# ENDOGATOR™ Hybrid Irrigation Tubing (K231418)

_Medivators · OCX · Aug 14, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K231418

## Device Facts

- **Applicant:** Medivators
- **Product Code:** [OCX](/productcode/OCX.md)
- **Decision Date:** Aug 14, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Pediatric

## Indications for Use

The ENDOGATOR™ Hybrid Irrigation Tubing (tubing and accessories to accommodate various endoscopes and irrigation pumps) is intended to provide irrigation via sterile water and to supply air (via an air pump) or CO2 (via a CO2 supply) along with sterile water during GI endoscopic procedures when used in conjunction with an irrigation pump or cautery unit.

## Device Story

ENDOGATOR™ Hybrid Irrigation Tubing is a single-use, sterile accessory for GI endoscopes. It consists of a threaded bottle cap, irrigation tubeset, coaxial tube for air/CO2, and backflow check valves. The device attaches to sterile water bottles and interfaces with irrigation pumps or cautery units. During procedures, the irrigation pump (actuated by a user-controlled foot pedal) siphons water from the bottle to the endoscope; the coaxial tube delivers pressurized air or CO2 from the endoscope processor or insufflator. The device acts as a passive conduit. Backflow valves prevent reverse flow to mitigate cross-contamination. It is used in clinical settings by healthcare professionals. The device benefits patients by enabling clear visualization and irrigation during endoscopy.

## Clinical Evidence

Bench testing only. Testing included water flow rate, air/CO2 flow rate, backflow valve pressure testing (10 PSI), and durability of bonded connections (10 lbs axial pull force). Biocompatibility testing was performed per ISO 10993-1:2018 for surface-mucosal membrane contact (<24 hours), including cytotoxicity, irritation, and sensitization. All acceptance criteria were met.

## Technological Characteristics

Materials: Nitrile Butadiene, ABS, Thermoplastic Elastomer, Polycarbonate, non-phthalate PVC, Polypropylene, Nickel-Plated Brass, Silicone, Mabs Terflux, UV cured glue. Passive fluid/gas delivery system. Sterilization: Ethylene Oxide (SAL 10^-6). Single-use, 24-hour multi-patient use. Connectivity: Luer connections to endoscope and irrigation pump.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- The Universal Irrigation Solution Hybrid ([K102855](/device/K102855.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 14, 2023

Medivators Disha Kabrawala Senior Regulatory Affairs Specialist 14605 28th Avenue North Minneapolis, Minnesota 55447

Re: K231418

Trade/Device Name: ENDOGATOR™ Hybrid Irrigation Tubing Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OCX Dated: May 11, 2023 Received: May 16, 2023

Dear Disha Kabrawala:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231418

Device Name ENDOGATOR™ Hybrid Irrigation Tubing

### Indications for Use (Describe)

The ENDOGATOR™ Hybrid Irrigation Tubing (tubing and accessories to accommodate various endoscopes and irrigation pumps) is intended to provide irrigation via sterile water and to supply air (via an air pump) or CO2 (via a CO2 supply) along with sterile water during GI endoscopic procedures when used in conjunction with an irrigation pump or cautery unit.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> |
|----------------------------------------------|-------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <input type="checkbox"/>            |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for CANTEL. The logo consists of a blue and green globe-like image on the left, followed by the word "CANTEL" in blue, sans-serif font. The globe is made up of curved lines in different shades of blue and green, giving it a three-dimensional appearance. The word "CANTEL" is written in all capital letters and is positioned to the right of the globe.

### 510(k) Summary For ENDOGATOR™ Hybrid Irrigation Tubing

 Medivators Inc 14605 28th Ave N, Minneapolis, MN 55447 Fax No: 844-348-5637

- Contact Person: Disha Kabrawala Senior Regulatory Affairs Specialist Phone: 732-319-7766 Email: disha kabrawala@steris.com
May 11, 2023 Summary Date:

{4}------------------------------------------------

# CANTEL MEDICAL TRADITIONAL PREMARKET NOTIFICATION [510(k)] ENDOGATOR™ Hybrid Irrigation Tubing

#### 1. Device Name

| Trade Name:            | ENDOGATOR™ Hybrid Irrigation Tubing        |
|------------------------|--------------------------------------------|
| Device Class:          | II                                         |
| Common/usual Name:     | ENDOGATOR™ Hybrid Irrigation Tubing        |
| Classification Name:   | Endoscope and Accessories                  |
| Classification Number: | 876.1500                                   |
| Product Code:          | OCX (endoscopic irrigation/suction system) |

#### 2. Predicate Devices

The Universal Irrigation Solution Hybrid, K102855

#### 3. Description of Devices

The ENDOGATOR™ HYBRID Irrigation Tubing device contains a threaded cap and tubeset designed to be attached to sterile water bottles. The proximal end of the irrigation tubing includes the threaded bottle cap for attachment to a sterile water bottle. The bottle cap has three tubes attached, one for receiving CO2 (and air in the case of Pentax SKU's), one coaxial tube which has an inner lens rinsing uptake tube surrounded by an air/CO2 delivery tube, and one for delivering water for irrigation which interfaces with an irrigation pump. The water delivery tubes, both irrigation as well as lens rinsing include backflow check valves.

