← Product Code [BTY](/productcode/BTY) · K231561

# Pulmonary Function Tester, Model: A9 (K231561)

_Guangzhou Homesun Medical Technology Co., Ltd. · BTY · Feb 21, 2024 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K231561

## Device Facts

- **Applicant:** Guangzhou Homesun Medical Technology Co., Ltd.
- **Product Code:** [BTY](/productcode/BTY.md)
- **Decision Date:** Feb 21, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1890
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Pediatric

## Indications for Use

The Pulmonary Function Tester (Model: A9) is intended to be used for measurement and data collection of lung function parameters. The system performs cooperation-dependent flow-volume measurements. Measurements shall be performed under the direction of a physician in the clinic, doctors' office or hospital. The Pulmonary Function Tester (Model: A9) is intended for use in patients from 4 years of age and older who can understand and perform instructions of the physician.

## Device Story

Hand-held pulmonary function tester; measures exhaled/inhaled air flow and volume via flow sensor; uses mouth-blowing method. Inputs: patient respiratory flow/volume. Processing: flow sensor data integration to calculate pulmonary ventilation indicators (e.g., FVC, SVC, MVV). Outputs: volume-time and flow-volume curves, calculated parameters displayed on LCD screen or connected computer/mobile software. Used in clinics/hospitals under physician direction. Benefits: enables assessment of lung function parameters for respiratory monitoring.

## Clinical Evidence

Bench testing only. Device performance verified against ISO 26782 (spirometers) and ISO 23747 (peak flow meters). Biocompatibility testing performed per ISO 10993-1, -5, -10. Gas pathway safety verified per ISO 18562-1, -2, -3, -4. Electrical safety and EMC verified per IEC 60601-1 and IEC 60601-1-2.

## Technological Characteristics

Hand-held device; ABS plastic and 304 stainless steel construction. Sensing: differential pressure sensor/pneumotach. Energy: 3.7V 1500mAh Li-ion battery or USB power. Connectivity: Bluetooth/USB. Software: Basic documentation level. Standards: ISO 26782, ISO 23747, IEC 60601-1, IEC 60601-1-2.

## Regulatory Identification

A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.

## Predicate Devices

- MASTERSCREEN PNEUMO & MASTERSCOPE ([K071753](/device/K071753.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 21, 2024

Guangzhou Homesun Medical Technology Co., Ltd % Tracy Che Registration engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90, Qianhai Road Shenzhen, Guangdong 518052 China

Re: K231561

Trade/Device Name: Pulmonary Function Tester, Model: A9 Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: Class II Product Code: BTY Dated: January 17, 2024 Received: January 17, 2024

Dear Tracy Che:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Rachana Visaria -S

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices

{2}------------------------------------------------

OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231561

Device Name Pulmonary Function Tester, Model: A9

#### Indications for Use (Describe)

The Pulmonary Function Tester (Model: A9) is intended to be used for measurement and data collection of lung function parameters. The system performs cooperation-dependent flow-volume measurements shall be performed under the direction of a physician in the clinic, doctors' office or hospital. The Pulmonary Function Tester (Model: A9) is intended for use in patients from 4 years of age and older who can understand and perform instructions of the physician.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510 (k) Summary

This "510(k) Summary" of 510(k) safety and effectiveness information is submitted in accordance with requirements of Title 21, CFR Section 807.92.

# (1) Applicant information

| 510 (k) owner's name:     | Guangzhou Homesun Medical Technology Co., Ltd                                                      |
|---------------------------|----------------------------------------------------------------------------------------------------|
| Address:                  | Floor 7th, Tianxiang Business Building, No.28, Li Fu Road<br>Haizhu District, Guangzhou, GD. China |
| Contact person:           | Suijie Huang                                                                                       |
| Phone number:             | +86 13826204942                                                                                    |
| Email:                    | hsj@huxijia.com                                                                                    |
| Date of summary prepared: | February 20, 2024                                                                                  |

## (2) Proprietary name of the device

| Trade name/Model:  | Pulmonary Function Tester, model: A9           |
|--------------------|------------------------------------------------|
| Regulation name:   | Predictive Pulmonary Function Value Calculator |
| Regulation number: | 21 CFR 868.1890                                |
| Product code       | BTY                                            |
| Review panel:      | Anesthesiology                                 |
| Regulation class:  | Class II                                       |

# (3) Predicate device

| Sponsor               | VIASYS HEALTHCARE GMBH            |
|-----------------------|-----------------------------------|
| Device Name and Model | MASTERSCREEN PNEUMO & MASTERSCOPE |
| 510(k) Number         | K071753                           |
| Product Code          | BTY                               |
| Regulation Number     | 21 CFR 868.1890                   |
| Regulation Class      | II                                |

# (4) Description/ Design of device

The Pulmonary Function Tester (Model: A9) is a hand-held pulmonary function testing device which can be used by patients older than 4 years old under the direction of a physician in the clinic, doctors' office or hospital. The Pulmonary Function Tester (Model: A9) is comprised of the main unit (display screen, sensor and communication module), USB cable, application software (computer software, mobile software), and a power supply. The device use is very simple, it adopts mouth blowing method. The device uses a flow sensor to measure the air flow and volume of the 

{5}------------------------------------------------

patient's exhaled or inhaled air. According to the volume-time curve and the flow-volume curve, the pulmonary ventilation indicators of human respiratory physiology, such as slow vital capacity, maximum minute ventilation and forced vital capacity, are estimated and reported on the LCD screen and computer/mobile software. The pulmonary Function Tester is powered by lithium battery (DC 3.7V, 1500mAh) or tablet/computer USB port, or the equipped power supply.

