Mobility Scooter (S1 SPORT, S1 PLUS, MAX SPORT)

K231472 · Suzhou Sweetrich Vehicle Industry Technology Co., Ltd. · INI · Jul 19, 2023 · Physical Medicine

Device Facts

Record IDK231472
Device NameMobility Scooter (S1 SPORT, S1 PLUS, MAX SPORT)
ApplicantSuzhou Sweetrich Vehicle Industry Technology Co., Ltd.
Product CodeINI · Physical Medicine
Decision DateJul 19, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3800
Device ClassClass 2

Indications for Use

The mobility Scoter is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Story

Motorized four-wheeled mobility scooter; provides mobility for disabled/elderly persons limited to seated position. Components: steel frame, seat, tiller console, electric motor, electromagnetic brake, two lead-acid batteries, off-board charger. Rider operates throttle lever, speed control dial, and handle on tiller console. Electromagnetic brake engages automatically when not in use. Used in indoor/outdoor environments. Provides mobility benefit to users. No clinical data included.

Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (ISO 10993-5, -10, -23) and performance testing per ISO 7176 series (stability, braking, energy consumption, dimensions, speed, obstacle climbing, EMC).

Technological Characteristics

Carbon steel frame; 24V 300W electric motor; electromagnetic braking; 24V/12AH lead-acid batteries; solid tires (8"). Dimensions: 1210-1260mm length. Connectivity: None. Sterilization: N/A. Software: None (mechanical/electrical control).

Indications for Use

Indicated for disabled or elderly persons limited to a seated position requiring indoor and outdoor mobility assistance.

