← Product Code [DSK](/productcode/DSK) · K231586

# Endophys Blood Pressure Monitor model BPM-30 (K231586)

_Endophys Technologies, LLC · DSK · Nov 3, 2023 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K231586

## Device Facts

- **Applicant:** Endophys Technologies, LLC
- **Product Code:** [DSK](/productcode/DSK.md)
- **Decision Date:** Nov 3, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1110
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Endophys Blood Pressure Monitor Model BPM-30 is intended to continuously provide systolic, diastolic and mean blood pressure based on the output of the Endophys PSS3 Pressure Sensing Sheath in patients undergoing therapeutic and/or diagnostic procedures involving percutaneous vascular access. The Endophys Blood Pressure Monitor Model BPM-30 monitor is additionally intended for use in transport situations within hospital environments.

## Device Story

BPM-30 is an electronic monitor for invasive blood pressure; inputs fiber optic signals from Endophys PSS3 Pressure Sensing Sheath; transforms optical data into electrical signals emulating conventional fluidic arterial transducers; outputs systolic, diastolic, and mean blood pressure values to standard physiological patient monitors and its own display; used in clinical settings (minimally invasive vascular procedures, critical care, hospital transport); operated by clinicians; provides real-time hemodynamic data to support clinical decision-making; benefits patients by enabling accurate, continuous pressure monitoring without requiring additional fluid lines or catheters.

## Clinical Evidence

Bench testing only. Evidence includes software verification, electrical safety and essential performance testing (IEC60601-1, IEC60601-1-8, IEC60601-2-34), EMC testing (IEC60601-1-2), packaging design verification, and functional testing. No clinical data provided.

## Technological Characteristics

Fiber optic pressure sensing; pressure range -30 to 300 mmHg; accuracy ±1 mmHg plus ±1% (≤50 mmHg) and ±3% (≥50 mmHg). Powered by AC adapter with internal battery (30-minute backup). Emulates electrical interface of conventional fluidic transducers. Connectivity: fiber optic input, electrical output to standard patient monitors. Standards: IEC60601-1, IEC60601-1-8, IEC60601-2-34, IEC60601-1-2.

## Regulatory Identification

A blood pressure computer is a device that accepts the electrical signal from a blood pressure transducer amplifier and indicates the systolic, diastolic, or mean pressure based on the input signal.

## Predicate Devices

- Endophys Blood Pressure Monitor model BPM-20 ([K160945](/device/K160945.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 3, 2023

Endophys Technologies LLC % Mike Winegar Principal Winegar Consulting, Inc. 7829 Ithaca Ln N Maple Grove, Minnesota 55311

Re: K231586

Trade/Device Name: Endophys Blood Pressure Monitor Model BPM-30 Regulation Number: 21 CFR 870.1110 Regulation Name: Blood Pressure Computer Regulatory Class: Class II Product Code: DSK Dated: May 30, 2023 Received: May 31, 2023

#### Dear Mike Winegar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Form Approved: OMB No. 0910-0120 Food and Drug Administration Expiration Date: 6/30/2023 See PRA Statement below. Indications for Use

510(k) Number (if known)

K231586

Device Name

#### Endophys Blood Pressure Monitor Model BPM-30

Indications for Use (Describe)

#### Indications for Use

The Endophys Blood Pressure Monitor Model BPM-30 is intended to continuously provide systolic, diastolic and mean blood pressure based on the output of the Endophys PSS3 Pressure Sensing Sheath in patients undergoing therapeutic and/or diagnostic procedures involving percutaneous vascular access. The Endophys Blood Pressure Monitor Model BPM-30 monitor is additionally intended for use in transport situations within hospital environments.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Endophys. The logo features the word "Endophys" in bold, with the "Endo" portion in black and the "phys" portion in red. A red line resembling a heartbeat is to the left of the word. Below the word, there is a tagline that reads "Measuring physiology where it happens."

## 510(k) SUMMARY

| Applicant           | Endophys Technologies, LLC<br>1933 E Levee St, Suite B<br>Dallas, TX 75207                                                                                              |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person      | Mike Winegar<br>Principal<br>Winegar Consulting, Inc.<br>763-639-0700<br>mwinegar@comcast.net                                                                           |
| Summary Date        | September 29, 2023                                                                                                                                                      |
| Proprietary Name    | Endophys Blood Pressure Monitor model BPM-30                                                                                                                            |
| Classification      | Class II                                                                                                                                                                |
| Classification Name | Computer, Blood-Pressure                                                                                                                                                |
| Regulation Number   | 21 CFR 870.1110                                                                                                                                                         |
| Product Code        | DSK                                                                                                                                                                     |
| Predicate Device    | The Endophys Blood Pressure Monitor model BPM-<br>30 is substantially equivalent to the currently<br>marketed Endophys Blood Pressure Monitor model<br>BPM-20 (K160945) |

{4}------------------------------------------------

#### Device Description:

The Endophys BPM-30 Blood Pressure Monitor is an electronic device that provides compatibility between a physiological fiber optic blood pressure sensor (transducer) and conventional invasive arterial blood pressure inputs to a standard physiological patient monitor. The device converts the optical transducer data to electrical signals that are interpreted by a conventional patient monitor and/or are displayed directly on the BPM-30. The BPM-30 accurately emulates a fluidic arterial blood pressure transducer and supplies electrical signals to its output that are indistinguishable from a conventional fluidic blood pressure transducer.

