LUMA LED patches (LUMA Blemish (TN2197) and LUMA Revive (TN2189))
K231555 · iSMART Developments, Ltd. · OHS · Jul 27, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231555
Device Name
LUMA LED patches (LUMA Blemish (TN2197) and LUMA Revive (TN2189))
Applicant
iSMART Developments, Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Jul 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LUMA Blemish is an over-the-counter device intended to emit energy in the blue, and red region of the light spectrum and is indicated to treat mild to moderate acne vulgaris of the face. LUMA Revive is an over-the-counter device intended to emit energy in the red and near infra-red region of the light spectrum and is indicated to treat full face wrinkles.
Device Story
LUMA LED patches are battery-operated, cordless, wearable phototherapy devices for home use. Devices consist of flexible silicone patches containing LEDs that emit narrow-band light: blue (415nm) and red (630nm) for acne; red (630nm) and near infra-red (830nm) for wrinkles. Patches are secured to the face using single-use adhesive hydrocolloid patches. Operation is controlled by a simple ON/OFF button; a 10-minute treatment timer automatically shuts off the device. Power is provided by an integrated 3.7V, 37mAh Lithium Polymer battery, rechargeable via a magnetic 2-pin to USB-A connector. The device includes a charging watchdog to prevent overcharge. Users operate the device independently; status is indicated by an LED on the patch surface. Photobiomodulation and endogenous photodynamic therapy mechanisms are utilized to treat skin conditions. The device does not perform measurements or diagnostic analysis.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety, electromagnetic compatibility, photobiological safety, battery safety, and biocompatibility (cytotoxicity, irritation, sensitization). Label comprehension study conducted with 26 subjects (mean age 31.7) confirmed safe and effective user understanding of instructions.
Indicated for individuals with mild to moderate acne vulgaris of the face (LUMA Blemish) or for the treatment of full-face wrinkles (LUMA Revive).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
MZ Skin LightMAX Supercharged LED Mask 2.0 (K213184)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
July 27, 2023
iSMART Developments Ltd Susan D'arcy Owner 129 Green Lanes, Boldmere, Sutton Coldfield Birmingham, West Midlands B73 5LT United Kingdom
Re: K231555
Trade/Device Name: LUMA LED patches (LUMA Blemish (TN2197) and LUMA Revive (TN2189)) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP, OHS Dated: May 30, 2023 Received: May 30, 2023
Dear Susan D'arcy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Jianting Wang -S" in a large, clear font. The text is horizontally oriented and appears to be a title or heading. The background is plain, with a watermark-like design faintly visible behind the text.
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231555
Device Name
LUMA LED light patches (LUMA Blemish and LUMA Revive)
### Indications for Use (Describe)
The LUMA LED light patches are over-the-counter devices intended to emit energy in the visible and near infra-red light spectrum.
LUMA Blemish is an over-the-counter device intended to emit energy in the red and blue region of the light spectrum, specifically indicated to treat mild to moderate acne vulgaris of the face.
LUMA Revive is an over-the-counter device intended to emit energy in the red and Near infra-red spectrum and is intended for the use in the treatment of facial wrinkles.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(k) Summary - K231555
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c).
### Submitter's Name: ISMART DEVELOPMENTS LTD
Submitter's Address: 129 Green Lanes, Sutton Coldfield, Birmingham B73 5LT
Contact Person: Susan D'Arcy
Telephone: +44 (0) 7880313315
Date Prepared: May 21st 2023
# Device Trade Name: LUMA LED patches (LUMA Blemish (TN2197) and LUMA Revive (TN2189))
Device Classification Information:
| Regulation<br>Number | Device<br>Classification name | Device<br>Class | Product<br>Code | Classification<br>Panel | Type |
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-----------------|------------------------------|------------------------|
| 21 CFR 878.4810<br>Laser surgical<br>instrument for<br>use in general<br>and plastic<br>surgery and in<br>dermatology. | LUMA Blemish<br>(TN2197)<br>Over-The-Counter<br>Powered Light-Based<br>Laser for Acne<br><br>LUMA Revive<br>(TN2189)<br>Light Based Over the<br>Counter Wrinkle<br>Reduction | Class 2 | OLP<br>OHS | General &<br>Plastic Surgery | Traditional<br>510 (k) |
{4}------------------------------------------------
#### 5.1. Device Description
The LUMA Patches (Blemish and Revive) are battery operated, cordless wearable Light emitting diode (LED) devices intended to emit an even, cool, narrow band of light in the blue (415nm), red (630nm) and near infra-red (830nm) spectrum for the treatment of mild to moderate acne vulgaris and facial wrinkles. The devices work through non-thermal mechanisms called photobiomodulation (wrinkles) and endogenous Photodynamic therapy (acne vulgaris).
LUMA Blemish and LUMA Revive consists of the following key components.