Regulatory Classification

Identification

A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it. July 19, 2023 Suzhou Sweetrich Vehicle Industry Technology Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China Re: K231472 Trade/Device Name: Mobility Scooter (S1 SPORT, S1 PLUS, MAX SPORT) Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: May 22, 2023 Received: May 22, 2023 Dear Ivy Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Heather L. Dean -S Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K231472 Device Name Mobility Scooter (S1 SPORT, S1 PLUS, MAX SPORT) Indications for Use (Describe) The mobility Scoter is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. Type of Use (Select one or both, as applicable) | <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | |-----------------------------------------------------------------------|---------------------------------------------------------------------------------| |-----------------------------------------------------------------------|---------------------------------------------------------------------------------| ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary #### l. Applicant Name: Suzhou Sweetrich Vehicle Industry Technology Co., Ltd. Address: No. 68 Xinfa Road, Suzhou Industrial Park, Suzhou, Jiangsu, China Name of contact person: Chen Lezhang Telephone: +86 139 1626 7666 Email: lezhang@sweetrich.cn; gc@sweetrich.cn Date prepared: 2023-05-18 Submission Correspondent ll. Ms. Ivy Wang Email: zxfda@sungoglobal.com Shanghai Sungo Management Consulting Company Limited Tel: +86-21-5881 7802 III. Device Device trade name: Mobility Scooter Model: SI SPORT, SI PLUS, MAX SPORT Regulatory Information: Classification name: Vehicle, motorized 3-wheeled Regulation class: 2 Regulation number: 21CFR 890.3800 Panel: Physical Medicine Product code: INI IV. Predicate device K182471 Device Name: Scooter Model: KPL001 Tianjin Kepler Vehicle Industry Co. Ltd. ## V. Device description {5}------------------------------------------------ The Mobility Scooter, Model S1 SPORT, S1 PLUS, MAX SPORT, has a base with Steel frame, two front wheels, two rear wheels, a seat, a tiller console, electric motor, electromagnetic brake, 2 rechargeable Lead-ac id Batteries with an off-board charger. The movement of the scooter is controlled by the rider who operates the throttle lever, speed control dial and handle on the tiller console. The device is installed with an electromagnetic brake that will engage automatically when the scooter is not in use and the brake cannot be used manually. The Scooter only can be operated on the flat road. The controller panel shape is slightly different among the three models, S1 SPORT, S1 PLUS, MAX SPORT. The S1 PLUS has a wind board on the triller, while the other two models do not have. The length of model MAX SPORT is 5cm longer than the other two models. #### VI. Indication for use The mobility scooter is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. #### VII. Non-Clinical Test Conclusion Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - . ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity - ISO 10993-10: 2010 Biological Evaluation of Medical Devices Part 10: Tests For . Irritation And Skin Sensitization - . ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation - ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability - . ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs - ISO 7176-3:2012 Wheelchairs Part 3: Determination of effectiveness of brakes . - . ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs {6}------------------------------------------------ and scooters for determination of theoretical distance range - . ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space - . ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs - . ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions - . ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatique strength - . ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs - . ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs - . ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies - . ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces. - . ISO 7176-14:2008 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods - . ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling. - . ISO 7176-16: 2012 Second edition 2012-12-01 Wheelchairs - Part 16: Resistance to ignition of postural support devices - . ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters - . ISO 7176-22: 2014 Wheelchairs - Part 22: Set-up procedures - . ISO 7176-25:2013 Wheelchairs - Batteries and charqers for powered wheelchairs - . Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2014. #### VIII. Comparison with the predicate device #### Table 1 General Comparison | Attribute | Subject device | Predicate device | Results | |--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Attribute | Subject device | Predicate device | Results | | Manufacturer | Suzhou Sweetrich Vehicle<br>Industry Technology Co.,<br>Ltd. | Tianjin Kepler Vehicle<br>Industry Co. Ltd. | / | | Proprietary name,<br>model | Electrical Scooter<br>SI SPORT, SI PLUS, MAX<br>SPORT | Scooter, KPL001 | / | | 510(k) number | K231472 | K182471 | / | | Device classification<br>name | Class II | Class II | Same | | Classification<br>regulations | 21 CFR 890.3800 | 21 CFR 890.3800 | Same | | Product code | INI | INI | Same | | Similarities | | | | | Indication for use | It is a motor driven, indoor<br>and outdoor transportation<br>vehicle with the intended<br>use to provide mobility to a<br>disabled or elderly person<br>limited to a seated<br>position. | It is a motor driven, indoor<br>and outdoor transportation<br>vehicle with the intended<br>use to provide mobility to a<br>disabled or elderly person<br>limited to a seated position. | Same | | Use condition | indoor and outdoor use | indoor and outdoor use | Same | | Number of wheels | 4, including two front<br>wheels and two rear<br>wheels | 4, including two pivoting<br>casters and two rear drive<br>wheels | Same | | Driving system | Direct drive on the rear<br>wheels | Direct drive on the rear<br>wheels | Same | | Brake | Electromagnetic | Electromagnetic | Same | | Time to brake | < 1 s | < 1 s | Same | | Main frame material | Carbon Steel | Carbon Steel | Same | | Frame style | Foldable seat, removable<br>battery