## (5) Indications for Use

The Pulmonary Function Tester (Model: A9) is intended to be used for measurement and data collection of lung function parameters. The system performs cooperation-dependent flow-volume measurements. Measurements shall be performed under the direction of a physician in the clinic, doctors' office or hospital. The pulmonary Function Tester (Model: A9) is intended for use in patients from 4 years of age and older who can understand and perform instructions of the physician.

| Item                                    | Subject device                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                   | Remark                                                                                                                                                                                                                                                 |                                                              |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Trade name                              | Pulmonary Function Tester<br>(Model A9)                                                                                                                                                                                                                                                                                                                                                                                      | MASTERSCREEN PNEUMO<br>& MASTERSCOPE                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                        |                                                              |
| 510 (k) number                          | K231561                                                                                                                                                                                                                                                                                                                                                                                                                      | K071753                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                        |                                                              |
| Regulation number                       | 21 CFR 868.1890                                                                                                                                                                                                                                                                                                                                                                                                              | 21 CFR 868.1890                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                   |                                                              |
| Regulation description                  | Predictive Pulmonary Function Value Calculator                                                                                                                                                                                                                                                                                                                                                                               | Predictive Pulmonary Function Value Calculator                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                   |                                                              |
| Product code                            | BTY                                                                                                                                                                                                                                                                                                                                                                                                                          | BTY                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                   |                                                              |
| Class                                   | II                                                                                                                                                                                                                                                                                                                                                                                                                           | II                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                   |                                                              |
| Indications for use/ Intended use       | The Pulmonary Function Tester (Model: A9) is intended to be used for measurement and data collection of lung function parameters. The system performs cooperation-dependent flow-volume measurements. Measurements shall be performed under the direction of a physician in the clinic, doctors' office or hospital. The Pulmonary Function Tester (Model: A9) is intended for use in patients from 4 years of age and older | The Master Screen Pneumo & Masterscope is intended to be used for measurement and data collection of lung function parameters. The system performs cooperation-dependent flow volume measurements. Mostly it will be used for COPD and Asthma patients. Measurements will be performed under the direction of a physician in the clinic, doctors office or hospital. It can be utilized for patients from 4 years of age and older | Similar                                                                                                                                                                                                                                                |                                                              |
|                                         | Item                                                                                                                                                                                                                                                                                                                                                                                                                         | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate device                                                                                                                                                                                                                                       | Remark                                                       |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | can understand<br>who<br>and                                                                                                                                                                                                                                                                                                                                                                                                       | as long as they can cooperate                                                                                                                                                                                                                          |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | perform instructions of the                                                                                                                                                                                                                                                                                                                                                                                                        | in the performance.                                                                                                                                                                                                                                    |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | physician.                                                                                                                                                                                                                                                                                                                                                                                                                         | The Master Screen Pneumo is                                                                                                                                                                                                                            |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    | powered from 100 - 240 V / 50                                                                                                                                                                                                                          |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    | - 60 Hz wall outlets. No energy                                                                                                                                                                                                                        |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    | is transferred to the patient.                                                                                                                                                                                                                         |                                                              |
|                                         | Patient population                                                                                                                                                                                                                                                                                                                                                                                                           | The device can be utilized for                                                                                                                                                                                                                                                                                                                                                                                                     | The device can be utilized for                                                                                                                                                                                                                         | Same                                                         |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | patients from 4 years of age                                                                                                                                                                                                                                                                                                                                                                                                       | patients from 4 years of age                                                                                                                                                                                                                           |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | and older as long as they can                                                                                                                                                                                                                                                                                                                                                                                                      | and older as long as they can                                                                                                                                                                                                                          |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | cooperate in the performance.                                                                                                                                                                                                                                                                                                                                                                                                      | cooperate in the performance.                                                                                                                                                                                                                          |                                                              |
|                                         | Environment of use                                                                                                                                                                                                                                                                                                                                                                                                           | Clinic, doctors' office<br>or                                                                                                                                                                                                                                                                                                                                                                                                      | Clinic.<br>doctors'<br>office<br>or                                                                                                                                                                                                                    | Same                                                         |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | hospital                                                                                                                                                                                                                                                                                                                                                                                                                           | hospital                                                                                                                                                                                                                                               |                                                              |
|                                         | Prescription or OTC                                                                                                                                                                                                                                                                                                                                                                                                          | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                       | Prescription                                                                                                                                                                                                                                           | Same                                                         |
|                                         | Basic unit specification                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                        |                                                              |
| Power supply                            |                                                                                                                                                                                                                                                                                                                                                                                                                              | Internal<br>battery:<br>3.7V,                                                                                                                                                                                                                                                                                                                                                                                                      | 100 - 240 V / 50 - 60 Hz                                                                                                                                                                                                                               | Different                                                    |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | 1500mAh                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                        |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | DC input: 5V, 1.2A                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                        |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | Power supply: input :                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                        |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | 100-240V~,50-60Hz,0.3A ;                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                        |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | output: 5V===1.2A                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                        |                                                              |
| Dimensions                              |                                                                                                                                                                                                                                                                                                                                                                                                                              | 213*78*28mm                                                                                                                                                                                                                                                                                                                                                                                                                        | 292 * 198 * 64 mm                                                                                                                                                                                                                                      | Different                                                    |