- 1. Silicone flexible patches
- 2. Adhesive hydrocolloid patches x14
- 3. Magnetic 2 pin, 4mm to USB A connector
- User manual 4.
The LUMA patches are home use wearable LED phototherapy devices composed of flexible patches containing Light emitting diodes (LEDs). The LEDs generate the light.
The LUMA patches are comprised of an endo skeleton made of silicone containing an upper surface made of Polyurethane (PU) that houses the ON/OFF button magnetic charging pins and a battery charging indicator and a lower surface manufactured from Polyethylene terephthalate (PET).
The LEDs produce blue, red and near infra-red (NIR) light in the visible spectrum.
LUMA Blemish (Blue: 415nm +/- 10nm and Red: 630nm +/- 10nm)
LUMA Revive (Red: 630nm +/- 10nm and NIR 830nm +/-10nm.).
The LEDs are driven by an integrated 3.7V, 37mAh Lithium Polymer battery.
The battery is charged via a magnetic 2 pin, 4mm to USB A connector that plugs into a standard 5v USB A power adaptor. The device cannot be operated while in charging mode. The device contains a charging watchdog that prevents overcharge. The battery status of the devices is relayed to the user by a single indicator LED on the upper surface of the LUMA patches (during charging) and by the treatment LEDs on the lower or inner surface (during treatment).
The LUMA patches contain a simple ON/OFF button on the upper surface of the patch that switches the LEDs ON. The treatment time is 10 minutes. The devices automatically turn OFF after 600 seconds (10 minutes). The user may stop the treatment program during the 10 minutes by pressing the ON/OFF button.
The LUMA patches are worn on the face and are held in place by an adhesive hydrocolloid patch.
{5}------------------------------------------------
The equipment is not used to make measurements of any sort, or to draw any conclusions regarding the indication to treat. The equipment does not require checks on the light output as the LEDs do not dim with age to any practical extent.
#### 5.2. Intended Use
LUMA Blemish is an over-the-counter device intended to emit energy in the blue, and red region of the light spectrum and is indicated to treat mild to moderate acne vulgaris of the face.
LUMA Revive is an over-the-counter device intended to emit energy in the red and near infra-red region of the light spectrum and is indicated to treat full face wrinkles.
#### 5.3. Substantial Equivalence
The LUMA Blemish and LUMA Revive are substantially equivalent to the primary predicate MZ Skin LightMAX Supercharged LED Mask 2.0 (K213184).
The LUMA Blemish and LUMA Revive are predicated against the MZ Skin LightMAX Supercharged LED Mask 2.0 (K213184) because they are both LED phototherapy devices intended to emit light in the red and blue and near infra-red region of the light spectrum and are indicated for the treatment of mild to moderate acne vulgaris and full-face wrinkles.
The LUMA Blemish and LUMA Revive are predicated against the LUSTRE PRO (K143713) as both devices are LED patches that are attached to the skin by way of an adhesive patch.
{6}------------------------------------------------
# 5.3.1. . Summary of Substantial Equivalence
| Description | Subject device | Primary | Secondary | Significant<br>differences |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| | LUMA Blemish (TN2197)<br>LUMA Revive (TN2189) | K213184<br>MZ Skin LightMAX Supercharged<br>LED Mask 2.0 | K143713<br>Lustre PRO Light<br>System | |
| Device<br>Manufacturer | ISMART Developments Ltd | MZ SKIN | Ambicare Health Ltd | na |
| Device Trade Name | LUMA patches<br>(Blemish & Revive) | MZ Skin LightMAX Supercharged LED<br>Mask 2.0 | Lustre PRO Light System | na |
| 510(K) Number | | K213184 | K143713 | na |
| Device<br>Classification<br>name | Over-The-Counter Powered Light-<br>Based Laser for Acne.<br>Light Based Over the Counter Wrinkle<br>Reduction | Over-The-Counter Powered Light-<br>Based Laser for Acne.<br>Light Based Over the Counter Wrinkle<br>Reduction | Over-The-Counter<br>Powered Light-Based<br>Laser for Acne. | Identical |
| Device Product<br>Code | OLP, OHS | OLP, OHS | GEX | Primary<br>predicate is<br>identical |
| Regulation<br>Number | 878.4810<br>Laser surgical instrument for use in<br>general and plastic surgery and in<br>dermatology. | 878.4810<br>Laser surgical instrument for use in<br>general and plastic surgery and in<br>dermatology. | 878.4810<br>Laser surgical instrument<br>for use in general and<br>plastic surgery and in<br>dermatology. | Identical |
| FDA Device<br>Classification | Class II | Class II | Class II | Identical |