pack, disassemble<br>for transport | Foldable seat, removable<br>battery pack, disassemble<br>for transport | Same | | Battery | lead-acid 24V/12AH | lead-acid 24V/12AH | Same | | Max loading weight | 120 kg/265 lbs approx | 120 kg/265 lbs approx | Same | | Charger | DC 24V/2A | DC 24V/2A | Same | | Rear wheel size/type | 8" (215 x 70 mm) Solid tire | 8" (215 x 70 mm) Solid tire | Same | | Differences | | | | | Controller | DR50 Dynamic | PG45A | Different. | | Maximum distance | 17 km | 15 km | Different. | | Attribute<br>of travel on the fully<br>charged battery | Subject device | Predicate device | Results | | Overall Dimension<br>(length*width*height) | 1210*662*925mm (Model:<br>S1 SPORT & S1 PLUS)<br>1260*662*925mm (Model:<br>MAX SPORT) | 1030mm*630mm * 920mm | Different | | Brake distance | ≤1m | 1m | Different | | Turning Radius | 1080-1200 mm | 1300 mm | Different | | Ground clearance | 40-50 mm | 45 mm | Different | | Maximum obstacle<br>climbing | 20 mm | 50 mm | Different | | Slope grade ability | 6 ° | 0-12° | Different | | Front wheel<br>size/type | 8" (215 x 70 mm) Solid tire | 7" (190 x 54 mm) Solid tire | Different | | Max speed | 6.4 km/h | 6.6 km/h | Different | | Motor | 24 V 300W | 24 V 180W | Different | | Base weight (not<br>including battery) | 46 kg | 50 kg | Different | | Battery Weight | 9 kg | 10.2 kg | Different | {7}------------------------------------------------ {8}------------------------------------------------ Difference Analysis: The design and technological characteristics of the subject device is basically similar to the predicate device chosen. There are minor differences between the devices including controller, travel distance, overall dimensions, front tire size, speed, turning radius, basic weight, battery weight, Brake distance, slope grade ability, obstacle climbing ability, Ground clearance and motor. There is no deleterious effect on safety and effectiveness due to the differences, and these minor differences do not influence the intended use function and use of the device. Moreover, the non-clinical tests and the predicate comparisons demonstrate that these differences in their technological characteristics do not raise any questions as to the safety and effectiveness. Therefore, the subject device is substantially equivalent to the Scooter (K182471). Table 2 Safety comparison | Attribute | Subject device | Predicate device | Results | |------------------|---------------------------------------------------------------|----------------------------------------------------------|---------| | Biocompatibility | All user directly contacting<br>materials are compliance with | All user directly contacting<br>materials are compliance | S.E. | {9}------------------------------------------------ | Attribute | Subject device | Predicate device | Results | |--------------------|-----------------------------------------------------------|--------------------------------------------------|---------| | | ISO10993-5,<br>ISO10993-10, ISO 10993-23<br>requirements. | with ISO10993-5 and<br>ISO10993-10 requirements. | | | EMC | ISO7176-21 | ISO7176-21 | S.E. | | Performance | ISO7176 series | ISO7176 series | S.E. | | Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | S.E. | ## Table 3 Safety comparison | Attribute | Subject device | Predicate device | Results | |------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | ISO7176-1 | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | S.E. | | ISO7176-2 | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification | S.E. | | ISO7176-3 | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | S.E. | | ISO7176-4 | The theoretical distance range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification. | The theoretical distance range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification. | S.E. | | ISO7176-5 | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | S.E. | | ISO7176-6 | The maximum speed, acceleration and deceleration of scooter has been determined after the testing according to the ISO 7176-6 | The maximum speed, acceleration and deceleration of scooter has been determined after the testing according to the ISO 7176-6 | S.E. | | ISO7176-7 | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7 | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7 | S.E. | | ISO7176-8 | All test results meet the | All test results meet the | S.E. | | Attribute | Subject device | Predicate device | Results | | ISO7176-9 | requirements in Clause 4 of ISO 7176-8<br>The test results shown that the device under tests could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9 | requirements in Clause 4 of ISO 7176-8<br>The test results shown that the device under tests could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9 | S.E. | | ISO7176-10 | The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10. | The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10. | S.E. | | ISO7176-11 | The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11. | The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11. | S.E. | | ISO7176-13 | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved | S.E. | | ISO7176-14 | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14 | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14 | S.E. | | ISO7176-15 | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15 | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15 | S.E. | | ISO7176-16 | The performance of resistance to ignition meets the requirements of ISO 7176-16. | The performance of resistance to ignition meets the requirements of ISO 7176-16. | S.E. | | ISO7176-21 | The EMC performance results meet the requirements of ISO 7176-21 | The EMC performance results meet the requirements of ISO 7176-21 | S.E. | | ISO7176-22 | All performed tests are set up as requirements of ISO 7176-22 | All performed tests are set up as requirements of ISO 7176-22 | S.E. | | ISO7176-25 | The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25 | The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25 | S.E. | {10}------------------------------------------------ {11}------------------------------------------------ ## IX. Summary of clinical testing No clinical study is included in this submission. #### X. Conclusions The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K182471.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...