The BPM-30 is implemented as a self-contained unit that has a fiber optic transducer connection as an input source and communicates with a patient monitor as its output. The BPM acts to directly emulate the electrical interface characteristics of conventional fluidic blood pressure transducers (that patient monitors are compatible with) while providing much more precise blood pressure data derived from a fiber optic transducer placed within an artery. Electrically emulating a conventional fluidic transducer uniquely allows a fiber optic pressure sensor to be used with a wide variety of existing physiological patient monitors without modification of those monitors. Systolic, diastolic, and mean blood pressure values are also displayed directly on the BPM-30 every four seconds.

The BPM-30 is powered by a standard AC power adapter. The BPM-30 also has an internal battery supply that will automatically recharge and maintain operation during brief (up to 30 minutes) power interruptions. The BPM-30 is used outside of the sterile environment and has standard alerts and alarms.

The BPM-30 has an operating pressure range of -30 to 300 mmHg with an accuracy of ±1 mmHg plus ±1% of reading for pressures ≤ 50 mmHg and within ±3% of reading for pressures ≥ 50 mmHg

#### Intended Use:

The Endophys BPM-30 Blood Pressure Monitor is meant to be used in minimally invasive vascular procedures and critical patient care situations where the accuracy and timeliness of arterial blood pressure measurements are important. It supports fiber optic transducers that are incorporated into the Endophys PressureSense™ hydrophilic coated Pressure Sensing Sheath (PSS3).

When used with the Endophys Pressure Sensing Sheath, the BPM-30 is designed to provide continuous and accurate systolic, diastolic and mean blood pressure data with no other catheter, pressure transducer or fluid line required.

{5}------------------------------------------------

### Indications for Use:

The Endophys Blood Pressure Monitor Model BPM-30 is intended to continuously provide systolic, diastolic and mean blood pressure based on the output of the Endophys PSS3 Pressure Sensing Sheath in patients undergoing therapeutic and/or diagnostic procedures involving percutaneous vascular access. The Endophys Blood Pressure Monitor Model BPM-30 monitor is additionally intended for use in transport situations within hospital environments.

## Summary of Technical Characteristics as Compared to the Predicate Device:

Blood Pressure Monitors, by nature, are instruments intended to provide clinicians with the physiological blood pressure of patients reported in milligrams of mercury (mmHg). Specialized computer based instruments have been developed which allow active monitoring of blood pressure during interventional procedures. Blood pressure monitoring using a fiber optic communication is the technological principle for both the subject and predicate devices. The technology requires use of a compatible intravascular pressure sensing sheath to provide this data to the clinician.

At a high level, the subject and predicate devices are based on the following same technological elements:

- . Calculates and displays hemodynamic (invasive blood pressure) data when attached to a compatible intravascular pressure sensing device (Endophys PSS3 Pressure Sensing Sheath)
- Optical signal (fiber optic) measurement of blood pressure ●
- Pressure measurement range -30 to 300 mmHg ●
- Pressure accuracy: System accuracy ±1 mmHg plus ±1% of reading for pressures ≤ 50 mmHg ● and within ±3% of reading for pressures ≥ 50 mmHg

The following technological differences exist between the subject and predicate devices:

- Modification of device design to be compatible with FISO Technologies Inc. Fabrey-Perot 0 fiber optic pressure sensors and circuitry incorporated into the Endophys Pressure Sensing Sheath (PSS3)
- Combine the functionality of the daughterboard onto the main circuit board, which were . separate in the predicate device
- Addition of an internal battery supply that will automatically recharge and maintain 0 operation during brief (up to 30 minutes) patient transport or power interruptions

{6}------------------------------------------------

## Performance Data

The following performance data were provided in support of the substantial equivalence:

- o Software Verification
- Electrical Safety and Essential Performance Testing (IEC60601-1, IEC60601-1-8, o IEC60601-2-34)
- EMC Testing (IEC60601-1-2)
- Packaging Design Verification ●
- Package, Device, and Labeling Inspection ●
- Functional Testing ●

The modified Endophys Blood Pressure Monitor met all specified criteria and did not raise new safety or performance questions. Based on the design verification performance the modified Endophys Blood Pressure Monitor was found to have the safety and effectiveness profile that is similar to the predicate device.

## Conclusion:

Endophys believes the proposed BPM-30 Blood Pressure Monitor is substantially equivalent to the legally marketed predicate device. The indications for use, methods of operation, design and materials used are either identical or substantially equivalent to the existing legally marketed predicate product.

---

**Source:** [https://fda.innolitics.com/device/K231586](https://fda.innolitics.com/device/K231586)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
