K231588 · ABBOTT MEDICAL · OWQ · Jun 30, 2023 · Cardiovascular
Device Facts
Record ID
K231588
Device Name
ViewFlex™ Eco Reprocessed ICE Catheter
Applicant
ABBOTT MEDICAL
Product Code
OWQ · Cardiovascular
Decision Date
Jun 30, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ViewFlex™ Eco Reprocessed ICE Catheter is a temporary intracardiac ultrasound catheter intended for use in patients to accurately visualize cardiac structures, blood flow and other devices within the heart when connected to a compatible intracardiac ultrasound console via the compatible ViewFlex™ Catheter Interface Module. Examples of the types of other devices that can be visualized include, and are not limited to, intracardiac catheters, septal occluders, delivery wires, delivery sheaths, sizing balloons and transseptal needles. The use of these images is limited to visualization with no direct or indirect diagnostic use.
Device Story
Intracardiac ultrasound catheter; 90cm length; 9F shaft; ultrasound transducer at tip; four-directional deflection (Left-Right, Posterior-Anterior, ≥120 degrees). Used in clinical settings (e.g., cardiac procedures) by physicians to visualize heart structures, blood flow, and interventional devices (e.g., septal occluders, transseptal needles). Connects to compatible consoles (ViewMate Z, ViewMate, ViewMate Multi, Philips CX50) via interface module. Provides real-time ultrasound imaging for visualization; no direct diagnostic use. Reprocessed by Abbott (max 2 cycles); proximal handle/connector markings identify status; device retired after max cycles. Benefits include intra-procedural guidance for complex cardiac interventions.
Clinical Evidence
No clinical investigations performed. Substantial equivalence demonstrated via non-clinical design verification, validation, and risk assessment.
Technological Characteristics
9F diagnostic intravascular ultrasound catheter; 90cm usable length; 4-way tip deflection (≥120°). Compatible with ViewMate and Philips CX50 consoles. Reprocessed by manufacturer (max 2 cycles). Complies with EN ISO 14971:2019 risk management standards.
Indications for Use
Indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 30, 2023
Abbott Medical Quynh Phuong Le Regulatory Affairs Manager 14901 Deveau Place Minnetonka, Minnesota 55345
Re: K231588
Trade/Device Name: ViewFlex™ Eco Reprocessed ICE Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OWQ, OBJ Dated: May 31, 2023 Received: June 1, 2023
Dear Ouynh Phuong Le:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## K231588 - Quynh Phuong Le
The following device is included in the scope of this 510(k) submission:
| Description | Reference<br>Model<br>Number<br>(Predicate) | Reprocessed<br>Model<br>Number<br>(Subject) | Usable<br>Length | French<br>Size | System<br>Compatibility |
|----------------------------------------------|---------------------------------------------|---------------------------------------------|------------------|----------------|----------------------------------------------------------------|
| ViewFlex™ Eco<br>Reprocessed ICE<br>Catheter | D087031-R | D087031-R | 90cm | 9F | ViewMate™Z<br>ViewMate™<br>ViewMate™<br>Multi<br>Philips CX 50 |
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K231588
Device Name ViewFlex™ Eco Reprocessed ICE Catheter
Indications for Use (Describe)
The ViewFlex™ Eco Reprocessed ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.
Type of Use (Select one or both, as applicable)
| <label><input checked="checked" type="checkbox"/> Commercial Use of a Pesticide for Producing an Agricultural Commodity</label> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> For Hire Structural Pest Control</label> |
|× | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Abbott logo. The logo consists of a blue square with a stylized "a" inside, followed by the word "Abbott" in black, bold font. The logo is simple and modern, and the colors are clean and professional.
# 510(k) Summary
This 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92.