# Air/CO2 and Lens Rinse Line Description

The threaded cap uses a coaxial tube to partition water and air/CO2 which is connected to the endoscope-interfacing connector. Pressurized air or CO2 provided by the endoscope's processor/light source or a CO2 insufflator is fed into the endoscope (either directly or first through the device & bottle system via luer connection.) The Endogator Hybrid enables the air/CO2 to pressurize the bottle and flow through the endoscope to the air/water valve. The Endogator Hybrid serves as a passive element in the system, allowing the delivery of pressurized air/CO2 to endoscopes.

### Irrigation Line Description

The irrigation tubing pump section is installed into peristaltic-roller-type irrigation pumps, which transfer water through means of applying force to the external surface of the tubeset via revolving rollers, actuated by a user-controlled foot pedal, causing water to be siphoned from the sterile water bottle and transferred

{5}------------------------------------------------

# CANTEL MEDICAL TRADITIONAL PREMARKET NOTIFICATION [510(k)] ENDOGATOR™ Hybrid Irrigation Tubing

downstream to the endoscope. The flow rate can be manipulated by userinteraction with the pump interface. The distal end of the irrigation tubing includes a backflow valve which allows water to travel only in one direction to mitigate cross-contamination. The irrigation check valve has a luer connection that enables the irrigation tubing to be connected to the ENDOGATOR™ auxiliary port or biopsy accessory adapters that interface with endoscopes (separate devices).

#### 4. Intended Use / Indications for Use

The ENDOGATOR™ Hybrid Irrigation Tubing (tubing and accessories to accommodate various endoscopes and irrigation pumps) is intended to provide irrigation via sterile water and to supply air (via an air pump) or CO2 (via a CO2 supply) along with sterile water during GI endoscopic procedures when used in conjunction with an irrigation pump or cautery unit.

#### ડ. Comparison of Technological Characteristics with the Predicate Device

A comparison of technical characteristics between the proposed and predicate devices is included in Table 1.

