NextDent Base is for dental professional use only and intended for the manufacturing of denture bases to support artificial teeth to form full or partial removable dentures.
Device Story
NextDent Base is an acrylate-based light-cure resin used for 3D printing denture bases. The device is intended for use in dental laboratories by dental professionals. The process involves using a VAT-based 3D printer (e.g., NextDent 5100 Figure 4) and 3D Sprint software to print the resin in layers, with each layer cured by UV light. Post-printing, parts are cured in a UV light chamber (e.g., NextDent LC-3DPrint Box) and finished using conventional dental methods. The device enables the fabrication of custom denture bases, supporting artificial teeth for removable dentures. The resin's mechanical properties and biocompatibility are verified to ensure safety and performance for permanent oral contact.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility per ISO 10993-1:2018 (cytotoxicity, sensitization, irritation, genotoxicity, systemic toxicity) and ISO 7405:2018. Mechanical performance was verified against ISO 20795-1:2013, including flexural strength, modulus, impact strength, sorption, solubility, and color stability. Results confirmed compliance with requirements for denture base polymers and substantial equivalence to the predicate.
Technological Characteristics
Pre-mixed acrylate-based UV light-cure resin containing pigments and photo-initiators. Polymerization via UV light source in a 3D printer and post-curing chamber. Compliant with ISO 20795-1:2013 and ISO 10993-1:2018. Non-sterile. Shelf life of two years.
Indications for Use
Indicated for dental professionals to manufacture denture bases supporting artificial teeth for full or partial removable dentures.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 20, 2023
Vertex-Dental B.V. % Patsy Trisler, Jd, Rac Consultant Oserve Group US Inc. 350 S. Main Street Doylestown, Pennsylvania 18901
Re: K231388
Trade/Device Name: NextDent Base Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: October 2, 2023 Received: October 2, 2023
Dear Patsy Trisler, Jd, Rac:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bobak Shirmohammadi -5
For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and
{2}------------------------------------------------
ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K231388
Device Name NextDent Base
Indications for Use (Describe)
NextDent Base is for dental professional use only and intended for the manufacturing of denture bases to support artificial teeth to form full or partial removable dentures.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### K231388 Section 5.0
## 510(k) SUMMARY— NextDent Base
| I. SUBMITTER | |
|--------------------|----------------------------------------------------------|
| Submitter Name: | Vertex-Dental B.V. |
| Submitter Address: | Centurionbaan 190<br>3769 AV Soesterberg The Netherlands |
| Contact Person: | N.C. - Peterse – van der Koppel |
| Telephone #: | +31 88 61 60 430 |
| Date Prepared: | October 20, 2023 |
| II. DEVICE | |
|------------------------------|-----------------------------------------------|
| Device Trade Name: | NextDent Base |
| Common Name(s): | Resin, denture |
| Classification Name | Resin, denture, relining, repairing, rebasing |
| Product Codes | EBI |
| Regulatory Class | Class 2 |
| Classification<br>Regulation | 21 CFR 872.3760 |
| III. PREDICATE | NextDent Denture 3D+, K191497 (Primary Predicate) |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DEVICE(s) | |
| | |
| IV. DEVICE<br>DESCRIPTION | |
| Device Identification: | Light-Cure Resin, provided in a container. |
| Device Characteristics: | NextDent Base is a combination of acrylate-based light-cure<br>resin with pigments, polymerized via photo initiators in a 3D<br>printer setting for fabrication of denture bases to support<br>artificial teeth to form full or partial removable dentures. |
| Environment of Use: | Healthcare facility/hospital Dental (technical) laboratory. |
| Summary (Description)<br>of Device: | Fabrication of denture bases with NextDent Base requires a<br>VAT based 3D printer that supports NextDent resins.<br>Printer and resin must be optimized to each other to get<br>complete and precise printed parts. |
{5}------------------------------------------------
| | Printing<br>Printer | Brand | Type | Software |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|------------------------------|-----------|
| | | 3D Systems | NextDent<br>5100 Figure<br>4 | 3D Sprint |
| | Post-Curing<br>Post-cure<br>unit | Brand | Type | Software |
| | | NextDent | LC-3DPrint<br>Box | n.a. |
| Fit of the parts strongly depends on design and library used.<br>Features like undercuts and spacers should be taken in<br>consideration when designing the parts. | | | | |
| Both the NextDentTM 5100 Figure4® 3D printer and the<br>post-curing lightbox NextDentTM LC-3DPrint Box make use<br>of a UV light source to polymerize the NextDent Base resin. | | | | |
| NextDentTM5100 Figure4® 3D printer or NextDentTM LC-<br>3DPrint Box is not included with the device. | | | | |
| Cured parts are finished using conventional dental methods<br>and instruments. NextDent 3D printed cured parts should be<br>cleaned with nonchemical products. | | | | |
| Materials of Use: | A combination of acrylate-based resins with photo-initiator<br>and pigments. | | | |
| V. INTENDED USE /<br>INDICATIONS FOR<br>USE | NextDent Base is for dental professional use only and<br>intended for the manufacturing of denture bases to<br>support artificial teeth to form full or partial removable<br>dentures. | | | |
