Advantage Ultra System, Advantage Fit Ultra System, Lynx Ultra System
K231549 · Boston Scientific Corporation · OTN · Jun 29, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231549
Device Name
Advantage Ultra System, Advantage Fit Ultra System, Lynx Ultra System
Applicant
Boston Scientific Corporation
Product Code
OTN · General, Plastic Surgery
Decision Date
Jun 29, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Advantage Ultra System and Advantage Fit Ultra System The mesh implant is intended for use as a suburethral sling for the treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The Advantage Ultra System and the Advantage Fit Ultra System delivery devices are intended for use as an aid in insertion, placement, fixation, and anchoring of the Advantage Ultra surgical mesh during urogynecological procedures. Lynx Ultra System The mesh implant is intended for use as a suburethral sling for the treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The Lynx Ultra System delivery device is intended for use as an aid in insertion, placement, fixation, and anchoring of the Lynx Ultra surgical mesh during urogynecological procedures.
Device Story
Systems consist of sterile, single-use polypropylene knitted mesh implants and delivery devices; used as suburethral slings to support urethra during stress/straining to prevent urine leakage. Delivery devices (needles with handles/pushers) facilitate insertion, placement, fixation, and anchoring of mesh through bodily tissues. Advantage/Advantage Fit Ultra systems designed for retropubic placement; Lynx Ultra system designed for suprapubic placement. Used in urogynecological procedures by physicians. Changes in this submission limited to electronic Instructions for Use (eIFU) regarding release/removal of disposable components; no changes to device design, materials, or principle of operation.
Clinical Evidence
Bench testing only. Conducted Design Validation and Summative Usability Evaluation to assess eIFU modifications. Results confirmed user needs were met and device functions as intended without new safety or effectiveness concerns compared to predicates.
Technological Characteristics
Materials: Polypropylene knitted mesh, polymer sleeves, polypropylene leader loops, polyester lead. Delivery device: Handle with curved needle and pusher. Sterile, single-use. No changes to materials or design from predicate.
Indications for Use
Indicated for female patients with stress urinary incontinence caused by urethral hypermobility and/or intrinsic sphincter deficiency.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Advantage Ultra System, Advantage Fit Ultra System, Lynx Ultra System (K211223)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
June 29, 2023
Boston Scientific Corporation Jotham Moon Senior Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752
Re: K231549
Trade/Device Name: Advantage™ Ultra System, Advantage Fit™ Ultra System, Lynx™ Ultra System Regulation Number: 21 CFR§ 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: OTN Dated: May 26, 2023 Received: May 30, 2023
Dear Jotham Moon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page
2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Angel A. Soler-garcia -S
for
Jessica Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K231549
Device Name Advantage™ Ultra System Advantage Fit™ Ultra System Lynx™ Ultra System
Indications for Use (Describe)
Advantage Ultra System and Advantage Fit Ultra System
The mesh implant is intended for use as a suburethral sling for the treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The Advantage Ultra System and the Advantage Fit Ultra System delivery devices are intended for use as an aid in insertion, placement, fixation, and anchoring of the Advantage Ultra surgical mesh during urogynecological procedures.
Lynx Ultra System
The mesh implant is intended for use as a suburethral sling for the treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The Lynx Ultra System delivery device is intended for use as an aid in insertion, placement, fixation, and anchoring of the Lynx Ultra surgical mesh during urogynecological procedures.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary for Advantage Ultra, Advantage Fit Ultra, and Lynx Ultra Systems
#### Date Prepared: June 29, 2023
#### A. Submitter
Boston Scientific Corporation Urology Division 100 Boston Scientific Way Marlborough, MA 01752
#### B. Contacts
Jotham Moon Senior Regulatory Affairs Specialist 508-683-4208 Jotham.Moon@bsci.com
Or
Brooke Cuddy Principal Regulatory Affairs Specialist 508-683-6434 Brooke.Cuddy@bsci.com
#### C. Device Names
| Trade name: | Advantage™ Ultra System |
|--------------------|-------------------------------------------------------------------------------------------------------------|
| Model Number: | M0068502060 |
| Common/Usual Name: | Mesh, surgical, synthetic, urogynecologic, for stress urinary<br>incontinence, retropubic or transobturator |
| Regulation Number: | 21 CFR §878.3300 |
| Regulation Name: | Surgical Mesh |
| Classification: | Class II |
| Product Code: | OTN |
| Trade name: | Advantage Fit™ Ultra System |
| Model Number: | M0068502160 |
| Common/Usual Name: | Mesh, surgical, synthetic, urogynecologic, for stress urinary<br>incontinence, retropubic or transobturator |
| Regulation Number: | 21 CFR §878.3300 |
| Regulation Name: | Surgical Mesh |
| Classification: | Class II |
| Product Code: | OTN |
| Trade name: | Lynx™ Ultra System |
| Model Number: | M0068503060 |
| Common/Usual Name: | Mesh, surgical, synthetic, urogynecologic, for stress urinary<br>incontinence, retropubic or transobturator |
| Regulation Number: | 21 CFR §878.3300 |
| Regulation Name: | Surgical Mesh |
| Classification: | Class II |
| Product Code: | OTN |
{5}------------------------------------------------
# D. Predicate Devices
For purposes of establishing substantial equivalence, the proposed Ultra Products were compared to the predicate devices shown in the table below. The predicate devices have not been subject to a design-related recall.
