K231633 · Alton (Shanghai) Medical Instruments Co., Ltd. · PKL · Jan 11, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K231633
Device Name
Disposable Hemoclip (AF series)
Applicant
Alton (Shanghai) Medical Instruments Co., Ltd.
Product Code
PKL · Gastroenterology, Urology
Decision Date
Jan 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Disposable Hemoclip is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of: 1. Endoscopic marking 2. Hemostasis for: - Mucosal/sub-mucosal defects < 3 cm - Bleeding ulcers - Arteries < 2 mm - Polyps < 1.5 cm in diameter - Diverticula in the colon 3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel; 4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively
Device Story
Disposable Hemoclip is a sterile, single-use endoscopic accessory used to deliver stainless steel clips to the GI tract. The device consists of a delivery system (handle and tube) and a pre-loaded clip. The delivery system allows for 360-degree rotation of the clip at the distal end. During an endoscopic procedure, the clinician deploys the clip to the target site; the clip jaws can be opened and closed up to five times prior to final deployment to allow for repositioning. The device is used by gastroenterologists or trained clinicians during endoscopic procedures. The output is the physical placement of a metal clip, which provides mechanical hemostasis, tissue marking, tube anchoring, or perforation closure. The device is MR Unsafe.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing only, including mechanical performance (e.g., clip retention force, rotation, open/close performance), biocompatibility (ISO 10993), and sterilization validation.
Indicated for patients requiring endoscopic clip placement in the GI tract for marking, hemostasis (mucosal/sub-mucosal defects <3cm, bleeding ulcers, arteries <2mm, polyps <1.5cm, colonic diverticula), anchoring jejunal feeding tubes, or supplemental closure of luminal perforations <20mm.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 11, 2024
Alton (Shanghai) Medical Instruments Co. Ltd Wei Song Project Engineer No.24 Building Jinshao Rd.1688. Baoshan District. 200949 Shanghai China
Re: K231633
Trade/Device Name: Disposable Hemoclip (AF series) Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: Class II Product Code: PKL Dated: December 11, 2023 Received: December 11, 2023
## Dear Wei Song:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
Page
3
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
## K231633
Device Name
Disposable Hemoclip (AF series)
Indications for Use (Describe)
The Disposable Hemoclip is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:
- 1. Endoscopic marking
- 2. Hemostasis for:
- · Mucosal/sub-mucosal defects < 3 cm
- · Bleeding ulcers
- · Arteries < 2 mm
- · Polyps < 1.5 cm in diameter
- · Diverticula in the colon
- 3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel;
- 4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively
Type of Use (Select one or both. as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
## I. SUBMITTER
Name: Alton (Shanghai) Medical Instruments Co. Ltd.
Address: No.24 Building, JinShao Rd. 1688, Baoshan District, 200949 Shanghai, P. R.
China
Name of contact person: Vivian Li
Position: Director of Quality Department
Tel: +86 21 56771811
Fax: +86 21 66307823
Mail: vivian@alton.com.cn
Date prepared: 2024-01-11
## II. Identification of Subjective Device
Device trade name: Disposable Hemoclip Regulation Name: Hemorrhoidal Ligator Regulation number: 21CFR 876.4400 Regulation class: 2 Review Panel: Gastroenterology/Urology Product code: PKL
## III. Identification of Predicate device
Predicate Submission Number: K172727 Trade/Device Name: Hemoclip Regulation Name: Hemorrhoidal Ligator Regulation Number: 21 CFR 876.4400 Regulatory Class: 2 Review Panel: Gastroenterology/Urology Product Code Description: Hemostatic Metal Clip For The Gi Tract Product Code: PKL
## IV. Device description
{5}------------------------------------------------
The Disposable Hemoclip is the key and common accessory in the diagnosis and treatment procedure of digestive endoscope. As a single use instrument, the disposable hemoclip does not allow to be used by twice, and cannot be separated by components except to be used as well. It is also a sterile device consisting of a pre-loaded, radiopaque, single-use, endoscopic clipping device consisting of two main components: the delivery system and the clip. The delivery system consists of a handle and tube. The delivery system is constructed using stainless steel and polyester materials. The delivery system will allow for the device to rotate at the distal end. The delivery system is offered in 1600mm, 1800mm, 2100mm and 2300mm working lengths.
