#### 6. Animal Study
Animal testing was not required for this submission.
#### 7. Clinical Testing
Clinical testing was not required for this submission.
### 8. Substantial Equivalent (SE) Conclusion
Based upon comparison to the predicate devices, the proposed device is Substantial Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.