| Feature                             | Proposed<br>ENDOGATOR Hybrid<br>Irrigation tubing                                                                                                                                                                                                                                                                                                                                                        | Predicate (K102855)<br>The Universal Irrigation<br>Solution Hybrid                                                                                                                                                                                                                                                                                                                                         | Comparison                                                                                                                                                                                               |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                        | The ENDOGATOR™ Hybrid<br>Irrigation Tubing (tubing and<br>accessories to accommodate<br>various endoscopes and<br>irrigation pumps) is intended to<br>provide irrigation via sterile<br>water and to supply air (via an<br>air pump) or CO2 (via a CO2<br>supply) along with sterile water<br>during GI endoscopic<br>procedures when used in<br>conjunction with an irrigation<br>pump or cautery unit. | The Universal Irrigation<br>Hybrid™ solution (tubing and<br>accessories to accommodate<br>various endoscopes and<br>irrigation pumps) is intended<br>to provide irrigation via sterile<br>water and to supply air (via an<br>air pump) or CO2 (via a CO2<br>supply) along with sterile<br>water during GI endoscopic<br>procedures when used in<br>conjunction with an irrigation<br>pump or cautery unit. | Identical                                                                                                                                                                                                |
| Sterile                             | Sterile                                                                                                                                                                                                                                                                                                                                                                                                  | Sterile                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                                |
| Sterilization<br>Method             | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                           | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                                |
| Sterilization<br>Assurance<br>Level | 10-6                                                                                                                                                                                                                                                                                                                                                                                                     | 10-6                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                                                |
| Feature                             | Proposed<br>ENDOGATOR Hybrid<br>Irrigation tubing                                                                                                                                                                                                                                                                                                                                                        | Predicate (K102855)<br>The Universal Irrigation<br>Solution Hybrid                                                                                                                                                                                                                                                                                                                                         | Comparison                                                                                                                                                                                               |
| Intended Use                        | 24hr/ Multi-patient use                                                                                                                                                                                                                                                                                                                                                                                  | 24hr/ Multi-patient use                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                                |
| Reprocessing                        | None; Daily disposal                                                                                                                                                                                                                                                                                                                                                                                     | None; Daily disposal                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                                                |
| Compatible<br>Endoscope(s)          | Olympus® 140/160/180/190<br>series GI endoscopes<br>PENTAX™ GI Endoscopes,<br>FUJI 500 series GI Endoscopes<br>FUJI 700 series GI Endoscopes                                                                                                                                                                                                                                                             | Olympus®<br>140/160/180/240/260 series GI<br>endoscopes                                                                                                                                                                                                                                                                                                                                                    | Different<br>Additional<br>part<br>numbers<br>implemente<br>d to for<br>Pentax, Fuji<br>endoscopes                                                                                                       |
| Patient<br>Contact                  | No direct contact                                                                                                                                                                                                                                                                                                                                                                                        | No direct contact                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                                |
| Target<br>Patient<br>Population     | Male/Female, Pediatric to<br>Adult<br>- Patients population<br>undergoing GI endoscopic<br>procedure                                                                                                                                                                                                                                                                                                     | Male/Female, Pediatric to<br>Adult<br>- Patients population<br>undergoing GI endoscopic<br>procedure                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                                                |
| Materials                           | Nitrile Butadiene (Buna-N)<br>acrylonitrile-butadiene-styrene<br>Acrylonitrile Butadiene Styrene<br>(ABS)<br>BLACK PEARL C<br>Thermoplastic Elastomer<br>Polycarbonate<br>Non-Phthalate Polyvinyl<br>Chloride (PVC)<br>Polypropylene<br>Buna-N Nitrile<br>Nickel-Plated Brass<br>Silicone<br>Mabs Terflux<br>UV cured Glue                                                                               | Nitrile Butadiene (Buna-N)<br>acrylonitrile-butadiene-styrene<br>Acrylonitrile Butadiene Styrene<br>(ABS)<br>BLACK PEARL C<br>Thermoplastic Elastomer<br>Polycarbonate<br>Phthalate Polyvinyl Chloride<br>(PVC)<br>Polypropylene<br>Buna-N Nitrile<br>Nickel-Plated Brass<br>Silicone<br>Mabs Terflux                                                                                                      | Similar<br>Use of non-<br>phthalate<br>polyvinyl<br>Chloride<br>material for<br>tubing                                                                                                                   |
| Usage                               | Single Use                                                                                                                                                                                                                                                                                                                                                                                               | Single Use                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                |
| Feature                             | Proposed<br>ENDOGATOR Hybrid<br>Irrigation tubing                                                                                                                                                                                                                                                                                                                                                        | Predicate (K102855)<br>The Universal Irrigation<br>Solution Hybrid                                                                                                                                                                                                                                                                                                                                         | Comparison                                                                                                                                                                                               |
| Construction                        | Threaded Cap,<br>Irrigation Tube set,<br>Coaxial Tube,<br>Backflow Check Valve                                                                                                                                                                                                                                                                                                                           | Threaded Cap,<br>Irrigation Tube set,<br>Coaxial Tube                                                                                                                                                                                                                                                                                                                                                      | Different<br>Backflow<br>check valve<br>is added to<br>prevent the<br>flow in<br>reverse<br>direction.<br>The<br>successful<br>results of<br>testing<br>prove that<br>device is<br>safe and<br>effective |
| Energy Used<br>/ Delivered          | None                                                                                                                                                                                                                                                                                                                                                                                                     | None                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                                                |

### Table 1: Physical Description and Technological Properties vs the Predicate Device

{6}------------------------------------------------

# CANTEL MEDICAL TRADITIONAL PREMARKET NOTIFICATION [510(k)] Page 4 of 6
ENDOC ATORIN Uybrid Friisotion Tubing ENDOGATOR™ Hybrid Irrigation Tubing

{7}------------------------------------------------

# CANTEL MEDICAL TRADITIONAL PREMARKET NOTIFICATION [510(k)] ENDOGATOR™ Hybrid Irrigation Tubing

### 6. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

| Testing                                  | Requirements                                                                                                                                                                            | Results |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Water Flow Rate                          | Lensing Flow Rate must be no lower<br>than 10% of Predicate after calling for<br>water for 20 seconds                                                                                   | Pass    |
| Insufflation Air & CO2 Flow              | Air & CO2 Flow Rate must be no<br>lower than 10% of the predicate                                                                                                                       | Pass    |
| Backflow Test                            | The backflow valve must withstand at<br>least 10 PSI                                                                                                                                    | Pass    |
| Lens Rinsing Flow Rate                   | Lensing Flow Rate must be no lower<br>than 10% of predicate after calling for<br>water for 20 seconds                                                                                   | Pass    |
| Durability of glue bonded<br>connections | The bonded connections must meet or<br>exceed 10 lbs of axial pull force and<br>must maintain strength and durability<br>requirements after being subjected to<br>2X EtO sterilization. | Pass    |

## Table 2: Non-clinical testing

{8}------------------------------------------------

# Biocompatibility testing

Biocompatibility of the ENDOGATOR Hybrid™ Irrigation Tubing was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "Surface -Mucosal Membrane" with a contact duration of "Limited (< 24 hours)". The following tests were performed: Cytotoxicity, Irritation, and Sensitization. All evaluation acceptance criteria were met.

### 7. Conclusion

Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well or better than the legally marketed predicate device (K102855), Class II (21 CFR 876.1500), product code OCX.

---

**Source:** [https://fda.innolitics.com/device/K231418](https://fda.innolitics.com/device/K231418)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).