The clip is made of stainless-steel material and deployed from the delivery system during use. The clip jaws are designed that they can be opened and closed up to five times prior to deployment, aiding in repositioning of the clip at the lesion site. Re-opening, closing, and rotation capability may be limited by clinical circumstances and patient anatomy.
#### V. Indication for use
The Disposable Hemoclip is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:
- 1. Endoscopic marking
- 2. Hemostasis for:
- Mucosal/sub-mucosal defects < 3 cm
- Bleeding ulcers
- Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- Diverticula in the colon
3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel;
4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively.
Precaution: It is recommended that healthcare providers distribute patient implant
{6}------------------------------------------------
cards with the name of the clip and date it was placed.
| Attribute | Subject device | Predicative device (K172727) | Discussion/<br>Conclusion |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Trade name | Disposable Hemoclip (AF series) | Hemoclip | / |
| Manufacturer | Alton (Shanghai) Medical<br>Instruments Co. Ltd | Hangzhou AGS MedTech CO., Ltd | / |
| Regulation name | Hemorrhoidal Ligator | Hemorrhoidal Ligator | Same |
| Regulatory Class | II | II | Same |
| Product code | PKL | PKL | Same |
| Clinical characteristics | | | |
| Indications for use | The Disposable Hemoclip is<br>indicated for clip placement<br>within the gastrointestinal (GI)<br>tract for the purpose of:<br>1. Endoscopic marking<br>2. Hemostasis for:<br>• Mucosal/sub-mucosal defects <<br>3 cm<br>• Bleeding ulcers<br>• Arteries < 2 mm<br>• Polyps < 1.5 cm in diameter<br>• Diverticula in the colon<br>3. Anchoring to affix jejunal<br>feeding tubes to the wall of the<br>small bowel;<br>4. As a supplemental closure<br>method of luminal perforations <<br>20 mm that can be treated<br>conservatively | The hemoclip is indicated for<br>endoscopic clip placement within<br>the gastrointestinal tract for the<br>purpose of:<br>1. Endoscopicmarking<br>2. Hemostasis for<br>• Mucosal/sub-mucosal defects<3cm<br>• Bleeding ulcers<br>• Arteries<2mm<br>• Polyps<1.5cm in diameter<br>• Diverticula in the colon<br>3. Anchoring to affix jejunal<br>feeding tubes to the wall of the<br>small bowel<br>4. As a supplementary method,<br>closure for GI tract luminal<br>perforations<20mm that can be<br>treated conservatively. | Same |
| General technological characteristics | | | |
| Principle of<br>Operation | Endoscopic accessory used to<br>deliver metal clips to the GI tract. | Endoscopic accessory used to<br>deliver metal clips to the GI tract. | Same |
| Rotation function | 360 ° rotatable | 360 ° rotatable | Same |
| Reposition<br>function | Yes | Yes and No (depending on model) | Same |
| Minimum<br>Endoscopic<br>Working Channel | Φ2.8mm | Φ2.8mm | Same |
| Outer Diameter of<br>the clip | Φ2.4mm; Φ2.6mm | Φ2.6mm | Difference,<br>see Discussion |
| Attribute | Subject device | Predicative device (K172727) | Discussion/Conclusion |
| Working Length | 1600 mm, 1800 mm, 2100 mm, 2300 mm | 1650 mm, 1950 mm, 2350 mm | Difference, see Discussion 1 |
| Clip Opening Width | 9 mm, 11 mm, 14 mm, 17 mm | 9 mm, 11 mm, 13 mm | Difference, See Discussion 2 |
| Clip material | Stainless Steel (SUS631) | Stainless Steel (SUS631) | Same |
| Sterilization | Method: Ethylene Oxide sterilization<br>SAL: 10-6 | Method: Ethylene Oxide sterilization<br>SAL: 10-6 | Same |
| Single use/reuse | For single use | For single use | Same |
| Mechanical performance | Surface roughness/Surface hardness/Clip Assembly Repeated<br>Open/Close performance/Clip retention force/Clip releasing force/360° rotation performance/Clamp Close Force/Clamp strength-stability/Separation Force/Clip approach/Metal part against corrosion | Open and Close/Clip releasing force/Clamping strength/Tensile Strength/Separation Force/ Rotation property/Corrosion | Difference, See Discussion 3 |
| Biocompatibility | FDA guidance and ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process” | ISO 10993 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing" | Same |
| MR Environment | MR Unsafe | Unknown | Difference, See Discussion 4 |