| 510(k) Number | K231588 |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Type | Special 510(k) |
| Date Prepared | 31 May 2023 |
| Manufacturer Name &<br>Address | Abbott Medical<br>14901 DeVeau Place<br>Minnetonka, MN 55345-2126, USA |
| Contact Person | Quynh Phuong Le<br>Regulatory Affairs Manager<br>949-769-5058 or 949-469-9779<br>quynphuong.le@abbott.com |
| Trade Name | ViewFlex™ Eco Reprocessed ICE Catheter |
| Common Name | ICE Catheter |
| Class | II |
| Classification Name | 870.1200, Diagnostic Intravascular Catheter |
| Product Code | OWQ |
| Subsequent Product Code | OBJ |
| Predicate Device | ViewFlex™ Eco Reprocessed ICE Catheter (K230934) |
| Device Description | The ViewFlex™ Eco Reprocessed ICE Catheter is a temporary<br>intracardiac ultrasound catheter intended for use in patients to<br>accurately visualize cardiac structures, blood flow and other<br>devices within the heart when connected to a compatible<br>intracardiac ultrasound console via the compatible ViewFlex™<br>Catheter Interface Module. Examples of the types of other devices<br>that can be visualized include, and are not limited to, intracardiac<br>catheters, septal occluders, delivery wires, delivery sheaths, sizing<br>balloons and transseptal needles. The use of these images is limited<br>to visualization with no direct or indirect diagnostic use. The<br>ViewFlex™ Eco Reprocessed ICE Catheter has a useable length<br>of 90 cm, with a 9 French (F) shaft with an ultrasound transducer.<br>A 10F introducer is recommended for use with this catheter for<br>insertion into the femoral or jugular veins. The catheter tip has<br>four-directional deflection allowing for Left-Right and Posterior-<br>Anterior deflection, with an angle of at least 120 degrees in each<br>direction.<br>The ViewFlex™ Eco Reprocessed ICE Catheter is compatible<br>with the ViewMate™ Z, ViewMate™, ViewMate™ Multi and<br>Philips CX50 ultrasound consoles.<br>The ViewFlex™ Eco Reprocessed ICE Catheter is reprocessed by<br>Abbott no more than two (2) times. Each catheter includes marking<br>on the proximal handle and connector that identify the catheter<br>status. The device is taken out of service after reaching the<br>maximum number of reprocessing cycles. Abbott restricts its<br>reprocessing to exclude devices previously reprocessed by other<br>reprocessors. |
| Indications for Use | The ViewFlex™ Eco Reprocessed ICE Catheter is indicated for<br>use in adult and adolescent pediatric patients to visualize cardiac<br>structures, blood flow and other devices within the heart. |
| Comparison | The subject and predicate devices have the same intended use,<br>indications for use and method of operation. There are no<br>changes to the claims, clinical applications, patient populations,<br>or performance specifications. The subject device has added an<br>additional reprocessing cycle after clinical use of the predicate<br>device cleared under K230934.<br>The differences between the subject device, ViewFlex™ Eco<br>Reprocessed ICE Catheter (2X), and the predicate device,<br>ViewFlex™ Eco Reprocessed ICE Catheter (1X), does not raise<br>any new questions of safety and effectiveness. |
| Non-Clinical Testing<br>Summary | Design verification activities were performed with their respective<br>acceptance criteria to ensure that the use of the subject device does<br>not affect the safety or effectiveness of the device. All testing<br>performed met the established performance specifications.<br><br><b>Testing</b><br>The ViewFlex™ Eco Reprocessed ICE Catheter was developed<br>and tested in accordance with the following industry guidance<br>documents and standards:<br>- Guidance for Industry and FDA staff: Medical Device<br>User Fee and Modernization Act of 2002, Validation Data<br>in Premarket Notification Submissions (510(k)s) for<br>Reprocessed Single-Use Medical Devices<br>- EN ISO 14971:2019 Medical Devices – Application of<br>Risk Management to Medical Devices<br><br><b>Types of Testing Performed</b><br>- Biocompatibility<br>- Design validation<br>- Design verification<br>- Cleaning Validation<br><br><b>Risk Management</b><br>Abbott has reviewed the applicable risk management files, and<br>there were no new or modified hazards identified as result of the<br>proposed modification. |
| Clinical Testing | No clinical investigations have been performed for the subject or<br>predicate devices. Clinical data is not required for the<br>demonstration of substantial equivalent based on the risk<br>assessment in Section 10 and verification in Section 12 and<br>validation in Section 13 summarized in this 510(k). |
| Conclusion | The technological characteristics for the subject device, and the<br>indications for use are the same as the predicate device. Based on<br>this and the data provided in this pre-market notification, the<br>subject device and predicate device have been demonstrated to be<br>substantially equivalent. |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Abbott logo. The logo consists of a blue abstract shape on the left and the word "Abbott" in black font on the right. The abstract shape is a stylized letter "A".
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Abbott, a healthcare company. The logo consists of a stylized blue symbol on the left, resembling a rounded square with a missing corner. To the right of the symbol is the word "Abbott" in bold, black font. The overall design is clean and corporate, reflecting the company's professional image.
The following device is included in the scope of this 510(k) submission:
| Description | Reference<br>Model<br>Number<br>(Predicate) | Reprocessed<br>Model<br>Number<br>(Subject) | Usable<br>Length | French<br>Size | System<br>Compatibility |
|----------------------------------------------|---------------------------------------------|---------------------------------------------|------------------|----------------|-----------------------------------------------------------------|
| ViewFlex™ Eco<br>Reprocessed ICE<br>Catheter | D087031-R | D087031-R | 90cm | 9F | ViewMate™ Z<br>ViewMate™<br>ViewMate™<br>Multi<br>Philips CX 50 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.