| VI. TECHNOLOGICAL<br>CHARACTERISTICS | NextDent Base is a pre-mixed acrylate-based UV light-cure<br>resin with pigments and polymerized via photo initiators in a 3D<br>printer setting. Automated printing of the resin in multiple<br>layers, each light-cured before adding next layer, are post-<br>cured in the UV light chamber. The cured parts are finished<br>using dental methods. | | | |
| VII. COMPARISON<br>TO THE PREDICATE<br>DEVICE: | Both NextDent Base and the predicate device are resins<br>intended for the manufacturing of 3D printed full and partial<br>removable denture bases. | | | |
| INTENDED USE: | Both devices are intended exclusively for professional<br>dental work. | | | |
| TECHNOLOGICAL<br>CHARACTERISTICS: | Both NextDent Base and the predicate device have the<br>following similar characteristics: | | | |
{6}------------------------------------------------
| | • Pre-mixed light-cure combination of acrylate-based<br>resin with photo-initiator<br>• Polymerization and post-curing by UV light<br>• Automated printing of resin with the same 3D printer<br>and software in multiple layers, each light-cured<br>before adding next layer, with post curing in lightchamber<br>• Finished by using dental methods | | | | | | |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|----------------------------------|------------------------------|---------------|-------------------------------|-----------|
| | Technological<br>characteristics | Predicate device | | New device:<br>NextDent Base | | | |
| | | Brand | Type | Software | Brand | Type | Software |
| | Design:<br>Printer | 3D<br>Systems | NextDent<br>5100<br>Figure<br>4® | 3D Sprint | 3D<br>Systems | NextDent<br>5100<br>Figure 4® | 3D Sprint |
| | Post-Curing: | | | | | | |
| | Post-cure<br>unit | NextDentTM | LC-<br>3DPrint<br>Box | n.a. | NextDentTM | LC-<br>3DPrint<br>Box | n.a. |
| VIII. SUMMARY OF<br>TESTING<br>[PERFORMANCE<br>DATA] | NextDent Base is tested for mechanical characteristics as<br>part of the product specification. The most applicable<br>standard for mechanical characteristics determination of<br>denture bases is ISO 20795-1:2013 Dentistry - Base<br>polymers - Part 1 Denture base polymers. | | | | | | |
| Biocompatibility<br>Testing: | According to both ISO 7405:2018 and ISO 10993-1: 2018<br>NextDent Base is considered a surface device, in contact<br>with intact oral mucosa, permanent (> 30 days), repeated<br>use.<br>The ISO 10993-1:2018 standard, including parts 3, 5, 10,<br>12, 17, 18 and 23 was followed and the following applicable<br>biological safety hazard endpoints were addressed:<br>• Chemical characterization<br>• Cytotoxicity<br>• Acute systemic toxicity<br>• Subacute/sub chronic toxicity<br>• Irritation or intracutaneous reactivity<br>• Sensitization<br>• Genotoxicity<br><br>In addition, the following risks were considered based on a<br>risk assessment, taking into account the specific nature and<br>duration of exposure to the device<br>• Material mediated pyrogenicity<br>• Chronic toxicity<br>• Presence of phthalates<br><br>In addition, ISO 10993-17 and -18 and ISO/TS 21726 were<br>followed to identify and quantitate the extractables and/or<br>leachables that may be released from the test article to | | | | | | |
{7}------------------------------------------------
| | NextDent Base is compliant to the applicable requirements<br>defined in ISO 10993 including parts 1, 3, 5, 10, 12, 17, 18<br>and 23, ISO 7405 and ISO/TS 21726 for permanent medical<br>devices and therefore biocompatible. |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bench Testing | NextDent Base is tested according to the methods in the<br>industry standard ISO 20795-1:2013 Dentistry – Base<br>polymers - Part 1 Denture base polymers. |
| | NextDent Base is compliant to the requirements defined in<br>ISO 20795-2:2013 for Type 4 light activated material except<br>for ultimate flexural strength, flexural modulus and water<br>solubility. Comparative testing showed that elasticity and<br>impact resistance properties are substantially equivalent to<br>the predicate. NextDent Base is regarding water solubility<br>compliant to requirements defined in ISO 20795-1:2013 for<br>type 2 autopolymerizable materials. The following bench<br>tests are conducted on NextDent Base:<br>Surface characteristics Polishability Color Color stability Translucency Freedom from porosity Ultimate flexural strength Flexural modulus Charpy impact strength Bonding to synthetic polymer teeth Sorption Solubility Dynamic fatigue Fracture work |
| | The mechanical performance of NextDent Base printed on<br>the NextDent 5100 printer is verified confirming its sufficient<br>repeatability and robustness regarding print orientation,<br>placement, and resin reuse. |
| Reprocessing, Sterility<br>and Shelf-Life Testing | The device is provided non-sterile.<br>From the Shelf-life testing, NextDent Base has a shelf life of<br>two years. |
| IX. CONCLUSIONS | NextDent Base and the predicate have the same intended<br>use and similar technological characteristics. |
|-----------------|---------------------------------------------------------------------------------------------------------------------|
| | The results of the performed tests show that NextDent Base<br>meets the applicable requirements of ISO 20795-1:2013 |
{8}------------------------------------------------
| Dentistry - Base polymers – Part 1 Denture base polymers<br>and confirm that the device performs similarly to the<br>predicate device. |
|----------------------------------------------------------------------------------------------------------------------------------------|
| It is therefore concluded that NextDent Base performs as<br>intended and is substantially equivalent to the predicate<br>device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.