| Components                              |                                                                                                                                                                                                                                                                                                                                                                                                                              | Mainly composed of main                                                                                                                                                                                                                                                                                                                                                                                                            | Mainly composed of Handle                                                                                                                                                                                                                              | Similar                                                      |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | unit, USB cable application                                                                                                                                                                                                                                                                                                                                                                                                        | with pneumotach and shutter,                                                                                                                                                                                                                           |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | software, and power supply.                                                                                                                                                                                                                                                                                                                                                                                                        | power unit and notebook                                                                                                                                                                                                                                |                                                              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    | (software).                                                                                                                                                                                                                                            |                                                              |
|                                         | Performance specification                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                        |                                                              |
| Technology                              | for                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                        | Same                                                         |
| measure<br>volume                       | flow<br>and                                                                                                                                                                                                                                                                                                                                                                                                                  | Differential pressure sensor                                                                                                                                                                                                                                                                                                                                                                                                       | Differential pressure                                                                                                                                                                                                                                  |                                                              |
| Flow                                    | Type                                                                                                                                                                                                                                                                                                                                                                                                                         | High Quality Pneumotach                                                                                                                                                                                                                                                                                                                                                                                                            | High Quality Pneumotach                                                                                                                                                                                                                                | Same                                                         |
| (PEF)                                   | Range                                                                                                                                                                                                                                                                                                                                                                                                                        | 0 – 16 L/s                                                                                                                                                                                                                                                                                                                                                                                                                         | 士 20 L/s                                                                                                                                                                                                                                               | Same                                                         |
|                                         | Accuracy                                                                                                                                                                                                                                                                                                                                                                                                                     | <=============================================================================================================================================================================                                                                                                                                                                                                                                                     | 0.1 to 14 L/s: ± 5% or ±                                                                                                                                                                                                                               | Similar,                                                     |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    | 0.2 L/s (whichever is greater)                                                                                                                                                                                                                         | conforms<br>to                                               |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                        | ISO 23747                                                    |
|                                         | Linearity                                                                                                                                                                                                                                                                                                                                                                                                                    | ≤5%                                                                                                                                                                                                                                                                                                                                                                                                                                | Not known                                                                                                                                                                                                                                              | Different but                                                |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                        | conforms<br>to                                               |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                        | ISO 23747                                                    |
|                                         | Frequency                                                                                                                                                                                                                                                                                                                                                                                                                    | ±0.25L/s<br>士<br>12%<br>or                                                                                                                                                                                                                                                                                                                                                                                                         | Not known                                                                                                                                                                                                                                              | Different but                                                |
|                                         | response                                                                                                                                                                                                                                                                                                                                                                                                                     | (whichever is greater)                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                        | conforms<br>to                                               |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                        | ISO 23747                                                    |
|                                         | Resolution                                                                                                                                                                                                                                                                                                                                                                                                                   | 0.01 L/s                                                                                                                                                                                                                                                                                                                                                                                                                           | 0.01 L/s                                                                                                                                                                                                                                               | Same                                                         |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              | < 0.15 kPa/(L/s) (not                                                                                                                                                                                                                                                                                                                                                                                                              | <0.05kPa/L/s (<0.5cm H2O/                                                                                                                                                                                                                              | Different but                                                |
|                                         | Resistance                                                                                                                                                                                                                                                                                                                                                                                                                   | including filter)                                                                                                                                                                                                                                                                                                                                                                                                                  | L/s) at 10 L/s                                                                                                                                                                                                                                         | conforms<br>to                                               |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                        | ISO 23747                                                    |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                        |                                                              |
|                                         | Item                                                                                                                                                                                                                                                                                                                                                                                                                         | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate device                                                                                                                                                                                                                                       | Remark                                                       |
|                                         | Type                                                                                                                                                                                                                                                                                                                                                                                                                         | Volume calculation through<br>integration, from flow<br>measurements                                                                                                                                                                                                                                                                                                                                                               | Volume calculation through<br>integration, from flow<br>measurements                                                                                                                                                                                   | Same                                                         |
|                                         | Range                                                                                                                                                                                                                                                                                                                                                                                                                        | 0 -10 L                                                                                                                                                                                                                                                                                                                                                                                                                            | $\pm$ 20 L                                                                                                                                                                                                                                             | Different but<br>conforms to<br>ISO 26782                    |
|                                         | Accuracy…

---

**Source:** [https://fda.innolitics.com/device/K231561](https://fda.innolitics.com/device/K231561)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