VI. Comparison of technological characteristics with the predicate device
{7}------------------------------------------------
## Discussion on differences between the subject device and the predicate device
Discussion 1: There are two specifications (Φ2.6mm) related to outer diameter of the clip for the subject device. \$2.6mm specification is same as the predicate device. Φ2.4mm specification is applicable for smaller endoscopic working channel, different from the predicate device. As the size of the outer diameter is designed to be compatible with the smaller size of the endoscopic working channel during the surgical procedure, in addition, the @2.4mm specification has been page 5 of 10
{8}------------------------------------------------
available on the other marketed device (K213338). Therefore, such differences will not affect clinical performance and safety of the subject device.
Discussion 2: The working length specification of the subject device is different from ones of the predicate device. The different working length designed is applicable for different endoscope length to reach the location of the lesion or the location as the operator expected, and the maximum working length of the subject device is shorter than the maximum one of the predicate device, therefore, such differences will not affect clinical performance and safety of the subject device.
Discussion 3: There are slightly different and more optional specifications of clip opening width on the subject device than the predicate device. The different opening width is designed appropriately for closure of tissues with different thickness and there are many optional specification available on the legally marketed devices, like 22 mm specification is available on Lockado™ Repositionable Hemostasis Clip (K202333). In addition, mechanical performance regarding the clamp between the subject device and the predicate device has been performed and the test results demonstrate substantial equivalence between the subject device and the predicate device. Therefore, such differences will not affect clinical performance and safety of the subject device.
Discussion 4: The bench performance tests, including mechanical performance, were carried out on both the subject device and the predicate device according to specific product technical specification and the test results demonstrate substantial equivalence between the subject device and the predicate device. therefore, such difference will not affect effectiveness and safety of the subject device.
Discussion 5: The disposable hemoclip is MR Unsafe. And the MR Environment for the predicate device is unknown. As warning related to MR Unsafe has been added in the IFU to protect the devices from use in MR environment, such difference will not affect effectiveness and safety of the subject device.
#### VII. Summary of substantial equivalence discussion
Based on the above comparison table as well as discussion on differences, the subject device and the predicate device have similar design features and performance page 6 of 10
{9}------------------------------------------------
specifications. Although there are some differences on size specification parameters (e.g. applicable endoscopic working channel, working length, outer diameter of the clip and clip opening width) on the subject device and predicate device, such differences will not affect the effectiveness and safety of the subject device. In addition, a performance comparison testing between the subject device and the predicate device is implemented and all performances including mechanical property are verified to confirm the performance of the subject device is substantially equivalent to the predicate device. These technological differences between the subject device and the predicate device do not affect the safety and effectiveness of the subject device when used as labeled.
#### VIII. Summary of Non-clinical Data
Non-clinical testing for Disposable Hemoclip was conducted to verify that the subject device met all design specifications, demonstrated safety and essential performance based on current applicable standards, and to demonstrate substantial equivalence to the predicate. The following tests were performed.
