← Product Code [PLF](/productcode/PLF) · K231496

# TITAN 3-D Wedge System (K231496)

_Paragon 28, Inc. · PLF · Aug 22, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K231496

## Device Facts

- **Applicant:** Paragon 28, Inc.
- **Product Code:** [PLF](/productcode/PLF.md)
- **Decision Date:** Aug 22, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The TITAN 3-D® Wedge System implants are indicated to be used for internal bone fixation for bone fractures, fusions, or osteotomies in the foot and ankle. The TITAN 3-D® Wedge System implants are intended for use with ancillary fixation. The TITAN 3-D® Wedge System implants are not indicated for use in the spine.

## Device Story

TITAN 3-D Wedge System consists of titanium alloy implants for small bone correction in foot and ankle; used for internal bone fixation during fractures, fusions, or osteotomies; requires ancillary fixation. Implants provided sterile; offered in various shapes/sizes. Device used by surgeons in clinical settings. Modifications include updated MR safety labeling per FDA guidance and addition of new contract suppliers. Performance verified via mechanical testing (dynamic/static compression, static compressive shear) and MR safety testing (image artifact, displacement force, torque, RF heating). Device labeled MR Conditional per ASTM F2503.

## Clinical Evidence

Bench testing only. Performance testing included dynamic compression, static compression, and static compressive shear. MR safety testing conducted per ASTM F2119, F2052, F2213, and F2182 to support MR Conditional labeling.

## Technological Characteristics

Titanium alloy implants. MR Conditional per ASTM F2503. Mechanical testing performed per ASTM standards for image artifacts (F2119), displacement force (F2052), torque (F2213), and RF heating (F2182).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- TITAN 3-D® Wedge System ([K162241](/device/K162241.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 22, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of the Department of Health & Human Services seal. To the right of the symbol is a blue square containing the acronym "FDA" in white letters, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.

Paragon 28 Inc Haylie Hertz Associate Manager of Regulatory Affairs 14445 Grasslands Drive Englewood, Colorado 80112

Re: K231496

Trade/Device Name: TITAN 3-D Wedge System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: PLF Dated: May 23, 2023 Received: May 24, 2023

Dear Haylie Hertz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K231496

Device Name TITAN 3-D® Wedge System

#### Indications for Use (Describe)

The TITAN 3-D® Wedge System implants are indicated to be used for internal bone fixation for bone fractures, fusions, or osteotomies in the foot and ankle. The TITAN 3-D® Wedge System implants are intended for use with ancillary fixation. The TITAN 3-D® Wedge System implants are not indicated for use in the spine.

| Type of Use (Select one or both, as applicable)                                         |                                                                                        |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="unicode-bidi:embed;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="unicode-bidi:embed;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K231496 510(K) SUMMARY

| 510(k) Number:                      | K231496                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:                       | Paragon 28, Inc.<br>14445 Grasslands Dr.<br>Englewood, CO 80112                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Contact:                            | Haylie Hertz<br>Associate Manager of Regulatory Affairs<br>Paragon 28, Inc.<br>14445 Grasslands Dr.<br>Englewood, CO 80112<br>Phone: 303-720-0017<br>hhertz@paragon28.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Date Prepared:                      | August 22, 2023                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Trade Name:                  | TITAN 3-D® Wedge System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Class:                       | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Predicate Device:                   | TITAN 3-D® Wedge System (K162241)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Description:                 | The TITAN 3-D® Wedge System contains a series of titanium alloy<br>implants used for the correction of small bones in the foot. It is<br>offered in varying shapes and sizes to accommodate a variety of<br>small bone applications. The implants are sold sterile.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification and<br>Product Code: | 21 CFR 888.3030; Single/multiple component metallic bone<br>fixation appliances and accessories; PLF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Indications for Use:                | The TITAN 3-D® Wedge System implants are indicated to be used<br>for internal bone fixation for bone fractures, fusions, or osteotomies<br>in the foot and ankle. The TITAN 3-D® Wedge System implants<br>are intended for use with ancillary fixation. The TITAN 3-D®<br>Wedge System implants are not indicated for use in the spine.                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Substantial<br>Equivalence:         | The intended use, principle of operation and fundamental scientific<br>technology of the modified device are identical to the predicate<br>device. The MR safety and compatibility performance was<br>assessed via testing in the MRI environment and labeling was<br>updated per the FDA Guidance “Testing and Labeling Medical<br>Devices for Safety in the Magnetic Resonance (MR) Environment”<br>issued May 20, 2021. The data demonstrate substantial equivalence<br>and the subject modifications do not raise new issues of safety or<br>effectiveness.<br>The subject device has the addition of new contract suppliers and<br>repeated performance testing has indicated that the supplier change<br>does not raise new issues of safety or effectiveness. There are no |

{4}------------------------------------------------

new chemicals or processes that the implant is exposed to when compared to the predicate therefore no new issues of safety and effectiveness are raised.

Performance MR Safety and Compatibility Testing has been completed and Testing: presented in the submission as recommended in the FDA Guidance "Testing and Labeling Medical Devices for Safetv in the Magnetic Resonance (MR) Environment" issued May 20, 2021, including FDA-recognized standards tests for the following potential hazards:

- . Image Artifact per ASTM F2119 Test Method for Evaluation of MR Image Artifacts from Passive Implants
- . Magnetically Induced Displacement Force per ASTM F2052 Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
- Magnetically Induced Torque per ASTM F2213 Test ● Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment
- . Radiofrequency (RF) Induced Heating per ASTM F2182 Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants during Magnetic Resonance Imaging

Based on the results, the TITAN 3-D® Wedge System will be labeled as "MR Conditional" with MRI Safety Information in the instructions for use as described in ASTM F2503 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment.

Performance testing for dynamic compression, static compression, and static compressive shear verified the change does not raise new questions of safety and effectiveness.

- Conclusion: The proposed device modifications do not raise new issues of safety or effectiveness and have been fully evaluated. The subject devices are substantially equivalent to the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K231496](https://fda.innolitics.com/device/K231496)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