| | Predicate Devices |
|--------------------------|---------------------------------------------------------------------------|
| Device Trade Name: | Advantage Ultra System<br>Advantage Fit Ultra System<br>Lynx Ultra System |
| Regulation Name: | Surgical Mesh |
| Regulation Number: | 21 CFR §878.3300 |
| Classification: | Class II |
| Product Code: | OTN |
| 510(k) Submitter/Holder: | Boston Scientific Corporation, Marlborough, MA |
| 510(k) #/ Clearance Date | K211223 / Cleared July 21, 2021 |
Predicate Devices for Establishing Substantial Equivalence
# E. Device Description
#### Advantage Ultra System and Advantage Fit Ultra System
The Advantage Ultra System and the Advantage Fit Ultra System are sterile, single-use systems each consisting of one delivery device and one blue mesh assembly. The mesh assembly is comprised of a polypropylene knitted mesh protected by two disposable polymer sleeves, two dilators, and one center tab. The sleeves and dilators are secured to the mesh by two polypropylene leader loops, and the center tab is secured to the mesh by a polyester lead. The two dilators at the distal ends of the mesh assembly are designed to be placed over the needle end of the delivery device.
The Advantage Ultra System and the Advantage Fit Ultra Systems are each packaged with their respective delivery device (Advantage Ultra or Advantage Fit Ultra delivery device). This is a single-use disposable delivery device that consists of a handle with a curved needle and a pusher component. The delivery device is designed to facilitate the passage of the mesh assembly through bodily tissues for retropubic placement.
The Advantage Fit Ultra System includes the same mid-urethral mesh implant as the Advantage Ultra System but differs in the design of the dilators and delivery devices. The needle portion of the Advantage Fit Ultra delivery device has a smaller outer diameter and a smaller bend radius and is offered to provide physicians with surgical options. The Advantage Fit Ultra dilators also have a smaller diameter to accommodate for the different needle size.
#### Lynx Ultra System
The Lynx Ultra System is a sterile, single-use system consisting of two delivery devices and one blue mesh assembly. The mesh assembly is comprised of a polypropylene knitted mesh protected by two disposable polymer sleeves, two dilators, and one center tab. The sleeves and dilators are secured to the mesh by two polypropylene leader loops, and the center tab is secured to the mesh by a polyester lead. At the distal ends of the mesh assembly there are association loops designed to be placed in the needle slot of the distal end of a delivery device.
{6}------------------------------------------------
The Lynx Ultra System is packaged with two Lynx System delivery devices. Each is a single-use disposable delivery device that consists of a handle with a curved needle with a needle slot for connection to the mesh assembly. The needle is designed to facilitate the passage of the mesh assembly through bodily tissues for suprapubic placement.
## F. Intended Use/Indications for Use
#### Advantage Ultra System and Advantage Fit Ultra System
The mesh implant is intended for use as a suburethral sling for the treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The Advantage Ultra System and the Advantage Fit Ultra System delivery devices are intended for use as an aid in insertion, placement, fixation, and anchoring of the Advantage Ultra surgical mesh during urogynecological procedures.
#### Lynx Ultra Svstem
The mesh implant is intended for use as a suburethral sling for the treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The Lynx Ultra System delivery device is intended for use as an aid in insertion, placement, fixation, and anchoring of the Lynx Ultra surgical mesh during urogynecological procedures.
#### G. Operating Principle
The mesh sling acts as a backboard to support the urethra during stress and straining events to prevent urine leakage.
#### H. Comparison of Key Technological/Performance Characteristics
The proposed Ultra Products are identical to the predicate devices cleared via K211223 in many aspects, as there are no changes to the intended use or indications for use, design, materials, sterilization, technological and performance characteristics, or principle of operation. The changes that are the subject of this premarket notification are limited to procedural steps within the Ultra Products' electronic Instructions for Use (eIFU)s. These changes modify the instructions and provide additional clarification related to release and removal of the disposable components of the mesh assembly. These changes impact the user interface by modifying the user workflow.
#### I. Substantial Equivalence
A direct comparison of key characteristics demonstrates that Ultra Products are substantially equivalent to the predicate devices in terms of intended use, principle of operation, technological and performance characteristics.
#### J. Performance Testing
Performance bench testing was conducted in the form of Design Validation and Summative Usability Evaluation to evaluate the eIFU modifications. The Design Validation provided objective evidence that all user needs impacted by the elFU modifications were met and overall performance for the intended clinical use was acceptable. The Summative Usability Evaluation demonstrated that the Ultra Products continue to function as intended and do not raise new issues of safety or effectiveness compared to the predicate devices.
{7}------------------------------------------------
## K. Conclusion
Based on the intended use/indications for use, comparison of key technological characteristics, and performance testing presented in this submission, it is concluded that the proposed Ultra Products are substantially equivalent to the established predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.