| | Subject device | K213184 | K143713 | Significant<br>differences |
| Description | LUMA Blemish (TN2197) | MZ Skin LightMAX Supercharged | Lustre PRO Light | |
| | LUMA Revive (TN2189) | LED Mask 2.0 | System | |
| Use | Over the Counter | Over the Counter | Over the Counter | Identical |
| Intended use and<br>Indications | LUMA Blemish is an over-the-counter<br>device intended to emit energy in the<br>red and blue region of the light<br>spectrum, specifically indicated to<br>treat mild to moderate acne vulgaris<br>of the face.<br>LUMA Revive is an over-the-counter<br>device intended to emit energy in the<br>red and Near Infra-red spectrum and<br>is intended for the use in the<br>treatment of full-face wrinkles. | The MZ Skin LightMAX Supercharged<br>LED Mask 2.0 is an over-the-counter<br>device intended to emit energy in the<br>red and blue region of the light<br>spectrum, specifically indicated to<br>treat mild to moderate acne vulgaris<br>of the face.<br>The MZ Skin LightMAX Supercharged<br>LED Mask 2.0 is an over-the-counter<br>device intended to emit energy in the<br>red and Near Infra-red spectrum and<br>is intended for the use in the<br>treatment of full-face wrinkles. | The Lustre PRO Light<br>system is<br>intended to<br>emit light in the blue<br>region<br>of<br>the<br>light<br>spectrum and is indicated<br>for the treatment of mild<br>to<br>moderate<br>acne<br>vulgaris. | Primary<br>predicate is<br>identical |
| Intended Location<br>of Use | Face | Face | Face | Identical |
| Energy Type | Light emitting diodes | Light emitting diodes | Light emitting diodes | Identical |
| Peak Wavelength<br>(FWHM) | Blue: 415nm +/- 10nm<br>Red: 630nm +/- 10nm<br>NIR 830nm +/-10nm | Blue: 415nm +/- 10nm<br>Red: 630nm +/- 10nm<br>NIR 830nm +/-10nm | 415nm +/-15nm | Primary<br>predicate is<br>identical |
| Total Intensity<br>(mW/cm²) | LUMA Blemish (TN2197)<br>44 mW/cm²<br>LUMA Revive (TN2189)<br>29 mW/cm² | 44 mW/cm²<br>29 mW/cm² | 5mW/cm² | Primary<br>predicate is<br>identical |
| Description | Subject device | K213184 | K143713 | Significant<br>differences |
| | LUMA Blemish (TN2197) | MZ Skin LightMAX Supercharged | Lustre PRO Light<br>System | |
| | LUMA Revive (TN2189) | LED Mask 2.0 | | |
| Total Dose | LUMA Blemish (TN2197)<br>26.4/cm² | 26.4/cm² | Blue 9J/cm² | Primary<br>predicate is<br>identical |
| | LUMA Revive (TN2189)<br>18J/cm² | 18J/cm² | | |
| Treatment protocol | Acne: 4 x weekly, 6 weeks | Acne: 4 x weekly, 6 weeks | Daily for 4 weeks | Primary<br>predicate is<br>identical |
| | Wrinkles: 5 x weekly, 6 weeks | Wrinkles: 5 x weekly, 6 weeks | | |
| Method of<br>attachment | Single use adhesive patch | Held in place by 2 silicone straps | Single use adhesive<br>patch | Substantially<br>equivalent to<br>secondary<br>predicate |
| Software<br>Controlled | Device uses a timer and software to<br>control treatment duration | Device uses a timer and software to<br>control treatment duration | Device uses a timer and<br>software to control<br>treatment duration | Identical |
| Power supply | 3.7V, 37mAh Lithium Polymer battery | 5V, 9.62Wh Lithium Polymer battery | 12V/1000mA Lithium-Ion<br>battery | Primary<br>predicate is<br>identical |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# 5.4. Substantial Equivalency and Comparison of Technological Similarities & Differences
### 5.4.1. Similarities between Subject devices and Primary predicate
From the comparative table above, the LUMA patches (LUMA Revive/LUMA Blemish) are substantially equivalent to the primary predicate device MZ Skin LightMAX Supercharged LED Mask 2.0 (K213184).
The key similarities are.
i. The intended use of the LUMA Blemish and LUMA Revive is equivalent to the MZ Skin LightMAX Supercharged LED Mask 2.0 an over-the-counter device that is intended for the use in the treatment of mild to moderate acne vulgaris and full-face wrinkles.
ii. The devices are phototherapy units utilizing light emitting diodes that emit in the red and blue spectrum for the treatment of mild to moderate acne vulgaris and the red and NIR spectrum for the treatment of full-face wrinkles.
The wavelength spectrum of the devices is identical. iii.
iv. The LUMA Patches have a similar power density and deliver similar doses of light that is substantially equivalent compared to the primary predicate device.
The LUMA patches have an identical treatment time and treatment protocol V. compared to the proposed primary predicate.
All devices use software to control the treatment time. vi.