#### > Biocompatibility
The biocompatibility evaluation for the Disposable Hemoclip was conducted in accordance with the FDA guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'" September 4, 2020, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
Clip part of the Disposable Hemoclip is released during the operation and will be retained and directly in contact with patient's tissue for a prolong contact duration (> 24 hours to 30 days), therefore, the following tests are considered and passed:
- MTT Cytotoxicity Test: ISO 10993-5:2009, Biological evaluation of medical . devices -- Part 5: Tests for In Vitro cytotoxicity
- Skin Sensitization Test, Intracutaneous Test: ISO 10993-10:2010, Biological ● evaluation of medical devices -- Part 10: Tests for irritation and sensitization
- Acute Systemic Toxicity Test, Subacute Systemic Toxicity Test, Subchronic .
{10}------------------------------------------------
Systemic Toxicity Test and Pyrogen Test: ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- Genotoxicity Test: ISO 10993-3:2014 Biological evaluation of medical . devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- Muscle Implant Test: ISO 10993-6: 2016 Biological evaluation of medical . devices - Part 6: Tests for local effects after implantation
- A Sterilization Validation
The EO sterilization of the Disposable Hemoclip has been validated according to the following applicable standards:
- . ISO11135:2014+A1:2018 Sterilization of medical device- validation and routine control of ethylene oxide sterilization
- ISO 11737-2:2019 Sterilization of Medical Device-Microbiological . methods-part 2: Tests of sterility performed in the validation of a sterilization process
- ISO 10993-7:2008 Biological evaluation of medical devices Part 7: ● Ethylene oxide sterilization residuals
- USP <85> Bacterial endotoxins test .
- A Shelf Life and Sterile Barrier System
Shelf Life and Sterile Barrier System of the Disposable Hemoclip has been validated according to the following applicable standards:
- ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier . Systems for Medical Devices
- . ISO11607-1:2019 Packaging for terminally sterilized Medical Device Part 1: Requirement for materials, sterile barrier systems and packaging systems
- . ISO11607-2:2019 Packaging for terminally sterilized Medical Device Part 2: Validation Requirement for forming, sealing and assembly process
- . ASTM F 1929-15: Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ASTM D 3078-02(2013): Standard Test Method for Determination of Leaks . page 8 of 10
{11}------------------------------------------------
in Flexible Packaging by Bubble Emission
- DIN 58593-6: 2016 Sterilization Sterile Supply Part 6: Microbial Barrier . Testing of Packaging Materials for Medical Devices Which Are to Be Sterilized
- . ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM D4169-16 Standard practice for performance testing of shipping . containers and systems (DC-13, Level II)
- > Performance Data – Bench
The performance tests were implemented on both the Disposable Hemoclip (subject device) and the predicate device (AGS Hemoclip) to demonstrate substantial equivalence according to the specific product specification with the following test items:
- Surface roughness & Surface hardness
- . Clip Assembly Repeated Open/Close Performance
- . Clip Retention Force
- . Clip Releasing Force
- . 360° Rotation performance
- Clip Close Force
- Clamp strength-durability
- . Clamp strength-stability
- . Separation Force
- . Clip approach
- . Metal Part Against Corrosion
- Performance Data Animal A
N/A, no animal studies are available for the subject device.
## IX. Summary of Clinical Data
N/A, no clinical studies are available for the subject device.
#### X. Conclusions
In conclusion, the technological characteristics, features, specifications, materials,
{12}------------------------------------------------
principle of operation, and intended use of the subject device substantially equivalent to the predicate devices quoted above. The differences between the subjective device and its predicate device do not introduce a new intended use and do not raise new issues of safety and effectiveness. Verification and Validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended. From the results of nonclinical testing described, it can be concluded that the subject device is substantially equivalent to the legally marketed predicate device.