The Precision-GI device is intended for use with an ultrasound endoscope (EUS) for fine needle biopsy (FNB) of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract.
Device Story
Precision GI is a battery-operated, motorized, automated rotational cutting biopsy needle; used with ultrasound endoscope for fine needle biopsy (FNB) of lesions/masses within or adjacent to the GI tract. Device consists of needle cannula, stylet, sheath, and handle; features echogenic tip for ultrasound visualization. Operated by physician through accessory instrument channel of endoscope. Motor rotates needle at 800 RPM to facilitate tissue acquisition. Output is a tissue sample for diagnostic evaluation. Benefits include improved tissue sampling via automated rotation. Used in clinical settings.
Clinical Evidence
Bench testing and preclinical study data provided. Bench tests included dimensional attributes, EUS compatibility, shaft insertion/withdrawal force, stopper lock operation, sample acquisition, echogenicity, needle deformation, endurance, handle resistance, torque, stylet/sheath/needle bonding strengths, tissue wrapping, cyclic loading durability, penetration force, biocompatibility, and sterilization. No human clinical trial data reported.
Technological Characteristics
Motorized, battery-operated rotational biopsy needle. 20G size. Materials: Nitinol needle with PTFE coating. Features echogenic tip pattern. Motor rotation speed: 800 RPM. Components: needle cannula, stylet, sheath, handle. Sterilization: Not specified.
Indications for Use
Indicated for fine needle biopsy (FNB) of submucosal lesions, mediastinal masses, lymph nodes, and intraperitoneal masses within or adjacent to the gastrointestinal tract using an ultrasound endoscope.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
EndoDrill® Model X biopsy instrument by Bibbinstruments AB (K212423)
Reference Devices
EchoTip ProCore HD Ultrasound Biopsy Needle Fine needle biopsy system, by Cook Medical (K210476)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 28, 2023
Limaca Medical Ltd % Susan Alpert Regulatory Affairs Consultant SFADC LLC 2425 L Street NW, Apt 307 Washington, District of Columbia 20037
Re: K231422
Trade/Device Name: Precision GI Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: II Product Code: FCG Received: July 27, 2023
Dear Susan Alpert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K231422
Device Name
Precision Gl
Indications for Use (Describe)
Precision GI is used with an ultrasound endoscope for fine needle biopsy (FNB) of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for LIMACA Medical. The logo is a stylized text design with the word "LIMACA" in a modern font. Below the logo is the text "For Better Endoscopic Biopsies".
# 510(k) SUMMARY
| Limaca's Precision GI | |
|---------------------------|-----------------------------------------------------------------------------------------------------|
| Submitter: | Limaca Medical Itd. |
| Address | 3 Ha' Rimon Street, |
| | Mevo-Carmel Science and Industry Park |
| | 1925000 Ein Ha Emeq Israel |
| Phone:<br>e-mail: | 972-4-6135050<br>info@limaca-medical.com |
| Contact Person: | Assaf Klein |
| Date Prepared: | 27 July 2023 |
| Name of Device: | Precision GI |
| Common or Usual Name: | Fine needle biopsy |
| Classification Name: | Kit, Needle, biopsy |
| Regulatory Class: | 21 CFR 876.1075, class II |
| Product Code: | FCG |
| Primary Predicate Device: | EndoDrill® Model X biopsy instrument by Bibbinstruments AB<br>(K212423) |
| Reference Device: | EchoTip ProCore HD Ultrasound Biopsy Needle Fine needle<br>biopsy system, by Cook Medical (K210476) |
## Device Description
The Precision GI Endoscopic ultrasound-guided Fine Needle Biopsy (EUS-FNB) device is a motorized, automated, rotational, cutting EUS-Biopsy needle with an echogenic tip inserted and operated through the instrument channel of an ultrasound imaging endoscope.
The Limaca Precision GI FNB needle is available in one size - 20G.
The Precision GI Endoscopic ultrasound-guided Fine Needle Biopsy (EUS-FNB) device is a battery-operated, motorized, automated rotational cutting biopsy device with an echogenic needle tip, advanced and operated by the physician through the accessory instrument channel of an ultrasound imaging endoscope.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for LIMACA Medical. The logo is a stylized text design with a blue accent mark above the letters. Below the logo, the text "For Better Endoscopic Biopsies" is displayed in a simple, sans-serif font.
#### Intended Use / Indications for Use
The Precision-GI device is intended for use with an ultrasound endoscope (EUS) for fine needle biopsy (FNB) of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract.
### Comparison of the Technological Characteristic with the Predicate Device
Precision GI is substantially equivalent to other legally marketed biopsy systems including the predicate: EndoDrill® Model X biopsy instrument by Bibbinstruments AB (k212423)
The Precision GI has the same intended use as its predicate and is used on the same anatomical sites.
The Precision GI device, like its predicate, the EndoDrill® Model X biopsy instrument by Bibbinstruments AB (k212423), is used in conjunction with an ultrasound endoscope for fine needle biopsy (FNB). Both devices perform tissue acquisition while rotating and advancing electrically. Both the Precision GI device and its predicate use a motor for rotation.
The motor rotation speed of the Precision GI device is 800RPM, while the motor speed of its predicate is higher. The reduced motor speed is safe and yet provides a good sample acquisition as demonstrated in bench tests and preclinical study.
Both devices have an echogenic pattern at the distal end of the needle to allow visualization of the needle tip under endoscopic ultrasound. The Precision GI device, like its predicate, EndoDrill® Model X biopsy instrument by Bibbinstruments AB (k212423), consists of needle cannula, stylet, sheath and handle.
The needle geometry of the Precision Gl and its predicate is similar. Both have flat circumferential sharpening, however the Precision GI sharpening is internal and the predicate is external. The Precision GI device needle size is 20GA and is made of Nitinol with PTFE coating, while its predicate size is 17GA needle made of Stainless steel. However, there are other FNB needles of different sizes and material, similar to the Precision GI that have been previously approved, such as EchoTip ProCore HD Ultrasound Biopsy Needle Fine needle biopsy system, by Cook Medical (K210476) (Reference device).
#### Performance Data
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Limaca Medical. The logo is a stylized version of the company name, with the letters "LIMA" in a light blue color and the letters "CA" in gray. Below the logo is the text "For Better Endoscopic Biopsi" in a sans-serif font.
The Precision GI device is determined to be substantial equivalent to the predicate device. Engineering bench testing as well as pre clinical study have been performed to support substantial equivalence with the predicate device. Various bench tests as well as pre- clinical data were included in the performance data, including: Dimensional Attributes and EUS compatibility, Shaft insertion/withdrawal force, Stopper lock operation force, Sample Acquisition, Echogenicity, Needle Deformation, Endurance to Distribution Environmental Hazards, Handle resistance force, Extender friction torque, Stylet withdrawal force, Stylet hold-in- position force, Sheath bonding strength, Axial stoper strength, Stylet bonding strength, Handle bending torque durability, Needle bonding torque strength, Tissue wrapping, Needle cyclic loading durability, Needle bonding durability, Needle tissue Penetration force, Needle bonding tensile strength, biocompatibility and sterilization.
#### Conclusions
The Precision GI is as safe and effective as the EndoDrill® Model X biopsy instrument by Bibbinstruments AB (k212423). The Precision GI has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the Precision GI and its predicate and reference devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Precision GI is as safe and effective as the EndoDrill® Model X biopsy instrument by Bibbinstruments AB (k212423). Thus, the Precision GI is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.