## 5.4.2. Differences between Subject devices and Primary predicate
The only difference between the LUMA patches and the primary predicate MZ Skin LightMAX Supercharged LED Mask 2.0 (K213184) is the method of attachment to the treatment area. The primary predicate is held in place by adjustable straps whereas the LUMA patches are held in place by an adhesive patch. The use method of attachment is identical to K143713 Lustre PRO Light System in that both devices use a single use double sided adhesive patch that adheres to the LED device and to the treatment area. The LUMA patches are therefore substantially equivalent to the K143713 Lustre PRO Light System in method of attachment.
Where there are differences between the subject devices and the predicates these have been addressed by non-clinical performance testing to the applicable standards identified in section 5 of this document.
{10}------------------------------------------------
#### 5.5. Non-clinical performance testing
The LUMA patches have been thoroughly evaluated for electrical safety and performance and has been found to conform to the following standards.
ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]. Medical electrical equipment. General requirements for basic safety and essential performance.
IEC 60601-1-2:2014 + IEC 60601-1-2:2014/A1:2020. Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances – Requirements and tests for: Home Healthcare Environment
47 CFR part 15b (FCC):2019 Electrical equipment for measurement, control, and laboratory use – EMC requirements - Part 1: General requirements. Conducted & Radiated Emissions- Class B
IEC 60601-1-11:2015 + IEC 60601-1-11:2015/A1:2021 Medical electrical equipment --Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
IEC 60601-2-83:2020+A11 :2021. Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
IEC 60601-2-57:2011. Medical electrical equipment – Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
IEC 62471:2006-07. Photobiological safety of lamps and lamp systems.
IEC 62133 -2:2017-02. Secondary Cells and Batteries Containing Alkaline or Other Non-Acid Electrolytes - Safety Requirements for Portable Sealed Secondary Cells, and for Batteries Made from Them, for Use in Portable Applications- Part 2: Lithium systems.
EN ISO 10993-1:2018. Biological evaluation of medical devices Part 1: Evaluation and testing
ISO 10993-5:2009. Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ISO 10993-10:2010 ISO 10993-10 Third Edition 2010. Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
{11}------------------------------------------------
EN/IEC 62304 :2006 + IEC 62304 :2006/A1 :2015Medical Device Software- Software Life Cycle Processes
In addition to the above standards the LUMA patch labelling was subject to label comprehension testing. With respect to medical devices available without the intervention of a physician, termed 'Over the Counter' (OTC).
To determine the effectiveness of labelling pertaining to a medical device, the labelling and LUMA Blemish and LUMA revive were tested with an appropriate sample of users.
A study was conducted demonstrating comprehension of the LUMA patch labelling. Twenty-six subjects took part in the study, 12:14 M: F. The mean age of the study group was 31.7 years of age, with a range of 14 to 55 years old. Thirty eight percent of the group (n=10) identified English as a second language. The average number of words incorrect in the REALM was 7, giving a mean reading ability of 59 (High school) (range 43-66).
No new use errors, hazards, hazardous situations, or hazard-related use scenarios were discovered during testing. Further improvement of the user interface design as it relates to safety was deemed unnecessary and there were no suggested revisions to the version of the user manual or box packaging tested.
The results of the human factors engineering process demonstrated that the packaging and labelling of the LUMA Blemish and LUMA Revive contained suitable information to allow the user to use the device safely and effectively for its intended purpose.
#### 5.6. Clinical Performance
Since the LUMA patches are substantially equivalent to the predicate devices and raise no new questions in terms of safety and efficacy, clinical data is not required.
#### 5.7. Statement of Substantial Equivalence:
513(i) of the FD&C Act (21 U.S.C. 360c(i) states that for substantial equivalence a proposed device is required to have the same intended use and similar technological characteristics as the predicate device. Where there are differences in technological characteristics, these can be negated by appropriate clinical or scientific data demonstrating that the proposed device is as safe and effective as the predicate device, and that the proposed device does not raise any different questions of safety and effectiveness than the predicate device for the same intended use.
ISMART Developments Ltd has demonstrated that the LUMA patches have an identical intended use, have the same generic classification and basic principles
{12}------------------------------------------------
and technologies as the primary predicate device (K213184). The devices utilize red and blue wavelengths of light with similar power densities and equivalent cumulative dose for the treatment of mild to moderate acne vulgaris and red and NIR of light with similar power densities and equivalent cumulative dose for the treatment of full-face wrinkles.
The LUMA patches have an identical mechanism of attachment to the treatment area as the secondary predicate device (K143713).
ISMART Developments has conducted non-clinical performance testing applicable to those general controls deemed necessary by the agency for this product classification and has determined that the LUMA Blemish and LUMA Revive raise no additional questions relating to safety. The LUMA Blemish and LUMA Revive are as safe and effective and performs as well as or better than the legally marketed predicate devices K213184 and K143713.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.